This file is an operator manual for the Reliance Vision Single-Chamber Washer/Disinfector manufactured by STERIS Corporation. The manual provides comprehensive instructions for operating and maintaining the washer/disinfector unit, which is designed for cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans, urinals, rubber/plastic goods, surgical instruments, theatre shoes, and EndoWrist instruments using specialized racks.

The manual details technical specifications, installation requirements, component identification, operating procedures, maintenance schedules, and troubleshooting guidance. Key features include automated chemical injection pumps for precise dispensing of cleaning agents, touch screen controls with multiple operating modes, emergency safety features, and optional systems like the SCS Conveyor System for automated loading/unloading. The unit can process loads up to 135 lbs using various specialized racks and accessories, with cycle options including Instruments, Utensils, Plastic Goods, Gentle Cycle, Rigid MIS, Orthopedics and Robotic cycles. The manual emphasizes proper loading techniques, chemical handling safety, and maintenance procedures required for optimal performance and equipment longevity.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

OPERATOR MANUAL

Reliance® Vision® Single-Chamber Washer/Disinfector

(2015-07-06) P920508-556

A WORD FROM STERIS CORPORATION

This manual contains important information on the proper use and routine maintenance of the Reliance® Vision® Single-Chamber Washer/Disinfector. All personnel involved in the use and maintenance of this equipment must carefully review and comply with the Safety Precautions and instructions contained in this manual. These instructions are important to protect the health and safety of personnel operating this washer/disinfector and should be retained in a conveniently accessible area for quick reference. This equipment is specifically designed only for the uses outlined in this manual.

Complete instructions for uncrating and connecting utilities, as well as the equipment drawing, have been provided. If they are missing, contact STERIS for replacement copies, providing the serial number and model of the washer/disinfector.

STERIS provides a line of accessories for use with this washer/ disinfector. A STERIS representative will gladly review these with you.

Please familiarize all users with this manual and store in an accessible location.

Also, refer to GENERAL NOTES – APPLICABLE TO STERILIZER AND WASHER EQUIPMENT DRAWINGS (P062941-091) for more information about utilities requirements.

Advisory IMPORTANT: A listing of the Safety Precautions to be observed when operating and servicing this washer/disinfector can be found in SECTION 1 of this manual. Do not operate the equipment until you have become familiar with this information.

STERIS does not intend, recommend or represent in any way that the Reliance Vision Single-Chamber Washer/Disinfector be used for the terminal disinfection or sterilization of any regulated medical device.

The Reliance Vision Single-Chamber Washer/Disinfector are intended only to perform an initial step in the processing of soiled, reusable medical devices. If medical devices will be contacting blood or compromised tissues, such devices must be terminally processed in accordance with device manufacturer's instructions and/or Good Hospital Practices before each use in human patients.

Any alteration of this equipment not authorized or performed by STERIS will void the warranty. Alteration of this equipment which could adversely affect its operation and efficacy may violate national, state and local laws or regulations.

©2015, STERIS Corporation. All rights reserved. Printed in Canada.

i Introduction Operator Manual 920508-556

The protection provided by the equipment may be impaired if the equipment is used in a manner not specified by STERIS.

To help ensure operators are adequately trained in the safe use of the equipment, STERIS recommends:

- all personnel who operate or maintain the equipment are trained in its operation and in its safe use;

- personnel working with toxic chemicals and vapors (if applicable) have comprehensive instructions in the washer/ disinfector process, relevant health hazards and methods to detect the escape of toxic materials;

- there is regular training of all personnel concerned with the operation and maintenance of the equipment; attendance records are maintained; and that evidence of understanding is demonstrated;

- current Safety Data Sheets (SDS) should be available to all users in the department. For current SDS, contact STERIS.

Indications For Use The Reliance Vision Single-Chamber Washer/Disinfector is intended for use in the cleaning and intermediate level disinfection of soiled reusable utensils, trays, glassware, bedpans and urinals, rubber and plastic goods, simple hard-surfaced rigid surgical instruments, such as forceps and clamps, theatre shoes, EndoWrist®1 instruments using da Vinci® Rack and other similar and related articles found in healthcare facilities.

Service Information A thorough preventive maintenance program is essential to help ensure safe and proper equipment operation. This manual contains maintenance schedules and procedures which should be followed for satisfactory equipment performance.

Customers are encouraged to contact STERIS concerning our comprehensive annual maintenance program. Under the terms of the program, preventive maintenance, adjustments and replacement of worn parts are performed on a scheduled basis to help ensure equipment performance at peak capability and to help avoid untimely and costly interruptions. STERIS maintains a global staff of well-equipped, qualified service technicians to provide this service, as well as professional repair services. Please contact STERIS for details.

1 EndoWrist is a registered trademark of Intuitive Surgical, Inc.

ii 920508-556 Operator Manual Introduction

Certification The Reliance Vision Single-Chamber Washer/Disinfector meets the applicable requirements of the following standards,:

• Underwriters Laboratories (UL) Standard UL 61010-1, Second Edition.

• Canadian Standards Association (CSA) CAN/CSA-C22.2 N° 61010-1, Second Edition.

• Canadian Standards Association (CSA) CAN/CSA-Z314.8-08, Third Edition.

Governing Directive for the affixing of the CE mark:

Medical Device Directive 93/42/EEC as amended by 2007/47/EC.

Standards applied to demonstrate conformity to the directive:

• EN/IEC 61010-1, Second Edition; EN/IEC 61326-1, 2005; EN/IEC 61010-2-040: Second Edition;

• ISO 15883-1* 2006 Washer-Disinfectors – General requirements, terms and definitions and tests;

• ISO 15883-2* 2006 Washer-Disinfectors – Requirements and tests for washer-disinfectors employing thermal disinfection for surgical instruments, anaesthetic equipment, bowls, dishes, receivers, utensils, glassware, etc.

NOTE: * When the optional Process Data Validation System (PDVS) or the optional Control Enhancement System is present.

iii Introduction Operator Manual 920508-556

Manufactured by:

Corporation STERIS Canada 490 boulevard Armand-Paris Québec (Québec) Canada G1C 8A3

The base language of this document is ENGLISH. Any translation must be made from the base language document.

STERIS Ltd.

Chancery House 190 Waterside Road Hamilton Industrial Park Leicester

LE5 1QZ

United Kingdom

Sales and Service -- Europe, Middle-East and Africa:

STERIS

116 Avenue de Magudas, 33185 Le Haillan

FRANCE

Support Service, by email:

Support_EMEA_IPT_SE@steris.com

All other requests, by phone: +33 5 56 93 94 94

Sales and Service -- Americas, Oceania and Asia:

STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060

440-354-2600 • 800-548-4873

Website:

www.steris.com iv 920508-556 Operator Manual Introduction

TABLE OF CONTENTS

Section Number Description Page A WORD FROM STERIS CORPORATION..................................................................i

Advisory............................................................................................................................................................ i

Indications For Use...........................................................................................................................................ii

Service Information...........................................................................................................................................ii

Certification...................................................................................................................................................... iii

1 SAFETY PRECAUTIONS ......................................................................................1-1

2 INSTALLATION VERIFICATION...........................................................................2-1

2.1 Technical Specifications..................................................................................................................... 2-1

2.1.1 Voltage, Amperage and Power Consumption ....................................................................... 2-1

2.1.2 Noise Level ............................................................................................................................ 2-2

2.1.3 Permissible Environmental Conditions .................................................................................. 2-2

2.1.4 Seismic Anchorage System (Accessory) .............................................................................. 2-2

2.2 Installation Verification ........................................................................................................................ 2-2

2.3 Chemicals and Cycles........................................................................................................................ 2-4

2.3.1 Using Prolystica Ultra Concentrate Chemicals...................................................................... 2-4

2.3.2 Using Other STERIS Chemicals............................................................................................. 2-7

2.3.3 Injection Units Range............................................................................................................. 2-9

3 COMPONENT IDENTIFICATION...........................................................................3-1

3.1 Washer/Disinfector – General ............................................................................................................. 3-1

3.1.1 Chamber ................................................................................................................................ 3-2

3.1.2 Chamber Door ....................................................................................................................... 3-2

3.1.3 Access Panels ....................................................................................................................... 3-2

3.1.4 Electrical Supply Box............................................................................................................. 3-2

3.1.5 Chemical Injection Pumps..................................................................................................... 3-2

3.1.6 Air Inlet Pre-Filter and HEPA Filter ......................................................................................... 3-3

3.1.7 Debris Screen........................................................................................................................ 3-3

3.1.8 Rotary Spray Arm Assemblies ............................................................................................... 3-3

3.2 Control System.................................................................................................................................... 3-5

3.2.1 Login Names and Passwords................................................................................................ 3-5

3.2.2 Ready Mode .......................................................................................................................... 3-6

3.2.3 Cycle Mode............................................................................................................................ 3-6

3.2.4 Supervisor Mode.................................................................................................................... 3-6

3.2.5 Service Mode......................................................................................................................... 3-6

3.2.6 Control System Touch Screens ............................................................................................. 3-6

(Continued...)

v Table of Contents Operator Manual 920508-556

TABLE OF CONTENTS (Cont’d)

Number Description Page 3 COMPONENT IDENTIFICATION (Cont’d)

3.2.7 Audible Signal........................................................................................................................ 3-9

3.2.8 Power (ON/OFF) Toggle Switches......................................................................................... 3-9

3.2.9 Printer................................................................................................................................... 3-10

3.2.10 Air Compressor (Accessory) ............................................................................................... 3-10

3.3 Emergency Safety Features.............................................................................................................. 3-11

3.3.1 Chamber Door Interlock Mechanism................................................................................... 3-11

3.3.2 Obstruction Sensor .............................................................................................................. 3-11

3.4 SCS Conveyor System (Option)........................................................................................................ 3-11

3.5 Control Enhancement System (Option) ............................................................................................ 3-12

3.6 Process Data Validation System (PDVS) (Option) ............................................................................ 3-12

4 OPERATING INSTRUCTIONS ..............................................................................4-1

4.1 Before Operating Equipment.............................................................................................................. 4-1

4.2 Accessories ........................................................................................................................................ 4-2

4.2.1 Hold-Down Screen, Instrument Tray and Instrument Stringers ............................................. 4-3

4.2.2 Vision Multi-Function Racks and General Purpose Vision Rack............................................ 4-5

4.2.3 Multi-Level Vision Manifold Racks ......................................................................................... 4-7

4.2.4 MIS and AN Vision SC Manifold Racks ............................................................................... 4-11

4.2.5 da Vinci Rack....................................................................................................................... 4-13

4.3 Flexible Hoses For Cannulated Instruments (Accessory) ................................................................ 4-14

4.3.1 Connecting Cannulated Instruments ................................................................................... 4-14

4.3.2 DOs...................................................................................................................................... 4-16

4.3.3 DON’Ts ................................................................................................................................ 4-16

4.4 Manual Loading ................................................................................................................................ 4-19

4.5 Automatic Loading............................................................................................................................ 4-21

4.6 Manual Entry of Cycle to be Processed ........................................................................................... 4-25

4.7 Typical Cycle Operation ................................................................................................................... 4-27

4.7.1 General ................................................................................................................................ 4-27

4.7.2 Pre-Wash Phase .................................................................................................................. 4-29

4.7.3 Wash 1 Phase – If Using Prolystica® Ultra Concentrate Chemicals .................................... 4-29

4.7.4 Pulsed Enzyme Phase – If Not Using Prolystica Ultra Concentrate Chemicals .................. 4-29

4.7.5 Wash Phase or Wash 1 Phase – If Not Using Prolystica Ultra Concentrate Chemicals ...... 4-30

4.7.6 Wash 2 Phase...................................................................................................................... 4-30

4.7.7 Rinse Phase......................................................................................................................... 4-30

(Continued...)

vi 920508-556 Operator Manual Table of Contents

Number Description Page 4 OPERATING INSTRUCTIONS (Cont’d)

4.7.8 Thermal Rinse Phase........................................................................................................... 4-31

4.7.9 Drying Phase ....................................................................................................................... 4-31

4.8 Unloading Washer/Disinfector .......................................................................................................... 4-32

4.9 Responding to a Warning Message ................................................................................................. 4-34

4.10 Responding to an Alarm................................................................................................................... 4-34

4.11 Pause a Cycle................................................................................................................................... 4-36

4.12 Abort a Cycle .................................................................................................................................... 4-36

4.13 Shutdown.......................................................................................................................................... 4-37

4.14 Power Failure .................................................................................................................................... 4-37

4.14.1 Door Obstruction ................................................................................................................. 4-38

4.14.2 Doors Operation During a Power Failure............................................................................. 4-39

4.15 Visual Inspection............................................................................................................................... 4-39

5 CYCLE AND CONTROL VALUE PROGRAMMING.................................................5-1

5.1 General ............................................................................................................................................... 5-1

5.2 Supervisor Mode................................................................................................................................. 5-9

5.2.1 Supervisor Mode Menus........................................................................................................ 5-9

5.2.2 Accessing Supervisor Mode................................................................................................ 5-10

5.3 Supervisors Menu ............................................................................................................................. 5-12

5.4 Cycles Menu..................................................................................................................................... 5-13

5.4.1 Modifying a Cycle ................................................................................................................ 5-13

5.4.2 Copying a Cycle .................................................................................................................. 5-18

5.5 Machine Setup Menu........................................................................................................................ 5-20

5.6 Print Menu......................................................................................................................................... 5-26

5.7 Process Verification – Chemical Injection Range ............................................................................. 5-27

5.8 Using Control Enhancement System (Option).................................................................................. 5-28

5.9 Using Process Data Validation System (PDVS) (Option).................................................................. 5-29

6 ROUTINE MAINTENANCE....................................................................................6-1

6.1 General ............................................................................................................................................... 6-1

6.2 Routine Maintenance .......................................................................................................................... 6-2

6.3 Debris Screen ..................................................................................................................................... 6-4

(Continued...)

vii

Table of Contents Operator Manual 920508-556

Number Description Page 6 ROUTINE MAINTENANCE (Cont’d)

6.4 Water Outlet Connection..................................................................................................................... 6-4

6.5 Clean Washer/Disinfector Exterior ...................................................................................................... 6-5

6.6 DECONTAM Cycle.............................................................................................................................. 6-6

6.7 Chemical Container Replacement...................................................................................................... 6-8

6.8 Changing Paper Roll and Printer Ink Cartridge .................................................................................. 6-9

6.8.1 Printer Paper Roll ................................................................................................................... 6-9

6.8.2 Printer Ink Cartridge............................................................................................................. 6-11

6.9 Inspect and Clean Rotary Spray Arm Assemblies ........................................................................... 6-12

6.10 Door Safety Test ............................................................................................................................... 6-15

6.11 Chamber Water Level Floats ............................................................................................................ 6-17

7 TROUBLESHOOTING...........................................................................................7-1

7.1 Warning Messages Chart .................................................................................................................. 7-2

7.2 Troubleshooting Chart ........................................................................................................................ 7-5

7.3 Alarm Messages Chart ....................................................................................................................... 7-8

8 REPLACEMENT PARTS AND CONSUMABLE PRODUCTS ...................................8-1 viii 920508-556 Operator Manual Table of Contents

LIST OF ILLUSTRATIONS

Figure Number Description Page Section 3: Component Identification

Figure 3-1. Washer/Disinfector ................................................................................................................ 3-1

Figure 3-2. Chemical Injection Pumps..................................................................................................... 3-3

Figure 3-3. Chamber and Service Access Interior .................................................................................. 3-4

Figure 3-4. Control System Location........................................................................................................ 3-5

Figure 3-5. Printer .................................................................................................................................. 3-10

Figure 3-6. Air Compressor (Accessory) ............................................................................................... 3-10

Figure 3-7. SCS Conveyor System (Option) .......................................................................................... 3-11

Section 4: Operating Instructions

Figure 4-1. Chemical Container ............................................................................................................... 4-2

Figure 4-2. Flexible Hold-Down Screen ................................................................................................... 4-3

Figure 4-3. Mesh Instrument Tray ............................................................................................................ 4-3

Figure 4-4. Instrument Stringer ................................................................................................................ 4-4

Figure 4-5. Vision Multi-Function Rack For Small Items........................................................................... 4-5

Figure 4-6. Vision Multi-Function Rack For Large Items .......................................................................... 4-5

Figure 4-7. General Purpose Vision Rack................................................................................................ 4-6

Figure 4-8. Two-Level Vision Manifold Rack and Vision Multi-Function Racks ....................................... 4-7

Figure 4-9. Three-Level Vision Manifold Rack and Vision Multi-Function Rack....................................... 4-8

Figure 4-10. Four-Level Vision Manifold Rack ........................................................................................... 4-9

Figure 4-11. Four-Level Vision SC Manifold Rack ..................................................................................... 4-9

Figure 4-12. Five-Level Vision Manifold Rack.......................................................................................... 4-10

Figure 4-13. MIS Vision SC Manifold Rack .............................................................................................. 4-11

Figure 4-14. AN Vision SC Manifold Rack ............................................................................................... 4-12

Figure 4-15. da Vinci Rack....................................................................................................................... 4-13

Figure 4-16. Flexible Hoses Connected to Irrigation Ports ...................................................................... 4-14

Figure 4-17. Examples of Cannulated Instruments.................................................................................. 4-14

Figure 4-18. Flexible Hose and Cannulated Instrument Connection....................................................... 4-15

Figure 4-19. Correct Placement of Flexible Hoses For Cannulated Instruments..................................... 4-17

Figure 4-20. Incorrect Placement of Flexible Hoses and Cannulated Instruments ................................. 4-18

Figure 4-21. Bar Code Tag Insertion ....................................................................................................... 4-22

Section 6: Routine Maintenance

Figure 6-1. Chemical Container ............................................................................................................... 6-8

Figure 6-2. Open Printer Door.................................................................................................................. 6-9

Figure 6-3. Tear Paper ............................................................................................................................. 6-9

Figure 6-4. Insert Paper Roll .................................................................................................................. 6-10

Figure 6-5. Ink Cartridge Location ......................................................................................................... 6-11

(Continued...)

ix List of Illustrations Operator Manual 920508-556

LIST OF ILLUSTRATIONS (Cont’d)

Figure Number Description Page Section 6: Routine Maintenance (Cont’d)

Figure 6-6. Cleaning Spray Arms and End Spray Jets .......................................................................... 6-12

Figure 6-7. Top Rotary Spray Arm Assembly ........................................................................................ 6-13

Figure 6-8. Bottom Rotary Spray Arm Assembly ................................................................................... 6-14

Figure 6-9. Detection Bar and Holes ..................................................................................................... 6-16

Figure 6-10. Chamber Water Level Float ................................................................................................. 6-17 x 920508-556 Operator Manual List of Illustrations

Table Number Description Page

LIST OF TABLES

Section 1: Safety Precautions

Table 1-1. Definition of Symbols ................................................................................................................ 1-4

Table 1-2. Definition of Symbols ................................................................................................................ 1-4

Section 2: Installation Verification

Table 2-1. Cycles Using Prolystica Ultra Concentrate Chemicals............................................................. 2-5

Table 2-2. Prolystica Ultra Concentrate Chemicals Specifications............................................................ 2-6

Table 2-3. Cycles Using Other STERIS Chemicals.................................................................................... 2-7

Table 2-4. Other STERIS Chemicals Specifications................................................................................... 2-8

Section 4: Operating Instructions

Table 4-1. Cycle Names and Bar Code Tag Numbers ............................................................................ 4-21

Section 5: Cycle and Control Value Programming

Table 5-1. Reliance Vision – Cycle Description Chart (With Prolystica Ultra Concentrate Chemicals) ........... 5-3

Table 5-2. Reliance Vision – Cycle Description Chart: ROBOTIC Cycle (With Prolystica Ultra

Concentrate Chemicals) .............................................................................................. 5-5

Table 5-3. Reliance Vision – Cycle Description Chart (With Other STERIS Chemicals) ............................ 5-6

Table 5-4. Chemical Injection Acceptance Range .................................................................................. 5-27

Table 5-5. Monitored Parameters (Alarm Generation) ............................................................................. 5-28

Table 5-6. Monitored Parameters (Record and Storage)......................................................................... 5-28

Table 5-7. Monitored Parameters (Alarm Generation) ............................................................................. 5-29

Table 5-8. Monitored Parameters (Record and Storage)......................................................................... 5-29

Section 6: Routine Maintenance

Table 6-1. Routine Maintenance - Check List ............................................................................................ 6-2

Section 8: Replacement Parts and Consumable Products

Table 8-1. Replacement Parts and Consumables ..................................................................................... 8-1 xi List of Tables Operator Manual 920508-556

1SAFETY PRECAUTIONS

The following Safety Precautions must be observed when operating or servicing the Reliance® Vision® Single- Chamber Washer/Disinfector. WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For emphasis, certain Safety Precautions are repeated throughout the manual.

It is important to review ALL Safety Precautions before operating or servicing the washer/disinfector.

Strictly following these Safety Precautions enhances the Customer’s ability to safely and effectively utilize the unit and helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own safety policies and procedures to enhance and complement these Safety Precautions.

WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:

WARNING – PERSONAL INJURY HAZARD:

WARNING – ELECTRIC SHOCK AND/OR BURN HAZARD:

Always load empty baskets on appropriate loading cart or surface.

Always use a hold-down screen to secure small or light items. While loading instrument racks, do not place miscellaneous articles on the sides of instrument trays. Incorrect rack loading/overloading could lead to injury and/or damage to the equipment or instruments.

If an obstruction is present in the chamber door, obstruction sensor will detect obstruction and door will not close. Wait until water flow has stopped before removing an obstruction.

Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment. Contact STERIS to schedule preventive maintenance or obtain the necessary maintenance manual if preventive maintenance is performed by Customer.

Maximum operating weight, including weight of basket and load, is 61 kg [135 lb] for Reliance Vision Single-Chamber Washer/Disinfector.

Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel. Repairs and adjustments performed by unqualified personnel or installation of unauthorized parts could cause personal injury, result in improper equipment performance, void the warranty or result in costly damage. Contact STERIS regarding service options.

Risk of pinch point between door and threshold when the door opens. Keep fingers away from threshold.

Risk of pinch point between door and upper panel. Do not push on top portion of doors; do not push on door when door is rising; do not push on door when door is jammed.

Fasteners and star washers are used to ensure protective bonding continuity. Always reinstall any star washer which may have been removed during installation or servicing.

1-1 Safety Precautions Operator Manual 920508-556

WARNING – ELECTRIC SHOCK AND/OR BURN HAZARD (Cont’d):

WARNING – POSSIBLE FIRE HAZARD:

WARNING – CHEMICAL BURN AND/OR EYE INJURY HAZARD:

WARNING – BURN HAZARD:

WARNING – SLIPPING HAZARD:

Pressing bottom portion of POWER (ON/OFF) toggle switch to turn off washer/disinfector DOES NOT cut off electrical power. Lockout/Tagout building electrical disconnect switch to turn off power to washer/ disinfector for maintenance or repairs.

Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel. Service personnel must disconnect all utilities to unit before servicing. No one should service unit unless all utilities have been properly locked out. Always follow local electrical codes and safety-related work practices.

When washer/disinfector is installed on a combustible floor, floor must be covered with a metal sheet extending to the outer edges of washer/disinfector.

Chemicals are caustic and can cause adverse effects to exposed tissues. Do not allow chemicals used in the washer/disinfector to contact eyes, skin, or do not attempt to swallow. Read and follow the precautions and instructions on the chemical label and in the Safety Data Sheet (SDS) prior to handling the chemical, refilling the chemical container or servicing the chemical injection pump. Wear appropriate Personal Protective Equipment (PPE) whenever handling the chemical or servicing the chemical injection pump and lines.

Wear gloves and eye protection when using a chemical. Avoid contact with the eyes or skin. If a chemical spills or splashes on you, flush the affected area with water for 15 minutes. If swallowed, DO NOT induce vomiting. Seek medical attention immediately.

Except for an emergency, do not open door while cycle is in progress. In an emergency, first stop cycle by pressing STOP followed by Abort on touch screen. Wait for water flow to stop. Once door is opened, wear appropriate Personal Protective Equipment (PPE) whenever reaching into the chamber.

If an obstruction is present in chamber door, DO NOT attempt to remove the object. Door automatically raises and remains open. Hot water and steam may be sprayed through door opening. Always wear appropriate Personal Protective Equipment (PPE) and wait for water flow to stop.

Inner surfaces of washer/disinfector are very hot after cycle is completed. Allow washer/disinfector to cool down before touching chamber. Always wear appropriate Personal Protective Equipment (PPE), including gloves as well as a face shield, and avoid all contact with inner walls when reaching into chamber.

To prevent slips, keep floors dry. Promptly clean up any spills or drippage. For spills or drippage of detergents or other chemicals, follow safety precautions and handling procedures set forth on detergent or chemical label and/or Safety Data Sheet (SDS).

1-2 920508-556 Operator Manual Safety Precautions

CAUTION – POSSIBLE EQUIPMENT DAMAGE:

Always select a cycle appropriate for the items being processed. Failure to do so may result in product damage.

Filters may be damaged from foreign oils and other contaminants. When replacing the filters, do not touch new filter with bare hands. Open plastic wrapping and hold filter through the wrapping when inserting into the filter housing.

Suction tips (as well as other items) must be fully enclosed within the accessory (curved suction tips facing inward) to prevent obstruction in conveyor system.

Use nonabrasive cleaners when cleaning unit. Follow directions on containers and rub in a back-and-forth motion (in same direction as surface grain). Abrasive cleaners will damage stainless steel.

Cleaners rubbed in a circular motion or applied with a wire brush or steel wool on door and chamber assemblies can be harmful to stainless steel. Do not use these cleaners on painted surfaces.

Use silicone lubricant to lubricate squeeze tubes. Petroleum-based lubricants, such as petroleum jelly or grease, will damage squeeze tubes.

When choosing a detergent, select one with a low-chloride content. Detergents with a high-chloride content must not be used, as such detergents may harm stainless steel.

Wipe all plastic surfaces or painted surfaces with 70% isopropyl alcohol cleaner (such as glass or countertop cleaners).

1-3 Safety Precautions Operator Manual 920508-556

The table below contains symbols which may be on your washer/disinfector components:

The table below contains symbols which may be on the identification nameplate of your washer/disinfector:

Table 1-1. Definition of Symbols

Symbol Definition

Protective Earth (Ground).

Warning! Risk of Electrical Shock.

Attention. Refer to Manual For Further Instructions.

ON (For Part of Equipment).

OFF (For Part of Equipment).

Transfer of Heat, Hot Surface.

Risk of Crushing Fingers or Hands.

Table 1-2. Definition of Symbols

Symbol Definition

MODEL Model of Washer/Disinfector.

S/N Serial Number of Washer/Disinfector.

k Power Rating of Washer/Disinfector.

V_~ Volt, Number of Phase (3 or 1), Alternating Current.

A Amperage Rating of Washer/Disinfector.

Year Year of Manufacture of Washer/Disinfector.

Hz Hertz – Frequency of Washer/Disinfector.

WIRE Number of Wires in Electrical Cable Excluding Protective Ground Wire.

1-4 920508-556 Operator Manual Safety Precautions

2INSTALLATION VERIFICATION

2.1 Technical

Specifications

These specifications are intended to describe the technical information located on the identification nameplate of the Reliance®

Vision® Single-Chamber Washer/Disinfector and to state other relevant information. Verify equipment drawing (920-508-564) or identification nameplate (located on the front bottom part of the load side) for proper voltage and amperage.

2.1.1 Voltage, Amperage and

Power Consumption

The washer/disinfectors operate on:

Reliance Vision Single-Chamber Washer/Disinfector:

• 200-208 V, three-phase, 60 Hz, three-wire + PE, steam-heated;

• 380-415 V, three-phase, 50 Hz, three-wire + PE, steam-heated;

• 380-415 V, three-phase, 50 Hz, three-wire + PE, electric-heated;

• 460-480 V, three-phase, 60 Hz, three-wire + PE, steam-heated;

• 460-480 V, three-phase, 60 Hz, three-wire + PE, electric-heated.

Optional low consumption:

• 200-208 V, three-phase, 50 Hz, three-wire + PE, steam-heated;

• 200-208 V, three-phase, 60 Hz, three-wire + PE, steam-heated.

• A protective ground conductor is required (Class 1 Equipment).

• The maximum currents and power consumptions are indicated on the identification nameplate (located on the front bottom part of the load side).

• Main supply voltage fluctuation not exceeding ±10% of nominal voltage.

• Installation Category: Overvoltage Category II.

IMPORTANT: Customer is responsible for compliance with applicable codes and regulations.

Refer to Uncrating/Installation Instructions (P920508-557) for proper connection.

2-1 Installation Verification Operator Manual 920508-556

2.1.2 Noise Level Equivalent Sound Pressure Level at work station (measured 36" [1.0 m] away from equipment and at 63" [1.6 m] from ground):

58.2 dB (A). (Results determined according to ISO-3746: 1995 Second Edition Standard: Acoustics Determination of Sound Power Levels of Noise Sources Survey Method.)

2.1.3 Permissible

Environmental Conditions

The Reliance Vision Single-Chamber Washer/Disinfector is designed to operate under the following conditions:

- Indoor use only;

- Altitude of operation up to 6560 ft (2000 m);

- Temperature of 41 to 104°F (5 to 40°C);

- Relative humidity of 30% to 80% non-condensing;

- Pollution Degree 2.

Pollution degree 2: Equipment must be installed in an environment where normally only non-conductive pollution occurs but where occasional, temporary conductivity caused by condensation can be expected (as determined according to International Standard EN/IEC 61010-1, Second Edition).

2.1.4 Seismic Anchorage

System (Accessory)

A seismic anchorage system (based on California requirements) is available for high-risk seismic zones.

2.2 Installation

Verification

An equipment drawing (920-508-564) showing all of the utility and space requirements was sent after the order for this equipment was received.

The clearance space specified on the equipment drawing is necessary for proper installation, operation and maintenance of the washer/ disinfector. Uncrating/Installation Instructions (P920508-557) are provided with the washer/disinfector. If these documents are missing or misplaced, contact STERIS, providing the serial number and model of the washer/disinfector. Replacement copies will be sent promptly.

After installing the equipment according to the Uncrating/Installation Instructions (P920508-557), complete the following checklist to help ensure complete and correct installation, or contact STERIS to schedule a service technician to test your installation and demonstrate proper equipment operation.

❑ Disconnect switch (not provided by STERIS), capable of being locked in OFF position for maintenance purposes, is installed in electrical supply lines near washer/disinfector. Disconnect switch must be compliant with occupational health and safety regulations, as well as electric codes. Disconnect switch must be marked as disconnecting device for equipment.

NOTE: If washer/disinfector is installed next to other equipment, disconnect switch should be located so service can be shut off to one piece of equipment at a time.

2-2 920508-556 Operator Manual Installation Verification

❑ Washer/disinfector is positioned as shown on equipment drawing (920-508-564), with required clearance space and in relation to building supply lines.

❑ Washer/disinfector is level. Use leveling legs if necessary.

❑ Building cold water line supplies water to washer/disinfector as specified on equipment drawing.

❑ Building hot water line supplies water to washer/disinfector as specified on equipment drawing.

❑ If required, building pure water line supplies water to washer/ disinfector as specified on equipment drawing.

❑ Building steam line provides steam to washer/disinfector as specified on equipment drawing (required on steam-heated washer/disinfectors only).

❑ Building condensate return line is connected to washer/ disinfector as specified on equipment drawing (required on steam-heated washer/disinfectors only).

❑ Building air supply to washer/disinfector is as specified on equipment drawing.

❑ Building waste line is connected to washer/disinfector as specified on equipment drawing.

❑ Plastic drain plate is installed over building floor drain.

❑ Excess of plastic drain plate is cut.

❑ Building ventilation system is connected to washer/disinfector as specified on equipment drawing (vented washer/disinfectors only).

❑ Electrical supply for washer/disinfector is as specified on equipment drawing.

❑ Pressure for air, water and steam supplies must be within specifications on equipment drawing.

❑ Type of chemicals used (ultra concentrate chemicals or other chemicals) is selected within Hardware menu in Service mode (refer to Uncrating/Installation Instructions [P920508-557] for more details).

❑ Chemical injection pumps are connected to proper chemical feed lines and each chemical pickup tube and low level sensor are placed in proper container.

❑ There are no loops in chemical feed lines and in chemical pickup tubes.

❑ Chemical feed line(s) excess is cut.

❑ Pickup tube(s) excess is cut.

❑ Floor underneath washer/disinfector has a noncombustible and nonslip surface or is covered with a metal sheet extending to the outer edges of the washer/disinfector.

❑ Walls surrounding washer/disinfector have a noncombustible surface.

IMPORTANT: After a few weeks of operation, inspect washer/ disinfector for leaks. Retighten all clamps and connections.

WARNING – POSSIBLE FIRE

HAZARD: When washer/ disinfector is installed on a combustible floor, floor must be covered with a metal sheet extending to the outer edges of washer/disinfector.

2-3

2.3 Chemicals and

Cycles

Maximizing the efficiency and the life of the washer/disinfector is predicated on the type of water and the cleaning chemistries used.

The corrosive effects of water (deionized, soft and hard water) and contaminants found in water (such as hardness cations, corrosive anions, metals, etc.) can shorten the useful life of both instrumentation and equipment. Hard water can result in scale formation as well, which can deposit on both equipment and instrumentation. It is also important for the cleaning chemistries utilized to have low and controlled foam to maintain the pressure in the washer/disinfector for adequate impingement and effective rinsing.

Recommendations for cleaning chemistries should take into consideration the make-up water characteristics, soil load and the type and composition (stainless steel, aluminum, plastics, etc.) of instrumentation or utensils to be cleaned. Cleaning chemistries used should maintain effectiveness regardless of water quality, type or the contaminants contained in the water and they should be formulated to help prevent corrosion of instrumentation and equipment due to the use of water in the cleaning process.

2.3.1 Using Prolystica Ultra

Concentrate Chemicals

Prolystica Ultra Concentrate Chemicals are used for processing loads within the washer/disinfector:

• Prolystica® Ultra Concentrate Enzymatic Cleaner – (liquid formula) a low-foaming, easy to rinse ultra concentrate using ten times less product than traditional enzymatic cleaners.

• Prolystica® Ultra Concentrate Neutral Detergent – (liquid formula) a low-foaming, easy to rinse ultra concentrate using ten times less product than traditional neutral detergents.

• Prolystica® Ultra Concentrate Alkaline Detergent – (liquid formula) a low-foaming, easy to rinse ultra concentrate using ten times less product than traditional alkaline detergents.

• Prolystica® Ultra Concentrate Lubricant – (non-silicone-based formula) an ultra concentrate for instrument lubrication using ten times less product than traditional instrument lubricants.

Reliance Vision Single-Chamber Washer/Disinfector

Using Prolystica® Ultra Concentrate Chemicals

(see SECTION 2.3.1 as well as Table 5-1 and Table 5-2)

Using Other STERIS Chemicals

(see SECTION 2.3.2 and Table 5-3) 2-4

Each load is processed using a specific cycle (see Table 2-1).

All the cycles use the Prolystica Ultra Concentrate Chemicals (refer to Tables 2-1 and 2-2 as well as Tables 5-1 and/or 5-2, Cycle Description Charts, for more details about the cycles and the Prolystica Ultra Concentrate Chemicals).

Table 5-1 and Table 5-2 – The Cycle Description Charts for the Reliance Vision Single-Chamber Washer/Disinfector give important information about the parameters within the cycles such as temperature, injection rate, etc. Use these charts as a reference while processing cycles.

Table 2-1. Cycles Using Prolystica Ultra Concentrate Chemicals

Cycle Used to Process... Chemicals That Can Be Used...

INSTRUMENTS surgical instruments.

Prolystica Ultra Concentrate Enzymatic Cleaner Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Alkaline Detergent Prolystica Ultra Concentrate Lubricant

UTENSILS

bedpans, trays, basins, bowls and theatre shoes.

Prolystica Ultra Concentrate Enzymatic Cleaner Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Alkaline Detergent

PLASTIC GOODS plasticware.

Prolystica Ultra Concentrate Enzymatic Cleaner Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Alkaline Detergent

GENTLE CYCLE delicate surgical instruments.

Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Lubricant

ORTHOPEDICS orthopedic instruments.

Prolystica Ultra Concentrate Enzymatic Cleaner Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Alkaline Detergent

RIGID MIS

rigid minimally invasive surgical instruments.

Prolystica Ultra Concentrate Enzymatic Cleaner Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Alkaline Detergent Prolystica Ultra Concentrate Lubricant

ANESTHESIA/

RESPIRATORY GOODS

anesthesia tubing, bags, masks and respiratory goods.

Prolystica Ultra Concentrate Enzymatic Cleaner Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Alkaline Detergent Prolystica Ultra Concentrate Lubricant

ROBOTIC EndoWrist®1 instruments.

Prolystica Ultra Concentrate Enzymatic Cleaner Prolystica Ultra Concentrate Neutral Detergent Prolystica Ultra Concentrate Alkaline Detergent

1EndoWrist is a registered trademark of Intuitive Surgical, Inc.

2-5 Installation Verification Operator Manual 920508-556 vhanynwattsb Highlight vhanynwattsb Highlight

IMPORTANT: STERIS does not promote, recommend or endorse the use of any other type of chemical additives and chemicals in the processing of articles in the Reliance…