BRAND NAME OR EQUAL EQUIPMENT

SALIENT CHARACTERISTICS

Equipment Configuration needed for following ORs: one light to be installed in each Operating Room (1, 2, 4, and 8), 4 lights total needed in this configuration.

Item: F Generation LED Surgical Light

Vendor: Stryker/Berchtold

Model: CHROMOPHARE F 628

Compatibility in ORs with existing equipment mount, with no additional modifications to in-room (below ceiling) structure besides addition of light.

Features: Low voltage direct drive DC, controlled by 12- 360 degree light head rotation, 104 LED configuration, 4 built in color option settings, built in guide light, built in wireless camera technology, optimum laminar flow compliance, ingress protection from dust and moisture, optional integrated ambient light, optional intuitive wall control, surgical lens material (surgical grade medical glass) on light cover.

LED units and light beams: Incorporate a minimum of 104 LEDs with at least 650 overlapping beams of light to minimize in-surgery adjustment

Light color temperature: 4 color options on light, including 3600K, 4000k, 4500k and 5000k

Guide light: Unit must allow clinical team to dim main LED surgical lights for minimally invasive cases, ambient light should be green

Wireless camera readiness: Unit must feature design readiness for implementing wireless 1080i resolution cameras at the light head

Required rotation: 360 degree rotation at the surgical light head Laminar flow around light: Rated at SWKI Class 5 for laminar flow compliance Materials: Surgical grade medical glass and aluminum casing Dust/Moisture Intrusion: Ingress Protection from dust & water of IP 53 standard Mounting plate from ceiling to arm: Distance must be less than 12 inches Arm length: 900mm Main mount tube Length: 230mm Focus ability: Focusable beam range no less than 7 inches – no greater than 12 inches when light is 1 meter from patient plane

Light intensity: Light must maintain brightness from at least 12 inches to 60 inches from the patient plane.

Power draw: No more than 2 Amps at 120 VAC

Ceiling mounting: Must be able to be mounted concurrently with existing below-ceiling infrastructure in the 4 ORs (Stryker/Berchtold ceiling mount)

Control for unit: Wall, with keypad option Electronics in circuit box on Main Mount: No.

Arm joint: Cardanic Classic

Life Cycle Support Characteristics:

• Design/utilization for operating room service

• Demonstrated reliability in clean environment

• Demonstrated maintainability, including parts/ service support, training, technical documentation

Equipment Configuration needed for following ORs: one light and one Video-Power Arm (VPA) needs to be installed in each Operating Room (3, 5, 6, and 7), 4 lights and 4 VPAs total needed in this configuration.

Item: F Generation LED Surgical Light

Vendor: Stryker/Berchtold

Model: CHROMOPHARE F 628

Features: Low voltage direct drive DC, controlled by 12- 360 degree light head rotation, 104 LED configuration, 4 built in color option settings, built in guide light, built in wireless camera technology, optimum laminar flow compliance, ingress protection from dust and moisture, optional integrated ambient light, optional intuitive wall control, surgical lens material (surgical grade medical glass) on light cover.

LED units and light beams: Incorporate a minimum of 104 LEDs with at least 650 overlapping beams of light to minimize in-surgery adjustment

Light color temperature: 4 color options on light, including 3600K, 4000k, 4500k and 5000k

Guide light: Unit must allow clinical team to dim main LED surgical lights for minimally invasive cases, ambient light should be green

Wireless camera readiness: Unit must feature design readiness for implementing wireless 1080i resolution cameras at the light head

Required rotation: 360 degree rotation at the surgical light head Laminar flow around light: Rated at SWKI Class 5 for laminar flow compliance Materials: Surgical grade medical glass and aluminum casing Dust/Moisture Intrusion: Ingress Protection from dust & water of IP 53 standard Mounting plate from ceiling to arm: Distance must be less than 12 inches Arm length: 900mm Main mount tube Length: 230mm Focus ability: Focusable beam range no less than 7 inches – no greater than 12 inches when the light is 1 meter from patient plane

Light intensity: Light must maintain brightness from at least 12 inches to 60 inches from the patient plane.

Power draw: No more than 2 Amps at 120 VAC Control for unit: Wall, with keypad option Electronics in circuit box on Main Mount: Yes.

Video Power Arm Characteristics:

Arm Length: 1100mm Number of Circuits: 2 Hospital power: Isolated Video bundle compatibility: Stryker equipment/instruments

Life Cycle Support Characteristics:

• Design/utilization for operating room service

• Demonstrated reliability in clean environment

Equipment Configuration needed for following ORs: one light and one equipment boom needs to be installed in each Operating Room (3, 5, 6, and 7), 4 lights and 4 equipment booms total needed in this configuration.

Item: F Generation LED Surgical Light, Teletom Equipment boom

Vendor: Stryker/Berchtold

Model: CHROMOPHARE F 628, TP-612-C/TC F628

F Generation LED Surgical Light:

Features: Low voltage direct drive DC, controlled by 12- 360 degree light head rotation, 104 LED configuration, 4 built in color option settings, built in guide light, built in wireless camera technology, optimum laminar flow compliance, ingress protection from dust and moisture, optional integrated ambient light, optional intuitive wall control, surgical lens material (surgical grade medical glass) on light cover.

LED units and light beams: Incorporate a minimum of 104 LEDs with at least 650 overlapping beams of light to minimize in-surgery adjustment

Light color temperature: 4 color options on light, including 3600K, 4000k, 4500k and 5000k

Guide light: Unit must allow clinical team to dim main LED surgical lights for minimally invasive cases, ambient light should be green

Wireless camera readiness: Unit must feature design readiness for implementing wireless 1080i resolution cameras at the light head

Required rotation: 360 degree rotation at the surgical light head Laminar flow around light: Rated at SWKI Class 5 for laminar flow compliance Materials: Surgical grade medical glass and aluminum casing Dust/Moisture Intrusion: Ingress Protection from dust & water of IP 53 standard Arm length: 900mm Focus ability: Focusable beam range no less than 7 inches – no greater than 12 inches when the light is 1 meter from the patient plane

Light intensity: Light must maintain brightness from at least 12 inches to 60 inches from the patient plane.

Power draw: No more than 2 Amps at 120 VAC

Teletom Boom Characteristics:

Features: All-electric brake system with vertical articulation, built in “Emergency Stop” button located on the Boom, vertical articulation capability throughout unit, combined suspension (Single Spindle) non-tandem option

Operating Range: 360 degree customization of services on the boom

Power Requirements on Boom: KEZR and KEXS Isolated Power per boom, 1047 UL listed

Extension pole: Top and bottom extensions Upper Arm Length Required: 1300mm Mid Tube Length Required: 150mm Column Tube length required: 250mm Weight load Capacity: At least 295 pounds (greater than is preferable) Brake system: All-Electric Front of boom panel: Connections to N2O, Medical Air, WAGD, O2 Right of boom panel: Emergency stop Rear of boom panel: 3 circuits (electric)

Note: Hoses for N2O, Medical Air, WAGD, and Oxygen must not kink in the boom arms and must accommodate existing above ceiling hookups without replumbing the hookups. Diameter of tubing must accommodate from the old hookups to new equipment.

Life Cycle Support Characteristics:

• Design/utilization for operating room service

• Demonstrated reliability in clean environment

Equipment Configuration needed for following ORs: one light and one equipment boom needs to be installed in each Operating Room (3, 5, 6, and 7), 4 lights and 4 equipment booms total needed in this configuration.

Item: F Generation LED Surgical Light, Teletom Equipment boom

Vendor: Stryker/Berchtold

Model: CHROMOPHARE F 628, TP-632-C/TC F628

F Generation LED Surgical Light:

Features: Low voltage direct drive DC, controlled by 12- 360 degree light head rotation, 104 LED configuration, 4 built in color option settings, built in guide light, built in wireless camera technology, optimum laminar flow compliance, ingress protection from dust and moisture, optional integrated ambient light, optional intuitive wall control, surgical lens material (surgical grade medical glass) on light cover.

LED units and light beams: Incorporate a minimum of 104 LEDs with at least 650 overlapping beams of light to minimize in-surgery adjustment

Light color temperature: 4 color options on light, including 3600K, 4000k, 4500k and 5000k

Guide light: Unit must allow clinical team to dim main LED surgical lights for minimally invasive cases, ambient light should be green

Wireless camera readiness: Unit must feature design readiness for implementing wireless 1080i resolution cameras at the light head

Required rotation: 360 degree rotation at the surgical light head Laminar flow around light: Rated at SWKI Class 5 for laminar flow compliance Materials: Surgical grade medical glass and aluminum casing Dust/Moisture Intrusion: Ingress Protection from dust & water of IP 53 standard Arm length: 900mm Focus ability: Focusable beam range no less than 7 inches – no greater than 12 inches when the light is 1 meter from patient plane

Light intensity: Light must maintain brightness from at least 12 inches to 60 inches from the patient plane.

Power draw: No more than 2 Amps at 120 VAC

Teletom Boom Characteristics:

Features: All-electric brake system with vertical articulation, built in “Emergency Stop” button located on the Boom, vertical articulation capability throughout unit, combined suspension (Single Spindle) non-tandem option

Operating Range: 360 degree customization of services on the boom

Power Requirements on Boom: KEZR and KEXS Isolated Power per boom, 1047 UL listed

Upper Arm Length Required: 1300mm Mid Tube Length Required: 150mm Column Tube length required: 250mm Extension pole: Top and bottom extensions Weight load Capacity: At least 200 pounds (greater than is preferable) Brake system: All-Electric Equipment Shelves: Four shelves on arm with LED backlight for increased visibility, size of one side of square platform should be greater than 750mm

Shelf control location 3rd shelf Front of boom panel: Top: 3 circuits, electric, Middle: CO2 connection, Bottom: 3 circuits, distribution board Right of boom panel: Top: blank, Middle: Emergency Stop, Bottom: 3 circuits Rear of boom panel: Top: Televac control panel, Televac filter Middle:

Nitrogen control panel, Bottom: 2 vacuum connections

Left of boom panel: Top: 3 circuits, electric, Middle: blank, Bottom: 3 vacuum

Note: Hoses for N2O, Medical Air, WAGD, and Oxygen must not kink in the boom arms and must accommodate existing above ceiling hookups without replumbing the hookups. Diameter of tubing must accommodate from the old hookups to new equipment.

Life Cycle Support Characteristics:

• Design/utilization for operating room service

• Demonstrated reliability in clean environment

Equipment Configuration needed for following ORs: one equipment boom needs to be installed in Operating Room 5, 1 equipment boom total needed in this configuration.

Item: Teletom Equipment boom

Vendor: Stryker/Berchtold

Model: Teletom 6 Series Electric Power TP-622

Features: All-electric brake system with vertical articulation, built in “Emergency Stop” button located on the Boom, vertical articulation capability throughout unit, combined suspension (Single Spindle) non-tandem option

Operating Range: 360 degree customization of services on the boom

Power Requirements on Boom: KEZR and KEXS Isolated Power per boom, 1047 UL listed

Equipment Shelves: One to four shelf capability with flexible mounting options

Upper Arm Length Required: 1000mm Mid Tube Length Required: 150mm Column Tube length required: 250mm Extension pole: Top and bottom extensions Weight load Capacity: At least 250 pounds Brake system: All-Electric Equipment Shelves: Three shelves on arm with LED backlight for increased visibility, size of one side of square platform should be greater than 750mm

Shelf control location 2nd shelf Front of boom panel: Top: Left: O2 and Air, Right: O2 and Air, Bottom:

Left: WAGD, CO2, Right: distribution board Right of boom panel: Top: Emergency Stop, Bottom: 2 circuits Rear of boom panel: Top: LV 29mm Bottom: 2 vacuum connections, one on right and one on left Left of boom panel: Top: blank, Bottom: blank

Note: Hoses for N2O, Medical Air, WAGD, and Oxygen must not kink in the boom arms and must accommodate existing above ceiling hookups without replumbing the hookups. Diameter of tubing must accommodate from the old hookups to new equipment.

Life Cycle Support Characteristics:

• Design/utilization for operating room service

• Demonstrated reliability in clean environment