Attachment 3 - VFR2627401A-1.10-11 Sep 2026.Bldg 24 Pharmacy USP Testing.Appendix.pdf
PDF 2 MB Posted
- Attached to
- VACCHCS Pharmacy Certification
- Solicitation number
- 36C26127Q0099
About this file
This file is an Appendix containing USP (United States Pharmacopeia) pharmaceutical compounding facility testing and certification reports for Building 24 Pharmacy at VA Fresno, dated September 10-11, 2026. The document compiles comprehensive testing results for the Service Order NC-VFR2627401A-1, conducted by TechSafety at the VA Fresno facility located at 2615 East Clinton Avenue, Fresno, California 93703, with contact Travis Horn.
The appendix documents testing of multiple hazardous and non-hazardous sterile pharmacy compounding areas, including two Class II, Type A2 Biological Safety Cabinets (Model NU-540-400), two unidirectional flow clean benches (Model NU-240-430), and associated support rooms. Testing encompassed biological safety cabinet certifications with inflow velocity measurements (103–105 FPM), HEPA filter integrity assessments (all passing with 0.000% max point leaks), airborne nonviable particle counts confirming ISO-5 and ISO-7 classifications, airflow visualization studies confirming unidirectional flow patterns, viable environmental sampling across multiple locations, and room-level pressurization and air change rate documentation. The facility includes a Hazardous Sterile Ante Room (ISO-7, 40 ACPH), Hazardous Sterile Buffer Room (ISO-7, 46 ACPH), Hazardous Storage Room (15 ACPH), Non-Hazardous Sterile Buffer Room (ISO-7, 180 ACPH combined with PEC), and an ancillary sterile compounding suite. All tests employed NIST-traceable equipment and adhered to USP, ANSI/NSF 49, CETA CAG-003, ISO 14644-1:2015, IEST-RP-CC standards, with retest dates scheduled for March 2027.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Biological Safety Cabinet Certification Report
4225 Executive Square, Suite 370
La Jolla, CA 92037
(800) 877-7742
Cert.: DV2691010465-02 Facility: VA Fresno ID: 5505820 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
Test Standard(s): MANUFACTURER, OSHA, ANSI/NSF 49, USP <797>, CETA CAG-003
Inspected Min Max Measured Result
Calculated Inflow Velocity (FPM) 100 110 105 Pass
Average Supply Velocity Zone1 (FPM) 55 65 60 Pass
Average Supply Velocity Zone2 (FPM) NA NA NA NA
Average Supply Velocity Zone3 (FPM) NA NA NA NA
Average Supply Velocity Zone4 (FPM) NA NA NA NA
Supply Uniformity NA NA Pass Pass
Downflow Smoke Pattern Test NA NA Pass Pass
View Screen Retention Smoke Pattern Test NA NA Pass Pass
Work Opening Edge Retention Smoke Pattern Test NA NA Pass Pass
Sash Seal Smoke Pattern Test NA NA Pass Pass
Site Assessment NA NA Pass Pass
At-Rest Unidirectional Airflow Visualization Test NA NA NA NA
Dynamic Unidirectional Airflow Visualization Test NA NA Pass Pass
Supply Delta Pressure ("W.C.) NA NA NA NA
Duct Pressure ("W.C.) NA NA NA NA
Supply Min Max Measured Results
HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass
HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA
Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass
Aerosol Concentration (ug/l) 10 100 23 Pass
Exhaust Min Max Measured Results
HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass
HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA
Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass
Aerosol Concentration (ug/l) 10 100 23 Pass
Comments: This certificate, DV2691010465-02, is supplement to test report DV2691010465. Added technician comments.
Fabian Olivares 9/16/2026 12:30:40 PM The space between the face of the BSC and permanent fixture directly opposite the access opening was less than 42 inches, used manufacturers alternative method to measure inflow readings.
**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial No. Cal. Due Date
005672 TSI 9535 T95351923010 05/2027
009016 TEC PH-4 4132 11/2026
012839 TEC AG-E3 4952 10/2026
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:33 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
Ver. 15 Oct 2026
La Jolla, CA 92037
(800) 877-7742
Cert.: DV2691010465-02 Facility: VA Fresno ID: 5505820 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
Inflow (FPM) Method: MFR spec: I4 - Constricted Access Calculated Inflow Velocity (FPM): 105
Workface: Width: 46.50" Height: 10.00" Workface Area (ft²): 3.23
Adj. Workface: Width: 46.50" Height: 3.00" Adj. Workface Area (ft²): 0.97
AK Factor: 1.00 Inflow Factor: 0 Volume (CFM): 340
Readings:
344 347 352 365 349 351 345 354
Supply (FPM) Method: MFR spec: S1 - Unidirectional (non-zoned) flow Average Supply Velocity(FPM): 60
Dimensions: Width: 46.50" Depth: 25.00" Supply Area (ft²): 8.07
Readings:
Supply Volume (CFM): 484
66 64 66 64 65 64 63
61 60 60 59 58 57 57
60 58 58 53 55 55 58
Site Assessment:
Exhaust System: Canopy: Negative: The Canopy Passed the performance test.
Exhaust System Alarm: Pass: The Exhaust System Alarm Passed the performance test.
Sash Alarm: Pass: The Sash Alarm Passed the performance test.
Interlock: NA: No Interlock Alarm present.
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:33 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert. No.: DV269109821 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-540-400
Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2
Fresno, CA 93703 Contact: Travis Horn SN: 199716071620
Airborne Nonviable Particle Count Test Report Clean Zone: 199716071620 Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).
Report Date: 09/2026 Class: ISO-5 Area: 8.07 ft² Flow Rate: 75.0 LPM
Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L
Testing is performed to satisfy ISO 14644-1:2015
Pass PPCM (um) ≥ 0.5
Loc Count 1 Count 2 Count 3 Average Result
1) 0 NA NA 0 Pass
T. C° % rH
21.1 54
Class Limit: 3,520 21.1 54
Sample Volume Test: Pass
Class Limit Location Test: Pass
Airborne Nonviable Particle Count Diagram:
www.techsafety.com
This record was electronically signed by:
Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:10 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Cert. No.: DV269109821 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-540-400
Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2
Fresno, CA 93703 Contact: Travis Horn SN: 199716071620
The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial # Calibration Due
013812 Govee H5075 EQ13812 10/2026
009464 Climet CI-750t 057234 11/2026 www.techsafety.com
This record was electronically signed by:
Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:10 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
Airflow Visualization Study
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910973 Facility: VA Fresno ID: 5505820 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
Service Type: Airflow Visualization Study
Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.
Doc. Title Doc. # Doc. Version Completed Date Originally Retained By
Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910973 Facility: VA Fresno ID: 5505820 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910104620-02 Facility: VA Fresno ID: 5505819 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
Test Standard(s): MANUFACTURER, OSHA, ANSI/NSF 49, USP <797>, CETA CAG-003
Inspected Min Max Measured Result
Calculated Inflow Velocity (FPM) 100 110 103 Pass
Average Supply Velocity Zone1 (FPM) 55 65 64 Pass
Average Supply Velocity Zone2 (FPM) NA NA NA NA
Average Supply Velocity Zone3 (FPM) NA NA NA NA
Average Supply Velocity Zone4 (FPM) NA NA NA NA
Supply Uniformity NA NA Pass Pass
Downflow Smoke Pattern Test NA NA Pass Pass
View Screen Retention Smoke Pattern Test NA NA Pass Pass
Work Opening Edge Retention Smoke Pattern Test NA NA Pass Pass
Sash Seal Smoke Pattern Test NA NA Pass Pass
Site Assessment NA NA Pass Pass
At-Rest Unidirectional Airflow Visualization Test NA NA NA NA
Dynamic Unidirectional Airflow Visualization Test NA NA Pass Pass
Supply Delta Pressure ("W.C.) NA NA NA NA
Duct Pressure ("W.C.) NA NA NA NA
Exhaust Min Max Measured Results
HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass
HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA
Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass
Aerosol Concentration (ug/l) 10 100 22 Pass
Supply Min Max Measured Results
HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass
HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA
Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass
Aerosol Concentration (ug/l) 10 100 22 Pass
Comments: This certificate, DV26910104620-02, is supplement to test report DV26910104620. Added technician comments.
Fabian Olivares 9/16/2026 12:25:49 PM The space between the face of the BSC and permanent fixture directly opposite the access opening was less than 42 inches, used manufacturers alternative method to measure inflow readings.
**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial No. Cal. Due Date
005672 TSI 9535 T95351923010 05/2027
009016 TEC PH-4 4132 11/2026
012839 TEC AG-E3 4952 10/2026
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:36 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910104620-02 Facility: VA Fresno ID: 5505819 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
Inflow (FPM) Method: MFR spec: I4 - Constricted Access Calculated Inflow Velocity (FPM): 103
Workface: Width: 46.50" Height: 10.00" Workface Area (ft²): 3.23
Adj. Workface: Width: 46.50" Height: 3.00" Adj. Workface Area (ft²): 0.97
AK Factor: 1.00 Inflow Factor: 0 Volume (CFM): 332
Readings:
351 348 344 337 331 336 342 347
Supply (FPM) Method: MFR spec: S1 - Unidirectional (non-zoned) flow Average Supply Velocity(FPM): 64
Dimensions: Width: 46.50" Depth: 25.00" Supply Area (ft²): 8.07
Readings:
Supply Volume (CFM): 516
66 67 67 64 64 65 66
67 63 63 63 64 62 65
63 61 63 62 61 63 64
Site Assessment:
Exhaust System: Canopy: Negative: The Canopy Passed the performance test.
Exhaust System Alarm: Pass: The Exhaust System Alarm Passed the performance test.
Sash Alarm: Pass: The Sash Alarm Passed the performance test.
Interlock: NA: No Interlock Alarm present.
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:36 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Cert. No.: DV269109845 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-540-400
Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2
Fresno, CA 93703 Contact: Travis Horn SN: 199733071720
Airborne Nonviable Particle Count Test Report Clean Zone: 199733071720 Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).
Report Date: 09/2026 Class: ISO-5 Area: 8.07 ft² Flow Rate: 75.0 LPM
Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L
Testing is performed to satisfy ISO 14644-1:2015
Pass PPCM (um) ≥ 0.5
Loc Count 1 Count 2 Count 3 Average Result
1) 0 NA NA 0 Pass
T. C° % rH
21.0 55
Class Limit: 3,520 21.0 55
Sample Volume Test: Pass
Class Limit Location Test: Pass
Airborne Nonviable Particle Count Diagram:
www.techsafety.com
This record was electronically signed by:
Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:24 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Cert. No.: DV269109845 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-540-400
Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2
Fresno, CA 93703 Contact: Travis Horn SN: 199733071720
The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial # Calibration Due
009464 Climet CI-750t 057234 11/2026
013812 Govee H5075 EQ13812 10/2026 www.techsafety.com
This record was electronically signed by:
Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:24 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV269109643 Facility: VA Fresno ID: 5505819 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
Service Type: Airflow Visualization Study
Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.
Doc. Title Doc. # Doc. Version Completed Date Originally Retained By
Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:06 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV269109643 Facility: VA Fresno ID: 5505819 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400
Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:
Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:06 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
Unidirectional Flow Device Certification Report
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910104641 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
Test Standard(s): USP <797>, CETA CAG-003, IEST-RP-CC002.4, USP <797> 2023, MANUFACTURER
Results
Inspected Min Max Measured Results
Outflow Velocity (FPM) 80 100 89 Pass
%RSD NA 15 5 Pass
Supply Delta Pressure ("W.C.) NA NA NA NA
Duct Pressure ("W.C.) NA NA NA NA
Backstreaming Test NA NA Pass Pass
Induction Leak Test NA NA Pass Pass
Smoke Source Glycerin and/or Glycol
Fluorescent Lighting (FC) NA NA NA NA
NAUltraviolet Lighting (μW/cm²) NA NA NA
NANoise Level (dBA) NA NA NA
NAVibration (in. RMS) NA NA NA
NAElectrical Current Leakage Test (mA) NA NA NA
NAElectrical Ground Resistance Test (Ohms) NA NA NA
NAElectrical Polarity NA NA NA
NAElectrical GCFI NA NA NA
Supply Min Max Measured Results
HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass
HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA
Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass
Aerosol Concentration (ug/l) 10 100 20 Pass
Comments:
**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial No. Cal. Due Date
005672 TSI 9535 T95351923010 05/2027
009016 TEC PH-4 4132 11/2026
012839 TEC AG-E3 4952 10/2026
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:40 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910104641 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
Velocity (FPM) Dimension: Width: 48.00" Height: 32.00" Average Velocity (FPM) 89
Area (ft²): 10.67
Readings: Volume (CFM): 950
79 84 87 89
88 90 86 89
95 96 92 88
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:40 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Cert. No.: DV26910103727 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-240-430
Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench
Fresno, CA 93703 Contact: Travis Horn SN: 199746071720
Airborne Nonviable Particle Count Test Report Clean Zone: 199746071720 Status: Operational/Dynamic Cust ID: Hood #2 Duration: 1.0 Min(s).
Report Date: 09/2026 Class: ISO-5 Area: 10.67 ft² Flow Rate: 75.0 LPM
Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L
Testing is performed to satisfy ISO 14644-1:2015
Pass PPCM (um) ≥ 0.5
Loc Count 1 Count 2 Count 3 Average Result
1) 0 NA NA 0 Pass
T. C° % rH
21.1 55
Class Limit: 3,520 21.1 55
Sample Volume Test: Pass
Class Limit Location Test: Pass
Airborne Nonviable Particle Count Diagram:
www.techsafety.com
This record was electronically signed by:
Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:37:53 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Cert. No.: DV26910103727 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-240-430
Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench
Fresno, CA 93703 Contact: Travis Horn SN: 199746071720
The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial # Calibration Due
013812 Govee H5075 EQ13812 10/2026
009464 Climet CI-750t 057234 11/2026 www.techsafety.com
This record was electronically signed by:
Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:37:53 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV269109753 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
Service Type: Airflow Visualization Study
Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.
Doc. Title Doc. # Doc. Version Completed Date Originally Retained By
Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:08 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV269109753 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:08 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910104632 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
Test Standard(s): IEST-RP-CC002.4, CETA CAG-003, USP <797> 2023, MANUFACTURER, USP <797>
Results
Inspected Min Max Measured Results
Outflow Velocity (FPM) 80 100 94 Pass
%RSD NA 15 8 Pass
Supply Delta Pressure ("W.C.) NA NA NA NA
Duct Pressure ("W.C.) NA NA NA NA
Backstreaming Test NA NA Pass Pass
Induction Leak Test NA NA Pass Pass
Smoke Source Glycerin and/or Glycol
Fluorescent Lighting (FC) NA NA NA NA
NAUltraviolet Lighting (μW/cm²) NA NA NA
NANoise Level (dBA) NA NA NA
NAVibration (in. RMS) NA NA NA
NAElectrical Current Leakage Test (mA) NA NA NA
NAElectrical Ground Resistance Test (Ohms) NA NA NA
NAElectrical Polarity NA NA NA
NAElectrical GCFI NA NA NA
Supply Min Max Measured Results
HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass
HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA
Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass
Aerosol Concentration (ug/l) 10 100 19 Pass
Comments:
**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial No. Cal. Due Date
005672 TSI 9535 T95351923010 05/2027
009016 TEC PH-4 4132 11/2026
012839 TEC AG-E3 4952 10/2026
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:38 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV26910104632 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
Velocity (FPM) Dimension: Width: 48.00" Height: 32.00" Average Velocity (FPM) 94
Area (ft²): 10.67
Readings: Volume (CFM): 1,003
90 90 83 88
90 97 103 97
87 92 101 106
Report Date: 09/2026 www.techsafety.com
This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:38 PM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Cert. No.: DV26910103620 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-240-430
Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench
Fresno, CA 93703 Contact: Travis Horn SN: 199751071720
Airborne Nonviable Particle Count Test Report Clean Zone: 199751071720 Status: Operational/Dynamic Cust ID: Hood #1 Duration: 1.0 Min(s).
Report Date: 09/2026 Class: ISO-5 Area: 10.67 ft² Flow Rate: 75.0 LPM
Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L
Testing is performed to satisfy ISO 14644-1:2015
Pass PPCM (um) ≥ 0.5
Loc Count 1 Count 2 Count 3 Average Result
1) 0 NA NA 0 Pass
T. C° % rH
21.0 56
Class Limit: 3,520 21.0 56
Sample Volume Test: Pass
Class Limit Location Test: Pass
Airborne Nonviable Particle Count Diagram:
www.techsafety.com
This record was electronically signed by:
Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:36:52 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Cert. No.: DV26910103620 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE
Facility: VA Fresno Building: 24 Model: NU-240-430
Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench
Fresno, CA 93703 Contact: Travis Horn SN: 199751071720
The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial # Calibration Due
013812 Govee H5075 EQ13812 10/2026
009464 Climet CI-750t 057234 11/2026 www.techsafety.com
This record was electronically signed by:
Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:36:52 AM eData v8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV269109722 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
Service Type: Airflow Visualization Study
Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.
Doc. Title Doc. # Doc. Version Completed Date Originally Retained By
Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
(800) 877-7742
Cert.: DV269109722 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE
Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430
Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:
Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM
Print Version: Service Manager v8.0.4.3 Field Service Technician
Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
Viable Sample Report
La Jolla, CA 92037
800.877.7742
Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA
Contact: Travis Horn Fresno, CA 93703 Model: NA
Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite
Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling
Area Status: Operational/Dynamic
Occupants: TSS: 1 Client: 2
15 - Active Air Samples
Sample ID Location Type
Sample Size Media Lot # Exp. Date
A-01 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
A-02 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
A-03 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
A-04 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
A-05 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027
A-06 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027
A-07 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027
A-08 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027
A-09 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027
A-10 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027
A-11 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
A-12 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
A-13 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
A-14 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027
Control NA NA TSA Settling Plate 691307P 01/2027
20 - Surface Samples
Sample ID Location Type
Sample Size Media Lot # Exp. Date
Control NA NA TSA Contact Plate 688281P 11/2026
S-01 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-02 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-03 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-04 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-05 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
S-06 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
S-07 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
S-08 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
S-09 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
S-10 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026 www.techsafety.com
09/2026Sample Date: This record was electronically signed by:
Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.
Print Date: 10/2026 Page: 1/4
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
800.877.7742
Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA
Contact: Travis Horn Fresno, CA 93703 Model: NA
Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite
Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling
20 - Surface Samples
Sample ID Location Type
Sample Size Media Lot # Exp. Date
S-11 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-12 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-13 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-14 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026
S-15 CNC 1 Plate TSA Contact Plate 688281P 11/2026
S-16 CNC 1 Plate TSA Contact Plate 688281P 11/2026
S-17 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
S-18 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
S-19 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026
Remarks: NA www.techsafety.com
09/2026Sample Date: This record was electronically signed by:
Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.
Print Date: 10/2026 Page: 2/4
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
800.877.7742
Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA
Contact: Travis Horn Fresno, CA 93703 Model: NA
Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite
Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling www.techsafety.com
09/2026Sample Date: This record was electronically signed by:
Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.
Print Date: 10/2026 Page: 3/4
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
La Jolla, CA 92037
800.877.7742
Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA
SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA
Contact: Travis Horn Fresno, CA 93703 Model: NA
Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite
Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling
The following NIST-Traceable equipment were used to collect these viable samples:
Equipment ID Manufacturer Model Serial # Calibration Due
012632 AP BUCK B30120 C103665 10/24/2026
012638 AP BUCK B30120 C103671 10/24/2026
www.techsafety.com
09/2026Sample Date: This record was electronically signed by:
Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1
Field Service Technician Print Version: Service Manager v8.0.4.3
I certify that this record is a true and accurate representation of work performed.
Print Date: 10/2026 Page: 4/4
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Test Report
General Information:
Project Site: VA Fresno
2615 East Clinton Avenue
Fresno, CA 93703
Room: Hazardous Sterile Ante Room
Test Report ID: DV26910104810-02
Project ID: NC-VFR2627401A-1
Project Contact: Travis Horn
VFR01 | VA Fresno
(559) 225-6100
Travis.Horn@va.gov
Project Lead Tech: Daniel Velazquez
823 Kifer Rd
Sunnyvale, CA 94086 tel: (800) 877-7742 fax: (510) 845-5595
Project Description: TH T3 Building 24 Pharmacy Testing Sept 2026 AL
Technician Remarks: This certificate, DV26910104810-02, is supplement to test report DV26910104810. Added test standard USP <797> 2023. Fabian Olivares
9/16/2026 12:18:20 PM
Test Standards: USP <797> 2023, CETA CAG-003, USP <800>, IEST-RP-CC006 www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 1/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02
Room Description:
Building: 24
Room: Hazardous Sterile Ante Room
Room Size:
Area (ft²): 227.26 Volume (ft³): 1,931
Area (m²): 21.11 Volume (m³): 55
Room Design:
ISO-7
Results
Inspected Min Max Measured Results
Room Airflow Volume (CFM) NA NA 1,281 FIO
Room Differential Pressure * * * Pass
Average Velocities (FPM) * * * FIO
Aerosol Concentration (ug/l) * * * Pass
Leak Test (%) * * * Pass
Filter Patching * * * Pass
Room Air Change Rate (ACPH) 30 NA 40 Pass
Airborne Nonviable Particle Count * * * Pass
* Results are aggregate of all units tested; individual unit criterion noted in the following pages.
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 2/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02
Individual Unit Measurements: Hazardous Sterile Ante Room
Terminal ID: TAF-03
Filter Area (ft²): 6.72 Kv: 1.00 Airflow Velocity (FPM): 96 Airflow Volume (CFM): 648
Inspected Min Max Measured Results
Airflow Volume (CFM) NA NA 648 FIO
Aerosol Concentration (ug/l) 10 100 29 Pass
HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass
Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass
Comments: NA
Terminal ID: TAF-04
Filter Area (ft²): 6.72 Kv: 1.00 Airflow Velocity (FPM): 94 Airflow Volume (CFM): 633
Inspected Min Max Measured Results
Airflow Volume (CFM) NA NA 633 FIO
Aerosol Concentration (ug/l) 10 100 30 Pass
HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass
Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass
Comments: NA www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 3/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02
Pressurization Data
Zone (+) Zone (-) Actual ΔP ("wc)
Min ΔP ("wc)
Max ΔP ("wc)
Directional Airflow
Door Gauge Readings ("wc) Meets CriteriaClosed ("wc) Open ("wc)
Hazardous Sterile Ante Room
Hazardous Storage Room
0.050 0.020 NA Pass NA NA Yes comments: NA
Hazardous Sterile Ante Room
Ambient Room 0.032 0.020 NA Pass NA NA Yes comments: NA www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 4/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02
Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Ante
Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).
Report Date: 09/2026 Class: ISO-7 Area: 227.26 ft² Flow Rate: 75.0 LPM
Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L
Testing is performed to satisfy ISO 14644-1:2015
Pass PPCM (um) ≥ 0.5
Loc Count 1 Count 2 Count 3 Average Result
1) 6480 NA NA 6480 Pass
2) 3746 NA NA 3746 Pass
3) 2786 NA NA 2786 Pass
4) 3693 NA NA 3693 Pass
5) 2173 NA NA 2173 Pass
6) 1453 NA NA 1453 Pass
T. C° % rH
21.2 55
21.2 56
21.2 56
21.3 57
21.4 57
21.3 57
Class Limit: 352,000 21.3 56
Sample Volume Test: Pass
Class Limit Location Test: Pass www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 5/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02
Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Ante
Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).
Report Date: 09/2026 Class: ISO-7 Area: 227.26 ft² Flow Rate: 75.0 LPM
Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L
Airborne Nonviable Particle Count Diagram:
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 6/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02
The following NIST-Traceable equipment were used to perform this test:
Equipment ID Manufacturer Model Serial # Calibration Due
006337 Evergreen Telemetry S-PVF-1 1900817C 10/2026
009016 TEC PH-4 4132 11/2026
009464 Climet CI-750t 057234 11/2026
009747 Evergreen Telemetry Wrist Reporter WR-401 2C03501 09/2026
009752 Evergreen Telemetry Three Pounder 2200368B 09/2026
012839 TEC AG-E3 4952 10/2026
013812 Govee H5075 EQ13812 10/2026 www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 7/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02
Cleanroom Diagram:
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 8/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Test Report
General Information:
Project Site: VA Fresno
2615 East Clinton Avenue
Fresno, CA 93703
Room: Hazardous Sterile Buffer Room
Test Report ID: DV26910104837-02
Project ID: NC-VFR2627401A-1
Project Contact: Travis Horn
VFR01 | VA Fresno
(559) 225-6100
Travis.Horn@va.gov
Project Lead Tech: Daniel Velazquez
823 Kifer Rd
Sunnyvale, CA 94086 tel: (800) 877-7742 fax: (510) 845-5595
Project Description: TH T3 Building 24 Pharmacy Testing Sept 2026 AL
Technician Remarks: This certificate, DV26910104837-02, is supplement to test report DV26910104837. Added test standard USP <797> 2023 & CETA CAG-003.
Fabian Olivares 9/16/2026 12:12:12 PM
Test Standards: USP <797> 2023, CETA CAG-003, USP <800>, IEST-RP-CC006 www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 1/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02
Room Description:
Building: 24
Room: Hazardous Sterile Buffer Room
Room Size:
Area (ft²): 110.00 Volume (ft³): 943
Area (m²): 10.22 Volume (m³): 27
Room Design:
ISO-7
Results
Inspected Min Max Measured Results
Room Airflow Volume (CFM) NA NA 722 FIO
Room Differential Pressure * * * Pass
Average Velocities (FPM) * * * FIO
Aerosol Concentration (ug/l) * * * Pass
Leak Test (%) * * * Pass
Filter Patching * * * Pass
Room Air Change Rate (ACPH) 30 NA 46 Pass
Airborne Nonviable Particle Count * * * Pass
* Results are aggregate of all units tested; individual unit criterion noted in the following pages.
www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 2/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02
Individual Unit Measurements: Hazardous Sterile Buffer Room
Terminal ID: TAF-01
Filter Area (ft²): 3.36 Kv: 1.00 Airflow Velocity (FPM): 116 Airflow Volume (CFM): 391
Inspected Min Max Measured Results
Airflow Volume (CFM) NA NA 391 FIO
Aerosol Concentration (ug/l) 10 100 48 Pass
HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass
Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass
Comments: NA
Terminal ID: TAF-02
Filter Area (ft²): 3.36 Kv: 1.00 Airflow Velocity (FPM): 99 Airflow Volume (CFM): 331
Inspected Min Max Measured Results
Airflow Volume (CFM) NA NA 331 FIO
Aerosol Concentration (ug/l) 10 100 57 Pass
HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass
Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass
Comments: NA www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 3/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02
Pressurization Data
Zone (+) Zone (-) Actual ΔP ("wc)
Min ΔP ("wc)
Max ΔP ("wc)
Directional Airflow
Door Gauge Readings ("wc) Meets CriteriaClosed ("wc) Open ("wc)
Hazardous Sterile Ante Room
Hazardous Sterile Buffer Room
0.024 0.010 0.030 Pass 0.021 0.000 Yes comments: NA www.techsafety.com
Report Date: 09/10/2026
This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1
Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3
Field Service TechnicianPrint Date: 10/13/2026 Page: 4/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable
Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026
Bldg 24 Pharmacy USP Testing Appendix
(800) 877-7742
Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room
Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02
Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Buffer
Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).
Report Date: 09/2026 Class: ISO-7 Area: 110.00 ft² Flow Rate: 75.0 LPM
Ret…