Addendum 001 Reference Only Drawings.pdf
PDF 8 MB Posted
- Solicitation number
- 36C26224B0012
About this file
This document appears to be an Addendum containing reference drawing details for a federal contract opportunity related to a Medical Oxygen Skid. The drawings provide product specifications, part numbers, and dimensional details for different size Medical Oxygen Skids. Key information includes:
The drawings include 8 different models of Medical Oxygen Skids, ranging from 4 to 16 high pressure oxygen bottles. The part numbers, tank pressures, and flow rates are provided for each model. Overall dimensions, weight, and material details are also specified. The drawings include schematic diagrams showing the component layout and piping configuration for the skids. Additional requirements are noted, such as the use of oxygen-cleaned parts, brazing by certified employees, and optional alarm panels.
View the file
Other files
Show all 21
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
FRONT VIEW
TOP VIEW
VIEW A-A
DETAIL A
A A
DETAIL B
SEE SHEET 2
A
D-20810866
APPVREV ECR REVISION DESCRIPTION BY DATE
# DESIGNATES A NOTE REFERENCE.
MAT'L
FINISH P/N
SEE TABLE
APPV'D DATE
JNB 2/2/2015
D-20810866
SEE B.O.M.
F
1/8"
N/A N/A
TABLE 1 - MEDICAL SKID PART NUMBERS AND SET POINTS
PART NUMBER DESCRIPTION RESERVE IN USE PS2 RESERVE LOW PS1 PRIMARY REG-R2 RESERVE REG -R1 TANK PSI
20810864 MEDICAL MB 4 BOTTLE O2 SKID 80 PSIG (PI-4) 1000 PSIG (PI-2) 50 - 55 PSIG (PI-3) 65 PSIG (PI-1) 125 20810865 MEDICAL MB 6 BOTTLE O2 SKID 80 PSIG (PI-4) 1000 PSIG (PI-2) 50 - 55 PSIG (PI-3) 65 PSIG (PI-1) 125 20810866 MEDICAL MB 8 BOTTLE O2 SKID 80 PSIG (PI-4) 1000 PSIG (PI-2) 50 - 55 PSIG (PI-3) 65 PSIG (PI-1) 125 20891908 MEDICAL MB 12 BOTTLE O2 SKID 80 PSIG (PI-4) 1000 PSIG (PI-2) 50 - 55 PSIG (PI-3) 65 PSIG (PI-1) 125 20891909 MEDICAL MB 16 BOTTLE O2 SKID 80 PSIG (PI-4) 1000 PSIG (PI-2) 50 - 55 PSIG (PI-3) 65 PSIG (PI-1) 125 20810867 MEDICAL MB 4 BOTTLE N2 SKID 180 PSIG (PI-4) 1000 PSIG (PI-2) 170 - 175 PSIG (PI-3) 185 PSIG (PI-1) 220 20810868 MEDICAL MB 6 BOTTLE N2 SKID 180 PSIG (PI-4) 1000 PSIG (PI-2) 170 - 175 PSIG (PI-3) 185 PSIG (PI-1) 220 20810869 MEDICAL MB 8 BOTTLE N2 SKID 180 PSIG (PI-4) 1000 PSIG (PI-2) 170 - 175 PSIG (PI-3) 185 PSIG (PI-1) 220 20894778 MEDICAL MB 12 BOTTLE N2 SKID 180 PSIG (PI-4) 1000 PSIG (PI-2) 170 - 175 PSIG (PI-3) 185 PSIG (PI-1) 220 20894774 MEDICAL MB 16 BOTTLE N2 SKID 180 PSIG (PI-4) 1000 PSIG (PI-2) 170 - 175 PSIG (PI-3) 185 PSIG (PI-1) 220 20840682 SKID MMB 8HPBU HYPERBARIC O2 80 PSIG (PI-4) 1000 PSIG (PI-2) 65 PSIG (PI-3) 75 PSIG (PI-1) 125 20987482 SKID MMB 4HPBU HYPERBARIC 02 80PSIG (PI-4) 1000 PSIG (PI-2) 65 PSIG (PI-3) 75 PSIG (PI-1) 125
5/8 3132 1/4
93 5/8
NOTE:
SKID IS DESIGNED FOR CHART
PERMA-CYL SIZES 1000-3000 LITER
SEE CHART SPEC DRAWINGS FOR DETAILS
NOTES:
1. WEIGHT OF SKID; 1,100 LBS
2. ALL PIPING, VALVES, FITTINGS TO BE CLEANED FOR OXYGEN SERVICE
3. ALL BRAZING TO BE COMPLETED ONLY WITH 56% SILVER SOLDER
4. ALL BRAZING TO BE COMPLETED ONLY BAG-7 CERTIFIED EMPLOYEES
5. PIPING MEETS REQUIREMENT OF NFPA55 AND NFPA99
6. NOT DESIGNED TO PICK UP OR MOVE WITH PRODUCT IN TANK
7. OPTIONAL LOCAL ALARM PANEL AVAILABLE (P/N 11896754)
8. ALL DIMENSIONS ARE ±1/2"
9. CARBON STEEL FRAME W/ POWDERCOATED
10. NON-SLIP STRIPS ON FLOOR
11. OVERSIZED METAL GRATING REQUIRED FOR SKIDS WITH 12 OR 16 HIGH PRESSURE BOTTLES
12. OVERSIZED AWNING REQUIRED FOR SKIDS WITH 12 OR 16 HIGH PRESSURE BOTTLES.
100 1/2
81 1/284
102 1/2
A CGT 209298 JNB 10/21/15
B CGT 209381 JNB 3/7/16
C CGT 209489 CGT 9/8/16
D CGT 209588 CKL 1/4/18
6 11
FOR OVERSIZED SKID ONLY
E CGT 210054 RLC 12/18/18
F CGT 210084 RLC 03/12/19
J.Iwanczyk Proprietary
SECTION B-B
SCALE .125
SECTION C-C
SCALE .125
A A
B
B
C
C
DRAWING
NO
D-20852344
CHART, INC.
BALL GROUND, GA 30107
APPVREV ECR REVISION DESCRIPTION BY DATE
THE MATERIALS AND INFORMATION, INCLUDING THE PRINCIPLES
OF DESIGN PRESENTED BY THIS PRINT, IS THE EXCLUSIVE
PROPERTY OF CHART INC., AND IS CONFIDENTIAL
INFORMATION. ACCORDINGLY, THIS INFORMATION IS SUBMITTED
TO YOU WITH THE AGREEMENT THAT IT IS NOT TO BE REPRODUCED,
COPIED, OR LOANED, IN PART OR IN WHOLE, NOR IS THE
INFORMATION TO BE RELAYED TO ANY COMPANY. ACCEPTANCE OF
THIS DRAWING WILL BE CONSTRUED AS AN AGREEMENT TO THE ABOVE.
DESIGNATES A NOTE REFERENCE.
MAT'L
FINISH P/N
20852344
APPV'D DATE
APPLICATION ORIGINAL ECR/ER
DRAWN
BY
CHECKED
BY
PROJ.
ENGR.
JNB 7/31/2015
CGT 10/6/2015
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN INCHES.
TOLERANCES:
FRACTIONS ±
ANGLES ± °
2 PLACE DECIMAL ±
3 PLACE DECIMAL ±
USED ON
SPEC DWG PERMA-CYL
3000HP PSTR FLEX FILL
TITLE:
DRAWING
NO.
SCALE:
DO NOT
SCALE
DRAWING
SHEET:
REV:
D-20852344
N/A 1 of 1
208999
SEE B.O.M.
C
1/2
N/A
N/A
GASEOUS EQUIVALENT: SCF AT 1 ATM & 70º F / Nm³ AT 1 ATM & 0ºC
GAS DELIVERY RATE **
PALLET BASE GALVANIZED
TYPE 304 STAINLESS STEEL
3,800 lbs / 1,724 kg
82,239 SCF / 2,161 Nm³
2,707 LITERS / 715 GALLONS
2,911 LITERS / 769 GALLONS
ASME SEC. VIII. DIV 1
1 % PER DAY NITROGEN
.62 % PER DAY OXYGEN
.62 % PER DAY ARGON
3000 HP FLEX FILL
1,350 SCF/H / 35.4 Nm³/h
250 PSI / 17.2 BAR
CODE
NET VOLUME
ARGON
NITROGEN
OXYGEN
NITROGEN
MATERIAL
NORMAL EVAPORATION RATE
TARE WEIGHT WITH LAB BASE
GROSS VOLUME
RELIEF SETTING MAWP
ITEM
DESIGN SPECIFICATION
80,425 SCF / 2,115 Nm³
66,592 SCF / 1,750 Nm³
OXYGEN
ARGON
PARTS LIST
1,350 SCF/H / 35.4 Nm³/h
1,350 SCF/H / 35.4 Nm³/h
DESCRIPTION
REGULATOR 1/4 FPT @ 140PSI
RV BRS 1/4MPT 550PSI
CYL-TEL GEN 5 0-200"H2O
VALVE CHECK BRS 1/2FPTX1/2FPT
RPD ASSY INLINE 1/2MPT 375PSI
MANUFACTURER P/N
011-0258
PRV9432T550
209105981
GENERANTCV-503B-T-5
12790-0140
REGO
76-603-57
BS&B
REGO
REGO
REGO
WORCESTER
CMM250A
T9454
REGO
APOLLO
REGO
REGO
PRV19434TP250 REGO
PRV9432T550 REGO
Chart
REGOT9454
T9454
VALVE SHUTOFF BRZ 1/2FPT SHORT
VALVE BALL BRS 1/2FPT
VALVE NEEDLE BRS 1/4MPT(ANGLE)
VALVE BALL DIV SS 1/2NPT
VALVE BRS SH 3/8FPTX3/8 SCH 10
VALVE SHUTOFF BRZ 1/2FPT SHORT
1 1712202
1812702
*R-1
RV-1
11051090
11501177
10907239
11905956
11773885
V-4
V-8
10502848
1V-9
V-11
V-1
V-13
V-12
V-10
V-7
RV-5
RV-4
V-2
10502848
*LI-1
PI-1
*R-2
*BD-1
MANUFACTURER
CV-1
11671281
PART NO.ITEM
QTY
RV-3
*RV-2 2
VALVE SHUTOFF 1/4FPT
RV BRS 1/4MPT 550PSI
RV BRS 1/4MPT 550PSI
RV BRS 1/4MPT 550PSI
11905999
10502848 VALVE SHUTOFF BRZ 1/2FPT SHORT
REGOT9452
REGOT9464DA
CN4416PMSE
PRV9432T550
PRV9432T550
CASH-ACME
20683719 VALVE 4-WAY BRS 1/8FPT B-43YTF2-WL8 SWAGELOK
WIKA52721099
DIFF PG 4.5" 0-100" H2O VITON (O2, N2)
DIFF PG 4.5" 0-150" H2O VITON (AR)
WIKA
20923162
20909588 52721125
1812702
1812702
1812702
11488574 RV BRS 1/2MPT 250PSI
20885664
REGULATOR 1/4 FPT @ 325PSI 12790-0325
CASH-ACME
20599868 RV BRS 1/2MPT 350PSI PRV19434TP250 REGO
11526569 RPD ASSY INLINE 1/2MPT 525PSI 011-0258 BS&B
20827654 PG 2-1/2" 0-600PSI/BAR 1/8MPT 52553914
WIKA
350 PSI / 24.1 BAR
500 PSI / 34.4 BAR
10502848 VALVE SHUTOFF BRZ 1/2FPT SHORT T9454
VM-1
REGO
MVE54
20820974 TOP & BTM FILL MANIFOLD
SCHEMATIC
TANK PRESSURE BUILD INLET ISO VALVE
TANK PRESSURE BUILD OUTLET ISO VALVE
TANK GAUGE ISOLATION VALVE
TANK PB INLET
TANK BOTTOM FILL & LIQUID USE
TANK TOP FILL
V-2
RV-5
RV-4
RV-3
RV-2
LI-1
RV-1
R-2
V-1
R-1
PV-1
PBV-1
CV-1
BD-1
TANK FILL ASSEMBLY RELIEF VALVE
TANK PRIMARY RELIEF VALVE
TANK PB INLET RELIEF VALVE
V-11
V-9
R-2
PI-1
LI-1
RV-3
A
B
C
D
E
F
G
H
J
K
L
RV-2
L
PBV-1
PV-1
TANK PB OUTLET
V-10
TANK FULL TRYCOCK / VENT VALVE
TANK GAS USE OUT
TANK PB OUTLET RELIEF VALVE
TANK ECON INLET RELIEF VALVE
TANK ECONOMIZER
TANK GAS USE IN
TANK ECONOMIZER INLET ISO VALVE
TANK GAS USE VALVE
PIPING
V-7
RV-2
V-7
RV-4
V-12
F
RV-5
G
D
J
H
V-8
BD-1
TANK DUAL RELIEF
TANK LIQUID PHASE
TANK VAPOR PHASE
K
V-4
V-1
BD-1
CV-1
TANK PB REGULATOR
TANK ECONOMIZER REGULATOR
TANK PROCESS VAPORIZER COIL
PI-1
TANK LIQUID LEVEL GAUGE
TANK FILL CHECK VALVE
TANK SECONDARY RELIEF DEVICE BURST DISC
TANK PRESSURE GAUGE 0-600 PSI
TANK PRESSURE BUILDING VAPORIZER COIL
NOMENCLATURE
TANK LIQUID WITHDRAWAL VALVE
TANK SAFETY LINE PURGE VALVE
TANK FULL TRYCOCK / VENT VALVE
TANK SAFETY 3-WAY DIVERTER VALVE
TANK ECONOMIZER VALVE
V-13
V-11
V-12
V-10
V-9
V-8
V-7
V-4
A
CB E
V-13
V-2
R-1
BTM FILL
TOP FILL
VENT/DRAIN
RV-1
VM-1
VM-1
TOP & BOTTOM FILL MANIFOLD W/DRAIN
E
C
O
N
S
A
F
E
T
Y
V
E
N
T
B
O
T
T
O
M
I L
L
T
O
P
I L
L
G
A
S
U
S
E
P
B
TOP VIEW
FRONT VIEW
75 1/2
(2 PLCS)
60 1/2
(2 PLCS)
56 1/2
(2 PLCS)
44 5/16
51 1/2
(2 PLCS)
120 5/8
RV-4
V-13
RV-3
V-7
BD-1
RV-2
V-8
RV-2
BD-1
V-7
R-2
V-10
R-1
RV-5
V-2
V-9
V-4
CV-1
VM-1
RV-1
V-1
PI-1
LI-1
V-12
V-11
NOTE:
* - PART NUMBERS ARE DEPENDENT ON THE PLUMBING OPTION KIT CHOSEN FOR THE VESSEL.
** - OPTIONAL 3,500 SCF/H (92 Mn
h) FLOW KIT AVAILABLE
DATA PLATE
5 3/8
A CGT 209500
UPDATE PARTS LIST, SCHEM, DIMS
RLB 07/11/17
B CGT 210007 CHANGED R-1 REGULATORS LAD 12/14/18
21267966
21267965 REGULATOR 1/2 NPT SET @ 300PSI
REGULATOR 1/2 NPT SET @ 125PSI PB504-125
PB504-300 REGO
REGO
C JE 211090
ADDED TABLE, CHANGED V-4 DATA
ZMB 4/8/2021
Pad Dimensions and Reinforcing:
A
Fence
(Area Must be Secure)
6"
5’-0"5’-0"
A Typical
Concrete:
Reinforcing:
Soil Bearing Capacity:
4000 PSI @ 28 Days.
ASTM A615 Grade 60
2000 PSF (minimum)
Pad Specifications
Tank Pad:
Note 1
5’-0" Opening Recommended.
Tank Pad Shall be Level and
3" Above Grade
450L 5’-8"
6’-9"
7’-7"
Model Height
Weight
Empty
Weight
Full N Tank Diameter
Width
4’-0"
2’-6"
3’-6"
Net Cap.
Liters
1000L
1500L
3200 4500
4800 6700
APPROXIMATE TANK DIMENSIONS & CAPACITIES
FOR PERMA-CYL MICROBULK SYSTEMS
Weights do not include lab base option (base option:265 lbs.)
TANK PAD
10’-0" x 10’-0" x 8" Thick
House Line
Location of Gate and Orientation of Equipment to Suit Site
MB1
#5 at 12" On Center Each Way
Depending upon exposure, this installation meets the standards set in National Fire
Protection Association booklet #55 and O.S.H.A. regulation 1910.104.
Pad Design Parameters: (Based Upon The International Building Code - 2012)
Soil Type D, Ultimate Wind Speed 180 mph, Risk Category 2, Exposure D
Seismic: 0.2 Sec. Spectral Response Acceleration of 306% gravity
MBP1- Pad Installation Procedure
MBP2- Standard Bollard Detail
MBE1- Tank Anchoring Procedure
3" Bottom Clearance
Weight
Full Ar
Weight
Full CO
(IBC OCCUPANCY CATEGORY 1 OR 2)
National Standard
For Industrial Use
Rolling Or Sliding Gate May Be Preferred.
Location Determined By Customer Based on Site Conditions to Protect Installation.
Bollards - See Drawing MBP2
10’-0"
Work This Drawing With 2015 Versions of Airgas National Microbulk Standard Drawings:
4/9/15
Tank Microbulk
Note 1
CONCRETE PAD INSTALLATION DETAILS
FOR PERMA-CYL MICROBULK SYSTEMS WITH
NITROGEN, ARGON, and CO
MODELS 450L, 1000L and 1500L
Tank Pad For Microbulk Systems
Date
CONFIDENTIAL NOTICE
Issued By
This document and the designs, specifications and engineering information disclosed hereon are the proprietary property of the Airgas company providing it and are not to be disseminated or reproduced without the prior express written consent of such Airgas company. This proprietary document is being sent in confidence and only for consideration of the matter hereon.
In accepting the loan of this document, recipient agrees to keep it and the matter disclosed hereon in confidence and shall not use or permit their use in any way detrimental to any Airgas companies.
The Airgas companies accept no responsibility for the use of this document by any recipient, this document is only for illustrative purposes. Date
C 2015 Airgas
S
A
F H
O
P F
SIO
S
E68652
T
T E O
O
I
R O
ES
NAL ENG
IN
E E
RR
E
G I T E
R
E D
DAVID
K.
CREMERS
Finish Grade
SECTION A-A
4" Compacted Crushed Stone
Tank Pad
6’-0" Chain Link Fence w/ 3 Strands of Barbed Wire
#5 at 12" On Center Each Way
Microbulk Tank Must Be Set On
Pad Before Fence Is Installed.
Firm Subgrade
For Pad Installation Procedure
See Drawing MBP1
3" Bottom Clearance
MB2
Depending upon exposure, this installation meets the standards set in National Fire
Protection Association booklet #55 and O.S.H.A. regulation 1910.104.
Pad Design Parameters: (Based Upon The International Building Code - 2012)
Soil Type D, Ultimate Wind Speed 180 mph, Risk Category 2, Exposure D
Seismic: 0.2 Sec. Spectral Response Acceleration of 306% gravity
National Standard
For Industrial Use 4/9/15
Location of Gate and Orientation of Equipment to Suit Site
(IBC OCCUPANCY CATEGORY 1 OR 2)
450L 5’-8"
6’-9"
7’-7"
Model Height
Weight
Empty
Tank Diameter
Width
4’-0"
2’-6"
3’-6"
Net Cap.
Liters
1000L
1500L
APPROXIMATE TANK DIMENSIONS & CAPACITIES
FOR PERMA-CYL MICROBULK SYSTEMS
Weights do not include lab base option (base option:265 lbs.)
Pad Dimensions and Reinforcing:
Concrete:
Reinforcing:
Soil Bearing Capacity:
4000 PSI @ 28 Days.
ASTM A615 Grade 60
2000 PSF (minimum)
Pad Specifications
Tank Pad: 10’-0" x 10’-0" x 8" Thick
#5 at 12" On Center Each Way
MBP1- Pad Installation Procedure
MBP2- Standard Bollard Detail
MBE1- Tank Anchoring Procedure
3" Bottom Clearance
Work This Drawing With 2015 Versions of Airgas National Microbulk Standard Drawings:
Weight
Full O
Transport Pad: 12’-0" x 12’-0" x 6" thick
#4 at 12" On Center Each Way
Centered In Slab
A
6"
5’-0"5’-0"
A Typical
Note 1
Tank Pad Shall be Level and
3" Above Grade
TANK PAD
House Line
Bollards
10’-0"
6’-0"6’-0"
12’-0"
Shall Be Clean and Level At Driveway Grade
Transport Pad (for oxygen service only)
Fence
(Area Must be Secure)
5’-0" Opening Recommended.
Rolling Or Sliding Gate May Be Preferred.
Location Determined By Customer Based on Site Conditions to Protect Installation.
- See Drawing MBP2
Tank Microbulk
Note 1
Microbulk Tank
CONCRETE PAD INSTALLATION DETAILS
FOR PERMA-CYL MICROBULK SYSTEMS WITH OXYGEN
MODELS 450L, 1000L and 1500L
For Joint Specifications
See Standard Drawing MBP1
Tank Pad For Microbulk Oxygen Systems
Date
CONFIDENTIAL NOTICE
Issued By
This document and the designs, specifications and engineering information disclosed hereon are the proprietary property of the Airgas company providing it and are not to be disseminated or reproduced without the prior express written consent of such Airgas company. This proprietary document is being sent in confidence and only for consideration of the matter hereon.
In accepting the loan of this document, recipient agrees to keep it and the matter disclosed hereon in confidence and shall not use or permit their use in any way detrimental to any Airgas companies.
The Airgas companies accept no responsibility for the use of this document by any recipient, this document is only for illustrative purposes. Date
C 2015 Airgas
S
A
F H
O
P F
SIO
S
E68652
T
T E O
O
I
R O
ES
NAL ENG
IN
E E
RR
E
G I T E
R
E D
DAVID
K.
CREMERS
Finish GradeTransport Pad
Tank Pad
6’-0" Chain Link Fence w/ 3 Strands of Barbed Wire
Microbulk Tank Must Be Set On
Pad Before Fence Is Installed.
#4 at 12" On Center Each Way Centered In Slab
For Pad Installation Procedure
See Drawing MBP1
SECTION A-A
4" Compacted Crushed Stone
Tank Pad
#5 at 12" On Center Each Way
Firm Subgrade
For Pad Installation Procedure
See Drawing MBP1
3" Bottom Clearance robert.scheehl Highlight
MB3
Protection Association booklet #55 and O.S.H.A. regulation 1910.104.
Pad Design Parameters: (Based Upon The International Building Code - 2012)
Soil Type D, Ultimate Wind Speed 180 mph, Risk Category 2, Exposure D
Seismic: 0.2 Sec. Spectral Response Acceleration of 306% gravity
National Standard
For Industrial Use 4/9/15
Pad Dimensions and Reinforcing:
A
Fence
(Area Must be Secure)
6"
6’-0"6’-0"
A Typical
Tank Pad:
Note 1
Note 1
5’-0" Opening Recommended.
Tank Pad Shall be Level and
3" Above Grade
Model Height
Weight
Empty
Weight
Full N Tank Diameter
Width
Net Cap.
Liters
APPROXIMATE TANK DIMENSIONS & CAPACITIES
FOR PERMA-CYL MICROBULK SYSTEMS
Weights do not include lab base option (base option:265 lbs.)*
TANK PAD
House Line
Location of Gate and Orientation of Equipment to Suit Site
#5 at 12" On Center Each Way
3" Bottom Clearance
Weight
Full Ar
Weight
Full CO
(IBC OCCUPANCY CATEGORY 1 OR 2)
Rolling Or Sliding Gate May Be Preferred.
12’-0"
2000L
3000L 10’-2"
9’-9" 4’-0"
4’-11" 2707
1945 2600
* 11,700
Location Determined By Customer Based on Site Conditions to Protect Installation.
Bollards - See Drawing MBP2
Microbulk Tank
12’-0" x 12’-0" x 10" Thick
CONCRETE PAD INSTALLATION DETAILS
FOR PERMA-CYL MICROBULK SYSTEMS WITH
NITROGEN, ARGON, and CO
MODELS 2000L and 3000L
Tank Pad For Microbulk Systems
Concrete:
Reinforcing:
Soil Bearing Capacity:
4000 PSI @ 28 Days.
ASTM A615 Grade 60
2000 PSF (minimum)
Pad Specifications
MBP1- Pad Installation Procedure
MBP2- Standard Bollard Detail
MBE1- Tank Anchoring Procedure
Work This Drawing With 2015 Versions of Airgas National Microbulk Standard Drawings:
Date
CONFIDENTIAL NOTICE
Issued By
This document and the designs, specifications and engineering information disclosed hereon are the proprietary property of the Airgas company providing it and are not to be disseminated or reproduced without the prior express written consent of such Airgas company. This proprietary document is being sent in confidence and only for consideration of the matter hereon.
In accepting the loan of this document, recipient agrees to keep it and the matter disclosed hereon in confidence and shall not use or permit their use in any way detrimental to any Airgas companies.
The Airgas companies accept no responsibility for the use of this document by any recipient, this document is only for illustrative purposes. Date
C 2015 Airgas
S
A
F H
O
P F
SIO
S
E68652
T
T E O
O
I
R O
ES
NAL ENG
IN
E E
RR
E
G I T E
R
E D
DAVID
K.
CREMERS
Finish Grade
SECTION A-A
4" Compacted Crushed Stone
Tank Pad
6’-0" Chain Link Fence w/ 3 Strands of Barbed Wire
#5 at 12" On Center Each Way
Microbulk Tank Must Be Set On
Pad Before Fence Is Installed.
Firm Subgrade
For Pad Installation Procedure
See Drawing MBP1
3" Bottom Clearance
MB4
Protection Association booklet #55 and O.S.H.A. regulation 1910.104.
Pad Design Parameters: (Based Upon The International Building Code - 2012)
Soil Type D, Ultimate Wind Speed 180 mph, Risk Category 2, Exposure D
Seismic: 0.2 Sec. Spectral Response Acceleration of 306% gravity
National Standard
For Industrial Use 4/9/15
Location of Gate and Orientation of Equipment to Suit Site
(IBC OCCUPANCY CATEGORY 1 OR 2)
Pad Dimensions and Reinforcing:
Concrete:
Reinforcing:
Soil Bearing Capacity:
4000 PSI @ 28 Days.
ASTM A615 Grade 60
2000 PSF (minimum)
Pad Specifications
Tank Pad:
#5 at 12" On Center Each Way
MBP1- Pad Installation Procedure
MBP2- Standard Bollard Detail
MBE1- Tank Anchoring Procedure
3" Bottom Clearance
Work This Drawing With 2015 Versions of Airgas National Microbulk Standard Drawings:
Transport Pad: 12’-0" x 12’-0" x 6" thick
#4 at 12" On Center Each Way
Centered In Slab
A
TANK PAD
6’-0"6’-0"
12’-0"
Shall Be Clean and Level At Driveway Grade
Transport Pad (for oxygen service only)
Model Height
Weight
Empty
Tank Diameter
Width
Net Cap.
Liters
APPROXIMATE TANK DIMENSIONS & CAPACITIES
FOR PERMA-CYL MICROBULK SYSTEMS
Weights do not include lab base option (base option:265 lbs.)*
2000L
3000L 10’-2"
9’-9" 4’-0"
4’-11" 2707
1945 2600
CONCRETE PAD INSTALLATION DETAILS
FOR PERMA-CYL MICROBULK SYSTEMS WITH OXYGEN
MODELS 2000L and 3000L
Weight
Full 0
10,200
6"
6’-0"6’-0"
A Typical
Note 1
Tank Pad Shall be Level and
3" Above Grade
House Line
12’-0"
Microbulk Tank
Fence
(Area Must be Secure)
Note 1
Bollards
5’-0" Opening Recommended.
Rolling Or Sliding Gate May Be Preferred.
Location Determined By Customer Based on Site Conditions to Protect Installation.
- See Drawing MBP2 12’-0" x 12’-0" x 10" Thick
For Joint Specifications
See Standard Drawing MBP1
Tank Pad For Microbulk Oxygen SystemsTank Pad For Microbulk Oxygen Systems
Date
CONFIDENTIAL NOTICE
Issued By
This document and the designs, specifications and engineering information disclosed hereon are the proprietary property of the Airgas company providing it and are not to be disseminated or reproduced without the prior express written consent of such Airgas company. This proprietary document is being sent in confidence and only for consideration of the matter hereon.
In accepting the loan of this document, recipient agrees to keep it and the matter disclosed hereon in confidence and shall not use or permit their use in any way detrimental to any Airgas companies.
The Airgas companies accept no responsibility for the use of this document by any recipient, this document is only for illustrative purposes. Date
C 2015 Airgas
S
A
F H
O
P F
SIO
S
E68652
T
T E O
O
I
R O
ES
NAL ENG
IN
E E
RR
E
G I T E
R
E D
DAVID
K.
CREMERS
Finish GradeTransport Pad
Tank Pad
6’-0" Chain Link Fence w/ 3 Strands of Barbed Wire
Microbulk Tank Must Be Set On
Pad Before Fence Is Installed.
#4 at 12" On Center Each Way Centered In Slab
For Pad Installation Procedure
See Drawing MBP1
SECTION A-A
4" Compacted Crushed Stone
Tank Pad
#5 at 12" On Center Each Way
Firm Subgrade
For Pad Installation Procedure
See Drawing MBP1
3" Bottom Clearance robert.scheehl Highlight
Protection Association booklet #55 and O.S.H.A. regulation 1910.104.
Pad Design Parameters: (Based Upon The International Building Code - 2012)
Soil Type D, Ultimate Wind Speed 200 mph, Risk Category 4, Exposure D
Seismic: 0.2 Sec. Spectral Response Acceleration of 230% gravity
4/9/15 MBHC1
Tank Pad For Microbulk Oxygen Systems
National Standard
For Healthcare Facility Use
450L 5’-8"
6’-9"
7’-7"
Model Height
Weight
Empty
Tank Diameter
Width
4’-0"
2’-6"
3’-6"
Net Cap.
Liters
1000L
1500L
APPROXIMATE TANK DIMENSIONS & CAPACITIES
FOR PERMA-CYL MICROBULK SYSTEMS
Weights do not include lab base option (base option:265 lbs.)
Weight
Full O
(IBC OCCUPANCY CATEGORY 4)
CONCRETE PAD INSTALLATION DETAILS FOR
PERMA-CYL MICROBULK HEALTHCARE SYSTEMS WITH OXYGEN
MODELS 450L, 1000L and 1500L
Location of Gate and Orientation of Equipment to Suit Site
Pad Dimensions and Reinforcing:
Concrete:
Reinforcing:
Soil Bearing Capacity:
4000 PSI @ 28 Days.
ASTM A615 Grade 60
2000 PSF (minimum)
Pad Specifications
Tank Pad: 12’-0" x 12’-0" x 8" Thick
#5 at 12" On Center Each Way
MBP1- Pad Installation Procedure
MBP2- Standard Bollard Detail
MBE1- Tank Anchoring Procedure
3" Bottom Clearance
Work This Drawing With 2015 Versions of Airgas National Microbulk Standard Drawings:
Transport Pad: 12’-0" x 12’-0" x 6" thick
#4 at 12" On Center Each Way
Centered In Slab
A
TANK PAD
6’-0"6’-0"
12’-0"
Shall Be Clean and Level At Driveway Grade
Transport Pad (for oxygen service only)
6"
6’-0"6’-0"
A
Typical
Note 1
Tank Pad Shall be Level and
3" Above Grade
12’-0"
Note 1
Bollards
Fence
(Area Must be Secure)
5’-0" Opening Recommended.
Rolling Or Sliding Gate May Be Preferred.
Location Determined By Customer Based on Site Conditions to Protect Installation.
- See Drawing MBP2
For Joint Specifications
See Standard Drawing MBP1
Tank Microbulk
Alarm Conduit
Electrical Conduit
House Line
Regulation
Stand
6 Cylinder
Reserve Manifold
Date
CONFIDENTIAL NOTICE
Issued By
This document and the designs, specifications and engineering information disclosed hereon are the proprietary property of the Airgas company providing it and are not to be disseminated or reproduced without the prior express written consent of such Airgas company. This proprietary document is being sent in confidence and only for consideration of the matter hereon.
In accepting the loan of this document, recipient agrees to keep it and the matter disclosed hereon in confidence and shall not use or permit their use in any way detrimental to any Airgas companies.
The Airgas companies accept no responsibility for the use of this document by any recipient, this document is only for illustrative purposes. Date
C 2015 Airgas
S
A
F H
O
P F
SIO
S
E68652
T
T E O
O
I
R O
ES
NAL ENG
IN
E E
RR
E
G I T E
R
E D
DAVID
K.
CREMERS
Finish GradeTransport Pad
Tank Pad
6’-0" Chain Link Fence w/ 3 Strands of Barbed Wire
Microbulk Tank Must Be Set On
Pad Before Fence Is Installed.
#4 at 12" On Center Each Way Centered In Slab
For Pad Installation Procedure
See Drawing MBP1
SECTION A-A
4" Compacted Crushed Stone
Tank Pad
#5 at 12" On Center Each Way
Firm Subgrade
For Pad Installation Procedure
See Drawing MBP1
3" Bottom Clearance
MBHC24/9/15
Protection Association booklet #55 and O.S.H.A. regulation 1910.104.
Pad Design Parameters: (Based Upon The International Building Code - 2012)
Soil Type D, Ultimate Wind Speed 200 mph, Risk Category 4, Exposure D
Seismic: 0.2 Sec. Spectral Response Acceleration of 230% gravity
Tank Pad For Microbulk Oxygen Systems
National Standard
For Healthcare Facility Use
(IBC OCCUPANCY CATEGORY 4)
Model Height
Weight
Empty
Tank Diameter
Width
Net Cap.
Liters
APPROXIMATE TANK DIMENSIONS & CAPACITIES
FOR PERMA-CYL MICROBULK SYSTEMS
Weights do not include lab base option (base option:265 lbs.)*
2000L
3000L 10’-2"
9’-9" 4’-0"
4’-11" 2707
1945 2600
Weight
Full 0
10,200
CONCRETE PAD INSTALLATION DETAILS FOR
PERMA-CYL MICROBULK HEALTHCARE SYSTEMS WITH OXYGEN
MODELS 2000L and 3000L
Location of Gate and Orientation of Equipment to Suit Site
Pad Dimensions and Reinforcing:
Concrete:
Reinforcing:
Soil Bearing Capacity:
4000 PSI @ 28 Days.
ASTM A615 Grade 60
2000 PSF (minimum)
Pad Specifications
Tank Pad:
#5 at 12" On Center Each Way
MBP1- Pad Installation Procedure
MBP2- Standard Bollard Detail
MBE1- Tank Anchoring Procedure
3" Bottom Clearance
Work This Drawing With 2015 Versions of Airgas National Microbulk Standard Drawings:
Transport Pad: 12’-0" x 12’-0" x 6" thick
#4 at 12" On Center Each Way
Centered In Slab
15’-0" x 12’-0" x 10" Thick
A
TANK PAD
12’-0"
Shall Be Clean and Level At Driveway Grade
Transport Pad (for oxygen service only)
6’-0"6’-0"
A
Note 1
12’-0"
Microbulk Tank
Note 1
Bollards
Fence
(Area Must be Secure)
5’-0" Opening Recommended.
Rolling Or Sliding Gate May Be Preferred.
Location Determined By Customer Based on Site Conditions to Protect Installation.
- See Drawing MBP2
For Joint Specifications
See Standard Drawing MBP1
6"
Typical
Tank Pad Shall be Level and
3" Above Grade
6 Cylinder
Reserve ManifoldRegulation
Stand
Alarm Conduit
Electrical Conduit
House Line
Date
CONFIDENTIAL NOTICE
Issued By
This document and the designs, specifications and engineering information disclosed hereon are the proprietary property of the Airgas company providing it and are not to be disseminated or reproduced without the prior express written consent of such Airgas company. This proprietary document is being sent in confidence and only for consideration of the matter hereon.
In accepting the loan of this document, recipient agrees to keep it and the matter disclosed hereon in confidence and shall not use or permit their use in any way detrimental to any Airgas companies.
The Airgas companies accept no responsibility for the use of this document by any recipient, this document is only for illustrative purposes. Date
C 2015 Airgas
S
A
F H
O
P F
SIO
S
E68652
T
T E O
O
I
R O
ES
NAL ENG
IN
E E
RR
E
G I T E
R
E D
DAVID
K.
CREMERS
Finish GradeTransport Pad
Tank Pad
6’-0" Chain Link Fence w/ 3 Strands of Barbed Wire
Microbulk Tank Must Be Set On
Pad Before Fence Is Installed.
#4 at 12" On Center Each Way Centered In Slab
For Pad Installation Procedure
See Drawing MBP1
SECTION A-A
4" Compacted Crushed Stone
Tank Pad
#5 at 12" On Center Each Way
Firm Subgrade
For Pad Installation Procedure
See Drawing MBP1
3" Bottom Clearance
REV DATE DWN CHK APVD DATE
JOB NO.
COMPUTER FILE
DRAWN BY
CHECKED BY
APPRD. / D.A.
DATE REV NO.
DWG NO.
an Air Liquide company
CONFIDENTIAL NOTICE
THIS DOCUMENT AND THE DESIGNS, SPECIFICATIONS AND ENGINEERING INFORMATION DISCLOSED
HEREON ARE THE PROPERTY OF AIRGAS, AN AIR LIQUIDE COMPANY, AND ARE NOT TO BE
DISSEMINATED OR REPRODUCED WITHOUT EXPRESS WRITTEN CONSENT OF AIRGAS, AN AIR LIQUIDE
COMPANY. THIS PROPRIETARY DOCUMENT IS BEING SENT IN CONFIDENCE AND ONLY FOR
CONSIDERATION OF THE MATTERS HEREON. IN ACCEPTING THE LOAN OF THIS DOCUMENT, RECIPIENT
AGREES TO KEEP IT AND THE MATTERS DISCLOSED HEREON I CONFIDENCE AND SHALL NOT USE
THEM NOR PERMIT THEIR USE IN ANY WAY DETRIMENTAL TO AIRGAS, AN AIR LIQUIDE COMPANY.
REV DATE DWN CHK APVD
S.M. HAW ASSOC.
T. MIRABELLA
C. LAFLIN
12-04-2020
Diameter
Slope Top For Drainage Concrete Fill
Grade
Slope For Drainage
3,000 PSI
Concrete Foundation
Schedule 40 Standard Wt. Pipe 4" Dia. x 5'-6" Lg.
12" Diameter 16" Lap 2-#5 Reinforcing Rings
Existing Paving or
Transport Pad
Crushed Stone 4" Compacted
Firm Subgrade
Grade FinishSee Typical Joint Detail
Typical Section Through Pads
Reinforcing
Clearance
Tank Pad
1'-4"
FOR CONSTRUCTION
C 2020 Airgas, an Air Liquide Company
Standard Bollard, Fence, and Joint Details
STF-DET-01
Standard Bollard Detail
Apply Per Manufacturers Recommendations.
Color to be Safety Yellow (206327).
VOC Alkyd Enamel
Finish Coat: RUST-OLEUM Industrial Choice 340
VOC Alkyd Enamel Primer.
Primer: RUST-OLEUM Industrial Choice 340 to Remove Rust and Scale
Clean Surface Per SSPC- SP3
Bollard Pipe Paint Specification
(Prevent Gates From Opening Fully).
Interfere with Any Access Gates
Location of Bollards Should Not
Maximum Distance Between Bollards is 4'-0" C/C.
On All Sides Exposed To Vehicle Traffic
Upon Site Conditions To Protect Installation
By Customer And Airgas Based
Bollard Locations To Be Verified
Bollard Design Per International Fire Code Section 312
1 "2
2"
(Level)
"3
"3 4x
Typical All Around or Tooled Edge Chamfer
Polyfoam Board"1
1 "2
Other Sealants Not Permitted Chem Seal CS 2727 Sealant
(818) 890-1401 www.flamemaster.com By The Flamemaster Corporation
Board is Prohibited Use of Asphalt Impregnated
For Oxygen Service
Argon Service For Nitrogen or
"Ceramar" www.wrmeadows.com
3" Clear
Bottom Reinforcing Typical
Typical
Top Reinforcing
1 "21 Clear Driveway Grade
Level and At Off-Load Pad
See Typical Joint Detail
High Performance Polyurethane Sealant
For Oxygen Service
For Nitrogen or Argon Service
Chem Seal CS 2727 Sealant
Sikaflex 1a or Other Approved Polyurethane Sealant High Performance
C/J
Typical Control Joint
Shown Thus On Plan
Typical Control Joint
1 "21 Deep Saw Cut Fill With Sealant
High Performance Polyurethane Sealant
For Oxygen Service
For Nitrogen or Argon Service
Chem Seal CS 2727 Sealant
For Crack Control
Reinforcing
3" Break
Fill With Sealant and Backer Rod
3" Deep Saw Cut
Pad Isolation Construction Joint
For Equipment Isolation
For Equipment Isolation
Construction Joint Continuous Pad
12-04-20
12-04-20
1/2 "
3.
2.
1.
T
DC
Typical Fence Detail
N.T.S.
Airgas Standard Reference Drawings:
STE-GEN-001 "Standard Grounding Guidelines for Bulk Systems"
3.
2.
1.
Bollard Locations
Typ. Sliding Gate - Plan
N.T.S.
SLIDING FILL GATE
36"
36"
36"
ENCASEMENT
12" DIA. CONC.
PIPE" SCH40 GALV.2
END OR CORNER POST,
OR SURFACE MATERIAL
TOP OF FINISH GRADE
" O.D.2
BRACE RAIL
END
RAIL
GALVANIZED BARB WIRE
(3) STRANDS - 4 POINT
6'
6'
11'
(TYPICAL)
O.D.2
BRACE RAIL
36"
36"
36" 36"
10'10'
7 GAUGE
WIRES
TENSION
TIES
WIRE
GALVANIZED BARB WIRE
(3) STRANDS - 4 POINT
PIPE" SCH40 GALV.2
END OR CORNER POST,
POST
LINE OR SURFACE MATERIAL
TOP OF FINISH GRADE
Typ. Sliding Gate - Elevation
N.T.S.
42"
8' MAX
32"
21"
STEEL
2"X2" STRUCTURAL
21"
4" TOEBOARD
HANDRAIL
See OSHA Standard 1910.23
Typ. Railing Detail
N.T.S.
ROD
TRUSS
" MIN.8
SCH40 GALV. PIPE
"2
GATE FRAME 1
8" DIA. CONC.
ENCASEMENT
ENCASEMENT
12" DIA. CONC.
TRUSS ROD
" MIN.8
END
RAIL
DIAMOND MESH
WOVEN WITH A 2"X2"
#9 GAUGE GALV. WIRE
CHAIN LINK FABRIC
SCH40 GALV. PIPE
LINE POST 2"
TM CL
Airgas Standard Reference Drawings:
FOR CONSTRUCTION
STF-DET-01 - Pad, Bollard, and Fence Installation Details STF-DET-02 - Pad and Anchoring Installation Notes
A
STF-DET-02
See Drawing Procedure For Pad Installation
Tank MicroBulk
See Table - This Drawing Specifications For Tank Anchoring
12.
11.
10.
9.
8.
7.
6.
5.
4.
3.
2.
1.
12'-0"
X
Ground
(Min. Height 8'-0") Lighting (1 Place)
Stand Regulation
See Drawing STF-DET-01 Area Must Be Secure w/ 3 Strands Of Barbed Wire.
6'-0" Chain Link Fence
Y
PLAN - TANK PAD
For Nitrogen, Oxygen, Argon or CO2 MicroBulk Tanks
ELEVATION
Tank Pad
3" Above Grade
Shall be Level and
12-04-20
12-04-20 0
Industrial LIN - LOX - LAR-CO2
MicroBulk VJ Vessels
C 2020 Airgas, an Air Liquide Company
Standard Foundation Reference tation of equipment.
Off-Load Pad, and orien-on location of gates, Airgas shall be consulted
GENERAL NOTES
STF-MB-003
A
By a Registered Professional Engineer.
Table. Larger Pads Must Have The Reinforcing Reviewed Direction Without Changing The Reinforcing Shown In the Concrete Pad Dimensions Can Be Increased By 2'-0" In Each
Association Booklet #55 And O.S.H.A. Regulation 1910.104.
Separation Requirements Defined In National Fire Protection Depending Upon Exposure, This Installation Shall Meet The
Design Seismic: Per Table.
Design Wind Velocity: Per Table.
Design For Risk Category II.
IBC 2018/ASCE7-16 and IBC 2015/ASCE7-10.
Design Loads are Based Upon the More Stringent of
Design Soil Capacity 2000 psf Minimum.
ASTM A615 Grade 60.
All Reinforcing Bars Shall Conform To
4000 Psi @ 28 Days.
Concrete Compressive Strength Shall Be
With The Latest Editions Of ACI-301, ACI-302 And ACI-318.
All Concrete Construction Shall Be In Accordance
Filled With Suitable Material & Compacted.
Any Soft Spots Shall Be Excavated, New Concrete, The Subgrade Shall Be Compacted.
Prior To Installing Granular Base & Pouring
Upon Between Site Owner and Aigas.
Exact Site Location And Orientation Shall Be Agreed
Verify All Dimensions And Conditions In Field.
Reference Building Line
STM-MB-001 "MicroBulk Vessel Decal Placement" STE-GEN-001 "Standard Grounding Guidelines for Bulk Systems" STA-GEN-006 "Courtyard Guidelines" STA-CO2-001 "Distance Exposure - Bulk Carbon Dioxide STA-GEN-004 "Distance Exposure - Liquid Oxygen" STA-GEN-003 "Distance Exposure - Cryogenic Systems" Turning Radius" STA-GEN-007 "MicroBulk Systems Delivery Guidelines and
Off-Load Pad
At Driveway Grade
Shall be Level and
Service Only)
(For Oxygen
Detail and Application STF-DET-01 for See Airgas Drawing Bollards
TM
See Drawing STF-DET-01 Area Must Be Secure w/ 3 Strands Of Barbed Wire.
6'-0" Chain Link Fence
Typ.
Double Gate 5'-0" Wide
Shown For Reference 3000L Tank an Air Liquide company
M a x
1'-6"
Typ.
N .T
.S
Centered In Slab
Bottom Reinforcing See Table
Firm Subgrade
3" Bottom Clearance
Each Way #4 at 12" O.C.
And Reinforcing See Table For Pad Size
Tank Pad
Crushed Stone 4" Compacted
SECTION A-A
Off-Load Pad
12'-0" x 12'-0" x 6" Thick
Not Shown on This Section See Drawing STF-DET-01 for Information
T
6"
DC CL
REV DATE DWN CHK APVD DATE
JOB NO.
COMPUTER FILE
DRAWN BY
CHECKED BY
APPRD. / D.A.
DATE REV NO.
DWG NO.
an Air Liquide company
CONFIDENTIAL NOTICE
THIS DOCUMENT AND THE DESIGNS, SPECIFICATIONS AND ENGINEERING INFORMATION DISCLOSED
HEREON ARE THE PROPERTY OF AIRGAS, AN AIR LIQUIDE COMPANY, AND ARE NOT TO BE
DISSEMINATED OR REPRODUCED WITHOUT EXPRESS WRITTEN CONSENT OF AIRGAS, AN AIR LIQUIDE
COMPANY. THIS PROPRIETARY DOCUMENT IS BEING SENT IN CONFIDENCE AND ONLY FOR
CONSIDERATION OF THE MATTERS HEREON. IN ACCEPTING THE LOAN OF THIS DOCUMENT, RECIPIENT
AGREES TO KEEP IT AND THE MATTERS DISCLOSED HEREON I CONFIDENCE AND SHALL NOT USE
THEM NOR PERMIT THEIR USE IN ANY WAY DETRIMENTAL TO AIRGAS, AN AIR LIQUIDE COMPANY.
REV DATE DWN CHK APVD
S.M. HAW ASSOC.
T. MIRABELLA
C. LAFLIN
12-04-2020
DETAIL B
DETAIL C
C
APPVREV ECR REVISION DESCRIPTION BY DATE
# DESIGNATES A NOTE REFERENCE.
MAT'L
FINISH P/N
SEE TABLE
APPV'D DATE
JNB 2/2/2015
APOLLO 1/2" 3-WAY DIVERTER VALVE
RESERVE IN USE PRESS SWITCH ALARM
RESERVE LOW PRESS SWITCH ALARM
PT-1
PS-2
OXYGEN PIGTAILS
WESTERN B-52 CGA 540 1/2"
UNITED BRASS 1/2" SWING CHECK VALVE
REGO 1/2 NPT 350 PSIG
REGO 1/2 NPT 75 PSIG ASME
VICTOR 1/4 NPT
REGO 1/2" 40-110 PSI MEDICAL REGULATOR
VICTOR HIGH PRESSURE REGULATOR SR450D
VICTOR 1/4 NPT
HIGH PRESSURE SUPPLY REGULATOR
PT4540 OXYGEN PIGTAILS - SUPERIOR
UNITED ELECTRIC 80-1700 PSI PRESS SWITCH
UNITED ELECTRIC 15-300 PSI PRESS SWITCH
1/4 NPT (4"DIAL) - WIKA
1/4 NPT (4"DIAL) - WIKA
PRESSURE GAUGE 0-4000 PSI
PRESSURE GAUGE 0-400 PSI
PRESSURE GAUGE 0-200 PSI
TITAN 1/2" W/100 MESH SS
APOLLO 1/2" 3-WAY DIVERTER VALVE
APOLLO 1/2" BALL VALVE W/UNION END
REGO 1/4 NPT 350 PSIGTHERMAL RELIEF VALVE FOR R-2
PRESSURE GAUGE 0-200 PSI
NOMENCLATURE
SS-XSS4-1 EXCESS FLOW VALVE - SWAGELOK
REGO 1/2" 9560CA MASTER VALVE
APOLLO 1/2" BALL VALVE W/UNION END HIGH PRESS SUPPLY REG ISO VALVE
HIGH PRESS SUPPLY ISO VALVE
MANUFACTURER
MAIN GAS OUTLET SAFETY 3-WAY VALVE
0-200 PSI PRESSURE GAUGE ISO VALVE APOLLO 1/4" BALL VALVE W/ 1/8 PURGE PORT
MAIN GAS SUPPLY REGULATOR
VALVE OUTLET ADAPTER W/ CHECK VALVE
EXCESS FLOW VALVE FOR PS-2
MAIN GAS INLET SAFETY 3-WAY VALVE
0-400 PSI PRESSURE GAUGE ISO VALVEV-8
V-7
TANK INLET CHECK VALVECV-2
V-9
CV-1
SS-XSS4-1 EXCESS FLOW VALVE - SWAGELOK
APOLLO 1/4" BALL VALVE W/ 1/8 PURGE PORT
MAIN GAS LINE INLET STRAINER
MAIN GAS OUTLET RELIEF VALVE ASME
R-1 S-1
RV-3 A&B
RV-2 A&B
MAIN GAS INLET RELIEF VALVE RV-1 A&B
DESCRIPTION
V-2
ITEM
V-1
R-2 A&B
PI-2
PI-1
PS-1
PI-4
PI-3
V-4 V-3
GAS SUPPLY REGULATOR ISO VALVE
V-6
V-5 A-D
EXCESS FLOW VALVE FOR PS-1
V-10 BRASS BLEED VALVE SUPERIOR 3000PSI 1/4"x1/4" BRASS VALVE
NOMENCLATURE (PARTS TO SUBSTITUTE FOR NITROGEN SKID)
DESCRIPTIONITEM MANUFACTURER
CV-1 VALVE OUTLET ADAPTER W/ CHECK VALVE WESTERN B-52 CGA 540 1/2"
RV-3 A&B MAIN GAS OUTLET RELIEF VALVE ASME REGO 1/2 NPT 275 PSIG ASME
R-2 A&B MAIN GAS SUPPLY REGULATOR REGO 1/2" 100-200 PSIG MEDICAL
PI-1 PRESSURE GAUGE 0-400 PSIG VICTOR 1/4 NPT
PI-3 PRESSURE GAUGE 0-400 PSIG VICTOR 1/4 NPT (4"DIAL) - WIKA
PT-1 NITROGEN PIGTAILS PT4580 NITROGEN PIGTAILS - SUPERIOR
V-11 PS-2 ISOLATION VALVE W/PORT EXHAUST SWAGELOK SS-42GVS4
CV-3 HIGH PRESS INLET CHECK VALVE APOLLO 1/2" 400WOG SPRING LOADED
PI-5 PRESSURE GAUGE 0-400 PSI 1/4 NPT (2"DIAL) - WIKA
PARTS TO SUBSTITUE FOR HYPERBARIC O2
ITEM DESCRIPTION MANUFACTURER
RV-3B BACK PRESSURE REGULATOR REGO ECL-140 SET 75 PSIG
RV-3A MAIN GAS OUTLET RELIEF VALVE ASME REGO 1/2" NPT 90PSIG
M MUFFLER IN BP REGULATOR PARKER LEGRIS COMECTIC P/N 0670-00-13
RV-4 HP REGULATOR RELIEF VALVE REGO NPT 1/4" 350PSIG
R-1CV-3
V-2
RV-1B
PI-1 PI-2
R-2B
V-5DV-5C
S-1
RV-2B
V-4
RV-1A
R-2A
V-3
RV-3B
V-5BV-5A
RV-2AV-8 RV-3AV-7
PS-2
PI-4 PI-3
MEDICAL MANIFOLD PIPING
PI-5
CV-1 CV-1CV-1 CV-1 CV-1
HIGH PRESSURE RESERVE
CV-2
CV-1
PIG-TAILS WESTERN P-63 CGA540 (PT-1)
FROM TANK GAS USE VALVE
V-10
CV-1CV-1
V-9
V-6
PS-1
TO FACILITY SOURCE VALVE
B4
NO
R5
C
Y
NC
C
NC
14 R
13 B
NO
Y
LOW LEVEL SWITCH ON TANK
- TERMINAL IN ELECTRICAL
IN JUNCTION BOX
NC
9 R
B8
NOC
Y 2
V-1
RESERVE IN USE
RESERVE LOW
V-11
RV-4
PI-4 PI-3
V-8
V-5A R-2A
V-7
V-5C
R-2B
V-5D
V-5BPS-2 PS-1
SEE
DETAIL C
RV-1A
V-4
RV-1B CV-3
V-2
PI-1 PI-2
R-1 V-1
V-6
RV-3A
CV-2
S-1
V-3
V-9
V-10PI-5
V-11 RV-2A
RV-2B
RV-4
C C C C
NONONONO
NCNCNCNC
G G G G
NCNCNCNC
NONONONO
C C C C
SPAR
E
LIQ
U
ID
LEVEL
R
ESER
VE IN
U
SE
R
ESER
VE LO
W
ALARM WIRES TO
FACILITY
TO HP MANIFOLD
TO GAS USE VALVE
ON LIQUID STORAGE TANK
TO FACILITY SOURCE VALVE
SEE PAGE 1, TABLE 1 FOR OTHER CONFIGURATIONS
RV-3B
APPVREV ECR REVISION DESCRIPTION BY DATE
# DESIGNATES A NOTE REFERENCE.
MAT'L
FINISH P/N
SEE TABLE
APPV'D DATE
JNB 2/2/2015
GAS USE
TO
P FILL
BO
TT
O M
F
IL
L
VENT
ECON OUT
SAFETY/ECON IN PB OUT
H
L
R E S E R V E
B O T T L E S
DIM A
102 1/2
81 1/2
100 1/2
56 1/2
45 3/8 52 3/8
3 7/8
PART NUMBER SPEC DRAWING MODEL DIM A (in.) WEIGHT W/FRAME (lbs.)
20814390 20852345 PERMA-CYL 1000HP FF 83 2600
20814391 20852342 PERMA-CYL 1500HP FF 83 2850
20814392 20852343 PERMA-CYL 2000HP FF 92 3300
20814394 20852344 PERMA-CYL 3000HP FF 92 3600
20755351 20852345 PERMA-CYL 1000VHP FF 120 3700
20755352 20852342 PERMA-CYL 1500VHP FF 121 4050
20793147 20852341 PERMA-CYL 2000VHP FF 123 4400
20793148 20852350 PERMA-CYL 3000VHP FF 124 5350
6 1/2
TYP2 3/4 TYP
13611
711 1/2
FOR OVERSIZED
SKIDS ONLY