Attachment 5 - PharmacySummary_NC-VFR2627401A-1.pdf

PDF 1 MB Posted

Attached to
VACCHCS Pharmacy Certification
Solicitation number
36C26127Q0099
Issued by

About this file

This is a pharmacy facility testing and certification report for Building 24 at VA Fresno, documenting compliance testing conducted on September 10, 2026, by Technical Safety Services (TSS).

The report encompasses comprehensive environmental and equipment testing across multiple rooms within the pharmacy facility. Testing includes certification of Biological Safety Cabinets (BSCs), unidirectional flow devices (clean benches), cleanroom certification, airborne nonviable particle count testing, viable sample collection, and pressurization data validation. Specific areas tested include the Hazardous Sterile Ante Room (ISO-7), Hazardous Sterile Buffer Room (ISO-7), Hazardous Storage Room, and Non-Hazardous Sterile Buffer Room (ISO-7). All tested BSCs achieved ISO-5 certification with passing results for airflow velocity, HEPA leak tests, and aerosol concentration measurements. Clean benches passed certification with measured outflow velocities of 89-94 FPM. Cleanroom testing confirmed compliance with applicable standards including USP <797>, USP <800>, CETA CAG-003, and IEST standards, with airborne particle counts well below ISO-7 and ISO-5 class limits. Viable sampling was conducted with 15 active air samples and 20 surface samples collected from various cleanroom locations. All pressurization tests passed, confirming proper directional airflow between adjacent areas. Equipment calibration records confirm use of NIST-traceable instruments with current calibration dates. The retest date is scheduled for March 2027.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

TH T3 Building 24 Pharmacy Testing Sept 2026 AL

Technical Safety Services Corporate Headquarters 4225 Executive Square, Suite 370 La Jolla, CA 92037 800.877.7742 www.techsafety.com

Project Site:

VA Fresno

2615 East Clinton Avenue

Fresno, CA 93703

Project Contact:

Travis Horn

2615 East Clinton Avenue

Fresno CA, 93703

Travis.Horn@va.gov

4225 Executive Square, Suite 370

La Jolla, CA 92037

(800) 877-7742

Pharmacy Summary Report

Project: NC-VFR2627401A-1 TH T3 Building 24 Pharmacy Testing Sept 2026 AL

Site: VA Fresno 2615 East Clinton Avenue Fresno, CA 93703

Contact: Travis Horn

(559) 225-6100 Travis.Horn@va.gov

Technician: Daniel Velazquez

Qualifications:

TSS Technicians are trained and competent in the methods and procedures in CETA CAG-003

Purpose:

This document serves as the preliminary report summary of testing performed at your facility on the date(s) listed. These tests were conducted to demonstrate compliance with USP <797>, CETA and applicable TSS SOPs and is subject to further review.

Summary of Test Results:

Below is a list of the services performed at your facility on the given test date(s). Where applicable, test results have been applied to the listed tasks. If results are pending, such as airborne and surface viable testing that requires post-test incubation and enumeration laboratory processing, these results will be sent to the client contact separately on a future date.

This document along with the attached field documentation can serve your auditing and internal assessment needs. If you require anything further, please contact TSS at (800) 877-7742.

NC-VFR2627401A-1 / September 11, 2026 1 of 3

Test Report Summary Table

Serial Number Cert. ID Test Description Unit Type

ISO

Class

Cert. Date Result

199716071620 DV2691010465 Certification A2 ISO-5 09/10/2026 Certification: Certified

199716071620 DV269109821 Non Viable Sample A2 ISO-5 09/10/2026 Non Viable Sample: Pass

199716071620 DV26910973 Special Project A2 ISO-5 09/10/2026 Airflow Visualization Study: Complete

199733071720 DV26910104620 Certification A2 ISO-5 09/10/2026 Certification: Certified

199733071720 DV269109845 Non Viable Sample A2 ISO-5 09/10/2026 Non Viable Sample: Pass

199733071720 DV269109643 Special Project A2 ISO-5 09/10/2026 Airflow Visualization Study: Complete

199746071720 DV26910104641 Certification Clean Bench ISO-5 09/10/2026 Certification: Certified

199746071720 DV26910103727 Non Viable Sample Clean Bench ISO-5 09/10/2026 Non Viable Sample: Pass

199746071720 DV269109753 Special Project Clean Bench ISO-5 09/10/2026 Airflow Visualization Study: Complete

199751071720 DV26910104632 Certification Clean Bench ISO-5 09/10/2026 Certification: Certified

199751071720 DV26910103620 Non Viable Sample Clean Bench ISO-5 09/10/2026 Non Viable Sample: Pass

199751071720 DV269109722 Special Project Clean Bench ISO-5 09/10/2026 Airflow Visualization Study: Complete

Building 24 Sterile Compounding Sui

DV26910964 Viable Sample Environmenta l Sampling Unit

NA 09/10/2026 Viable Samples (35): Sampled

Hazardous Sterile Ante Room

DV26910104810 Certification HD Sterile Anteroom

ISO-7 09/10/2026

Airflow: FIO HEPA Leak Test: Pass Room Air Exchange Rate: Pass Room Pressures: Pass Non-Viable Sample: Pass

Hazardous Sterile Buffer Room

DV26910104837 Certification HD Sterile Buffer Room

ISO-7 09/10/2026

Airflow: FIO HEPA Leak Test: Pass Room Air Exchange Rate: Pass Room Pressures: Pass Non-Viable Sample: Pass

Hazardous Storage Room

DV26910104915 Certification HD Storage NA 09/10/2026 Airflow: FIO Room Air Exchange Rate: Pass Room Pressures: Pass

Non-Hazardous Sterile Buffer Room

DV26910104742 Certification Non-HD Sterile Buffer Room

ISO-7 09/10/2026

Airflow: FIO HEPA Leak Test: Pass Room Air Exchange Rate: Pass Room Air Exchange Rate + PEC: Pass Room Pressures: Pass Non-Viable Sample: Pass

Definitions / Acronyms

ACPH Air changes per hour

BSC Biological Safety Cabinet

CAI Compounding Aseptic Isolator

CACI Compounding Aseptic Containment Isolator

CETA Controlled Environment Testing Association

CER Controlled Environment

CFM Cubic Feet per Minute (air volume measurement)

CSP Compounded Sterile Preparation

DCA Direct Compounding Area

FPM Feet per Minute (air velocity measurement)

HEPA High Efficiency Particulate Air (filter)

NC-VFR2627401A-1 / September 11, 2026 2 of 3

HVAC Heating Ventilation and Air Conditioning

IEST Institute for Environmental Sciences and Technology

ISO International Organization for Standardization

LAFW Laminar Air Flow Workstation

NSF/ANSI NSF International/ American National Standards Institute

PEC Primary Engineering Control (ISO5 Environment, e.g., BSC, LAFW, CAI, CACI)

RAER Room Air Exchange Rate (ACPH)

SCA Segregated Compounding Area

SEC Secondary Engineering Control (Cleanroom)

TAF Terminal Air Filter (Ceiling/Room Side HEPA Filter)

UAV Unidirectional Airflow Visualization um Micrometer 1 x 10-6 meters (particle size measurement)

VLF Vertical Laminar Flow

WC Water Column (pressure measurement)

NC-VFR2627401A-1 / September 11, 2026 3 of 3

Biological Safety Cabinet Certification Report

La Jolla, CA 92037

(800) 877-7742

Cert.: DV2691010465 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Test Standard(s): MANUFACTURER, OSHA, ANSI/NSF 49, USP <797>, CETA CAG-003

Inspected Min Max Measured Result

Calculated Inflow Velocity (FPM) 100 110 105 Pass

Average Supply Velocity Zone1 (FPM) 55 65 60 Pass

Average Supply Velocity Zone2 (FPM) NA NA NA NA

Average Supply Velocity Zone3 (FPM) NA NA NA NA

Average Supply Velocity Zone4 (FPM) NA NA NA NA

Supply Uniformity NA NA Pass Pass

Downflow Smoke Pattern Test NA NA Pass Pass

View Screen Retention Smoke Pattern Test NA NA Pass Pass

Work Opening Edge Retention Smoke Pattern Test NA NA Pass Pass

Sash Seal Smoke Pattern Test NA NA Pass Pass

Site Assessment NA NA Pass Pass

At-Rest Unidirectional Airflow Visualization Test NA NA NA NA

Dynamic Unidirectional Airflow Visualization Test NA NA Pass Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 23 Pass

Exhaust Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 23 Pass

Comments:

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:33 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

La Jolla, CA 92037

(800) 877-7742

Cert.: DV2691010465 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Inflow (FPM) Method: MFR spec: I4 - Constricted Access Calculated Inflow Velocity (FPM): 105

Workface: Width: 46.50" Height: 10.00" Workface Area (ft²): 3.23

Adj. Workface: Width: 46.50" Height: 3.00" Adj. Workface Area (ft²): 0.97

AK Factor: 1.00 Inflow Factor: 0 Volume (CFM): 340

Readings:

344 347 352 365 349 351 345 354

Supply (FPM) Method: MFR spec: S1 - Unidirectional (non-zoned) flow Average Supply Velocity(FPM): 60

Dimensions: Width: 46.50" Depth: 25.00" Supply Area (ft²): 8.07

Readings:

Supply Volume (CFM): 484

66 64 66 64 65 64 63

61 60 60 59 58 57 57

60 58 58 53 55 55 58

Site Assessment:

Exhaust System: Canopy: Negative: The Canopy Passed the performance test.

Exhaust System Alarm: Pass: The Exhaust System Alarm Passed the performance test.

Sash Alarm: Pass: The Sash Alarm Passed the performance test.

Interlock: NA: No Interlock Alarm present.

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:33 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 2/2

(800) 877-7742

Cert. No.: DV269109821 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199716071620

Airborne Nonviable Particle Count Test Report Clean Zone: 199716071620 Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 8.07 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.1 54

Class Limit: 3,520 21.1 54

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:10 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate representation of work performed.

Print Date: 09/10/2026

(800) 877-7742

Cert. No.: DV269109821 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199716071620

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

013812 Govee H5075 EQ13812 10/2026

009464 Climet CI-750t 057234 11/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:10 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

Airflow Visualization Study

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910973 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 1/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910973 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 2/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104620 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Test Standard(s): MANUFACTURER, OSHA, ANSI/NSF 49, USP <797>, CETA CAG-003

Inspected Min Max Measured Result

Calculated Inflow Velocity (FPM) 100 110 103 Pass

Average Supply Velocity Zone1 (FPM) 55 65 64 Pass

Average Supply Velocity Zone2 (FPM) NA NA NA NA

Average Supply Velocity Zone3 (FPM) NA NA NA NA

Average Supply Velocity Zone4 (FPM) NA NA NA NA

Supply Uniformity NA NA Pass Pass

Downflow Smoke Pattern Test NA NA Pass Pass

View Screen Retention Smoke Pattern Test NA NA Pass Pass

Work Opening Edge Retention Smoke Pattern Test NA NA Pass Pass

Sash Seal Smoke Pattern Test NA NA Pass Pass

Site Assessment NA NA Pass Pass

At-Rest Unidirectional Airflow Visualization Test NA NA NA NA

Dynamic Unidirectional Airflow Visualization Test NA NA Pass Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 22 Pass

Exhaust Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 22 Pass

Comments:

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:36 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 1/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104620 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Inflow (FPM) Method: MFR spec: I4 - Constricted Access Calculated Inflow Velocity (FPM): 103

Workface: Width: 46.50" Height: 10.00" Workface Area (ft²): 3.23

Adj. Workface: Width: 46.50" Height: 3.00" Adj. Workface Area (ft²): 0.97

AK Factor: 1.00 Inflow Factor: 0 Volume (CFM): 332

Readings:

351 348 344 337 331 336 342 347

Supply (FPM) Method: MFR spec: S1 - Unidirectional (non-zoned) flow Average Supply Velocity(FPM): 64

Dimensions: Width: 46.50" Depth: 25.00" Supply Area (ft²): 8.07

Readings:

Supply Volume (CFM): 516

66 67 67 64 64 65 66

67 63 63 63 64 62 65

63 61 63 62 61 63 64

Site Assessment:

Exhaust System: Canopy: Negative: The Canopy Passed the performance test.

Exhaust System Alarm: Pass: The Exhaust System Alarm Passed the performance test.

Sash Alarm: Pass: The Sash Alarm Passed the performance test.

Interlock: NA: No Interlock Alarm present.

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:36 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 2/2

(800) 877-7742

Cert. No.: DV269109845 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199733071720

Airborne Nonviable Particle Count Test Report Clean Zone: 199733071720 Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 8.07 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.0 55

Class Limit: 3,520 21.0 55

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:24 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

(800) 877-7742

Cert. No.: DV269109845 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199733071720

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

013812 Govee H5075 EQ13812 10/2026

009464 Climet CI-750t 057234 11/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:24 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109643 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:06 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 1/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109643 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:06 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 2/2

Unidirectional Flow Device Certification Report

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104641 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Test Standard(s): USP <797>, CETA CAG-003, IEST-RP-CC002.4, USP <797> 2023, MANUFACTURER

Results

Inspected Min Max Measured Results

Outflow Velocity (FPM) 80 100 89 Pass

%RSD NA 15 5 Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Backstreaming Test NA NA Pass Pass

Induction Leak Test NA NA Pass Pass

Smoke Source Glycerin and/or Glycol

Fluorescent Lighting (FC) NA NA NA NA

NAUltraviolet Lighting (μW/cm²) NA NA NA

NANoise Level (dBA) NA NA NA

NAVibration (in. RMS) NA NA NA

NAElectrical Current Leakage Test (mA) NA NA NA

NAElectrical Ground Resistance Test (Ohms) NA NA NA

NAElectrical Polarity NA NA NA

NAElectrical GCFI NA NA NA

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 20 Pass

Comments:

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:40 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 1/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104641 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Velocity (FPM) Dimension: Width: 48.00" Height: 32.00" Average Velocity (FPM) 89

Area (ft²): 10.67

Readings: Volume (CFM): 950

79 84 87 89

88 90 86 89

95 96 92 88

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:40 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 2/2

(800) 877-7742

Cert. No.: DV26910103727 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199746071720

Airborne Nonviable Particle Count Test Report Clean Zone: 199746071720 Status: Operational/Dynamic Cust ID: Hood #2 Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 10.67 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.1 55

Class Limit: 3,520 21.1 55

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:37:53 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

(800) 877-7742

Cert. No.: DV26910103727 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199746071720

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

013812 Govee H5075 EQ13812 10/2026

009464 Climet CI-750t 057234 11/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:37:53 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109753 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:08 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 1/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109753 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:08 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 2/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104632 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Test Standard(s): IEST-RP-CC002.4, CETA CAG-003, USP <797> 2023, MANUFACTURER, USP <797>

Results

Inspected Min Max Measured Results

Outflow Velocity (FPM) 80 100 94 Pass

%RSD NA 15 8 Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Backstreaming Test NA NA Pass Pass

Induction Leak Test NA NA Pass Pass

Smoke Source Glycerin and/or Glycol

Fluorescent Lighting (FC) NA NA NA NA

NAUltraviolet Lighting (μW/cm²) NA NA NA

NANoise Level (dBA) NA NA NA

NAVibration (in. RMS) NA NA NA

NAElectrical Current Leakage Test (mA) NA NA NA

NAElectrical Ground Resistance Test (Ohms) NA NA NA

NAElectrical Polarity NA NA NA

NAElectrical GCFI NA NA NA

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 19 Pass

Comments:

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:38 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 1/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104632 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Velocity (FPM) Dimension: Width: 48.00" Height: 32.00" Average Velocity (FPM) 94

Area (ft²): 10.67

Readings: Volume (CFM): 1,003

90 90 83 88

90 97 103 97

87 92 101 106

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:38 PM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/2026 Page: 2/2

(800) 877-7742

Cert. No.: DV26910103620 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199751071720

Airborne Nonviable Particle Count Test Report Clean Zone: 199751071720 Status: Operational/Dynamic Cust ID: Hood #1 Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 10.67 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.0 56

Class Limit: 3,520 21.0 56

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:36:52 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

(800) 877-7742

Cert. No.: DV26910103620 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199751071720

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

013812 Govee H5075 EQ13812 10/2026

009464 Climet CI-750t 057234 11/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:36:52 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109722 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 1/2

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109722 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.3.1 Field Service Technician

Print Date: 09/10/2026 Page: 2/2

Viable Sample Report

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling

Area Status: Operational/Dynamic

Occupants: TSS: 1 Client: 2

15 - Active Air Samples

Sample ID Location Type

Sample Size Media Lot # Exp. Date

A-01 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-02 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-03 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-04 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-05 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-06 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-07 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-08 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-09 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-10 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-11 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-12 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-13 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-14 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

Control NA NA TSA Settling Plate 691307P 01/2027

20 - Surface Samples

Sample ID Location Type

Sample Size Media Lot # Exp. Date

Control NA NA TSA Contact Plate 688281P 11/2026

S-01 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-02 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-03 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-04 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-05 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-06 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-07 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-08 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-09 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-10 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026 www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

Print Date: 09/2026 Page: 1/4

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling

20 - Surface Samples

Sample ID Location Type

Sample Size Media Lot # Exp. Date

S-11 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-12 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-13 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-14 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-15 CNC 1 Plate TSA Contact Plate 688281P 11/2026

S-16 CNC 1 Plate TSA Contact Plate 688281P 11/2026

S-17 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-18 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-19 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

Remarks: NA www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

Print Date: 09/2026 Page: 2/4

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

Print Date: 09/2026 Page: 3/4

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling

The following NIST-Traceable equipment were used to collect these viable samples:

Equipment ID Manufacturer Model Serial # Calibration Due

012632 AP BUCK B30120 C103665 10/24/2026

012638 AP BUCK B30120 C103671 10/24/2026

www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.3.1

I certify that this record is a true and accurate

Print Date: 09/2026 Page: 4/4

(800) 877-7742

Test Report

General Information:

Project Site: VA Fresno

2615 East Clinton Avenue

Fresno, CA 93703

Room: Hazardous Sterile Ante Room

Test Report ID: DV26910104810

Project ID: NC-VFR2627401A-1

Project Contact: Travis Horn

VFR01 | VA Fresno

(559) 225-6100

Travis.Horn@va.gov

Project Lead Tech: Daniel Velazquez

823 Kifer Rd

Sunnyvale, CA 94086 tel: (800) 877-7742 fax: (510) 845-5595

Project Description: TH T3 Building 24 Pharmacy Testing Sept 2026 AL

Technician Remarks: NA

Test Standards: CETA CAG-003, USP <800>, IEST-RP-CC006 www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 1/9 I certify that this record is a true and accurate

FIO: For Information Only NA: Not Applicable

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

Room Description:

Building: 24

Room: Hazardous Sterile Ante Room

Room Size:

Area (ft²): 227.26 Volume (ft³): 1,931

Area (m²): 21.11 Volume (m³): 55

Room Design:

ISO-7

Results

Inspected Min Max Measured Results

Airflow Volume (CFM) NA NA 1,281 FIO

Grand Mean (FPM) NA NA NA NA

Airflow Uniformity (%RSD) NA NA NA NA

Average Velocities (FPM) * * * FIO

Individual Unit Uniformity (%RSD) * * * NA

Leak Test (%) * * * Pass

Patch (% of Filter Area) * * * Pass

Room Air Exchange Rate (AC/H) 30 NA 40 Pass

Airborne Nonviable Particle Count * * * Pass

Differential Pressure * * * Pass

* Footnote – Results are aggregate of all units tested; individual unit criterion noted in the following pages.

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 2/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

TAF AIRFLOW MEASUREMENTS: Hazardous Sterile Ante Room

Filter ID Individual Measurements Lower Limit

Upper Limit

Avg Vel (fpm)

Area (ft²)

Kv Avg Vol (cfm)

%RSD

Limit

%RSD ΔP Result

Cert. ID: 146DB627-CD0F-4E66-9432-B4C1012562E1

NA NA 96 6.72 1.00 648 NA NA NA FIOTAF-03 (CFM) 648

comments: NA

Cert. ID: B40B8CF2-0D18-45D9-B67B-B4C1012562E1

NA NA 94 6.72 1.00 633 NA NA NA FIOTAF-04 (CFM) 633

comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 3/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

TAF INSTALLATION LEAK TEST RESULTS: Hazardous Sterile Ante Room

PAO (μg/l)% Patched AreaFilter ID

Initial Final

Leakage Diagram (Magnitude, Location, and Dimension)

% Leak Conc. Result

Conc.Max Min Max Initial Final Max

TAF-03 No Diagram 0.00 0.00 3.00 29 10 100 0.000 NA 0.010 Pass comments: NA

TAF-04 No Diagram 0.00 0.00 3.00 30 10 100 0.000 NA 0.010 Pass comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 4/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

Pressurization Data

Zone (+) Zone (-) Actual ΔP ("wc)

Min ΔP ("wc)

Max ΔP ("wc)

Directional Airflow

Door Gauge Readings ("wc) Meets CriteriaClosed ("wc) Open ("wc)

Hazardous Sterile Ante Room

Hazardous Storage Room

0.050 0.020 NA Pass NA NA Yes comments: NA

Hazardous Sterile Ante Room

Adjacent Room 0.032 0.020 NA Pass NA NA Yes comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 5/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Ante

Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-7 Area: 227.26 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 6480 NA NA 6480 Pass

2) 3746 NA NA 3746 Pass

3) 2786 NA NA 2786 Pass

4) 3693 NA NA 3693 Pass

5) 2173 NA NA 2173 Pass

6) 1453 NA NA 1453 Pass

T. C° % rH

21.2 55

21.2 56

21.2 56

21.3 57

21.4 57

21.3 57

Class Limit: 352,000 21.3 56

Sample Volume Test: Pass

Class Limit Location Test: Pass www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 6/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Ante

Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-7 Area: 227.26 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 7/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

Cleanroom Diagram:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 8/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

006337 Evergreen Telemetry S-PVF-1 1900817C 10/2026

009016 TEC PH-4 4132 11/2026

009464 Climet CI-750t 057234 11/2026

009747 Evergreen Telemetry Wrist Reporter WR-401 2C03501 09/2026

009752 Evergreen Telemetry Three Pounder 2200368B 09/2026

012839 TEC AG-E3 4952 10/2026

013812 Govee H5075 EQ13812 10/2026 www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 9/9 I certify that this record is a true and accurate

(800) 877-7742

Test Report

General Information:

Project Site: VA Fresno

2615 East Clinton Avenue

Fresno, CA 93703

Room: Hazardous Sterile Buffer Room

Test Report ID: DV26910104837

Project ID: NC-VFR2627401A-1

Project Contact: Travis Horn

VFR01 | VA Fresno

(559) 225-6100

Travis.Horn@va.gov

Project Lead Tech: Daniel Velazquez

823 Kifer Rd

Sunnyvale, CA 94086 tel: (800) 877-7742 fax: (510) 845-5595

Project Description: TH T3 Building 24 Pharmacy Testing Sept 2026 AL

Technician Remarks: NA

Test Standards: USP <800>, IEST-RP-CC006 www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 1/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837

Room Description:

Building: 24

Room: Hazardous Sterile Buffer Room

Room Size:

Area (ft²): 110.00 Volume (ft³): 943

Area (m²): 10.22 Volume (m³): 27

Room Design:

ISO-7

Results

Inspected Min Max Measured Results

Airflow Volume (CFM) NA NA 722 FIO

Grand Mean (FPM) NA NA NA NA

Airflow Uniformity (%RSD) NA NA NA NA

Average Velocities (FPM) * * * FIO

Individual Unit Uniformity (%RSD) * * * NA

Leak Test (%) * * * Pass

Patch (% of Filter Area) * * * Pass

Room Air Exchange Rate (AC/H) 30 NA 46 Pass

Airborne Nonviable Particle Count * * * Pass

Differential Pressure * * * Pass

* Footnote – Results are aggregate of all units tested; individual unit criterion noted in the following pages.

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 2/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837

TAF AIRFLOW MEASUREMENTS: Hazardous Sterile Buffer Room

Filter ID Individual Measurements Lower Limit

Upper Limit

Avg Vel (fpm)

Area (ft²)

Kv Avg Vol (cfm)

%RSD

Limit

%RSD ΔP Result

Cert. ID: AEBB45B6-C6CE-44EB-9BE1-B4C10125928B

NA NA 116 3.36 1.00 391 NA NA NA FIOTAF-01 (CFM) 391

comments: NA

Cert. ID: 3399BEFC-E1DB-4F89-940C-B4C10125928B

NA NA 99 3.36 1.00 331 NA NA NA FIOTAF-02 (CFM) 331

comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 3/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837

TAF INSTALLATION LEAK TEST RESULTS: Hazardous Sterile Buffer Room

PAO (μg/l)% Patched AreaFilter ID

Initial Final

Leakage Diagram (Magnitude, Location, and Dimension)

% Leak Conc. Result

Conc.Max Min Max Initial Final Max

TAF-01 No Diagram 0.00 0.00 3.00 48 10 100 0.000 NA 0.010 Pass comments: NA

TAF-02 No Diagram 0.00 0.00 3.00 57 10 100 0.000 NA 0.010 Pass comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 4/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837

Pressurization Data

Zone (+) Zone (-) Actual ΔP ("wc)

Min ΔP ("wc)

Max ΔP ("wc)

Directional Airflow

Door Gauge Readings ("wc) Meets CriteriaClosed ("wc) Open ("wc)

Hazardous Sterile Ante Room

Hazardous Sterile Buffer Room

0.024 0.010 0.030 Pass 0.021 0.000 Yes comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 5/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837

Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Buffer

Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-7 Area: 110.00 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 21066 NA NA 21066 Pass

2) 6440 NA NA 6440 Pass

3) 5680 NA NA 5680 Pass

4) 3400 NA NA 3400 Pass

5) 10026 NA NA 10026 Pass

6) 19026 NA NA 19026 Pass

T. C° % rH

21.0 55

21.0 55

21.2 56

21.2 57

21.1 57

21.1 56

Class Limit: 352,000 21.1 56

Sample Volume Test: Pass

Class Limit Location Test: Pass www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 6/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837

Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Buffer

Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-7 Area: 110.00 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 7/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837

Cleanroom Diagram:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.3.1

Field Service TechnicianPrint Date: 09/10/2026 Page: 8/9 I certify that this record is a true and accurate

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID:…

This is the start of the file's text. The full file is on GovTribe.