Attachment J-5.pdf

PDF 43 KB Posted

Attached to
Calibration and Metrology Services (CAMS) IV
Solicitation number
80JSC021R0051
Issued by

View the file

Other files

File Type Posted
CAMS IV final SSS.pdf PDF
80JSC021R0051 - Amendment 4.pdf PDF
80JSC021R0051 T and Cs - Amendment 4.docx DOCX document
CAMS IV Final RFP Amendment 3 QAs.pdf PDF
80JSC021R0051 - Amendment 3 SF30.pdf PDF
80JSC021R0051 - Amendment 2 SF30.pdf PDF
80JSC021R0051 - Amendment 1 SF30.pdf PDF
Attachment_L3.xlsx XLSX spreadsheet
CAMS IV Final RFP Amendment 2 QAs.pdf PDF
CAMS IV Preproposal Bulletin.pdf PDF
CAMS IV Final RFP Cover Letter.pdf PDF
Attachment L-5.xlsx XLSX spreadsheet
Attachment J-4.pdf PDF
Attachment J-6.xlsx XLSX spreadsheet
Attachment J-1.pdf PDF
Attachment L-3.xlsx XLSX spreadsheet
CAMS IV Draft RFP QAs.pdf PDF
CAMS IV Draft RFP Changes to Final RFP.pdf PDF
Attachment J-3.xls XLS spreadsheet
80JSC021R0051.pdf PDF
Show all 20

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Parameter Range Standard Acceleration - Linear 1 to 10 G's ± 0.50% Rate Table

Acceleration - Dynamic 1 to 10 G's, 30 Hz-900 Hz ± 1% to ± 3% Standard Accelerometers 1 to 10 G's, 900 Hz-10 kHz ± 1% to ± 3%

Attribute Gages Radius ± 25 seconds Optical Comparator, Overlay Angle ± 1 minute

Capacitance 1000 pF, 100 pF, 10 pF ± 5 ppm Standard Capacitors

0.00001 µF to 1µF, 1 kHz ± 0.001% to ± 0.3% Capacitance Bridge

0.00001 µF to 10 mF, 12 Hz to 100 kHz ± 0.02% to 5% LCR Meter

Crimp Tools Dimension ± 100 microinches Pin Gages Crimp Height Digital Point Micrometer

Current-DC 100 µA to 10 A ± 0.015% to 0.05% Current Calibrator 10 A to 100 A ± 0.04% Potentiometer and Precision Shunts 100 A to 500 A ± 0.05% Potentiometer and Precision Shunts 100 nA to 1A ± 0.01% Precision DMM

Current-AC 100 µA to 10 A, 40 Hz to 10 kHz ± 0.04% to 0.40% Current Calibrator 10 to 100 A, 60 Hz to 400 Hz ± 0.05% Precision Shunts 100 µA to 1A, 10 Hz to 50 kHz ± 0.065% to 0.45% Precision DMM

Dial Indicators to 4 inches ± 50 microinches Gage Blocks Lab Master Universal

Durometers Type A ± 0.3 duro Durocalibrator Type D ± 0.4 duro

Fiber Optic Attenuation 0-70 dB ± 0.5 dB Fiber Optic Power Meter 850 & 1300 nm Up to 100 um Core

Fiber Optic Power 850 nm, 1300 nm, 1310 nm & 1550 nm ± 1.30% Fiber Optic Power Meter 100 uW Level Up to 100 um Core

Fiber Optic Wavelength 600 to 1750 nm ± 1nm Optical Analyzer

Flow-Gas 0 to 1000 SLPM ± 0.2% reading Mol-Blox

Flow-Gas [Oxygen Clean] 0 to 1000 SLPM ± 0.20% Laminar Flow Elements, Mol-Blox

Accuracy/Uncertainty

Parameter Range StandardAccuracy/Uncertainty

Flow-Liquid 0.4 to 400 GPM (1.5 to 1500 L/min) ± 0.05% Water Flow Calibrator

.001 to 1 GPM (3.8 x10-3 to 3.8 L/min) ± 0.05%

Flow-Linear 20 to 10,000 ft/min ± 3% Wind Tunnel

Frequency 1, 5 & 10 MHz ± 7 x 10-12 Cesium Beam Standard 1 GHz to 26 GHz ± 3 x 10-11 High Frequency Counter

Force 0 to 500 lbs (0 to 2224 Newtons) ± 0.01% Class-S Mass Standard & Moorehouse Dead Weight Tester Compression 0 to 100,000 lbs ± 0.25% Hydraulic Force Cal Sub-System Tension 0 to 50,000 lbs ± 0.25% Hydraulic Force Cal Sub-System

Torque - Wrench 50 to 1000 lbf-ft (68 to 1355 N-m) ± 0.0065 lbf-ft/lbf-ft + 0.1 lbf-ft Torque Standard 1 to 50 lbf-ft (1.36 to 68 N-m) ± 0.0012 lbf-ft/lbf-ft + 0.032 lbf-ft

1.2 to 100 lbf-in (136 mN-m to 11300mN-m) ± 0.0023 lbf-in/lbf-in + 0.086 lbf-in

Torque - Transducer 50 to 1000 lbf-ft (68 to 1355 N-m) ± 0.00032 lbf-ft/lbf-ft + 0.065 lbf-ft Torque Arms, Mass 1 to 50 lbf-ft (1.36 to 68 N-m) ± 0.00017 lbf-ft/lbf-ft + 0.008 lbf-ft

1.2 to 100 lbf-in (136 mN-m to 11300mN-m) ± 0.0002 lbf-in/lbf-in + 0.056 lbf-in

Humidity-Absolute -70 Deg C to -30 Deg C ± 0.3°C Low Humidity Generator -30 Deg C to 10 Deg C ± 0.13°C 10 Deg C to 25 Deg C ± 0.15°C 25 Deg C to 90 Deg C ± 0.3°C

Humidity-Relative 0 to 90% RH ± 0.6% (15% to 90% RH) Humidity Standard

Inductance 1 x 10-4 to 10 H ± 0.02% to ± 0.1% RCL Digibridge

1 x 10-4 to 1200 H @ 1 KHz ± 0.10% to ± 0.3% Standard Inductors to 4 inches ± 25 microinches Gage Blocks, Lab Master Universal to 24 inches ± 60 microinches to 72 inches ± 100 microinches to 4 inches ± 15 microinches Gage Blocks to 24 inches ± 45 mciroinches to 72 inches ± 100 microinches to 12 inches ± 25 microinches Labmaster Universal, Gage Blocks to 24 inches ± 60 microinches Supermicrometer, Gage Blocks

Liquids:

Inside Dimensional Measuring Devices

Outside Dimensional Measuring Devices

Length/End Standards/Pins & Plugs

Buffer Solution pH 2-10 ± 0.01 pH pH Standards Specific Gravity .001 to 2.11 ± 0.50% Troemner Balance

Mass to 1 lb ± 9 micropounds Standard Weights 5 lb ± 51 micropounds Mass Comparators 10 lb ± 80 micropounds 20 lb ± 130 micropounds 50 lb ± 510 micropounds

100 mg - 1 mg ± 0.004 mg 1 g - 100 mg ± 0.007 mg 1 kg -1 g ± 0.13 mg 2 kg ± 2.6 mg 5 kg ± 5.8 mg 10 kg ± 16 mg 25 kg ± 34 mg

Magnetics 100 to 10 K Gauss (0.01 to 1 T) ± 1% Standard Magnets

Optical Comparators Stage Travel ± 50 microinches Glass Scales Angle Measurement ± 30 seconds Angle Blocks

Surface Plate Flatness Class AA ± 10 microinches Dynamic Laser Measuring System

Phase 0 to 360 Deg (0 to 6.28 Rad) Phase Bridges 1 KHz ± 0.005 Deg. (8.7 x10-5 Rad) < 10 kHz ± 0.030 Deg. (5.2 x10-4 Rad) < 100 kHz ± 0.100 Deg. (1.7 x10-3 Rad) < 1 MHz ± 0.500 Deg. (8.7 x10-3 Rad) < 2 MHz ± 1.000 Deg. (1.7 x10-2 Rad) 0 to 360 Deg, 10 Hz to 50 kHz ± 0.05 Deg. Phase Meter

Power 1 to 2000 Watts, DC to 400 Hz ± 0.10% Precision Voltage and Current Sources

Protractors, Levels to 90 degrees ± 30 seconds Sine Bar, Gage Blocks

Pressure-Static 1e-07 to 4e-3 Torr ± 2.2% reading Stabil Ion Gauges 0 to 104" Hg (0 to 210 kPa) ± 0.004" Hg (±6.75 Pa) Pressure Standard 0 to 10,000 PSI ± 0.02% Pressure Calibration System, Piston Gauge 0 to 1,000 PSI ± 30 ppm RDG Piston Gauge

Pressure [Oxygen Clean] 0 to 1000 PSI ± 0.02% Pressure Calibration System 1000 to 10,000 PSI ± 0.02% Standard Gage

Ratio-DC 1:1 to 1:1000 ± 10 ppm Divider, Resistive

Ratio-AC 1:1 to 1:1000 ± 10 ppm Ratio Transformers

Resistance 1.0 Ohm ± 1 ppm Thomas 1 Ohm Resistors 10 K Ohms ± 3 ppm Resistors

1 x 10-5 to 1 GOhms ± 0.005% Resistance Bridge

RF-Attenuation 1 to 100 db at 30 MHz ± 0.01 db/10 db Waveguide Beyond Cutoff Attenuator 1 to 20 db, DC to 18 GHz ± 0.2 db Coaxial Attenuators 0 to 100 db, 2.5 MHz to 18 GHz ± 0.3 db maximum IF Substitution

RF-Power .5 to 10 mW, 10 MHz to 18 GHz ± 2% Thermistor Mount 10 to 30 W, 10 MHz to 18 GHz ± 5% Power Sensor/Power Meter 1 to 50W, 10 MHz to 500 MHz ± 5% Calorimetric Test Set and Power Meters

Ring Gages to 3.5 inches ± 20 microinches Labmaster Universal, Gage Blocks, Standard Rings

Rise Time 17.5 pS ± 2.5 pS Rise Time Standard

Sound PressureLevel Sensitivity (Ref. 250 Hz) Referenced to 250 Hz ± 0.08 db to 0.20 db SPL 1", 1/2" and 1/4" Microphones Calibration System Acoustic Response 250 Hz to 25 kHz ± 0.10 db to 0.50 db 1", 1/2" and 1/4" Microphones Calibration System

Steel Rules to 72 inches ± 100 microinches Optical Comparator Dynamic Measuring System

Temperature Ref. -195.881 Deg C (Comparison) ± 0.04° C LN2 Comparator Source Ref. -38.8344 Deg C ± 0.014°C Mercury Triple Point Cell

Ref. 0.01 Deg C ± 0.013° C Water Triple Point Cell Ref. 29.7646 Deg C ± 0.014° C Gallium Melting Point Cell Ref. 231.928 Deg C ± 0.018° C Tin Freeze Point Cell Ref. 419.5276 Deg C ± 0.029° C Zinc Freeze Point Cell

Measure -200 Deg to 660 Deg C ± 0.01° C Platinum Resistance Thermometer 800 Deg C to 2300 Deg C ± 1.5° C to 3.5° C Thermocouple/Thermocouple Simulation

Infrared (IR Thermometry) -15°C to 500°C ± 0.53° C to 2.4° C Black Body Calibrator

Thread Plugs 2-80 TPI, to 3 inches ± 100 microinches Measuring Thread Wires, Supermicrometer, Gage Blocks

Thread Rings 2-80 TPI, to 3 inches ± 200 microinches Setting Plugs

Voltage-DC 10 V ± 0.04 ppm Programable Josephson Voltage Standard 100 mV to 1 kV ± 2 ppm Reference divider / Null Detector

1x10-6 to 1.5 kV ± 0.001% Voltage Divider

0.5 kV to 30 kV ± 0.01% Voltage Divider

Voltage-AC 2 mV to 70 V, 20 Hz to 1 MHz ± 25 ppm to 1500 ppm AC Measurement Standard 70 V to 1 kV, 20 Hz to 100 kHz ± 50 ppm to 850 ppm AC Measurement Standard 500 mV to 3V, to 1 kHz to 100 MHz ± 0.01% to 1.0% Thermal Voltage converter

VSWR 10MHz to 18 GHz, 40 dB Return Loss ± 2% Network Analyzer

Vacuum 4x10-7 to 1 x10-3 Torr (2.7 x 10-7Pa to 0.13 Pa) ± 5% Ion Gages

1 x 10-5 to 1 Pa. ± 2% Spinning Rotor Gage

1 x 10-3 to 1000 Torr (0.13 Pa to 1.3 x105 Pa) ± 0.08% Vacuum System

Summary