SOLICITATION QUESTIONS BREAKOUT.xlsx
XLSX spreadsheet 14 KB Posted
- Solicitation number
- 36C26320Q0386
About this file
This solicitation seeks proposals for fume hood testing, inspection, maintenance, and repair services for facilities in Nebraska and Western Iowa. The Department of Veterans Affairs intends to award a firm fixed price contract covering services for fume hoods and biological safety cabinets at locations in Omaha, Grand Island, and Lincoln, Nebraska from July 2020 through June 2025. The procurement is set aside for service-disabled veteran owned small businesses. Offerors must be verified in the VA Vendor Information Pages database and have an active SAM registration reflecting NAICS code 541380 for testing laboratories. The contract will be governed by Department of Labor service contract labor standards and wage determinations. Proposals are due by May 29, 2020 and will be evaluated using lowest price technically acceptable criteria.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Follow up questions and answers 5-20-2020.pdf | ||
| 36C26320Q0386 AMEND 0001.pdf | ||
| SOLICITATION QUESTIONS.pdf | ||
| 36C24720R0008_6.docx | DOCX document | |
| 36C24720R0008 0005.pdf | ||
| REVISED COST SCHEDULE A00001.pdf | ||
| 36C26320Q0386 0001_1.docx | DOCX document | |
| Non-Emergent Wheelchair Van Services PWS -- 5-20-2020.pdf | ||
| 36C26320Q0386-000.docx | DOCX document | |
| 36C26320Q0386-001.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet1
| Question #3 | ||||||||
| OMAHA | ||||||||
| Research Bld 15 | ||||||||
| Room # | Room Type | Special Instructions | Hepa Filters | HEPA in Return, Exhaust, Ceiling or AHU | Pressure Monitors | Make | Hood Type | Model |
| R-106 | Lab | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2 BSC | NU425-300 | |
| R-106 | Animal Research | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-106 | Animal Research | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2 BSC | NU425-400 | |
| R-114 | Animal Research | Inspected Once Per Year | 2 | Supply and Exhaust | Bellco | Class II A2 BSC | 8011-74000 | |
| R-114 | Animal Research | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-130 | Animal Research | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-130 | Research | Inspected Once Per Year | 1 | Exhaust | Labconco Corporation | Animal Isolator | .38400xx | |
| R-133 | Research | Inspected Once Per Year | 1 | Exhaust | Allentown Caging Equip. Co. | Animal Unit | .3840005-18 | |
| R-135 | Research | Inspected Once Per Year | 1 | Exhaust | Labconco Corporation | Animal Isolator | .38400xx | |
| R-141 | Research | Inspected Once Per Year | 1 | Exhaust | Labconco Corporation | Animal Isolator | .38400xx | |
| R-148 | Animal Research | Inspected Once Per Year | 2 | Supply and Exhaust | Lab Products Inc. | Horzontal Clean Beach | .30909B | |
| R-202 | Endocrine Research | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2 BSC | NU425-400 | |
| R-202 | Endocrine Research | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2 BSC | NU425-300 | |
| R-202 | Endocrine Research | Inspected Once Per Year | 2 | Supply and Exhaust | Forma Scientific | Class II A1 BSC | 1100 | |
| R-207 | Lab | Inspected Once Per Year | 2 | Supply and Exhaust | Baker Company | Class II A2 BSC | SG-600 | |
| R-208 | Endocrine Research | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-208 | Endocrine Research | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-211 | Nueroendocrine Lab | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-212 | Lab | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-212 | Research | Inspected Once Per Year | 2 | Supply and Exhaust | Baker Company | Class II A2 BSC | .SG403A-HE | |
| R-214A | Research | Inspected Once Per Year | 2 | Supply and Exhaust | Baker Company | Class II A2 BSC | .SG403A-HE | |
| R-214A | Research | Inspected Once Per Year | 1 | Supply | Baker Company | Horizontal Clean Bench | EG-4320 | |
| R-214C | Surgery Research | Inspected Once Per Year | 0 | N/A | Fisher Hamilton | Fume Hood - Standard | .54L-28x51 | |
| R-217 | Gastrointestinal Phys. | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .5' Fume Hood | |
| R-225 | Infectious Disease | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-227 | Molecular Biophysics | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-302 | Lab | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-310 | Molecular Biology Core | Inspected Once Per Year | 2 | Supply and Exhaust | Bellco | Class II A1 BSC | 8010-76000 | |
| R-311 | Molecular Biology | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .5' Fume Hood | |
| R-313 | Liver Study | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-314 | Lab | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-316 | Liver Study | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .5' Fume Hood | |
| R-317 | Liver Study | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-318 | Protein Chemistry | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-320 | Hybridoma Culture | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2 BSC | NU425-400 | |
| R-320 | Hybridoma Culture | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2-600 | NU425-600 | |
| R-321 | Expermental Imm. | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood | |
| R-325 | Expermental Imm. | Inspected Once Per Year | 0 | N/A | Labconco Corporation | Fume Hood - Standard | 70702-90-591190 | |
| R-326 | Lab | Inspected Once Per Year | 2 | Supply and Exhaust | Baker Company | Class II A2 BSC | B40-112-08 | |
| R-327 | Lab | Inspected Once Per Year | 0 | N/A | Brown/Morse | Fume Hood - Standard | .3' Fume Hood |
| Question #3 | ||||||||
| OMAHA | ||||||||
| Room # | Room Type | Special Instructions | Hepa Filters | HEPA in Return, Exhaust, Ceiling or AHU | Pressure Monitors | Make | Hood Type | Model |
| Main Hospital | ||||||||
| Basement | ||||||||
| B-506 | Pulmonary Research | Inspected Once Per Year | 0 | N/A | Labconco Corporation | Fume Hood - Standard | 22552-S-9013 | |
| B-508 | Tissue Culture | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2 BSC | NU425-400 | |
| B-514 | Pulmonary Research | Inspected Once Per Year | 0 | N/A | Labconco Corporation | Fume Hood - Standard | 22552-S-9013 | |
| B-545 | Pulmonary Research | Inspected Once Per Year | 2 | Supply and Exhaust | Labconco Corporation | Class II A2 BSC | 36209 | |
| B-702 | Pharmacy- Pos Buffer | Inspected Twice Per Year | 1 | Supply | Yes | NuAire | Horizontal Clean Bench | .NU301-436 |
| B-702 | Pharmacy-Pos Buffer | Inspected Twice Per Year | 1 | Supply | Yes | NuAire | Horizontal Clean Bench | .NU301-636 |
| B-702 | Pharmacy - Neg Buffer | Inspected Twice Per Year | 2 | Supply and Exhaust | Yes | NuAire | Class II B2 BSC | .NU-S430-600 |
| Second Floor | ||||||||
| 2303 | Histology | Inspected Once Per Year | 2 | Supply and Exhaust | Shandon Lipsaw Inc. | Goss Hood | GrossLab Senior | |
| 2277 | Urinalysis | Inspected Once Per Year | 2 | Supply and Exhaust | Labconco Corporation | Class II A2 BSC | 34308 Logic | |
| Sixth Floor | ||||||||
| 6603-1 | Mircrobiology | Inspected Once Per Year | 2 | Supply and Exhaust | NuAire | Class II A2 BSC | .NU424-400 | |
| 6603-1 | Mircrobiology | Inspected Once Per Year | 2 | Supply and Exhaust | Labconco Corporation | Class II A2 BSC | 36208 DELTA | |
| 6605A-1 | Mircrobiology BSL-TB | Inspected Once Per Year | 2 | Supply and Exhaust | Yes | Labconco Corporation | Class II A2 BSC | .3460801 6' Logic |
| 6605A-1 | Mircrobiology BSL Lab | Certification | 0 | Yes | Custom | Custom | Custom | |
| GRAND ISLAND | ||||||||
| Room # | Room Type | Special Instructions | Hepa Filters | HEPA in Return, Exhaust, Ceiling or AHU | Pressure Monitors | Make | Hood Type | Model |
| SE101 | Pharmacy-Pos Buffer | Inspected Twice Per Year | 2 | Supply and Exhaust | Yes | NuAire | Horizontal Clean Bench | NU301-636 |
| SE101 | Pharmacy- Neg Buffer | Inspected Twice Per Year | 2 | Supply and Exhaust | Yes | NuAire | ClassII B2 BSC | NU-S430-600 |
| LINCOLN | ||||||||
| Room # | Room Type | Special Instructions | Hepa Filters | HEPA in Return, Exhaust, Ceiling or AHU | Pressure Monitors | Make | Hood Type | Model |
| 79 | Pharmacy - Pos Buffer | Inspected Once Per Year | 1 | Supply | Yes | NuAire | Horizontal Clean Bench | NU301-436 |
| 79 | Pharmacy- Neg Buffer | Inspected Twice Per Year | 2 | Supply | Yes | NuAire | Class II B2 BSC | NU-S430-600 |
| Question #4 | I don’t know if any of our Fume Hoods have air flow monitoring. There is no report that we have to show this information. |
| Question #5 | Only in the Pharmacy and Microbilogy would have these monitors. We have never been asked this question and I can't find any report that would have this information on it. |
| Question #9 | Yes |