This document is a detailed technical specification and equipment drawing for the AMSCO® 7052HP Single-Chamber Washer/Disinfector manufactured by STERIS Corporation. The device is a cabinet-type mechanical washer intended for cleaning and intermediate-level disinfection of soiled reusable medical instruments, utensils, and related articles in healthcare facilities. It features a touch screen control system, eight factory-loaded cleaning cycles, a patented spray arm system, and can be configured with steam or electric heating. The washer has dimensions of 42 x 80-3/4 x 32 inches, weighs approximately 1,400 pounds when shipping, and is designed to conform to various international safety and electromagnetic compatibility standards.
The washer requires specific utility connections including electricity (200-208V, 380-415V configurations), hot and cold water, pure water, compressed air, and steam inlet. It has comprehensive safety features such as vertical chamber doors with obstruction sensors, a safety lockout feature, and door interlock to prevent cross-contamination. The unit offers optional accessories like a printer, Acu-Rinse Reservoir, and RAS Cycle for robotic-assisted surgery instruments. The specification includes detailed requirements for water quality, pressure ranges, flow rates, and environmental operating conditions. The washer is designed for indoor use at altitudes up to 6,562 feet, with operating temperatures between 41-104°F and a maximum relative humidity of 80% at 88°F, decreasing linearly to 50% at 104°F.
View the file
Other files
Show all 11
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
AMSCO® 7052HP— 10283376 Rev. J 1
AMSCO® 7052HP
Single-Chamber Washer/Disinfector
APPLICATION
The AMSCO 7052HP Single-Chamber Washer/Disinfector with touch screen control is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans and urinals, trays, bowl, basin, kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
DESCRIPTION
The AMSCO 7052HP Single-Chamber Washer/Disinfector is a cabinet-type mechanical washer that includes:
• Color touch screen control system and visual progress indicator
• Audible warning system with adjustable volume
• Space saving, power vertical sliding doors
• Eight factory-loaded (Instruments, Orthopedic
Instruments, Utensils, Plastic Goods, Rigid M.I.S., Anesthesia / Respiratory Goods, Gentle and Decontamination) and up to 20 Custom cycles. Each cycle is effective, fast, energy efficient and may be customized (within allowed range of parameters) to meet specific operating requirements
• LED status indicator lighting
• Compact space saving footprint
• Conforms to seismic building code requirements
• Patented spray arm system providing entire chamber coverage
• 180 degree reverse installable
SPECIFICATIONS
Configurations Steam-Heated, Electric-Heated
Size (W x H x L) Operating dimensions: 1067 x 2051 x 813 mm (42 x 80-3/4 x 32") Chamber dimensions: 673 x 667 x 648 mm (26-1/2 x 26-1/4 x 25-1/2")
Loading height: 787 mm (31") above finished floor
Weight Shipping Weight: Approximately 635 kg (1400 lb) Operating Weight: 700 kg (1543 lb)
AMSCO® 7052HP— 10283376 Rev. J 2
Utility Requirements
Refer to proper equipment drawing for installation details and specifications:
Electric-Heated Unit: 10292222, 10292223, 10292224 or 10292225 Steam-Heated Unit: 10292218, 10292219, 10292220 or 10292221
Electricity • 200 to 208 V, 60Hz (Steam)
• 200 to 208 V, 50 Hz (Steam)
• 380 to 400 V, 60 Hz
• 380 to 415 V, 50 Hz
• 460 to 480 V, 60 Hz
STANDARDS
These AMSCO 7052HP Single-Chamber Washer/Disinfector and AMSCO 7053HP Single-Chamber Washer/Disinfector meet the applicable requirements of the following:
Safety Standards - General Requirements:
• IEC 61010-1:2010+A1:2016, Ed. 3.1
• UL 61010-1:2012 Ed.3+R:19Jul2019
• CSA C22.2 No. 61010-1-12 Ed.3+U1;U2;A1 Safety Standards - Particular Requirements:
• IEC 61010-2-040:2020 Ed 3
• UL 61010-2-040:2021 Ed. 3
• CSA C22.2 No. 61010-2-040:2021 Ed 3 Electromagnetic Compatibility (EMC) Standards:
• IEC 61326-1:2020
• Immunity test reference:
– IEC 61000-4-2: 2008
– IEC 61000-4-3: 2006
– IEC 61000-4-4: 2004
– IEC 61000-4-5: 2005
– IEC 61000-4-6: 2008
– IEC 61000-4-8: 2009
– IEC 61000-4-11: 2004-03
• Emissions test reference:
– CISPR 11:2015+AMD1:2016 CSV (EN 55011:2016
+A1:2017) United States of America / Conducted and Radiated Emission testing
• Federal Communications Commission (FCC) CFR 47, Part
15, Subpart B, Class A Unintentional Radiators for use in Commercial and Industrial Areas.
INFORMATION TO USER:
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provided reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
WARNING: Changes or modifications not expressly approved by STERIS could void the user's authority to operate the equipment.
Canada / Conducted and Radiated Emission testing
• Industry Canada ICES-001, Issue 4, Class A, Group 1
LABELING REQUIREMENTS:
Self-Declaration of Compliance (SDoC) to Industry Canada
ICES-001.
Guidelines and other regulations followed:
• Performance requirements:
– ISO 15883-1:2006/A2014 (EN ISO 15883-1:2009
+A1:2014)
– ISO 15883-2:2006 (EN ISO 15883-2:2009)
– ANSI/AAMI ST15883-1:2009 (R2014)
– ANSI/AAMI ST15883-2:2013 (R2015)
– CAN/CSA-Z15883-1-09 (R2019)
– CAN/CSA-Z15883-2-09 (R2019)
Building Construction Code:
• United States of America
– California Building Standards Code (California Code of
Regulations, Title 24) for Seismic Anchoring Requirements
The software life cycle processes of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector meets the following standard:
• IEC 62304:2006/Amd 1:2015 Governing Directive for the affixing of the CE mark:
AMSCO® 7052HP— 10283376 Rev. J 3
• Medical Device Regulation (EU) 2017/745 (MDR)
– EC Certificate: MDR 761318.
Conformity to other applicable directives:
• Machinery (Directive 2006/42/EC)
• Electromagnetic compatibility (Directive 2014/30/EU)
FEATURES
Vertical sliding power door:
• Door is constructed of double pane tempered glass to allow operator to view chamber interior with door closed.
• Color LED light illuminates top portion of each door to indicate machine status.
• Door is pneumatically activated using touch screens located on the control panel (one each side of unit) and is equipped with a built-in safety system.
• Each door is mounted on a compressed seal that reduces heat loss and increases heating capability.
• If a power failure occurs, door can be opened manually. A door interlock feature is provided to prevent cross-contamination.
• Unit can be configured in single- or double-door operation.
Stainless-steel pump is powered by a dual-speed motor.
High pump speed provides equivalent capacity of a 10.4 kW (14 hp) motor, 625 L/min at 60 m (165 U.S. gal/min at 196 ft) head pressure. Low pump speed provides equivalent capacity of a 2.6 kW (3.5 HP) motor, 341 L/min at 14 m (90 U.S. gal/ min at 46 ft) head pressure. Pump impeller is mounted directly on motor shaft and does not require additional bearings. Pump motor is equipped with a drip-proof frame, magnetic starter, overload protection and sealed bearings (not requiring periodic lubrication).
Pump, spray system and all recirculating piping are of stainless-steel construction.
Wash chamber:
• Constructed of 16-gauge, 316L stainless steel (No. 4 finish), argon-welded and polished. Chamber is of sanitary-type design (horizontal fixed surfaces are sloped, without overlapping sheet metal) for complete drainability and reducing hard-to-clean locations.
• Single-walled, insulated construction of chamber exterior reduces heat loss and noise level to the work area.
• Rotary spray assemblies are positioned (one at top and one at bottom of chamber, each measuring 620 mm [24- 3/8"] long) to reach all load surfaces. Spray assemblies are easily removable without tools to aid in cleaning and maintenance.
Sump:
• Constructed of 316L stainless steel with 27 L (7.2 gal) capacity with removable stainless-steel filter preventing debris from entering pump and piping system.
• Heating coil (steam or electric) at the bottom of the wash chamber (sump) raises and maintains water temperature up to 85°C (185°F) during the Wash phase and up to 90°C (194°F) during the Thermal Rinse phase.
Chemical peristaltic injection pumps:
• Pre-Programmed cycles and pumps are optimized to use
Prolystica® Ultra Concentrate (PUC) HP chemicals. Each ultra concentrated product is 10 times the concentration of a traditional product; so, 10 times less chemical is injected per cycle.
• Washer/Disinfector comes with one enzyme and one detergent injection peristaltic pump which automatically add a selected quantity of chemical from 0.2 to 16 mL/L (1/40 to 2 oz/gal).
• One pump is dedicated to automatically add a selected quantity of lubricant during Thermal Rinse phase.
• Pumps are positioned near chemical containers. A low-level sensor is included to indicate when detergent level is low, or when insufficient chemical is available for next cycle.
• 30 m (100’) extension tubes are included to pump chemicals from canisters installed in a remote location to unit.
• Control monitors volume of chemicals injected and indicates if this parameter meets specified criteria during all specific phases.
• Featuring the HP Pick-Up Tubes, capable delivering chemistry from a single container to one or two washers.
Refer to Technical Data Sheet PS00000540, HP Pick-Up Tube, for details.
Top Utility Connections (except drain connections) facilitate installation. All utilities (including vent, steam, electric, water and compressed air) are connected on washer top. Top water injection gives a secure air gap of 800 mm (31-1/2").
Backflow preventers are not required for hot and pure water.
Front Service Panels provide easy access to piping, valves, electrical components and wiring. Unit can be configured as either a Vented System that exhausts chamber vapors to building exhaust system through 76 mm (3.0") OD vent connection located at top of washer or as a Non-Vented Drying System where chamber vapors are exhausted through cold water condenser to facility room.
Drying System uses uniquely designed four-sided inflow drying pattern to produce high-flow air curtain. Air curtain provides broad, efficient, drying within wash chamber.
Recycled and non-recycled air is manifolded and circulated through piping and accessory providing energy efficient system while ensuring complete chamber air coverage. Fresh air is drawn through a HEPA filter. Drying system includes a
2.2 kW (3 hp) blower, to remove vapor from chamber prior to opening doors, and three electric heaters totaling 11.4 kW to heat and maintain chamber air temperature. When Acu- Drying is selected, drying time self-adjusts to meet drying time of loaded rack and items.
AMSCO® 7052HP— 10283376 Rev. J 4
Pure Water Stainless-Steel Supply Valve is supplied on washer/disinfector so Pure Water Rinse Supplymay be used with factory programmed cycles and optional Acu-Rinse Reservoir.
Drain Discharge Cool Down ensures water drained at end of each phase, from chamber sump to building drain system, does not exceed 60°C (140°F). If water temperature in sump is higher than 60°C (140°F), cold water is automatically added to reduce temperature of water discharged into building drain system.
Sampling Valve is supplied in washer/disinfector recirculation piping for North American Customers that are not able to install sampling valves on facility supply lines near washer/ disinfector.
Barrier Wall Kit Flanges for one wall (clean side) installation come standard. An optional kit to enclose second wall (dirty side) is also available (FD351).
Cycle Graph enables monitoring of injection volume, water circulating pressure and water temperature during a cycle.
Graphs are visible during cycle and can also be displayed at cycle completion. Data can be exported using ConnectAssure Technology or USB port.
SAFETY FEATURES
Vertical chamber doors are equipped with an obstruction sensor to detect any door obstruction. If obstruction is present, door automatically opens.
The washer/disinfector is equipped with a safety lockout feature so a cycle cannot start unless the door is fully closed.
If the door is opened during a cycle, all utility services to chamber are shut off, and the cycle stops.
Door interlock feature is provided to prevent cross-contamination. Door interlock feature allows only one door to be opened at a time whenever power is on. When the cycle is in process, door interlock prevents either door from being opened. Access to the load is then restricted.
Safety stop pushbutton(s), one on the load and one on the unload side, automatically stop all unit operation when pressed.
Building electrical supply disconnect switch must be used to shut off power to the unit before servicing.
CYCLE DESCRIPTIONS
NOTE: STERIS does not intend, recommend or represent in any way that this AMSCO 7052HP Single-Chamber Washer/ Disinfector be used for the terminal disinfection or sterilization of any regulated medical device. AMSCO 7052HP Washer/Disinfector is intended only to perform an initial step in the processing of soiled, reusable medical devices. If medical devices contact blood or compromised tissues, such devices must be terminally processed in accordance with good hospital practices before each use in human patients.
Once treatment cycle is selected, washer/disinfector automatically processes load through standard phases (additional phases are included in certain treatment cycles depending on unit configuration) as described on Cycle Charts and in supplied Operator Manual.
AMSCO 7052HP Single-Chamber Washer/Disinfector enables Customers to use Dual Wash phase combining PUC HP enzyme and PUC HP detergent in the same filling which accelerates cycle time and reduces water consumption.
Custom cycles are available but must be validated by the Customer (not STERIS responsibility).
CONTROL SYSTEM
The AMSCO 7052HP Single-Chamber Washer/Disinfector is equipped with a touch screen Customer interface. This 213 mm (8.4") color touch screen is mounted at eye level beside chamber door (both sides of unit) and tilted for better visibility, allowing easy monitoring of all wash cycles. Control system monitors and controls all phases of each programmed cycle.
The control system features:
• Cycle selection is simplified through use of Icons representing devices and cycle confirmation (on-screen) to ensure correct cycle is selected.
• Service mode for preventive maintenance testing and to facilitate troubleshooting.
• Built-in service diagnostic program to permit system calibration and verification of component operations.
• Security lock-out feature that enables cycles and temperatures to be locked and unchangeable without proper access code.
• Cycle data is stored as a protection against power disruption. Data may be downloaded from controller using supplied USB ports.
• Permits operator to monitor current washer/disinfector status (including current chamber temperature, circulating pressure and time remaining in phase).
• Indicates any abnormal conditions with audible and visual alarm system.
INSTALLATION
The washer/disinfector is designed as a fully enclosed cabinet for freestanding or recessed installation. Clearance between top of the unit and the ceiling must be at least 356 mm (14").
If the system is recessed through one or two barrier walls, stainless-steel barrier flanges are included to provide a finished wall appearance.
Once installed, system is designed to allow for easy access for maintenance purposes.
All configuration modifications are available upon request.
Contact STERIS to receive a quotation.
ACCESSORIES AND OPTIONS
Printer (FD094 or FD012), if provided, produces an easy-to-read printed record of whether load was properly processed at preset temperature, as well as a complete list of alarm and abort in-cycle messages. A paper out LED is included.
Acu-Rinse Reservoir (if equipped) pre-heats Pure Water for Rinse and Thermal Rinse phases significantly reducing cycle times. Constructed of 316L stainless steel.
AMSCO® 7052HP— 10283376 Rev. J 5
Acu-Wash Reservoir recovers a portion of water used in Thermal Rinse treatment and stores it for use in Wash treatment of next cycle reducing overall cycle water consumption.
RAS Cycle (Robotic Assisted Surgery Cycle) (FD000057 or FD258) is used with RAS 12 Rack (FD256) for effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci X/Xi and S/Si EndoWrist1 instruments.
Steam Condensate Return To Drain (FD353) allows for connection of steam condensate return outlet to drain when steam condensate return line is not available in building. Cold water is also injected in drain piping when condensate return water temperature is too high. Condensate return cool down keeps temperature in drain piping below 60°C (140°F).
Seismic Tie-Down Kit (FD354) includes a seismic report for proper installing and securing of washer to building floor. Unit is designed to comply with seismic building code.
Additional Chemical Pump (FD349; up to two pumps) is available, giving flexibility to wash with neutral process, alkaline process, or to vary chemical used, depending on load type.
Drain Acu-CoolDown (FD361) precisely injects correct amount of potable water into facility drain system, minimizing volume of cold water needed.
Offset Installation Air Management Kit (FD352) controls airflow at top of door assemblies.
ConnectAssure Data Export (CAT3030) exports XML cycle records from equipment to Customer tracking system.
ConnectCare Technology Platform enriches the lives of equipment operators by empowering them with technology and data to help transform the way they work. In addition to putting data into your hands, it enables your STERIS Equipment Service Technician to utilize equipment performance data to take proactive steps to prevent component failure and help you avoid unplanned downtime.
High-quality STERIS service coupled with innovative technology reduces downtime, enhances operator experience, and delivers great outcomes. Ask your local Equipment Service Technician for more information.
PREVENTIVE MAINTENANCE
Regularly servicing your equipment with STERIS-recommended preventive maintenance helps to extend the useful life of the equipment and minimizes equipment breakdown. The STERIS Preventive Maintenance program includes equipment adjustments, replacement of worn parts and other maintenance activities performed on regularly scheduled time-based intervals to help ensure optimal equipment performance and to help avoid untimely and costly interruptions. With an average of 12 years of experience and over 200 hours of training, STERIS Equipment Service Technicians provide the knowledge and expertise to perform thorough Preventive Maintenance on your equipment. Each Equipment Service Technician's trunk is stocked with the right STERIS OEM parts to provide on-time preventive maintenance and industry-leading first-call fix rates. Contact your local District Service Manager for more information.
NOTES
1. Machine segments (unit is bolted onto a wooden skid) are shipped in one cardboard carton (W x H x L): 1270 x 2311 x 1067 mm (50 x 91 x 42").
2. Customer must ensure the washer/disinfector stands on a noncombustible floor (floor should be level).
3. Customer must provide utility connections with shutoff disconnects within 0.6 m (2’) of equipment perimeter and below ceiling deck or purchase Installation Kit from
STERIS.
4. STERIS recommends vacuum breakers (not provided by STERIS) be installed on service lines, and disconnect switches (with lockout in OFF position; not provided by STERIS) be installed in electric supply lines near equipment.
5. For all ventilation ducting from the washer/disinfector, STERIS recommends installation of a dedicated, corrosion-proof, flexible watertight duct (76 mm [3"] OD) to the exterior of the building, sloped toward the washer/disinfector.
6. Minimum ceiling height for door removal is 2388 mm (94").
7. STERIS recommends illumination of the service area along with providing a convenience outlet for maintenance.
8. Noise level of 60.6 dB; Operating weight of 700 kg (1543 lb).
9. Unit is shipped with an Operator Manual. Maintenance Manual is available for purchase.
10. ConnectAssure Data Export requires an Ethernet drop and Ethernet connection within 10 ft of the equipment.
The Customer must supply the network cable.
Refer to the Following Equipment Drawings for Installation Details
Equipment Drawing Number Equipment Drawing Title
Electric-Heated Configurations
10292222 AMSCO 7052HP Electric heated, W/ Acu-Rinse Res., Unit flush to wall
10292223 AMSCO 7052HP Electric heated, W/ Acu-Rinse Res., Unit offset from wall
1. da Vinci® and EndoWrist® are registered trademarks of Intuitive Surgical, Inc.
AMSCO® 7052HP— 10283376 Rev. J 6
Refer to the Following Equipment Drawings for Installation Details
Equipment Drawing Number Equipment Drawing Title
10292224 AMSCO 7052HP Electric heated, W/O Acu-Rinse Res., Unit flush to wall
10292225 AMSCO 7052HP Electric heated, W/O Acu-Rinse Res., Unit offset from wall
Steam-Heated Configrations
10292218 AMSCO 7052HP steam heated, W/ Acu-Rinse Res., Unit flush to wall
10292219 AMSCO 7052HP steam heated, W/ Acu-Rinse Res., Unit offset from wall
10292220 AMSCO 7052HP steam heated, W/O Acu-Rinse Res., Unit flush to wall
10292221 AMSCO 7052HP steam heated, W/O Acu-Rinse Res., Unit offset from wall
Selections Checked Below Apply To This Equipment
POWER2
Steam-Heated Unit Electric-Heated Unit
VOLTAGE2
200 to 208 V, 60 Hz (Steam) 200 to 208 V, 50 Hz (Steam) 380 to 400 V, 60 Hz 380 to 415 V, 50 Hz 460 to 480 V, 60 Hz
ACCESSORIES3
Impact Printer (FD094) Thermal Printer (FD012) Seismic Tie-Down Kit (FD354) Condensate Return To Drain (FD353) Two-Level Manifold Rack with Equal Space (FD79-100)) Three-Level Manifold Rack (FD74- 900) Four-Level Single-Chamber Manifold Rack (FD75-000)
Four-Level Manifold Rack (FD75- 100) Five-Level Manifold Rack (FD75-200) Multi-Function Rack for Large Items
(FD75-500)
Multi-Function Rack for Small Items
(FD75-600)
General Purpose Basket (FD77-100) RAS 12 Rack (FD256) Self Docking Transfer Cart (FD199) Transfer Cart Adapter (FD209) Set of Dry Contacts (FD362) Handheld Bar Code Reader (FD363) Conductivity Probe (FD350) Additional Wall Flange (FD351) Installation Kit – Steam (FD088) Installation Kit – Electric (FD037) Installation Kit – Steam with Canadian Registration Number (CRN)
(FD057)
Additional Chemical Pump (FD349)
Offset Installation Air Management Kit (FD352) IQ-OQ-PQ Protocols (FD359) Machine Identifier (11000866) Drain Acu-CoolDown (FD361) ConnectAssure Technology for Data Export (CAT3030) ConnectCare Technology Platform (Remote Monitoring Services, Equipment Performance Reports, Service Documentation, Troubleshooting Support and Equipment Install Base Details.
Available in U.S. and Canada only)
OPTIONS
Acu-Rinse Reservoir (FD000054) Acu-Wash Reservoir (FD000055) 40A circuit retrofit compatibility option (only available on 208V configuration without AcuRinse option unit) (FD000056) RAS Cycle (FD000057)
Item:
Locations:
2. Careful consideration must given to voltage selection prior to ordering. Later changes require substantial field modification.
3. Refer to SD867 for Material Handling Accessories. Refer to 10726323 for AMSCO® Top Enclosure Panels Accessories. Refer to SD857 for SCS
Conveyor System Accessories. Refer to 10370003 for ATLAS WAV System Accessories.
dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight dacciarri Highlight
AMSCO® 7052HP— 10283376 Rev. J 7
For Further Information, contact:
STERIS Corporation 5960 Heisley Rd.
Mentor, OH 44060–1834 • USA 440–354–2600 • 800–548–4873 www.steris.com
The base language of this document is ENGLISH. Any translations must be made from the base language document.
CUSTOMER IS RESPONSIBLE FOR
COMPLIANCE WITH APPLICABLE LOCAL
AND NATIONAL CODES AND
REGULATIONS.
©2024, STERIS Corporation.
All rights reserved.
This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.
www.steris.com
42 [1067]
UNIT WIDTH
80-3/4 [2051]
UNIT
HEIGHT
31 [787]
LOADING
HEIGHT
26-1/4 [667]
LOADING
CLEARANCE
HEIGHT
8-1/2 [216]
[381]
91 [2311]
RAISED DOOR
HEIGHT
34 [864]
MAX LENGTH
32 [813]
UNIT LENGTH
90 [2286]
SERVICE
AREA
HEIGHT
45 [1143]
WALL OPENING
*(NOTE F)
1-1/2 [38]
MINIMUM
94 [2388] *(NOTE G)
RECOMMENDED
98 [2489] *(NOTE H)
MAX 1 [25]
ADJUSTMENT WITH
WASHER LEGS
(UNIT HEIGHT 81-3/4 [2077] MAX)
MINIMUM
94 [2388] *(NOTE G)
RECOMMENDED
96 [2438] *(NOTE H)
81-7/8 [2080]
MAX
HEIGHT
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
SERVICE AREA
REQUIRED FOR
DRYING MOTOR
REMOVAL
TYPICAL INSTALLATION
NOTES:
(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].
(B) SERVICE ACCESS SIDE OF UNIT IS RECOGNIZED BY THE TOUCH SCREEN AT THE RIGHT OF
WASH CHAMBER. ELECTRICAL AND CONTROL COMPONENTS, AS WELL AS MOST OF
SERVICABLE PARTS ARE ACCESSIBLE FROM THAT SIDE OF UNIT. NON-SERVICE SIDE IS
EQUIPPED WITH THE TOUCH SCREEN ON THE LEFT OF THE WASH CHAMBER.
(C) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.
(D) WEIGHTED EQUIVALENT SURFACE SOUND PRESSURE LEVEL: 64.0 dB(A) CALCULATED AS DESCRIBED IN ISO 3746.
(E) VERTICAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 3-1/2 in [89 mm] WIDTH X 80 in [2032 mm] LONG.
(F) HORIZONTAL BARRIER WALL FLANGES ARE PROVIDED WITH UNIT TO FILL GAP BETWEEN
UNIT AND WALL ON BOTH SIDES. DIMENSIONS ARE 49 in [1245 mm] WIDTH X 16 in [406 mm] HEIGHT.
(G) MINIMUM REQUIRED DIMENSION FOR DOOR REMOVAL WITH LIMITED ADJUSTMENT AVAILABLE: +0 in [0 mm] THE STANDARD BARRIER WALL FLANGES CAN'T BE USED AT THAT DIMENSION. -3/4 in [19.1 mm]
(H) RECOMMENDED DIMENSION FOR DOOR REMOVAL WITH FULL LEVELING ADJUSTMENT AVAILABLE: +1 in [25.4 mm] THE STANDARD BARRIER WALL FLANGES CAN BE USED AT THAT DIMENSION. *(NOTE E AND F) -3/4 in [19.1 mm]
FRONT VIEW
SERVICE ACCESS
SIDE OF UNIT
*(NOTE B)
TOP VIEW
PASS THROUGH
DOOR
SIDE VIEWLEVEL FLOOR
*(NOTE C)
W A
LL
O
P E
N
IN
G H
E
IG
H T
C E
IL
IN
G H
E
IG
H T
FO
R
D O
O R
R E
M O
V A
L
Page 1 of 10 10292220.D
Released
2X 33-7/16 [849]
2X 27-11/16 [704]
5-11/16 [144]
7-1/8 [181]
4X 15 [381]
2-1/8 [54]
17-3/8 [441]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) CONCRETE FLOOR MUST REMAIN FREE OF DEFECTS IN THESE AREA, AS PER ESR-1917 & ESR-3187.
FOR INSTALLATION OF SEISMIC ANCHORS (NOT PROVIDED BY STERIS).
(B) THE FLOOR SINK AND DRAIN SHALL NOT OCCUPY THESE AREAS. 5 in x 5 in [127 mm x 127 mm] FLOOR SINK WITH Ø4 in [102 mm] DRAIN IS RECOMMENDED WHEN USING SEISMIC MOUNTS.
SEISMIC MOUNTS
SEISMIC ANCHORAGE
FOR DETAIL SEE FD354
QTY 4 ANCHORS
FINISHED WALL
PREFERENCE
LOCATION OF
FLOOR SINK
*(NOTE A AND B)
Page 2 of 10 10292220.D
80-3/4 [2051]
7-1/8 [181]
[381]
31-1/8 [791]
[533]
23-1/32 [585]
20-17/32 [521]
19-1/2 [495]
3-7/8 [99]
[457]
15-1/2 [394]
32-11/32 [822]
HW HOT WATER
CW COLD WATER
PW PURE WATER
A COMPRESSED AIR
D FLOOR DRAIN LOCATION
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (PLUMBING)
CONNECTION POSITION
FRONT VIEW
A
D
PW
HW
CW
SIDE VIEW
HW
PW
CW
A
D
CR
STOP VIEW
HW
CW
PW
A
CR
S
D
S STEAM
CR CONDENSATE RETURN
S
CR
Page 3 of 10 10292220.D
79-1/4 [2013]
35-13/16 [910]
22-3/8 [568]
[636]
38-1/8 [967]
39-5/8 [1007]
E ELECTRICAL
N NETWORK
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (ELECTRICITY)
CONNECTION POSITION
FRONT VIEW SIDE VIEW
E
N
E
N
TOP VIEW
E
N
Page 4 of 10 10292220.D
6-5/32 [156]
[178]
81-29/32 [2080]
EX VENTILATION EXHAUST
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY (VENTILATION)
CONNECTION POSITION (OPTIONAL)
FRONT VIEW SIDE VIEW
EX
SUGGESTED DUCTWORK
(BY OTHERS)
EX
VAPOR CONDENSATE
COLLECTING MANAGED
BY THE UNIT
TOP VIEW
EX
VENTILATION CONNECTION (OPTIONAL)
(SEE PAGE 9 *(NOTE 1)
DRY CONTACT (SEE PAGE 10 *(NOTE 3))
EX
BY OTHERS
Page 5 of 10 10292220.D
60 [1524]
REF
8-1/2 [216]
132 [3350]
MAX.
17 [432]
INSTRUMENT
(With PUC HP)
UTENSIL
(With PUC HP)
CHEMICAL
(CH)
ENZYME CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
DETERGENT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.74 oz [22 ml] 0.74 oz [22 ml]
LUBRICANT CONSUMPTION PER CYCLE
(SEE PAGE 7; *(NOTE 4)) 0.2 oz [6 ml] N/A
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
*NOTES:
(A) DOSING STATION AND CHEMICAL CONTAINERS MUST BE ON THE SAME FLOOR OF THE BUILDING AS THE WASHER/DISINFECTOR.
CHEMICAL CONTAINER MUST BE LOCATED AT A LOWER HEIGHT THAN DOSING STATION. MAXIMUM CONTAINER INCLINATION
IS 15 DEG. FROM FLOOR. AMBIENT TEMPERATURE SHALL NOT EXCEED 32 °C [90 °F]
(B) HP PICKUP TUBE SYSTEM CAN PROVIDE UP TO TWO WASHERS WITH CHEMISTRY FROM THE SAME CONTAINER.
DOSING STATION
CEILING
ROUTE CHEMICAL LINES
AND SIGNAL CABLE
THROUGH PIPE TO REACH
BOTTOM OF UNIT
CHEMICAL CONTAINERS
*(NOTE A)
12 ft [3.6 m] OF INJECTION LINE
(FROM CONTAINERS TO
DOSING STATION)
*(NOTE B)
DOSING STATION
WEIGHT : 15 lb [6.8 kg]
2X FD349 SHOWN
UP TO 5 TUBES
100 ft [30 m] OF INJECTION LINE
AND SIGNAL CABLE
DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.
STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND
SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.
EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO
WASHER/STATION TO AVOID FORMING BUNDLES THAT COULD
AFFECT THE CHEMICAL INJECTION SYSTEM'S PERFORMANCE
(DO NOT CUT SIGNAL CABLE).
Page 6 of 10 10292220.D
UTILITY REQUIREMENTS CHART (PLUMBING)
UTILITY PARAMETER REQUIREMENT NOTES
HOT WATER
(HW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
1. STERIS REQUIRES TO INSTALL WATER HAMMER
ARRESTOR, UNION & PRESSURE GAUGE.
2. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
*STERIS OFFERS FLEXIBLE HOSES
ACCESSORY (FD088 OR FD037)
WITH INSTRUCTIONS.
3. SAMPLING VALVES INSTALLATION AS CLOSE AS
POSSIBLE TO WASHER WATER CONNECTIONS IS
RECOMMENDED.
PRESSURE RANGE
15 - 50 psig DYNAMIC
[103-345 kPa] MAX. 90 psig STATIC [620 kPA]
OPERATING FLOW RATE 6.8 US GPM [25.7 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
9.9 US GAL [37.5 L]
WITH ACU-WASH RESERVOIR:
6.6 US GAL [25.0 L]
TEMPERATURE RANGE 110 °F [43 °C] TO 150 °F [65.5 °C]
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
COLD WATER
(CW)
*(NOTES 6)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
4. CONSUMPTION BASED ON INSTRUMENT CYCLE WITH
PROLYSTICA ULTRA CONCENTRATE HIGH
PERFORMANCE CHEMICALS (PUC HP) WITH
THE FOLLOWING UTILITY CONDITIONS:
THREE-LEVEL MANIFOLD RACK
(CW) COLD WATER PRESSURE: 30 psig DYNAMIC [207 kPa]
(CW) COLD WATER TEMPERATURE: 70 °F [21 °C]
(HW) HOT WATER TEMPERATURE: 150 °F [65 °C]
(PW) PURE WATER TEMPERATURE 70 °F [21 °C]
(S) STEAM SUPPLY: 80 psig DYNAMIC [552 kPa]
UTENSIL CYCLE CONSUMPTION:
TWO-LEVEL MANIFOLD RACK
CW: 10.4 US GAL [39.4 L]
HW: 9.6 US GAL [36.3 L]
PW: 6.4 US GAL [24.2 L]
S: 14.1 lb [6.4 kg]
5. STERIS RECOMMENDS THE FOLLOWING WATER
QUALITY PARAMETERS:
TOTAL DISSOLVED SOLIDS: 100-200 ppm (MAX. 500 ppm) TOTAL ALKALINITY (CaCO3) : 70-120 ppm (MAX. 180 ppm) pH : 6.8-7 (MAX. 7.5) TOTAL SILICA : 0.1-0.5 ppm (MAX. 1 ppm)
PRESSURE RANGE
30 - 50 psig DYNAMIC
[206-345 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 10.7 US GPM [40.5 L/min] TO
14.1 US GPM [53.4 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
*(NOTE 4)
FOR SUMP FILLING: 10.5 US GAL [39.8 L]
(ADD THE FOLLOWING CONSUMPTION
FOR OPTIONS/CONFIGURATIONS)
CONDENSATE RETURN COOL DOWN:
6.2 US GAL [23.5 L]
NON-VENTED SYSTEM:
12.9 US GAL [48.8 L]
DRAIN COOLDOWN:
14.5 US GAL [54.9 L]
DRAIN ACU-COOLDOWN:
10.9 US GAL [41.3 L]
TEMPERATURE RANGE 70 °F [21 °C] MAXIMUM
WATER QUALITY *(NOTE 5) HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)
PURE WATER
(PW)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
6. BACK FLOW PREVENTER AGAINST BACK
SIPHONAGE MAY BE REQUIRED (LOCAL
PLUMBING CODE)
7. PW IS OPTIONAL. IF UNAVAILABLE, UNIT WILL
PROCESS THERMAL RINSE USING HOT WATER
SUPPLY.
PRESSURE RANGE 5 - 30 psig DYNAMIC [30 - 206 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING FLOW RATE 5.5 US GPM [21 L/min] TO
12.6 US GPM [48 L/min]
TOTAL CONSUMPTION PER
INSTRUMENT CYCLE
* (NOTE 4 AND NOTE 7 )
6.6 US GAL [25.0 L]
TEMPERATURE RANGE 70 °F [21.1 °C] TO 150 °F [65.5 °C]
RESISTIVITY
(CONDUCTIVITY)
0.1-0.5 M -cm (2-10 µS/cm) pH 6.8 - 7.5
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 7 of 10 10292220.D
UTILITY REQUIREMENTS CHART (PLUMBING)
COMPRESSED
AIR
(A)
CONNECTION SIZE / TYPE 1/8 in / FEMALE NPT [1/8 in / MALE BSPT]
1. AIR PURITY PER ISO-8573-1: 2010 [7:4:4]
MAX. PARTICLE DENSITY: 10 mg/m³
MAX. DEW POINT FOR WATER CONTENT:
37 °F [3 °C]
MAX. OIL CONCENTRATION FOR THE OIL
CONTENT: 5 mg/m³
2. STERIS RECOMMENDS A REFRIGERATED AIR
DRYER WHERE ENVIRONMENTAL DEW POINT
CONDITIONS ARE HIGHER THAN
RECOMMENDED.
3. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE MIN 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]
OPERATING FLOW RATE 1.75 SCFM [50 L/min]
DRAIN
(D)
FLOOR DRAIN SIZE MIN 4 in [102 mm] O.D. WITH 8 in x 8 in [204 mm x 204 mm] FLOOR SINK
4. WASHER DRAIN OUTLET SIZE 1-1/2 in / MALE NPT
5. MAX OFFSET OF 3/4 in [19 mm] FROM
OUTLET (PAGE 3, D).
6. FOR DRAIN REQUIREMENTS WHEN USING
SEISMIC MOUNTS, SEE PAGE 2.
OPERATING FLOW RATE GRAVITY DRAIN : 50 US GPM
[189.3 L/min]
CHEMICAL
DOSING
SYSTEM
CEILING ROUTING
CONDUIT 3 in CONDUIT
11. STERIS RECOMMENDS TO ROUTE CHEMICAL
LINES AND SIGNAL CABLE THROUGH A
3 in [76 mm] ID CONDUIT (BY OTHERS).
100 ft [30 m] OF INJECTION LINE AND
SIGNAL CABLE ARE PROVIDED WITH THE UNIT.
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
STEAM INLET
(S)
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
7. STERIS REQUIRES AT LEAST 97% SATURATED
DRY PLANT STEAM, FREE OF DIRT, RESIDUES
AND CONTAMINANTS.
8. STERIS RECOMMENDS TO INSTALL A STEAM
TRAP AS CLOSE AS POSSIBLE TO UNIT
CONNECTION.
9. PROPER OPERATION OF THE WASHER
REQUIRES A MINIMUM PRESSURE DIFFERENTIAL
OF 21.5 psig [148 kPa] BETWEEN STEAM
INLET AND CONDENSATE RETURN. IN CASES OF
LOW STEAM INLET PRESSURE, CONDENSATE
CAN BE ROUTED TO DRAIN TO INCREASE
DIFFERENTIAL.
*STERIS OFFERS CONDENSATE RETURN
TO DRAIN ACCESSORY (FD353)
10. CUSTOMER MUST PROVIDE UTILITY
CONNECTIONS WITH SHUTOFFS DISCONNECTS
WITHIN 2 FEET OF THE PERIMETER OF THE
EQUIPMENT AND BELOW THE CEILING DECK.
PRESSURE RANGE
30 - 80 psig DYNAMIC
[206-552 kPa] MAX. 90 psig STATIC [620 kPa]
OPERATING
MASS FLOW RATE
169 lb/h [76.5 kg/h] AT 30 psig [207 kPa]
271 lb/h [123 kg/h AT 80 psig [550 kPa]
TOTAL CONSUMPTION
PER INSTRUMENT CYCLE
(SEE PAGE 7, *(NOTE 4))
14.1 lb [6.4 kg]
CONDENSATE
RETURN
(CR)
PRESSURE RANGE MAX. 8.5 psig [68.9 kPa] BACK PRESSURE (17 ft [5.2 m] ELEVATION)
OPERATING FLOW RATE PEAK : 0.54 US GPM [2.05 L/min]
CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT [1/2 in / MALE BSPT]
Page 8 of 10 10292220.D
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
UTILITY REQUIREMENTS CHART (ELECTRIC)
UTILITY
PARAMETER AND REQUIREMENT NOTES
CONNECTION
SIZE/TYPE
NOMINAL
AMPERAGE
RECOMMENDED
CIRCUIT
ELECTRICITY
(EL)
200-208V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
36 A 50 A
1. ALL CONNECTIONS SHOULD BE IN
ACCORDANCE WITH LOCAL CODES.
2. NEUTRAL WIRE NOT REQUIRED.
3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES
AND SAFETY-RELATED WORK PRATICES TO
DETERMINE THE VALUE OF PROTECTION.
4. WIRING ON THE EQUIPMENT TERMINATES AT
A JUNCTION BOX. WIRING BETWEEN JUNCTION
BOX AND BUILDING SERVICE LINES ARE NOT
SUPPLIED BY STERIS.
5. DISCONNECT SWITCHES WITH OFF POSITION
LOCKOUT (SUPPLIED BY OTHERS) MUST BE
INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE
EQUIPMENT.
6. STERIS RECOMMENDS A LIGHT IN SERVICE
AREA.
200-208V 60 Hz, 3-Ph
WITH 40A
MIN. CIRCUIT
PROTECTION
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
28.5 A 40 A
200-208V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
35 A 50 A
460-480V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
16.5 A 30 A
380-400V 60 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
19 A 30 A
380-400-415V 50 Hz, 3-Ph
1 in [27 mm] 1-1/4 in [35 mm]
CONDUIT
18.5 A 30 A
NETWORK
(N)
OPTIONAL
100/1000 BASE T SUPPORTING
TCP/IP PROTOCOL
RJ45F CONNECTION
7. NETWORK CONNECTION (IF APPLICABLE)
- ACTIVE NETWORK DROP WITH CAT 5/6 CABLE
WITHIN 10 ft [3 m] OF UNIT
- STATIC OR DHCP RESERVED IP ADDRESS
(FOR EACH UNIT)
- NETWORK CABLE (SUPPLIED BY OTHER)
Page 9 of 10 10292220.D
UTILITY REQUIREMENTS CHART
VENTILATION
EXHAUST
(EX)
OPTIONAL
CONNECTION SIZE 3 in [76 mm]
OUTSIDE DIAMETER
1. NOT REQUIRED WITH NON VENTED UNIT.
2. STERIS REQUIRES CORROSION-RESISTANT AND
WATER-TIGHT DUCT:
- MIN OPERATING TEMPERATURE 210 °F [99 °C]
- SLOPED TOWARDS THE WASHER AND NO DEAD LEGS
- INSTALL DRAIN PART AT LOWEST POINT NEAR UNIT
3. EXHAUST FAN NORMALLY OPEN-AUXILIARY CONTACTS
IS AVAILABLE TO SUPPLY A SIGNAL TO THE FACILITY HVAC
CONTROL SYSTEM WHENEVER THE WASHER VENTILATION
EXHAUST FAN IS IN OPERATION.
PRESSURE RANGE 1/2 in [50 mm]
VACUUM AT CONNECTION
OPERATING FLOW RATE 60 SCFM [1.7 m³/min]
TEMPERATURE RANGE MAX. 240 °F [116 °C]
PHYSICAL SPECIFICATIONS
SHIPPING DIMENSION W x H x L 50 in x 91 in x 42 in [1270 mm x 2311 mm x 1067 mm]
1. LEVELING FEET ARE PROVIDED FOR PROPER INSTALLATION.
2. PERMISSIBLE ENVIRONMENTAL CONDITIONS.
THIS EQUIPMENT IS DESIGNED TO GIVE OPTIMAL RESULTS
UNDER THE FOLLOWING CONDITIONS:
- INDOOR USE ONLY;
- ALTITUDE OF OPERATION UP TO 6562 ft [2000 m];
- TEMPERATURE OF 41 TO 104 °F [5 to 40 °C];
- MAXIMUM RELATIVE HUMIDITY IS 80% FOR TEMPERATURE UP
TO 88 °F [31 °C], DECREASING LINEARLY TO 50% RELATIVE
HUMIDITY AT 104 °F [40 °C];
- POLLUTION DEGREE 2: EQUIPMENT MUST BE INSTALLED IN AN
ENVIRONMENT WHERE NORMALLY ONLY NON-CONDUCTIVE
POLLUTION OCCURS BUT WHERE OCCASIONAL, TEMPORARY
CONDUCTIVITY CAUSED BY CONDENSATION CAN BE EXPECTED
(AS DETERMINED ACCORDING TO INTERNATIONAL STANDARD
EN/IEC 61010-1).
3. STERIS ASSUMES NO RESPONSIBILITY FOR CHANGES MADE TO,
OR FAILURE TO OBSERVE THE SPECIFICATIONS ON EQUIPMENT
DRAWING AND NOTE PAGES. SPECIFICATIONS AND
DESCRIPTIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
SHIPPING WEIGHT 1268 lb [575 kg]
OPERATING DIMENSION W x H x L 42 in x 80.75 in x 32 in [1067 mm x 2051 mm x 813 mm]
OPERATING WEIGHT
(WITH WATER AND FULLY LOADED) 1254 lb [569 kg]
AVERAGE FLOOR LOAD
OVER WASHER FOOTPRINT 0.93 psi [6.43 kPa]
MAX. FLOOR LOAD
AT WASHER FOOT PADS
(3 in [76.2 mm] DIAMETER PADS) 67 psi [462 kPa]
NOISE LEVEL:
A-WEIGHTED EQUIVALENT
SURFACE SOUND PRESSURE LEVEL
60.6 dB(A)
CALCULATED AS DESCRIBED
IN ISO 3746.
OPENING AREA FROM
CLEAN SIDE TO SOILED SIDE
46.5 in² [300 cm²]
*(NOTE 4)
4. OPENING AREA CAN BE REDUCED TO 13 in² [84 cm²]
WITH INSTALLATION OF AIR MANAGEMENT PANELS
ACCESSORY (FD352)
CHAMBER DIMENSION W x H x L
VOLUME
26.5 in x 26.25 in x 25.5 in [673 mm x 667 mm x 648 mm]
10.3 ft³ [0.29 m³]
TYPICAL HEAT LOSS AT 75 °F [24 °C]
CONTINUOUS OPERATION
NON-VENTED UNIT
WITHOUT ACU-RINSE RES.
VENTED UNIT (OPTIONAL)
WITHOUT ACU-RINSE RES.
CLEAN SIDE 2049 Btu/h [600 W]
2049 Btu/h [600 W]
SOILED SIDE 6557 Btu/h [1920 W]
4747 Btu/h [1390 W]
TOTAL 8606 ± 854 Btu/h [2520 ± 250 W]
6796 ± 854 Btu/h [1990 ± 250 W]
VALUES INCLUDE HEAT LOSS DURING DOOR OPENING FOR LOADING AND UNLOADING OPERATIONS. "TOTAL" WILL VARY DEPENDING ON DISINFECTION TIME AND TEMPERATURE, DRYING TIME AND TEMPERATURE AND USE RATE OF UNIT.
AN ADDITIONAL HEAT LOSS FOR LOAD COOL DOWN IN CLEAN SIDE CAN BE ADDED TO THE TOTAL HEAT LOSS OF THE WASHER:
MAXIMUM 1713 Btu [1807 kJ]
T I T L E
EQUIPMENT DRAWING OF
AMSCO® 7052HP SINGLE-CHAMBER WASHER/DISINFECTOR
STEAM HEATED, W/O ACU-RINSE RES., UNIT FLUSH TO WALL
Corporation
SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER
& REVISIONA N/A Jean Tanguay 10292220
This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.
Page 10 of 10 10292220.D
| HL-TD-AMSCO 7052HP - TECH DATA SHEET (EN) (21) |
| AMSCO 7052HP SINGLE-CHAMBER WASHER-DISINFECTOR STEAM HEATED, W-O ACU-RINSE RES., UNIT FLUSH TO WALL (6) |
| Feuille 1 |
| new_view_35 |
| new_view_37 |
| new_view_57 |
| new_view_66 |
| A1 |
| Feuille 2 |
| new_view_55 |
| A1 |
| Feuille 3 |
| new_view_38 |
| new_view_39 |
| new_view_40 |
| A1 |
| Feuille 4 |
| new_view_44 |
| new_view_45 |
| new_view_46 |
| A1 |
| Feuille 5 |
| new_view_41 |
| new_view_42 |
| new_view_43 |
| new_view_65 |
| A1 |
| Feuille 6 |
| new_view_47 |
| A1 |
| Feuille 7 |
| A1 |
| Feuille 8 |
| A1 |
| Feuille 9 |
| new_view_64 |
| A1 |
| Feuille 10 |
| A1 |
| check_0: Yes |
| check_1: Off |
| check_2: Yes |
| check_3: Off |
| check_4: Off |
| check_5: Off |
| check_6: Off |
| check_7: Off |
| check_8: Yes |
| check_9: Off |
| check_10: Yes |
| check_11: Off |
| check_12: Yes |
| check_13: Off |
| check_14: Off |
| check_15: Off |
| check_16: Off |
| check_17: Off |
| check_18: Off |
| check_19: Off |
| check_20: Off |
| check_21: Off |
| check_22: Off |
| check_23: Off |
| check_24: Off |
| check_25: Off |
| check_26: Yes |
| check_27: Off |
| check_28: Off |
| check_29: Off |
| check_30: Off |
| check_31: Off |
| check_32: Off |
| check_33: Off |
| check_34: Off |
| check_35: Off |
| check_36: Off |
| check_37: Off |
| check_38: Off |
| check_39: Off |
File details come from the government source that posted it. Updated .