Performance Work Statement Summary

This Performance Work Statement (PWS) describes recurring testing, certification, environmental monitoring, and HEPA filtration services required at the Richard L. Roudebush VA Medical Center in Indianapolis, Indiana. The contract covers a 12-month base period with four optional 12-month renewal periods. Services support patient safety, employee safety, infection prevention, pharmaceutical compounding compliance, and reliable operation of critical ventilation and containment systems across four operational areas: Pharmacy, Laboratory (P&LMS), Research, and Engineering.

Pharmacy Services require semiannual complete certification of five Primary Engineering Controls (PECs)—two Class II Type A2 biosafety cabinets (BSCs), two horizontal laminar airflow workstations (LAFWs), and one compounding aseptic isolator (CAI)—to ISO Class 5 specifications, plus certification of three ISO Class 7 rooms with 48 viable air and surface field samples per event. Additional tasks include semiannual environmental monitoring, LAFW prefilter replacement, diffuser cleaning, and HEPA filter integrity testing for 10 ceiling/pass-through filters. Certifications must be conducted 8:00 to 17:00 Monday through Friday, with at least two personnel required due to ceiling filter test port locations.

Laboratory Services require annual NSF/ANSI 49 field certification of 13 Class II Type A2 BSCs, annual operational testing of four chemical fume hoods, and annual capture/exhaust certification of one grossing station. Research Services cover all 60 inventory entries: annual NSF/ANSI 49 certification for 20 BSCs and 36 chemical fume hoods, semiannual HEPA/airflow certification for two LFWS units, annual manufacturer preventive maintenance and functional certification for one hypoxia chamber, and annual classification and containment testing for one mini BSC/special device. Engineering Services require annual inspection and in-place aerosol-photometer leak testing of 24 HEPA filters in AHU 37A/37B, six HEPA filters in AHU 46, and two terminal HEPA filters in Operating Room 5, plus quarterly replacement of 23 isolation-exhaust prefilters and annual replacement of 23 isolation HEPA filters using safe-change methods.

Contractor qualifications are stringent: Pharmacy PEC/SEC certification must be performed by a CNBT Registered Certification Professional for Sterile Compounding Facilities (RCP-SCF); Class II BSC field certification requires NSF Enhanced BSC Field Certifier accreditation; the lead fume-hood tester must hold current CNBT credentials or equivalent documented training with ASHRAE 110; the lead HEPA/cleanroom tester must hold current CNBT cleanroom-testing credentials or equivalent training with in-place aerosol-photometer testing. All measuring instruments must be maintained in calibration and traceable to recognized standards. A mandatory pre-proposal site visit is required; offers will not be accepted from companies that do not attend. Post-award submittals include quality control and safety plans within 10 business days, and a reconciled electronic inventory within 30 calendar days identifying each asset with manufacturer, model, serial number, equipment classification, service frequency, and filter specifications.

Work shall be scheduled through the Contracting Officer's Representative (COR) during normal hours (0700–1800 Monday–Friday, excluding Federal holidays), though early morning, evening, weekend, and shutdown work may be required to protect patient care. The Contractor shall not shut down air handlers, exhaust fans, or critical systems without approved outage and explicit authorization. Pharmacy semiannual certification intervals shall not exceed six months; other semiannual events should be approximately six months apart; annual events must be completed within 12 months. All work requires compliance with facility infection-control, interim life-safety, fall-protection, lockout/tagout, and access requirements. Aerosol challenge must be contained so material is not introduced into occupied clinical spaces or other systems, coordinated with the COR/COTR and affected departments. Operating Room 5 testing shall occur during approved operating-room outages with coordination for diffuser removal, aerosol challenge, purge/clearance, cleaning, reassembly, and return to service.

Reports must be submitted within five business days after service (10 calendar days for Pharmacy reports after incubation and laboratory results are complete). Each report package shall include one searchable PDF and one editable electronic data file containing asset identification, date/time, operating condition, technician credentials, applicable standards/methods, raw measurements, calculations, acceptance criteria, pass/fail determination, and current calibration certificates for all test instruments used. HEPA test reports must include filter-bank maps, unique filter IDs, upstream concentration and uniformity data, scan methods, and leakage results. Fume hood reports must show sash position, traverse points, individual and average velocities, variability, smoke results, and monitor/alarm results. Pharmacy reports must include whether testing occurred under dynamic or static conditions, room diagrams with sampling locations, equipment and room dimensions, pressure differentials, total air changes per hour (ACPH), personnel present, viable-sample media details, colony-forming-unit counts, and smoke-study videos using clearly visible, neutrally buoyant smoke. Operating Room 5 reports must document each terminal filter's identifier and location, challenge concentration and uniformity, complete scan results, terminal airflow, total room supply airflow, room volume, total air changes per hour, pressure relationships, temperature, humidity, and final pass/fail status.

Deficiencies are classified as Critical (immediate or significant patient, personnel, infection-control, containment, or critical-system risk requiring immediate verbal notice and written notice within 24 hours), Major (failure of applicable acceptance criterion or inability to complete required certification requiring prompt COR/COTR notice and written deficiency in final report), or Minor (condition requiring correction or monitoring without failure of required performance, documented in routine report). Retesting after Contractor error, damage, improper adjustment, incomplete work, or invalid test is at no additional cost. Contractor personnel assigned to unescorted work shall obtain and maintain PIV cards and facility credentials in accordance with VA Directive 0735 and current credentialing procedures. The Government will designate a COR, provide reasonable coordinated access to listed equipment and spaces, furnish available inventories and prior reports, and coordinate required outages and escorts, but does not furnish labor, tools, test instruments, calibration, transportation, personal protective equipment, consumables, replacement filters, sampling materials, or laboratory analysis.

Pricing shall be firm-fixed-price for recurring scheduled services under one or more contract line items for each contract period, with condition-based HEPA replacement and other variable work separately priced as firm-fixed-price option contract line items. Invoices shall be submitted through VA-designated electronic invoicing processes and identify contract line item, service date, department/system, quantity, authorization reference for variable work, accepted report, and amount. Unauthorized, incomplete, failed, or unaccepted work shall not be invoiced as complete, and invoiced quantities for variable or optional work shall not exceed quantities authorized in writing by the Contracting Officer and accepted by the Government.

View the file

Other files

File Type Posted
36C25027Q0065 0003.pdf PDF
1A.jpeg JPEG file
2A.jpeg JPEG file
36C25027Q0065 0002.pdf PDF
36C25027Q0065 0001.pdf PDF
Attachment II - VAAR 852.219-75.pdf PDF
36C25027Q0065.pdf PDF
Attachment I - SCA WD.pdf PDF
36C25027Q0065.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

PERFORMANCE WORK STATEMENT

Testing, Certification, Environmental Monitoring, and HEPA Filtration

Services

1.0 BACKGROUND

The Richard L. Roudebush VA Medical Center (RLRVAMC) requires recurring inspection, testing, certification, environmental monitoring, preventive service, and filter work for laboratory containment equipment, Pharmacy engineering controls and controlled environments, Research equipment, and selected Engineering HEPA filtration systems. The services support patient safety, employee safety, infection prevention, pharmaceutical compounding compliance, and reliable operation of critical ventilation and containment systems.

2.0 OBJECTIVE AND SCOPE

The Contractor shall furnish all qualified personnel, supervision, labor, test equipment, calibrated instruments, materials, consumables, transportation, reports, labels, videos, disposal services, and incidentals required for scheduled recurring work. Pharmacy: certify five primary engineering controls

(PECs) to ISO Class 5 specifications or better, certify three ISO Class 7 rooms, perform environmental monitoring, replace prefilters, service diffusers, and test 10 ceiling/pass-through HEPA filters.

Laboratory (P&LMS): service 18 active assets—13 Class II Type A2 BSCs, four chemical fume hoods, and one grossing station.

Research: service all 60 inventory entries—36 chemical fume hoods, 20 BSCs, two LFWS units, one hypoxia chamber, and one mini BSC/special containment device.

Engineering: annually inspect and test 24 HEPA filters in AHU 37A/37B, six HEPA filters in AHU 46, and two terminal HEPA filters in Operating Room 5 served by AHU 38.

The equipment inventories in Appendices A through D are part of the scope. Unlisted equipment is outside the scope of this contract.

2.1 Explicit Scope Boundaries

The AHU 37A/37B filters are the 24-filter bank serving the 9 operating rooms; they are not an additional set of 24 filters in each operating room.

The AHU 46 filters are the six 30-inch by 30-inch by 12-inch-deep HEPA filters serving the Cardiology procedure room.

Operating Room 5 is served by AHU 38. AHU 38 contains no central HEPA filters; the two HEPA filters in scope are terminal filters located in Operating Room 5 and are separate from the AHU 37A/37B filter bank.

Operating room and Cardiology room relationships are provided for system identification and outage coordination. Except for the terminal-filter and room-performance measurements expressly required for

Operating Room 5 in Section 8.5, general room certification, air-change verification, pressure testing, or cleanliness certification is excluded unless expressly stated elsewhere in this PWS.

Isolation-room listings identify systems served. Room-level pressure, ACPH, door/seal, or infection-control certification is excluded from the isolation-exhaust scope.

Repairs and replacement parts not expressly included are outside scope unless ordered under a funded line item or contract modification. Contractor-caused damage and retesting remain the Contractor's responsibility.

3.0 PERIOD AND PLACE OF PERFORMANCE

The anticipated period of performance is one 12-month base period and four 12-month option periods.

Each exercised period includes the service frequencies stated in this PWS.

Place of performance: Richard L. Roudebush VA Medical Center, 1481 West 10th Street, Indianapolis, Indiana 46202, including roofs, mechanical rooms, interstitial spaces, controlled Pharmacy areas, laboratories, Research spaces, and other listed locations.

4.0 APPLICABLE REQUIREMENTS AND ORDER OF PRECEDENCE

The following editions are the technical basis for this PWS. The edition is incorporated by solicitation or contract controls. A later revision does not change contract scope or price unless incorporated by the

Contracting Officer. The Contractor shall promptly notify the COR/COTR and Contracting Officer of conflicts before proceeding.

29 CFR 1910.1450, Occupational Exposure to Hazardous Chemicals in Laboratories, including the facility

Chemical Hygiene Plan requirements applicable to hood operation and maintenance.

VHA Directive 1108.12, Management and Monitoring of Pharmaceutical Compounded Sterile Preparations, dated February 14, 2024.

USP General Chapter <797>, Pharmaceutical Compounding—Sterile Preparations, and USP General

Chapter <800>, Hazardous Drugs—Handling in Healthcare Settings, official versions in effect for the facility.

CETA CAG-002 (2026), CAG-003 (2022), CAG-005 (2025), CAG-006 (2023), CAG-008 (2022), CAG-009

(2023), CAG-014 (2022), and CAG-015 (2025), as applicable to the specific equipment and service.

ISO 14644-1:2015, ISO 14644-2:2015, and ISO 14644-3:2019.

NSF/ANSI 49-2024, Biosafety Cabinetry: Design, Construction, Performance, and Field Certification.

ANSI/ASHRAE Standard 110-2016 (RA 2025), Methods of Testing Performance of Laboratory Fume

Hoods.

ANSI/ASHRAE Standard 111-2024, Measurement, Testing, Adjusting, and Balancing of Building HVAC

Systems, where airflow system measurement is required.

ANSI/ASHRAE/ASHE Standard 170-2021 and adopted addenda applicable to health care ventilation systems.

IEST-RP-CC034.5 (2022), HEPA and ULPA Filter Leak Tests, and the applicable current edition of IEST-

RP-CC001 for HEPA filter performance and construction.

NFPA 45-2024, Standard on Fire Protection for Laboratories Using Chemicals, where applicable to chemical hood installation and operation.

CDC/NIH Biosafety in Microbiological and Biomedical Laboratories, 6th edition, as advisory biosafety practice were incorporated by facility policy.

VA HVAC Design Manual, VA Pharmacy Service Design Guide, VHA technical guidance, facility policies, infection-control requirements, and manufacturer instructions furnished or identified by the Government.

Where the PWS prescribes a more frequent service interval than a referenced minimum, the PWS frequency controls. An asset shall not be represented as certified as a standard that does not apply to its equipment type or listing.

5.0 CONTRACTOR QUALIFICATIONS

Pharmacy PEC/SEC certification shall be performed by a CNBT Registered Certification Professional for

Sterile Compounding Facilities (RCP-SCF), or by a firm employing an RCP-SCF who is responsible for the work, as required by VHA Directive 1108.12.

Class II BSC field certification shall be performed by personnel holding current NSF Enhanced BSC Field

Certifier accreditation or another qualification specifically accepted by the applicable standard and the

Government.

The lead fume-hood tester shall hold a current CNBT fume-hood credential or demonstrate equivalent documented training and experience with ASHRAE 110 field testing acceptable to the Government.

The lead HEPA/cleanroom tester shall hold a current CNBT cleanroom-testing credential or demonstrate equivalent documented training and experience with in-place aerosol-photometer testing and the cited

IEST/CETA methods.

Personnel conducting viable environmental monitoring shall be trained in aseptic sampling, chain of custody, incubation/media controls, result interpretation, and applicable USP/CETA requirements. Analysis shall be performed by a qualified microbiology laboratory using documented methods.

The Contractor shall submit technician names, roles, credentials, certificate numbers, expiration dates, relevant experience, and proposed assignments before work begins and whenever personnel change.

All measuring instruments shall be appropriate for the range and accuracy required, maintained in calibration at the manufacturer-prescribed interval, and traceable to recognized standards where applicable.

6.0 PRE-PERFORMANCE SUBMITTALS AND INVENTORY VERIFICATION

6.1 Mandatory Pre-Proposal Site Visit

Offerors shall attend the Government-scheduled site visit. Offerors shall inspect accessible equipment, housings, service clearances, filter banks, test ports, roof/interstitial access, shutdown constraints, and controlled-area entry requirements. Offerors shall verify conditions that materially affect labor, materials, access, safety, scheduling, and pricing.

AHU 37A/37B contains eighteen 24-inch by 24-inch by 11.5-inch deep and six 12-inch by 24-inch by 11.5-inch deep. Offerors shall field-verify dimensions, construction, efficiency and gasket orientation.

AHU 46 contains six nominal 30-inch by 30-inch by 12-inch-deep HEPA filters. Offerors shall verify actual dimensions, construction, efficiency, and gasket orientation.

Operating Room 5 contains two terminal HEPA filters served by AHU 38, which have no central HEPA filters. Offerors shall verify each terminal filter's location, dimensions, manufacturer, model, efficiency rating, housing and seal type, access method, test ports, upstream aerosol-injection access and downstream scan access.

Pharmacy contains 10 HEPA filters identified collectively as ceiling and pass-through HEPA filters. The

Government does not have the final filter-by-filter allocation; offerors shall verify the number, location, equipment association, access method, and filter identifiers.

Offers will not be accepted from companies that do not attend the site visit. Performance issues in the past made this a necessary restriction to apply to the procurement.

6.2 Post-Award Submittals

Within 10 business days after award: Quality Control Plan, safety plan, credential roster, instrument/calibration list, sample report forms, and proposed annual schedule.

Within 30 calendar days after award: reconciled electronic inventory identifying each asset, room, asset number, manufacturer, model, serial number, equipment classification, service frequency, filter quantity/size/efficiency, and discrepancies.

Before each controlled-area event: work plan, proposed tests, sampling map, shutdown/access needs, aerosol and containment method, and infection-control controls.

No newly discovered asset, repair, or replacement may be performed or invoiced without written authorization under the contract.

7.0 SCHEDULING, ACCESS, AND PROTECTION OF OPERATIONS

Normal performance hours are 0700 through 1800, Monday through Friday, excluding Federal holidays.

Early morning, evening, weekend, and shutdown work may be required to protect patient care and departmental operations. All work shall be scheduled through the COR.

Pharmacy semiannual certifications shall be scheduled so the interval does not exceed six months. Other semiannual events shall be approximately six months apart; annual events shall be completed within the applicable 12-month certification interval.

The Contractor shall not shut down an air handler, exhaust fan, PEC, or critical system without an approved outage and explicit authorization to proceed.

Work in sterile compounding, operating-room, procedural, laboratory, roof, or interstitial areas shall comply with facility infection-control, interim life-safety, fall-protection, lockout/tagout, and access requirements.

Aerosol challenge shall be planned and contained so challenge material is not introduced into occupied clinical spaces or other systems. The Contractor shall coordinate isolation, shutdown, purge, and return-to-service criteria with the COR/COTR and affected department.

Operating Room 5 testing shall be performed during an approved operating-room outage. The Contractor shall coordinate diffuser removal, aerosol challenge, purge or clearance, cleaning, reassembly, and return to service with the COR/COTR and affected operating-room personnel.

Photography and video are limited to required equipment documentation and smoke studies. Images must exclude patients, protected health information, computer screens, labels containing patient data, and unauthorized personnel.

8.0 TECHNICAL PERFORMANCE REQUIREMENTS

8.1 Pharmacy PECs, SECs, Environmental Monitoring, and HEPA Filters

The Contractor shall perform a complete semiannual certification event for all five Pharmacy PECs and all three ISO Class 7 rooms. The five PECs consist of two Class II Type A2 BSCs, two LAFWs, and one CAI.

Semiannual PEC testing shall include the applicable field-certification tests; it is not a limited visual inspection.

Perform nonviable particle counting and classification; measure and document supply and exhaust airflow;

calculate and report total air changes per hour (ACPH) for each room; and verify pressure relationships and segregation, temperature, relative humidity, airflow visualization, HEPA filter integrity, and applicable alarms and monitors, as required by USP, CETA, VA, and the approved room design. For the non-HD classified room, the certification report shall separately state: (1) ACPH provided by HVAC supply air and

(2) total ACPH, including the airflow contribution from the LAFWs located in that room. The report shall identify the room volume and measured airflow values used in each calculation. BSCs: inflow and downflow measurements, supply and exhaust HEPA leak testing, dynamic airflow smoke patterns with video, alarms/interlocks, site installation assessment, and other NSF/ANSI 49 and CETA tests applicable to the model and installation.

LAFWs: HEPA leak testing, airflow velocity and uniformity, nonviable particle count, induction-leak/back-streaming evaluation, dynamic airflow smoke visualization with video, alarms/controls, and diffuser inspection.

CAI: HEPA leak testing, airflow and particle performance, chamber and transfer-chamber pressure, ingress/egress or recovery testing, dynamic airflow smoke visualization with video, alarms/interlocks, and glove/sleeve integrity when required by the manufacturer or governing procedure.

Pass-through HEPA filter(s): test in accordance with CETA CAG-015, the manufacturer method, and the applicable room/containment requirements. Report each pass-through separately.

Perform 48 viable air and surface field samples of PECs, pass-throughs, and ISO Class 7 rooms at each semiannual event. Samples should be collected and incubated in accordance with the dual plate method, each site will have a separate bacterial and fungal sample. Required positive/negative media controls and blanks are additional and shall not be counted among the 48 field samples.

Provide complete laboratory analysis, colony counts, alert/action-level comparison, required organism identification, excursion assessment, and follow-up recommendations. Maintain chain of custody and media/incubation records.

Replace four LAFW prefilters (two per LAFW) semiannually and dispose of removed prefilters. Remove and clean five laminar-flow hood/PEC diffusers semiannually in accordance with manufacturer instructions.

Inspect and perform filter integrity-test for all 10 Pharmacy ceiling/pass-through HEPA filters semiannually.

Certification activities must be conducted during the hours of 8:00 to 17:00, Monday through Friday during standard business days. Certifications will not be conducted on weekends nor on Federal Holidays without express authorization of Compounded Sterile Preparations Pharmacy Program Manager and COR.

Contractor shall send at least two personnel to perform certification activities due to the locations of ceiling

HEPA filter test ports.

8.2 Laboratory (P&LMS)

The active Laboratory inventory is 13 Class II Type A2 BSCs, four chemical fume hoods, and one grossing station. Missing model, asset, or certification-history fields shall be verified during inventory reconciliation.

BSCs: certify annually using NSF/ANSI 49 field-certification tests applicable to each cabinet. Apply a certification label showing date, status, certifier, and next due date.

Chemical fume hoods: test annually at the designated operating sash position, including condition inspection, face-velocity traverse and uniformity, airflow visualization, sash operation, monitor/alarm verification, fan/exhaust response, and comparison with the facility/manufacturer design criterion.

Grossing station: annually verify exhaust capture using manufacturer criteria and applicable ventilation methods, including velocity/capture measurements, smoke visualization, controls/alarms, lighting and work-surface condition, and drainage/utility observations relevant to safe operation.

8.3 Research

The Research inventory contains 60 entries: 36 chemical fume hoods, 20 BSCs, two LFWS units, one hypoxia chamber, and one mini BSC/special containment device. All 60 entries are included. The hypoxia chamber is in room D3014.

Standard BSCs: certify annually using the applicable NSF/ANSI 49 field tests and manufacturer criteria.

Chemical fume hoods: test annually using the routine operational tests stated in Section 8.2.

LFWS units: certify semiannually for HEPA integrity, airflow velocity/uniformity, particle performance, airflow visualization, alarms/controls, and applicable manufacturer criteria.

Mini BSC/special containment device: determine the actual equipment classification, listing, manufacturer, and intended function. Perform the applicable manufacturer and containment tests annually. Do not label the device NSF-certified unless it is eligible and all applicable criteria are met.

Hypoxia chamber: annually perform manufacturer-required preventive maintenance and functional certification, including oxygen/atmosphere sensor calibration, control and alarm verification, door/seal condition, purge and gas-supply safety functions, exhaust/containment observations, and documented setpoint accuracy. Do not represent this chamber as a BSC or fume hood.

8.4 AHU 37A/37B and AHU 46 HEPA Filter Banks

Annual service is inspection and applicable integrity/performance testing. HEPA filter and filter-bank position shall receive a unique durable identifier that is carried forward in reports.

Inspect filter media, frames, gaskets, seals, clamps, holding frames, housing surfaces, penetrations, access doors, test ports, and evidence of moisture, corrosion, bypass, damage, loading, or microbial contamination.

Record filter manufacturer, model, serial/lot if available, nominal/actual dimensions, efficiency/performance level, installation date if known, orientation, design airflow, measured differential pressure, and observed condition.

Perform in-place aerosol-photometer leak testing under operating conditions using IEST-RP-CC034.5.

Establish and document upstream challenge concentration and uniformity; scan the downstream filter face, perimeter, gasket, frame, and accessible penetrations; and apply the acceptance criterion for the installed filter's performance level and approved test method.

Measure/document airflow, face velocity, differential pressure, and other system performance needed to establish that testing occurred under representative conditions. Identify limitations that prevent a complete scan; a limited test shall not be reported as a full certification.

Following test activity, restore the system, verify normal airflow and controls, remove challenge aerosol in accordance with the approved plan, and obtain authorization before returning the system to clinical service.

8.5 Operating Room 5 Terminal HEPA Filters

AHU 38 serves Operating Room 5 and contains no central HEPA filters. Operating Room 5 contains two terminal HEPA filters. The Contractor shall inspect and test each terminal filter annually.

Inspect the filter media, frame, terminal housing, gasket or gel seal, retaining devices, diffuser, ceiling interface, accessible penetrations, and evidence of damage, deterioration, moisture, loading, or bypass.

Assign a durable unique identifier to each terminal filter and record its location, manufacturer, model, serial or lot number when available, nominal and actual dimensions, efficiency rating, installation date if known, housing and seal type, and observed condition.

Perform an in-place aerosol-photometer integrity and leak test under representative operating airflow using

IEST-RP-CC034.5. Establish and document a safe upstream challenge concentration and uniformity for each terminal; scan the entire accessible downstream filter face, perimeter, frame, seal, housing joints, and penetrations; and apply the acceptance criterion for the installed filter's performance level, approved method, manufacturer requirements, and facility design.

Measure and report terminal airflow volume or face velocity and uniformity, as applicable to the terminal design, and differential pressure when a manufacturer-provided test port, gauge, or approved measurement method is available.

Under the representative operating mode, document total supply airflow to Operating Room 5, the room dimensions and calculated volume, total air changes per hour, room pressure relationship and airflow direction relative to adjacent spaces, temperature, and relative humidity. Compare results with the facility-approved design criteria and the contract-adopted VA and ASHRAE/ASHE healthcare ventilation requirements.

Identify any surface or component that could not be accessed or scanned. A limited scan or incomplete room-performance test shall not be reported as a complete certification.

After testing, reinstall and clean diffusers and disturbed surfaces as required, remove challenge aerosol under the approved plan, verify normal airflow and controls, document final operating status, and obtain authorization before returning Operating Room 5 to service.

8.6 Isolation Exhaust Filter Systems

Provide and replace 23 isolation-exhaust prefilters quarterly and 23 HEPA filters annually using the bag-in/bag-out or other housing-specific safe-change method. Verify quantities, sizes, filter-bank relationships, and housing access during the site visit and initial inventory reconciliation.

Coordinate exhaust shutdowns and room impact before work; follow facility infection-control and respiratory-protection requirements.

Contain, label, transport, and dispose of removed filters in accordance with the facility waste determination and applicable regulations. Do not assume all filters have the same waste classification.

After HEPA replacement, verify seating and seals, perform an in-place integrity test where the housing permits and the system requirement call for it, document differential pressure and operation, and obtain release to service.

9.0 INTENTIONALLY LEFT BLANK

10.0 SERVICE SUMMARY AND FREQUENCIES

Area Quantity / Asset Service Frequency

Pharmacy 5 PECs (2 BSC, 2 LAFW, 1

CAI)

Complete certification including but not limited to:

Dynamic smoke-pattern testing with video; non-viable particle count; HEPA filter challenge/leak testing; airflow testing; applicable certification testing; ISO Class 5 classification

Semiannual; interval not to exceed 6 months

Pharmacy 3 ISO Class 7 rooms

Room/SEC certification including but not limited to:

USP room performance testing; ISO 7 environmental classification/particle testing; pressure/airflow/ACPH

[VHA requires 35 ACPH for ISO Class 7 rooms]

Semiannual

Pharmacy 48 field samples per event Viable bacterial/fungal air and surface sampling;

positive/negative controls additional; laboratory analysis; action-level comparison Semiannual

Pharmacy 10 HEPA filters HEPA filter challenge/leak testing for ceiling and pass-through HEPA filters. Semiannual

Pharmacy 4 LAFW prefilters Change prefilters; dispose of removed filters (2 prefilters per LAFW) Semiannual

Pharmacy 5 PEC diffusers Remove, clean, and reinstall Semiannual

Laboratory 13 BSCs NSF/ANSI 49 field certification Annual

Laboratory 4 fume hoods Operational certification/testing Annual

Laboratory 1 grossing station Capture/exhaust certification Annual

Research 20 BSCs NSF/ANSI 49 field certification Annual

Research 36 fume hoods Operational certification/testing Annual

Research 2 LFWS units HEPA/airflow certification Semiannual

Research 1 hypoxia chamber Manufacturer PM and functional certification Annual

Research 1 mini BSC/special device Classification and applicable containment certification Annual

Engineering 24 AHU 37A/37B HEPA filters Inspect and integrity/performance test Annual

Engineering 6 AHU 46 HEPA filters Inspect and integrity/performance test Annual

Engineering 2 OR 5 terminal HEPA filters Integrity/leak testing; terminal airflow and room-performance verification Annual

Engineering 23 isolation prefilters Provide, replace, dispose Quarterly

Engineering 23 isolation HEPA filters Provide, replace, dispose; post-test as applicable Annual

11.0 REPORTING AND DELIVERABLES

The Contractor shall provide preliminary pass/fail information before leaving the work area. Critical deficiencies shall be reported immediately to the COR/COTR and affected department, followed by written notice within 24 hours. Final routine reports other than Pharmacy reports shall be submitted electronically within five business days after service. Final Pharmacy reports, including completed incubation and laboratory results, shall be submitted within 10 calendar days after service and within the VHA reporting limit incorporated into the contract.

One searchable PDF report package and one editable electronic inventory/report data file for each service event.

Asset/system identification, room, department, date/time, operating condition, technician, credentials, and work performed.

Applicable standard/method and edition, manufacturer criterion, instrument identification, calibration date/status, raw measurements, calculations, acceptance criteria, and pass/fail determination. The final report shall include a current calibration certificate for each test instrument used. The certificate shall match the instrument identified in the report, be valid on the testing date, and document traceability to

NIST or another recognized standard, as applicable.

For HEPA tests: filter-bank or terminal-filter map, unique filter ID and location, upstream concentration and uniformity, scan method, leakage or penetration results, accessible surfaces scanned.

For fume hoods: sash test position, traverse points, individual and average velocities, variability, smoke result, monitor/alarm result, observed cross-drafts/room conditions, and certification status.

For Pharmacy: whether testing occurred under dynamic or static conditions; room diagrams showing viable and nonviable sampling locations; equipment and room dimensions; supply and exhaust filters; pressure differentials; total ACPH for each classified room; for the non-hazardous-drug room, supply-air ACPH and total ACPH including LAFW airflow; number of personnel present during certification; viable-sample media, lot, expiration, certificates of analysis, and incubation conditions; bacterial and fungal colony-forming-unit counts; genus and species identification when an action level is exceeded and identification is practicable;

alert and action-level comparisons; as-found and as-left conditions for adjusted equipment; PEC-specific reports that reference manufacturer-approved values; calibration documentation; and smoke-study videos in a VA-acceptable format using clearly visible, neutrally buoyant smoke.

For Operating Room 5: each terminal filter's identifier and location, challenge concentration and uniformity, complete scan results, terminal airflow, total room supply airflow, room volume, calculated total air changes per hour, pressure relationship and airflow direction, temperature, relative humidity, test operating mode, access limitations, acceptance criteria, and final pass/fail status.

For filter replacement: old/new filter data, reason, authorization reference, disposal/manifest information as applicable, installation details, post-test data, and final return-to-service status.

Certification labels shall not be applied to failed equipment. Restricted/conditional labels shall plainly state the limitation and refer to the report.

Incomplete or inaccurate reports shall be corrected and resubmitted within three business days after

Government notice. The Contractor shall retain service documentation for at least six years after final payment, or longer if the contract or law requires.

12.0 DEFICIENCIES, CORRECTIVE ACTION, AND RETESTING

Class Definition Notification

Critical Immediate or significant patient, personnel, infection-control, containment, or critical-system risk Immediate verbal notice; written within 24 hours

Class Definition Notification

Major Failure of an applicable acceptance criterion or inability to complete a required certification Prompt COR/COTR notice and written deficiency in final report

Minor Condition requiring correction or monitoring without failure of required performance Document in routine report

Document the failed parameter, measured result, acceptance criterion, likely impact, and recommended corrective action.

Minor adjustments customarily required to complete certification are included. Repairs, parts, and condition-based replacements require written Contracting Officer authorization under the applicable priced option contract line item or another authorized contractual action.

Retesting after Contractor error, damage, improper adjustment, incomplete work, or invalid test is at no additional cost.

13.0 QUALITY CONTROL AND GOVERNMENT SURVEILLANCE

The Contractor is responsible for quality control. The Quality Control Plan shall address personnel qualification, method control, calibration, sampling controls, report review, data integrity, nonconforming work, corrective action, and document retention. The Government may observe work, review records, independently verify measurements, reject incomplete deliverables, and require correction of nonconforming work.

Government observation, review, or acceptance does not relieve the Contractor of responsibility for compliance. The COR/COTR may coordinate and inspect performance but may not change contract terms, price, quantity ceilings, or funding.

14.0 SECURITY, PIV, AND FACILITY ACCESS

Because performance requires routine and continuing access to VA facilities, Contractor and subcontractor personnel assigned to unescorted work shall be eligible for, promptly apply for, obtain, and maintain the credential determined by VA, including a PIV card when required. The Contractor shall comply with FAR

52.204-9, VAAR 852.204-70, VA Directive 0735, and current VA credentialing procedures incorporated into the contract.

Initiate sponsorship, identity proofing, fingerprinting, background investigation, training, enrollment, and issuance actions promptly after award.

Do not assign personnel who cannot obtain or maintain required eligibility to work requiring that access.

Display/carry credentials as directed, comply with check-in and escort rules, protect keys/badges, and return Government credentials when access ends.

PIV eligibility does not independently authorize entry into Pharmacy, Research, laboratories, roofs, interstitial spaces, mechanical rooms, or clinical areas; area-specific authorization remains required.

No access to VA information systems, patient records, or VA sensitive information is anticipated.

Immediately report inadvertent exposure to the COR/COTR and follow VA privacy/security direction.

The C&A requirements do not apply and a Security Accreditation Package is not required

15.0 GOVERNMENT-FURNISHED ACCESS AND INFORMATION

The Government will designate a COR; provide reasonable coordinated access to listed Government equipment and spaces; furnish available inventories, prior reports, and facility procedures; and coordinate required outages, escorts, and departmental contacts. Access remains subject to patient care, security, safety, infection control, and operational restrictions.

The Government does not furnish Contractor labor, tools, test instruments, calibration, transportation, PPE, routine consumables, replacement filters, sampling materials, laboratory analysis, or other resources required by this PWS.

16.0 FFP REQUIREMENTS

Recurring scheduled services shall be priced under one or more firm-fixed-price contract line items for each contract period. Condition-based HEPA replacement and other variable work shall be separately priced as firm-fixed-price option contract line items. The Government is not obligated to purchase any optional quantity unless the Contracting Officer exercises the applicable option in writing. At a minimum, pricing shall distinguish:

17.0 INVOICING

Invoices shall be submitted in arrears through the VA-designated electronic invoicing process and shall identify the contract line item, service date, department/system, quantity, authorization reference for variable work, accepted report, and amount. Unauthorized, incomplete, failed, or unaccepted work shall not be invoiced as complete. The Contracting Officer will incorporate applicable FAR/VAAR invoicing clauses into the solicitation and award. For variable or optional work, the invoiced quantity shall not exceed the quantity authorized in writing by the Contracting Officer and accepted by the Government.

APPENDIX A — PHARMACY INVENTORY

Equipment / System Location. EE/MX Make/Model SN

ISO Class 5 PEC

Class II, Type A2 Biological

Safety Cabinet (BSC)

A-7086

Right side of room X1826296 Baker BSC SG404 127090

ISO Class 5 PEC

Class II, Type A2 Biological

Safety Cabinet (BSC)

A-7086

Left side of room

MX182629

3 Baker BSC SG404 127091

ISO Class 5 PEC

Horizontal Laminar Airflow

Workstation (LAFW)

A-7086

Right Side of room

MX249301

6 NuAire-340-630 199005060820

ISO Class 5 PEC

Horizontal Laminar Airflow

Workstation (LAFW)

A-7086

Left side of room

MX249301

NuAire-340-630 198953060420

ISO Class 5 PEC

Restricted Access Barrier

System – Compounding

Aseptic Isolator (CAI)

D-2035 MX271780

NuAire

NU-PR797-0400

209374111521

ISO Class 7

Positive Pressure

Cleanroom A-7086

ISO Class 7

Negative Pressure

Cleanroom A-7086

ISO Class 7 Anteroom A-7086

APPENDIX B — LABORATORY (P&LMS) INVENTORY

Location EE/MX Type Model Serial Number

Cytology A2064 MX3546715 Class II Type A2 NU-540-400 224778021324

Micro A2044 EE51772 Fume Hood 54L590P 54L590P

Micro A2044 MX3610896 Class II Type A2 NU-540-400 224971022624

Micro A2044 MX3610898 Class II Type A2 NU-540-300 225067022924

Histology A2067 MX1895323 Fume Hood EH6V 126690

Frozen Room A2066 MX1630107 Class II Type A2 NU-540-300 182346100917

Frozen Room A2066 MX1705534 Fume Hood FH5C 123429

Molecular A2042 MX3400517 Class II Type A2 NU-540-400 216884110822

Molecular A2042 MX3400516 Class II Type A2 NU-540-400 216882110822

Molecular A2042 MX3400474 Class II Type A2 NU-540-400 216887110822

Molecular A2043 MX3400462 Class II Type A2 NU-540-400 216883110822

Molecular A2043 MX3400519 Class II Type A2 NU-540-400 216886110822

Micro A2044 MX3688203 Class II Type A2 NU-540-300 228747111124

Histology A2067 MX1826106 Grossing Station Not provided MB670H15001

Micro A2049 EE4241 Class II Type A2 NU-425-600 22454WN

Processing A2132 EE46944 Class II Type A2 3620400 070874972

Mohs C-1094 MX1982304 Class II Type A2 Not provided N/A

Cytology A2064 No EE Assigned Fume Hood Not provided 231056295

APPENDIX C — RESEARCH INVENTORY

Line Room Inventory Description EE/MX / Identifier

1 D3014 1 Hypoxia Chamber EE82869/EE63841

2 C3077 1 Fume Hood EE24581

3 C3078 1 Fume Hood EE24583

4 C3078 1 BSC IU 000125

5 C3092 1 Fume Hood EE9351

6 C3111 1 Fume Hood EE9361

7 C3113 1 Fume Hood EE9362

8 C3113 1 Fume Hood EE67976

9 C3113 1 LFWS EE63937

10 C3113 1 LFWS EE67766

11 C3116 1 BSC EE79416

12 C3117 1 BSC EE65512

13 C6064 1 Fume Hood EE52550

14 C6071 1 BSC MX3644482

15 C6072 1 BSC EE69558

16 C6077 1 Fume Hood EE51336

17 C6078 1 Fume Hood EE51322

18 C6080 1 Fume Hood Not provided

19 C6082 1 Fume Hood Not provided

20 C6091 1 Fume Hood EE52364

21 C6114 1 Fume Hood EE59407

Line Room Inventory Description EE/MX / Identifier

22 C6115 Mini BSC EE86988

23 C6115 1 Fume Hood EE52475

24 C6133 1 BSC EE34270

25 C6134 1 BSC EE4240

26 C6134 1 BSC EE34271

27 C6139 1 Fume Hood EE59204

28 C7093 1 Fume Hood EE28216

29 C7095 1 Fume Hood EE28215

30 C7098 1 BSC EE51409

31 C7106 1 BSC MX2054121

32 C7110 1 BSC MX2062903

33 D2004 1 BSC EE4544

34 D2005 1 BSC MX1944880

35 D2006 1 Fume Hood EE52041

36 D2007 1 Fume Hood EE51719

37 D2007 1 BSC MX1633251

38 D2010 1 BSC EE67995

39 D2012 1 Fume Hood No MX/EE# Vectaire

40 D3002 1 BSC EE26708

41 D3004 1 Fume Hood No MX/EE# Vectaire #26

42 D3005 1 Fume Hood No MX/EE# Vectaire #21

43 D3006 1 Fume Hood MX1633270

44 D3007 1 Fume Hood No MX/EE# Vectaire #23

45 D3008 1 Fume Hood No MX/EE# Vectaire #24

46 D3014 1 BSC MX1633240

47 D3015 1 BSC MX2031906

48 D3016 1 BSC EE51541

49 D3016 1 Fume Hood No MX/EE# Vectaire #25

50 C3102 1 Fume Hood EE67978

51 C3113 1 Fume Hood EE63937

52 C338A 1 Fume Hood EE73176

53 C340A 1 Fume Hood EE72534

54 C6082 1 Fume Hood EE51329

55 C6136 1 Fume Hood EE746

56 C6136 1 BSC MX2026750

57 D2001 1 Fume Hood EE21352

58 E220 1 Fume Hood EE37551

59 C2192 1 Fume Hood EE9353

60 AB007 1 Fume Hood MX3624750

APPENDIX D — ENGINEERING SYSTEM INVENTORY

System Access Served Area / Rooms Filter / Housing Frequency

AHU 37A/37B 5th Floor MER 9 Surgery operating rooms 24 HEPA filters total; (18)

24 x 24 x 11.5 in, (6) 12 x

24 x 11.5 in

Annual test;

condition-based replacement

System Access Served Area / Rooms Filter / Housing Frequency

AHU 46 6th Floor MER Cardiology procedure room 6 HEPA filters; nominal 30 x 30 x 12 in

Annual test;

condition-based replacement

OR 5 terminal

HEPA filters

OR 5 ceiling terminals;

upstream challenge access to verify Operating Room 5; served by AHU 38

2 terminal HEPA filters; no central HEPA filters in

AHU 38

Annual test;

condition-based replacement

EF-3 A-Wing roof; Stairwell 10 A-7076, A-7074, A-7072, A-8076, A-8074, A-

8072 Housing location to verify Quarterly prefilter;

annual HEPA

EF-4 A-Wing roof; Stairwell 10 A-7050, A-7052, A-7054, A-8050, A-8052, A-

8054 Housing location to verify Quarterly prefilter;

annual HEPA

EF-5 A-Wing roof; Stairwell 10 A-7070, A-7068, A-8070, A-8068 Housing location to verify Quarterly prefilter;

annual HEPA

EF-6 A-Wing roof; Stairwell 10 A-7056, A-7058, A-8056, A-8058 Housing location to verify Quarterly prefilter;

annual HEPA

EF-79 Roof; Stairwell 1, 9th-floor C-

Wing C-6133 Housing location to verify Quarterly prefilter;

annual HEPA

EF-80 Roof; Stairwell 1, 9th-floor C-

Wing C-6134 Housing location to verify Quarterly prefilter;

annual HEPA

EF-113 Roof A-4040, A-4042, A-4056, A-4031, A-4018 A-5001 interstitial Quarterly prefilter;

annual HEPA

EF-163 Roof A-6016, A-6033 A-5001 interstitial Quarterly prefilter;

annual HEPA

EF-164 Roof A-6035, A-6034 A-5001 interstitial Quarterly prefilter;

annual HEPA

EF-165 Roof A-6017, A-6015 A-5001 interstitial Quarterly prefilter;

annual HEPA

EF-199 Roof; Stairwell 1, 3rd-floor C-

Wing A-1156 Housing location to verify Quarterly prefilter;

annual HEPA

APPENDIX E — ACCEPTANCE AND REPORTING CHECKLIST

Review Item Minimum Evidence

Inventory Asset and filter identifiers reconcile to the approved inventory; discrepancies documented.

Method Applicable standard, edition, test method, operating condition, and limitations are stated.

Instrumentation Instrument ID, range, accuracy, calibration date/status, and traceability are provided.

Measurements Raw readings and calculations support the result; no summary-only pass/fail statement.

Criteria Equipment-specific acceptance criterion and its source are stated.

Status PASS/CERTIFIED, FAIL/NOT CERTIFIED, or RESTRICTED/CONDITIONAL is unambiguous.

Deficiency Failure, consequence, notification, recommendation, and authorization need are documented.

Retest Initial result, corrective action/replacement, post-test, and final status are traceable.

Deliverables Report, editable data, labels, videos, sampling results, and filter logs are complete.