36C25819B0023-002.pdf
PDF 150 KB Posted
- Attached to
- 36C258-19-AP-0947 / 605-19-3-6012-0126 - 605-19-416 ICU 2SE Bridge Completion (VA-19-00054195)
- Solicitation number
- 36C25819B0023
About this file
36C25819B0023 S02 Bridge 1 120 V 20 Amp pwr Med Gas Panel r.pdf
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C25819B0023-0003002.pdf | ||
| 36C25819B0023-0003003.pdf | ||
| 36C25819B0023-0003000.docx | DOCX document | |
| 36C25819B0023-0003001.pdf | ||
| 36C25819B0023-0002011.pdf | ||
| 36C25819B0023-0002012.pdf | ||
| 36C25819B0023-0002010.docx | DOCX document | |
| 36C25819B0023-0002013.pdf | ||
| 36C25819B0023-0001005.pdf | ||
| 36C25819B0023-0001004.pdf | ||
| 36C25819B0023-0001003.pdf | ||
| 36C25819B0023-0001002.pdf | ||
| 36C25819B0023-0001001.pdf | ||
| 36C25819B0023-0001000.docx | DOCX document | |
| 36C25819B0023-001.docx | DOCX document | |
| 36C25819B0023-011.pdf | ||
| 36C25819B0023-007.pdf | ||
| 36C25819B0023-006.pdf | ||
| 36C25819B0023-003.pdf | ||
| 36C25819B0023-010.pdf | ||
| 36C25819B0023-004.pdf | ||
| 36C25819B0023-008.pdf | ||
| 36C25819B0023-005.pdf | ||
| 36C25819B0023-009.pdf | ||
| 36C25819B0023-000.docx | DOCX document |
Show all 25
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
R.F.I.
(REQUEST FOR INFORMATION)
Project #: 605-331 RFI #: C – 122
Project Name: Intensive Care Unit Consolidation Date Requested: 4/24/17
Solicitation No.: VA262-15-J-6178 Reference: Med gas Alarm panels
Drawing:
Drawing PP101
Specification Section:
DESCRIPTION OF PROBLEM OR INFORMATION REQUESTED
ISSUE:
1- Electrical drawing doesn't show any power for Med gas Alarm panels, as per the cut sheet is required 115 volt or 230 VAC primary power .Please specified the circuit will use for that.
2- This is not enough space between studs in are showing on the plan to be install.
3- This is no height for the Alarm panels
PROPOSED SOLUTION:
For number 2 Stacked or side by side when relocate to column #13.
OWNER RESPONSE
OWNER TRACKING No.: AMEND No.:
VA Project Engineer/Manager: Date:
Please provide circuit 2EX2-13 for alarm panel, proposed solution for number 2 is acceptable, Nasser Nejati @IMEG-TTG 4/24/17
Panels can be mounted between 42" to 48" max. a.f.f.
KAL, 4/25/2017
Description of Work
POWER FOR - MED GAS PANEL
As directed by RFI# C-122 provide 120 volt 20 amp power for Med Gas panel.
Furnish and install conduit, supports, wire and back box and connection to panel.
Using circuit 2EX2-13