STX-MED-028.pdf
PDF 277 KB Posted
- Solicitation number
- 36C26224B0012
About this file
This document appears to be an engineering drawing for a standard medical bulk installation of a liquid main/cylinder reserve supply oxygen USP system. It provides a detailed process and instrumentation diagram, including nozzle legend and typical set points for pressure control valves, alarms/switches, and pressure safety devices. The drawing is confidential and proprietary to Airgas, an Air Liquide company.
The related federal contract opportunity is for a project to upgrade Building 80 at the VA New Mexico Healthcare System to meet the current needs for physical therapy and hospice patients. The work includes general construction, mechanical and electrical work, utility systems, demolition, and other tasks such as reconfiguring existing spaces, adding a large-scale permanent oxygen tank, modifying the oxygen storage room, and completing landscaping and irrigation. The solicitation is open to contractors, who must furnish all supervision, tools, labor, training, materials, equipment, and transportation necessary to complete the work in accordance with the terms, conditions, specifications, and drawings.
View the file
Other files
Show all 21
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
an Air Liquide company
SYSTEM SHALL BE INSTALLED
PER THIS ENGINEERED
DESIGN. ANY MODIFICATIONS
MUST BE APPROVED BY
AIRGAS BULK ENGINEERING.
AutoCAD SHX Text
O.PARKER
AutoCAD SHX Text 4/16/2021
AutoCAD SHX Text
J.KUKSA
AutoCAD SHX Text
C.LAFLIN
AutoCAD SHX Text
AutoCAD SHX Text
STX-MED-028
AutoCAD SHX Text
STANDARD MEDICAL BULK INSTALLATION
AutoCAD SHX Text
LIQUID MAIN/CYLINDER RESERVE SUPPLY
AutoCAD SHX Text
OXYGEN USP
AutoCAD SHX Text
OP
AutoCAD SHX Text
DRAWING CREATION
AutoCAD SHX Text 4/16/21
AutoCAD SHX Text
JK
AutoCAD SHX Text
CL
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
REV
AutoCAD SHX Text
DATE
AutoCAD SHX Text
DWN
AutoCAD SHX Text
CHK
AutoCAD SHX Text
APVD
AutoCAD SHX Text
DATE
AutoCAD SHX Text
JOB NO.
AutoCAD SHX Text
COMPUTER FILE
AutoCAD SHX Text
DRAWN BY
AutoCAD SHX Text
CHECKED BY
AutoCAD SHX Text
APPRD. / D.A.
AutoCAD SHX Text
PAGE
AutoCAD SHX Text
REV NO.
AutoCAD SHX Text
DWG NO.
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
CONFIDENTIAL NOTICE
AutoCAD SHX Text THIS DOCUMENT AND THE DESIGNS, SPECIFICATIONS AND ENGINEERING INFORMATION DISCLOSED HEREON ARE THE PROPERTY OF AIRGAS, AN AIR LIQUIDE COMPANY, AND ARE NOT TO BE DISSEMINATED OR REPRODUCED WITHOUT EXPRESS WRITTEN CONSENT OF AIRGAS, AN AIR LIQUIDE COMPANY. THIS PROPRIETARY DOCUMENT IS BEING SENT IN CONFIDENCE AND ONLY FOR CONSIDERATION OF THE MATTERS HEREON. IN ACCEPTING THE LOAN OF THIS DOCUMENT, RECIPIENT AGREES TO KEEP IT AND THE MATTERS DISCLOSED HEREON I CONFIDENCE AND SHALL NOT USE THEM NOR PERMIT THEIR USE IN ANY WAY DETRIMENTAL TO AIRGAS, AN AIR LIQUIDE COMPANY.
AutoCAD SHX Text
PROCESS AND INSTRUMENT DIAGRAM
AutoCAD SHX Text
REV
AutoCAD SHX Text
DATE
AutoCAD SHX Text
DWN
AutoCAD SHX Text
CHK
AutoCAD SHX Text
APVD
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text
STANDARD
AutoCAD SHX Text
INSTALLATION INFORMATION
AutoCAD SHX Text
CUSTOMER NAME
AutoCAD SHX Text
CUSTOMER CITY / STATE
AutoCAD SHX Text
TANK SIZE(S) / PRODUCT
AutoCAD SHX Text
INSTALL DATE
AutoCAD SHX Text
LEAD TECHNICIAN
AutoCAD SHX Text
TYPICAL SET POINTS FOR MEDICAL OXYGEN APPLICATIONS
AutoCAD SHX Text
PRESSURE CONTROL VALVES
AutoCAD SHX Text
ALARMS/SWITCHES
AutoCAD SHX Text
PRESSURE SAFETY DEVICES
AutoCAD SHX Text
PCV1 (PB) 130 PSIG
AutoCAD SHX Text
LIS1 1+ DAY SUPPLY
AutoCAD SHX Text
PSV1/2 250 PSIG
AutoCAD SHX Text
PCV2 (ECON) 150 PSIG
AutoCAD SHX Text
PS102 1500 PSIG
AutoCAD SHX Text
PSE1/2 350-375 PSIG
AutoCAD SHX Text
PCV3 200 PSIG
AutoCAD SHX Text
PS103 105 PSIG
AutoCAD SHX Text
PSV3A/B/C/D 350 PSIG
AutoCAD SHX Text
PCV101 95 PSIG
AutoCAD SHX Text
PSV3A1 (FILL) 400 PSIG
AutoCAD SHX Text
PCV102A 55 PSIG RESPIRATORY 65 PSIG HYPERBARICS
AutoCAD SHX Text
PSV101A 475-500 PSIG
AutoCAD SHX Text
PCV102B 50 PSIG RESPIRATORY 60 PSIG HYPERBARICS
AutoCAD SHX Text
PSV111/112 300 PSIG
AutoCAD SHX Text
PSV113A/B 75 PSIG RESPIRATORY 100 PSIG HYPERBARICS
AutoCAD SHX Text
LIS
AutoCAD SHX Text
AutoCAD SHX Text
PI
AutoCAD SHX Text
AutoCAD SHX Text D
AutoCAD SHX Text A
AutoCAD SHX Text L
AutoCAD SHX Text G
AutoCAD SHX Text C
AutoCAD SHX Text E
AutoCAD SHX Text V1
AutoCAD SHX Text B
AutoCAD SHX Text F
AutoCAD SHX Text N
AutoCAD SHX Text
PSE3
AutoCAD SHX Text E1
AutoCAD SHX Text
PCV2
AutoCAD SHX Text C2
AutoCAD SHX Text C1
AutoCAD SHX Text
VENT
AutoCAD SHX Text
TO ATMOSPHERE
AutoCAD SHX Text
LOCKED OPEN
AutoCAD SHX Text G100
AutoCAD SHX Text
REMOVE CV
AutoCAD SHX Text
INTERNALS
AutoCAD SHX Text
DUAL SAFETIES AND RUPTURE DISCS.
AutoCAD SHX Text
%%UTYPICAL VESSELS AS SUPPLIED BY MANUFACTURER
AutoCAD SHX Text
MANUFACTURERS, HOWEVER, AS A
AutoCAD SHX Text
MINIMUM, V1 SHALL HAVE
AutoCAD SHX Text
PIPING ARRANGEMENT DUE TO THE CHOICE OF
AutoCAD SHX Text
THERE MAY BE SOME DIFFERENCES IN THE VESSELS
AutoCAD SHX Text
VAPOR PHASE CONTENTS GAUGE
AutoCAD SHX Text
LIQUID PHASE CONTENTS GAUGE
AutoCAD SHX Text
VAPOR RETURN (CAPPED)
AutoCAD SHX Text
LIQUID WITHDRAWAL
AutoCAD SHX Text
LIQUID USE (CAPPED)
AutoCAD SHX Text
FULL TRYCOCK
AutoCAD SHX Text
TOP FILL
AutoCAD SHX Text N
AutoCAD SHX Text L
AutoCAD SHX Text G
AutoCAD SHX Text F
AutoCAD SHX Text E
AutoCAD SHX Text D
AutoCAD SHX Text C
AutoCAD SHX Text
%%UVESSEL NOZZLE LEGEND
AutoCAD SHX Text
BOTTOM FILL
AutoCAD SHX Text
VENT
AutoCAD SHX Text A
AutoCAD SHX Text B
AutoCAD SHX Text
AutoCAD SHX Text
FC1A
AutoCAD SHX Text
FILL ASSEMBLY
AutoCAD SHX Text
CV1A
AutoCAD SHX Text
PSV3A1
AutoCAD SHX Text
STR2
AutoCAD SHX Text
HV17A
AutoCAD SHX Text
ANTENNA
AutoCAD SHX Text
ALARM SIGNAL:
AutoCAD SHX Text
LOW LEVEL MAIN VESSEL
AutoCAD SHX Text
SOLAR PANEL
AutoCAD SHX Text
AutoCAD SHX Text
DPT
AutoCAD SHX Text
AutoCAD SHX Text
PT
AutoCAD SHX Text
BAM BUDDY OR UNIT
AutoCAD SHX Text
CV2
AutoCAD SHX Text
HV2
AutoCAD SHX Text
HV1
AutoCAD SHX Text
PSV3B
AutoCAD SHX Text
PSV3C
AutoCAD SHX Text
PSV3D
AutoCAD SHX Text
PSV3A
AutoCAD SHX Text
HV11
AutoCAD SHX Text
HV10
AutoCAD SHX Text
HV5
AutoCAD SHX Text
HV17
AutoCAD SHX Text
HV7
AutoCAD SHX Text
HV8
AutoCAD SHX Text
HV9
AutoCAD SHX Text
HV15
AutoCAD SHX Text
HV13
AutoCAD SHX Text
HV14
AutoCAD SHX Text
HV20
AutoCAD SHX Text
PCV1
AutoCAD SHX Text
PCV3
AutoCAD SHX Text
PSE2
AutoCAD SHX Text
PSE1
AutoCAD SHX Text
PSV1
AutoCAD SHX Text
HV12
AutoCAD SHX Text
PSV2
AutoCAD SHX Text
HV18
AutoCAD SHX Text
HV19
AutoCAD SHX Text
HV4
AutoCAD SHX Text
STR1
AutoCAD SHX Text
HV3
AutoCAD SHX Text
HV16
AutoCAD SHX Text
HV6
AutoCAD SHX Text
AMBIENT VAPORIZER
AutoCAD SHX Text
AutoCAD SHX Text
AutoCAD SHX Text E100
AutoCAD SHX Text
MEDICAL PRESSURE CONTROL MANIFOLD
AutoCAD SHX Text S101
AutoCAD SHX Text S102
AutoCAD SHX Text
AutoCAD SHX Text
PS
AutoCAD SHX Text
AutoCAD SHX Text
PI
AutoCAD SHX Text
ALARM SIGNAL:
AutoCAD SHX Text
RESERVE IN USE
AutoCAD SHX Text
FOR AUX MANIFOLD
AutoCAD SHX Text 102A
AutoCAD SHX Text
PI
AutoCAD SHX Text 102B
AutoCAD SHX Text
PI
AutoCAD SHX Text
AutoCAD SHX Text
PI
AutoCAD SHX Text
CUSTOMER
AutoCAD SHX Text
SOURCE
AutoCAD SHX Text
VALVE
AutoCAD SHX Text
AUXILIARY
AutoCAD SHX Text
SOURCE
AutoCAD SHX Text
CONNECTION
AutoCAD SHX Text
HOSPITAL OWNED
AutoCAD SHX Text
HI/LOW LINE
AutoCAD SHX Text
PRESSURE ALARM
AutoCAD SHX Text
SYSTEM SUPPLIED
AutoCAD SHX Text
AND INSTALLED
AutoCAD SHX Text
BY AIRGAS
AutoCAD SHX Text
PIPING FROM THIS POINT
AutoCAD SHX Text
SUPPLIED AND INSTALLED
AutoCAD SHX Text
BY CUSTOMER
AutoCAD SHX Text
PS
AutoCAD SHX Text
AutoCAD SHX Text
HV109
AutoCAD SHX Text
HV121
AutoCAD SHX Text
HV120
AutoCAD SHX Text
HV119B
AutoCAD SHX Text
HV119A
AutoCAD SHX Text
PSV113B
AutoCAD SHX Text
PSV113A
AutoCAD SHX Text
PCV102B
AutoCAD SHX Text
PCV102A
AutoCAD SHX Text
HV118B
AutoCAD SHX Text
HV118A
AutoCAD SHX Text
HV115
AutoCAD SHX Text
HV116
AutoCAD SHX Text
HV117
AutoCAD SHX Text
HV113
AutoCAD SHX Text
HV112
AutoCAD SHX Text
HV111
AutoCAD SHX Text
PSV111
AutoCAD SHX Text
CV111
AutoCAD SHX Text
HV114
AutoCAD SHX Text
CV112
AutoCAD SHX Text
PSV112
AutoCAD SHX Text
AutoCAD SHX Text
PS
AutoCAD SHX Text
PI
AutoCAD SHX Text
AutoCAD SHX Text
CV50F
AutoCAD SHX Text
CV50E
AutoCAD SHX Text
CV50D
AutoCAD SHX Text
CV50C
AutoCAD SHX Text
CV50B
AutoCAD SHX Text
CV50A
AutoCAD SHX Text 101B
AutoCAD SHX Text
PI
AutoCAD SHX Text 101A
AutoCAD SHX Text
PI
AutoCAD SHX Text V50B
AutoCAD SHX Text V50A
AutoCAD SHX Text V50C
AutoCAD SHX Text V50E
AutoCAD SHX Text V50D
AutoCAD SHX Text V50F
AutoCAD SHX Text
CYLINDER RESERVE
AutoCAD SHX Text
NUMBER OF CYLINDERS DEPENDS ON SITE
AutoCAD SHX Text
ALARM SIGNAL:
AutoCAD SHX Text
RESERVE LOW PRESSURE
AutoCAD SHX Text 1/2"x1"
AutoCAD SHX Text
AutoCAD SHX Text
PSV101A
AutoCAD SHX Text
HV53
AutoCAD SHX Text
PCV101
AutoCAD SHX Text
HV52
AutoCAD SHX Text
HV51
AutoCAD SHX Text
%%UNOTES
AutoCAD SHX Text
1. SEE SYSTEM INFORMATION SHEET FOR: SEE SYSTEM INFORMATION SHEET FOR: CUSTOMER INFORMATION DESIGN CONDITIONS VESSEL INFORMATION BILL OF MATERIALS 2. SEE INSTALLATION DATA SHEET FOR: SEE INSTALLATION DATA SHEET FOR: LINE SPECIFICATION AND LEAK TEST TABLE GENERAL CONSTRUCTION NOTES 3. SEE INSTALLATION DATA SHEET FOR LINE SIZE SEE INSTALLATION DATA SHEET FOR LINE SIZE AND MATERIAL. 4. VESSEL MAY HAVE SINGLE OR DUAL FILL VESSEL MAY HAVE SINGLE OR DUAL FILL VALVE CONFIGURATION (HV3/HV4). TO BE CONFIRMED AT INSTALL AND REDLINED ON AS-BUILT DRAWING AS NEEDED. 5. EXTENDED FILL MANIFOLD MUST BE EXTENDED FILL MANIFOLD MUST BE COMPLIANT WITH DOT AND INTERNAL REQUIREMENTS.
| Sheets and Views |
| Model |