Attachment 3 - VFR2627401A-1.10-11 Sep 2026.Bldg 24 Pharmacy USP Testing.Appendix.pdf

PDF 2 MB Posted

Attached to
VACCHCS Pharmacy Certification
Solicitation number
36C26127Q0099
Issued by

About this file

This file is an Appendix containing USP (United States Pharmacopeia) pharmaceutical compounding facility testing and certification reports for Building 24 Pharmacy at VA Fresno, dated September 10-11, 2026. The document compiles comprehensive testing results for the Service Order NC-VFR2627401A-1, conducted by TechSafety at the VA Fresno facility located at 2615 East Clinton Avenue, Fresno, California 93703, with contact Travis Horn.

The appendix documents testing of multiple hazardous and non-hazardous sterile pharmacy compounding areas, including two Class II, Type A2 Biological Safety Cabinets (Model NU-540-400), two unidirectional flow clean benches (Model NU-240-430), and associated support rooms. Testing encompassed biological safety cabinet certifications with inflow velocity measurements (103–105 FPM), HEPA filter integrity assessments (all passing with 0.000% max point leaks), airborne nonviable particle counts confirming ISO-5 and ISO-7 classifications, airflow visualization studies confirming unidirectional flow patterns, viable environmental sampling across multiple locations, and room-level pressurization and air change rate documentation. The facility includes a Hazardous Sterile Ante Room (ISO-7, 40 ACPH), Hazardous Sterile Buffer Room (ISO-7, 46 ACPH), Hazardous Storage Room (15 ACPH), Non-Hazardous Sterile Buffer Room (ISO-7, 180 ACPH combined with PEC), and an ancillary sterile compounding suite. All tests employed NIST-traceable equipment and adhered to USP, ANSI/NSF 49, CETA CAG-003, ISO 14644-1:2015, IEST-RP-CC standards, with retest dates scheduled for March 2027.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Biological Safety Cabinet Certification Report

4225 Executive Square, Suite 370

La Jolla, CA 92037

(800) 877-7742

Cert.: DV2691010465-02 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Test Standard(s): MANUFACTURER, OSHA, ANSI/NSF 49, USP <797>, CETA CAG-003

Inspected Min Max Measured Result

Calculated Inflow Velocity (FPM) 100 110 105 Pass

Average Supply Velocity Zone1 (FPM) 55 65 60 Pass

Average Supply Velocity Zone2 (FPM) NA NA NA NA

Average Supply Velocity Zone3 (FPM) NA NA NA NA

Average Supply Velocity Zone4 (FPM) NA NA NA NA

Supply Uniformity NA NA Pass Pass

Downflow Smoke Pattern Test NA NA Pass Pass

View Screen Retention Smoke Pattern Test NA NA Pass Pass

Work Opening Edge Retention Smoke Pattern Test NA NA Pass Pass

Sash Seal Smoke Pattern Test NA NA Pass Pass

Site Assessment NA NA Pass Pass

At-Rest Unidirectional Airflow Visualization Test NA NA NA NA

Dynamic Unidirectional Airflow Visualization Test NA NA Pass Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 23 Pass

Exhaust Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 23 Pass

Comments: This certificate, DV2691010465-02, is supplement to test report DV2691010465. Added technician comments.

Fabian Olivares 9/16/2026 12:30:40 PM The space between the face of the BSC and permanent fixture directly opposite the access opening was less than 42 inches, used manufacturers alternative method to measure inflow readings.

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:33 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

Ver. 15 Oct 2026

La Jolla, CA 92037

(800) 877-7742

Cert.: DV2691010465-02 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Inflow (FPM) Method: MFR spec: I4 - Constricted Access Calculated Inflow Velocity (FPM): 105

Workface: Width: 46.50" Height: 10.00" Workface Area (ft²): 3.23

Adj. Workface: Width: 46.50" Height: 3.00" Adj. Workface Area (ft²): 0.97

AK Factor: 1.00 Inflow Factor: 0 Volume (CFM): 340

Readings:

344 347 352 365 349 351 345 354

Supply (FPM) Method: MFR spec: S1 - Unidirectional (non-zoned) flow Average Supply Velocity(FPM): 60

Dimensions: Width: 46.50" Depth: 25.00" Supply Area (ft²): 8.07

Readings:

Supply Volume (CFM): 484

66 64 66 64 65 64 63

61 60 60 59 58 57 57

60 58 58 53 55 55 58

Site Assessment:

Exhaust System: Canopy: Negative: The Canopy Passed the performance test.

Exhaust System Alarm: Pass: The Exhaust System Alarm Passed the performance test.

Sash Alarm: Pass: The Sash Alarm Passed the performance test.

Interlock: NA: No Interlock Alarm present.

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:33 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert. No.: DV269109821 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199716071620

Airborne Nonviable Particle Count Test Report Clean Zone: 199716071620 Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 8.07 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.1 54

Class Limit: 3,520 21.1 54

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:10 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Cert. No.: DV269109821 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199716071620

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

013812 Govee H5075 EQ13812 10/2026

009464 Climet CI-750t 057234 11/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:10 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

Airflow Visualization Study

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910973 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910973 Facility: VA Fresno ID: 5505820 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199716071620Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104620-02 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Test Standard(s): MANUFACTURER, OSHA, ANSI/NSF 49, USP <797>, CETA CAG-003

Inspected Min Max Measured Result

Calculated Inflow Velocity (FPM) 100 110 103 Pass

Average Supply Velocity Zone1 (FPM) 55 65 64 Pass

Average Supply Velocity Zone2 (FPM) NA NA NA NA

Average Supply Velocity Zone3 (FPM) NA NA NA NA

Average Supply Velocity Zone4 (FPM) NA NA NA NA

Supply Uniformity NA NA Pass Pass

Downflow Smoke Pattern Test NA NA Pass Pass

View Screen Retention Smoke Pattern Test NA NA Pass Pass

Work Opening Edge Retention Smoke Pattern Test NA NA Pass Pass

Sash Seal Smoke Pattern Test NA NA Pass Pass

Site Assessment NA NA Pass Pass

At-Rest Unidirectional Airflow Visualization Test NA NA NA NA

Dynamic Unidirectional Airflow Visualization Test NA NA Pass Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Exhaust Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 22 Pass

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 22 Pass

Comments: This certificate, DV26910104620-02, is supplement to test report DV26910104620. Added technician comments.

Fabian Olivares 9/16/2026 12:25:49 PM The space between the face of the BSC and permanent fixture directly opposite the access opening was less than 42 inches, used manufacturers alternative method to measure inflow readings.

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:36 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104620-02 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Inflow (FPM) Method: MFR spec: I4 - Constricted Access Calculated Inflow Velocity (FPM): 103

Workface: Width: 46.50" Height: 10.00" Workface Area (ft²): 3.23

Adj. Workface: Width: 46.50" Height: 3.00" Adj. Workface Area (ft²): 0.97

AK Factor: 1.00 Inflow Factor: 0 Volume (CFM): 332

Readings:

351 348 344 337 331 336 342 347

Supply (FPM) Method: MFR spec: S1 - Unidirectional (non-zoned) flow Average Supply Velocity(FPM): 64

Dimensions: Width: 46.50" Depth: 25.00" Supply Area (ft²): 8.07

Readings:

Supply Volume (CFM): 516

66 67 67 64 64 65 66

67 63 63 63 64 62 65

63 61 63 62 61 63 64

Site Assessment:

Exhaust System: Canopy: Negative: The Canopy Passed the performance test.

Exhaust System Alarm: Pass: The Exhaust System Alarm Passed the performance test.

Sash Alarm: Pass: The Sash Alarm Passed the performance test.

Interlock: NA: No Interlock Alarm present.

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:36 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Cert. No.: DV269109845 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199733071720

Airborne Nonviable Particle Count Test Report Clean Zone: 199733071720 Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 8.07 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.0 55

Class Limit: 3,520 21.0 55

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:24 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Cert. No.: DV269109845 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-540-400

Address: 2615 East Clinton Avenue Room: HD Sterile Buffer Room Unit Type: A2

Fresno, CA 93703 Contact: Travis Horn SN: 199733071720

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

009464 Climet CI-750t 057234 11/2026

013812 Govee H5075 EQ13812 10/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:38:24 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109643 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:06 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109643 Facility: VA Fresno ID: 5505819 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-540-400

Phone: (559) 225-6100 Ext.5735 24 SN: 199733071720Bld:

Email: Travis.Horn@va.gov HD Sterile Buffer Room Class: II Type: A2Rm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:06 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

Unidirectional Flow Device Certification Report

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104641 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Test Standard(s): USP <797>, CETA CAG-003, IEST-RP-CC002.4, USP <797> 2023, MANUFACTURER

Results

Inspected Min Max Measured Results

Outflow Velocity (FPM) 80 100 89 Pass

%RSD NA 15 5 Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Backstreaming Test NA NA Pass Pass

Induction Leak Test NA NA Pass Pass

Smoke Source Glycerin and/or Glycol

Fluorescent Lighting (FC) NA NA NA NA

NAUltraviolet Lighting (μW/cm²) NA NA NA

NANoise Level (dBA) NA NA NA

NAVibration (in. RMS) NA NA NA

NAElectrical Current Leakage Test (mA) NA NA NA

NAElectrical Ground Resistance Test (Ohms) NA NA NA

NAElectrical Polarity NA NA NA

NAElectrical GCFI NA NA NA

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 20 Pass

Comments:

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:40 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104641 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Velocity (FPM) Dimension: Width: 48.00" Height: 32.00" Average Velocity (FPM) 89

Area (ft²): 10.67

Readings: Volume (CFM): 950

79 84 87 89

88 90 86 89

95 96 92 88

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:40 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Cert. No.: DV26910103727 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199746071720

Airborne Nonviable Particle Count Test Report Clean Zone: 199746071720 Status: Operational/Dynamic Cust ID: Hood #2 Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 10.67 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.1 55

Class Limit: 3,520 21.1 55

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:37:53 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Cert. No.: DV26910103727 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199746071720

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

013812 Govee H5075 EQ13812 10/2026

009464 Climet CI-750t 057234 11/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:37:53 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109753 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:08 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109753 Facility: VA Fresno ID: 5505818 Cust ID: Hood #2

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199746071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:08 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104632 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Test Standard(s): IEST-RP-CC002.4, CETA CAG-003, USP <797> 2023, MANUFACTURER, USP <797>

Results

Inspected Min Max Measured Results

Outflow Velocity (FPM) 80 100 94 Pass

%RSD NA 15 8 Pass

Supply Delta Pressure ("W.C.) NA NA NA NA

Duct Pressure ("W.C.) NA NA NA NA

Backstreaming Test NA NA Pass Pass

Induction Leak Test NA NA Pass Pass

Smoke Source Glycerin and/or Glycol

Fluorescent Lighting (FC) NA NA NA NA

NAUltraviolet Lighting (μW/cm²) NA NA NA

NANoise Level (dBA) NA NA NA

NAVibration (in. RMS) NA NA NA

NAElectrical Current Leakage Test (mA) NA NA NA

NAElectrical Ground Resistance Test (Ohms) NA NA NA

NAElectrical Polarity NA NA NA

NAElectrical GCFI NA NA NA

Supply Min Max Measured Results

HEPA As Found Max Point Leak (%) 0.000 0.010 0.000 Pass

HEPA As Left Max Point Leak (%) 0.000 0.010 NA NA

Patch (%) 0 3 Pre:0.00/New:0.00/Tot:0.00 Pass

Aerosol Concentration (ug/l) 10 100 19 Pass

Comments:

**** Unit Certified **** The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial No. Cal. Due Date

005672 TSI 9535 T95351923010 05/2027

009016 TEC PH-4 4132 11/2026

012839 TEC AG-E3 4952 10/2026

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:38 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV26910104632 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Velocity (FPM) Dimension: Width: 48.00" Height: 32.00" Average Velocity (FPM) 94

Area (ft²): 10.67

Readings: Volume (CFM): 1,003

90 90 83 88

90 97 103 97

87 92 101 106

Report Date: 09/2026 www.techsafety.com

This record was electronically signed by:Retest Date: 03/2027 eData: v8.0.3.1 Daniel Velazquez 09/10/2026 12:38 PM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Cert. No.: DV26910103620 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199751071720

Airborne Nonviable Particle Count Test Report Clean Zone: 199751071720 Status: Operational/Dynamic Cust ID: Hood #1 Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-5 Area: 10.67 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 0 NA NA 0 Pass

T. C° % rH

21.0 56

Class Limit: 3,520 21.0 56

Sample Volume Test: Pass

Class Limit Location Test: Pass

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

This record was electronically signed by:

Page: 1/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:36:52 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Cert. No.: DV26910103620 Svc. Order: NC-VFR2627401A-1 Make: NUAIRE

Facility: VA Fresno Building: 24 Model: NU-240-430

Address: 2615 East Clinton Avenue Room: Non-HD Sterile Buffer Room Unit Type: Clean Bench

Fresno, CA 93703 Contact: Travis Horn SN: 199751071720

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

013812 Govee H5075 EQ13812 10/2026

009464 Climet CI-750t 057234 11/2026 www.techsafety.com

This record was electronically signed by:

Page: 2/2 web: www.techsafety.com Daniel Velazquez 9/10/2026 10:36:52 AM eData v8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.Print Date: 10/13/2026

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109722 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

Service Type: Airflow Visualization Study

Comments: (Dynamic) Tracer moves in a unidirectional manner without reflux. Actions of the personnel do not affect the general trajectory of tracer.

Doc. Title Doc. # Doc. Version Completed Date Originally Retained By

Diagram NA NA 09/10/2026 Daniel Velazquez www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 1/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

(800) 877-7742

Cert.: DV269109722 Facility: VA Fresno ID: 5505817 Cust ID: Hood #1

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NUAIRE

Contact: Travis Horn Fresno, CA 93703 Model: NU-240-430

Phone: (559) 225-6100 Ext.5735 24 SN: 199751071720Bld:

Email: Travis.Horn@va.gov Non-HD Sterile Buffer Room Class: NA Type: Clean BenchRm:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: NA eData v8.0.3.1 Daniel Velazquez 09/10/2026 09:07 AM

Print Version: Service Manager v8.0.4.3 Field Service Technician

Print Date: 10/13/2026 Page: 2/2 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

Viable Sample Report

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling

Area Status: Operational/Dynamic

Occupants: TSS: 1 Client: 2

15 - Active Air Samples

Sample ID Location Type

Sample Size Media Lot # Exp. Date

A-01 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-02 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-03 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-04 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-05 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-06 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-07 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-08 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-09 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-10 ISO-7 1000 Liters TSA Settling Plate 691307P 01/2027

A-11 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-12 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-13 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

A-14 ISO-5 1000 Liters TSA Settling Plate 691307P 01/2027

Control NA NA TSA Settling Plate 691307P 01/2027

20 - Surface Samples

Sample ID Location Type

Sample Size Media Lot # Exp. Date

Control NA NA TSA Contact Plate 688281P 11/2026

S-01 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-02 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-03 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-04 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-05 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-06 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-07 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-08 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-09 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-10 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026 www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.

Print Date: 10/2026 Page: 1/4

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling

20 - Surface Samples

Sample ID Location Type

Sample Size Media Lot # Exp. Date

S-11 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-12 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-13 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-14 ISO-5 1 Plate TSA Contact Plate 688281P 11/2026

S-15 CNC 1 Plate TSA Contact Plate 688281P 11/2026

S-16 CNC 1 Plate TSA Contact Plate 688281P 11/2026

S-17 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-18 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

S-19 ISO-7 1 Plate TSA Contact Plate 688281P 11/2026

Remarks: NA www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.

Print Date: 10/2026 Page: 2/4

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.

Print Date: 10/2026 Page: 3/4

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

La Jolla, CA 92037

800.877.7742

Cert.: DV26910964 Facility: VA Fresno ID: 5507045 Cust ID: NA

SO: NC-VFR2627401A-1 Addr: 2615 East Clinton Avenue Make: NA

Contact: Travis Horn Fresno, CA 93703 Model: NA

Phone: (559) 225-6100 Ext.5735 Bld: 24 SN: Building 24 Sterile Compounding Suite

Email: Travis.Horn@va.gov Rm: Building 24 Sterile Compoundi Class: NA Type: Environmental Sampling

The following NIST-Traceable equipment were used to collect these viable samples:

Equipment ID Manufacturer Model Serial # Calibration Due

012632 AP BUCK B30120 C103665 10/24/2026

012638 AP BUCK B30120 C103671 10/24/2026

www.techsafety.com

09/2026Sample Date: This record was electronically signed by:

Daniel Velazquez 09/10/2026 10:35 AM eData: v 8.0.3.1

Field Service Technician Print Version: Service Manager v8.0.4.3

I certify that this record is a true and accurate representation of work performed.

Print Date: 10/2026 Page: 4/4

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Test Report

General Information:

Project Site: VA Fresno

2615 East Clinton Avenue

Fresno, CA 93703

Room: Hazardous Sterile Ante Room

Test Report ID: DV26910104810-02

Project ID: NC-VFR2627401A-1

Project Contact: Travis Horn

VFR01 | VA Fresno

(559) 225-6100

Travis.Horn@va.gov

Project Lead Tech: Daniel Velazquez

823 Kifer Rd

Sunnyvale, CA 94086 tel: (800) 877-7742 fax: (510) 845-5595

Project Description: TH T3 Building 24 Pharmacy Testing Sept 2026 AL

Technician Remarks: This certificate, DV26910104810-02, is supplement to test report DV26910104810. Added test standard USP <797> 2023. Fabian Olivares

9/16/2026 12:18:20 PM

Test Standards: USP <797> 2023, CETA CAG-003, USP <800>, IEST-RP-CC006 www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 1/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02

Room Description:

Building: 24

Room: Hazardous Sterile Ante Room

Room Size:

Area (ft²): 227.26 Volume (ft³): 1,931

Area (m²): 21.11 Volume (m³): 55

Room Design:

ISO-7

Results

Inspected Min Max Measured Results

Room Airflow Volume (CFM) NA NA 1,281 FIO

Room Differential Pressure * * * Pass

Average Velocities (FPM) * * * FIO

Aerosol Concentration (ug/l) * * * Pass

Leak Test (%) * * * Pass

Filter Patching * * * Pass

Room Air Change Rate (ACPH) 30 NA 40 Pass

Airborne Nonviable Particle Count * * * Pass

* Results are aggregate of all units tested; individual unit criterion noted in the following pages.

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 2/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02

Individual Unit Measurements: Hazardous Sterile Ante Room

Terminal ID: TAF-03

Filter Area (ft²): 6.72 Kv: 1.00 Airflow Velocity (FPM): 96 Airflow Volume (CFM): 648

Inspected Min Max Measured Results

Airflow Volume (CFM) NA NA 648 FIO

Aerosol Concentration (ug/l) 10 100 29 Pass

HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass

Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass

Comments: NA

Terminal ID: TAF-04

Filter Area (ft²): 6.72 Kv: 1.00 Airflow Velocity (FPM): 94 Airflow Volume (CFM): 633

Inspected Min Max Measured Results

Airflow Volume (CFM) NA NA 633 FIO

Aerosol Concentration (ug/l) 10 100 30 Pass

HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass

Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass

Comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 3/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02

Pressurization Data

Zone (+) Zone (-) Actual ΔP ("wc)

Min ΔP ("wc)

Max ΔP ("wc)

Directional Airflow

Door Gauge Readings ("wc) Meets CriteriaClosed ("wc) Open ("wc)

Hazardous Sterile Ante Room

Hazardous Storage Room

0.050 0.020 NA Pass NA NA Yes comments: NA

Hazardous Sterile Ante Room

Ambient Room 0.032 0.020 NA Pass NA NA Yes comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 4/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02

Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Ante

Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-7 Area: 227.26 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Testing is performed to satisfy ISO 14644-1:2015

Pass PPCM (um) ≥ 0.5

Loc Count 1 Count 2 Count 3 Average Result

1) 6480 NA NA 6480 Pass

2) 3746 NA NA 3746 Pass

3) 2786 NA NA 2786 Pass

4) 3693 NA NA 3693 Pass

5) 2173 NA NA 2173 Pass

6) 1453 NA NA 1453 Pass

T. C° % rH

21.2 55

21.2 56

21.2 56

21.3 57

21.4 57

21.3 57

Class Limit: 352,000 21.3 56

Sample Volume Test: Pass

Class Limit Location Test: Pass www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 5/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02

Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Ante

Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-7 Area: 227.26 ft² Flow Rate: 75.0 LPM

Retest Date: 03/2027 Occupants: 1 Total Sample Vol.: 75.0 L

Airborne Nonviable Particle Count Diagram:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 6/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02

The following NIST-Traceable equipment were used to perform this test:

Equipment ID Manufacturer Model Serial # Calibration Due

006337 Evergreen Telemetry S-PVF-1 1900817C 10/2026

009016 TEC PH-4 4132 11/2026

009464 Climet CI-750t 057234 11/2026

009747 Evergreen Telemetry Wrist Reporter WR-401 2C03501 09/2026

009752 Evergreen Telemetry Three Pounder 2200368B 09/2026

012839 TEC AG-E3 4952 10/2026

013812 Govee H5075 EQ13812 10/2026 www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 7/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Ante Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104810-02

Cleanroom Diagram:

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:57 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 8/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Test Report

General Information:

Project Site: VA Fresno

2615 East Clinton Avenue

Fresno, CA 93703

Room: Hazardous Sterile Buffer Room

Test Report ID: DV26910104837-02

Project ID: NC-VFR2627401A-1

Project Contact: Travis Horn

VFR01 | VA Fresno

(559) 225-6100

Travis.Horn@va.gov

Project Lead Tech: Daniel Velazquez

823 Kifer Rd

Sunnyvale, CA 94086 tel: (800) 877-7742 fax: (510) 845-5595

Project Description: TH T3 Building 24 Pharmacy Testing Sept 2026 AL

Technician Remarks: This certificate, DV26910104837-02, is supplement to test report DV26910104837. Added test standard USP <797> 2023 & CETA CAG-003.

Fabian Olivares 9/16/2026 12:12:12 PM

Test Standards: USP <797> 2023, CETA CAG-003, USP <800>, IEST-RP-CC006 www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 1/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02

Room Description:

Building: 24

Room: Hazardous Sterile Buffer Room

Room Size:

Area (ft²): 110.00 Volume (ft³): 943

Area (m²): 10.22 Volume (m³): 27

Room Design:

ISO-7

Results

Inspected Min Max Measured Results

Room Airflow Volume (CFM) NA NA 722 FIO

Room Differential Pressure * * * Pass

Average Velocities (FPM) * * * FIO

Aerosol Concentration (ug/l) * * * Pass

Leak Test (%) * * * Pass

Filter Patching * * * Pass

Room Air Change Rate (ACPH) 30 NA 46 Pass

Airborne Nonviable Particle Count * * * Pass

* Results are aggregate of all units tested; individual unit criterion noted in the following pages.

www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 2/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02

Individual Unit Measurements: Hazardous Sterile Buffer Room

Terminal ID: TAF-01

Filter Area (ft²): 3.36 Kv: 1.00 Airflow Velocity (FPM): 116 Airflow Volume (CFM): 391

Inspected Min Max Measured Results

Airflow Volume (CFM) NA NA 391 FIO

Aerosol Concentration (ug/l) 10 100 48 Pass

HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass

Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass

Comments: NA

Terminal ID: TAF-02

Filter Area (ft²): 3.36 Kv: 1.00 Airflow Velocity (FPM): 99 Airflow Volume (CFM): 331

Inspected Min Max Measured Results

Airflow Volume (CFM) NA NA 331 FIO

Aerosol Concentration (ug/l) 10 100 57 Pass

HEPA Leak Concentration (%) NA 0.010 As Found:0.000/As Left:NA Pass

Patch (%) 0.00 3.00 Pre:0.00/New:0.00/Tot:0.00 Pass

Comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 3/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02

Pressurization Data

Zone (+) Zone (-) Actual ΔP ("wc)

Min ΔP ("wc)

Max ΔP ("wc)

Directional Airflow

Door Gauge Readings ("wc) Meets CriteriaClosed ("wc) Open ("wc)

Hazardous Sterile Ante Room

Hazardous Sterile Buffer Room

0.024 0.010 0.030 Pass 0.021 0.000 Yes comments: NA www.techsafety.com

Report Date: 09/10/2026

This record was electronically signed by:Retest Date: 03/2027 eData: v 8.0.3.1

Daniel Velazquez 09/10/2026 12:29 PM Print Version: Service Manager v8.0.4.3

Field Service TechnicianPrint Date: 10/13/2026 Page: 4/8 I certify that this record is a true and accurate representation of work performed.FIO: For Information Only NA: Not Applicable

Project ID: NC-VFR2627401A-1 Test Dates: 10-11 Sep 2026

Bldg 24 Pharmacy USP Testing Appendix

(800) 877-7742

Client: VA Fresno Test Date(s): 09/2026 Room: Hazardous Sterile Buffer Room

Proj. ID: NC-VFR2627401A-1 Technician: Daniel Velazquez Report ID: DV26910104837-02

Airborne Nonviable Particle Count Test Report Clean Zone: Hazardous Sterile Buffer

Room Status: Operational/Dynamic Cust ID: NA Duration: 1.0 Min(s).

Report Date: 09/2026 Class: ISO-7 Area: 110.00 ft² Flow Rate: 75.0 LPM

Ret…