Copy of JUN 2026 EQ CONSOLIDATION - Schedule of Line Items with QA _Updated_HTM.xlsx

XLSX spreadsheet 55 KB Posted

Attached to
6525--JUN 2026 Equipment Only Consolidation
Solicitation number
36A79726Q0011
Issued by

About this file

This is a Schedule of Line Items with Quality Assurance documentation for the VA's June 2026 Equipment Only Consolidation procurement (Solicitation 36A79726Q0011), administered by the Department of Veterans Affairs National Acquisition Center. The document outlines medical imaging and radiological equipment requirements across multiple VA facilities organized by VISN (Veterans Integrated Service Network), with detailed specifications, vendor questions, and government responses.

The procurement encompasses approximately 200+ line items distributed across 23 VISNs and multiple VA medical centers nationwide, including equipment such as X-ray systems (portable and fixed), ultrasound systems (point-of-care and advanced diagnostic), computed tomography (CT), magnetic resonance imaging (MRI), positron emission tomography/CT (PET/CT), nuclear medicine, interventional radiology C-arms, and mammography systems. Equipment quantities range from single units to multiple systems per location, with specifications detailing technical requirements including detector sizes, pixel specifications, frequency ranges, battery life, weight limitations, and software capabilities. The QA section documents vendor inquiries and government responses regarding equipment specifications, clinical equivalencies, and requirements modifications. Significant modifications include weight increases for portable ultrasound systems (from 240-265 lbs), screen size reductions to enhance competition (19-21 inch monitors), removal of vendor-specific features (such as PureWave and xPlane imaging), and relocation of advanced features to value-added sections. The government consistently maintained clinical requirements while accepting commercially equivalent alternatives where appropriate, particularly for detector specifications, spatial resolution, and imaging modes, to promote fair and open competition among multiple vendors.

View the file

Other files

File Type Posted
Updated Tech Specs.zip ZIP file
36A79726Q0011 0001.docx DOCX document
662B60031-757B60015.zip ZIP file
Vendor Workup Spreadsheet - 36A79726Q0011.xlsx XLSX spreadsheet
VA Directive 6550 - New Appendix A - published Jun. 2023.pdf PDF
JUN 2026 Equipment Only Consolidation Schedule.pdf PDF
36A79726Q0011_1.docx DOCX document
EQ JUN 2026 CONSOLIDATION - Schedule of Line Items with QA Template-.xlsx XLSX spreadsheet
589B69013- 662B60015.zip ZIP file
570B60012- 589B63022.zip ZIP file
402B61001-570B60011.zip ZIP file
Vendor Folders.zip ZIP file
Memorandum for Training.pdf PDF
JUN 2026 EQ Consolidation - RFO.pdf PDF
Show all 14

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SCHEDULE OF LINE ITEMS

HTME VISNSTATIONEQUIPMENT POEQUIPMENT TYPEQTY
VISN 1: VA New England Healthcare System402 AUGUSTA ME402B61001XR US1
VISN 1: VA New England Healthcare System523 JAMAICA PLAIN MA523B61010XR US2
VISN 10: VA Healthcare System of Ohio506 ANN ARBOR MI506B60039XR US2
VISN 10: VA Healthcare System of Ohio538 CHILLICOTHE OH538B60001XR US2
VISN 10: VA Healthcare System of Ohio552 DAYTON OH552B50018XR US4
VISN 10: VA Healthcare System of Ohio583 INDIANAPOLIS IN583B60039XR US3
VISN 10: VA Healthcare System of Ohio757 COLUMBUS OH757B60015XR US1
VISN 12: VA Great Lakes Health Care System578 HINES IL578B68013XR US1
VISN 12: VA Great Lakes Health Care System578 HINES IL578B68014XR US1
VISN 15: VA Heartland Network589 TOPEKA KS589B62002XR PORT C ARM1
VISN 15: VA Heartland Network589 COLUMBIA MO589B63016XR US1
VISN 15: VA Heartland Network589 COLUMBIA MO589B63022XR US2
VISN 16: South Central VA Health Care Network564 FAYETTEVILLE AR564B60002XR US5
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60005XR US3
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60006XR US2
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60015XR US1
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60019XR US7
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60020XR US1
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60022XR US2
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60023XR US1
VISN 16: South Central VA Health Care Network580 HOUSTON TX580B60025XR US1
VISN 16: South Central VA Health Care Network598 LITTLE ROCK AR598B60018XR MAMMO1
VISN 16: South Central VA Health Care Network598 LITTLE ROCK AR598B60020XR MAMMO1
VISN 17: VA Heart of Texas Health Care Network549 DALLAS TX549B60505XR US9
VISN 17: VA Heart of Texas Health Care Network671 SAN ANTONIO TX671B60042XR PORT C ARM1
VISN 17: VA Heart of Texas Health Care Network756 EL PASO TX756B65006XR US1
VISN 17: VA Heart of Texas Health Care Network756 EL PASO TX756B65007XR PORT C-ARM2
VISN 19: VA Rocky Mountain Network554 COLORADO SPRING CO554B60021XR-PORT C-ARM1
VISN 19: VA Rocky Mountain Network635 OKLAHOMA CITY OK635B60002XR US1
VISN 19: VA Rocky Mountain Network635 OKLAHOMA CITY OK635B60002XR US1
VISN 19: VA Rocky Mountain Network635 OKLAHOMA CITY OK635B60003XR PORT C ARM1
VISN 2: VA Healthcare Network Upstate New York526 BRONX NY526B60005XR US1
VISN 2: VA Healthcare Network Upstate New York632 NORTHPORT NY632B61006XR US1
VISN 20: VA Northwest Network648 PORTLAND OR648B63006XR US1
VISN 20: VA Northwest Network648 PORTLAND OR648B63007XR US1
VISN 20: VA Northwest Network668 SPOKANE WA668B63003XR US2
VISN 21: VA Sierra Pacific Network459 HONOLULU HI459B60029XR US3
VISN 21: VA Sierra Pacific Network570 FRESNO CA570B60010XR RAD1
VISN 21: VA Sierra Pacific Network570 FRESNO CA570B60011XR RAD1
VISN 21: VA Sierra Pacific Network570 FRESNO CA570B60012XR RAD1
VISN 21: VA Sierra Pacific Network612 MARINEZ CA612B69010XR US1
VISN 21: VA Sierra Pacific Network640 PALO ALTO CA640B64053XR US5
VISN 21: VA Sierra Pacific Network640 PALO ALTO CA640B64054XR US1
VISN 21: VA Sierra Pacific Network640 PALO ALTO CA640B64056XR US1
VISN 21: VA Sierra Pacific Network640 PALO ALTO CA640B64057XR PORT C ARM1
VISN 21: VA Sierra Pacific Network640 PALO ALTO CA640B64058XR PORT C ARM1
VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO CA662B60014XR US3
VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO CA662B60015XR US2
VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO CA662B60025XR US1
VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO CA662B60031XR PORT C ARM1
VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO CA662B60032XR US4
VISN 22: VA Desert Pacific Healthcare Network644 PHOENIX AZ644B68023XR PORT C ARM2
VISN 22: VA Desert Pacific Healthcare Network644 PHOENIX AZ644B68024XR PORT C ARM1
VISN 22: VA Desert Pacific Healthcare Network644 PHOENIX AZ644B68025XR PORT C ARM1
VISN 22: VA Desert Pacific Healthcare Network664 SAN DIEGO CA664B60058XR US1
VISN 22: VA Desert Pacific Healthcare Network678 TUCSON AZ678B61031XR CT1
VISN 22: VA Desert Pacific Healthcare Network678 TUCSON AZ678B61032XR MRI1
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES CA691B60052XR PORT C ARM1
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES CA691B60053XR US2
VISN 23: VA Midwest Health Care Network499 SIOUX FALLS SD499B69000XR US1
VISN 23: VA Midwest Health Care Network499 IOWA CITY IA499B69004XR URO1
VISN 23: VA Midwest Health Care Network499 IOWA CITY IA499B69005XR US2
VISN 4: VA Stars & Stripes Healthcare Network460 WILMINGTON DE460B68008XR US1
VISN 4: VA Stars & Stripes Healthcare Network460 WILMINGTON DE460B68009XR US1
VISN 4: VA Stars & Stripes Healthcare Network595 LEBANON PA595B69013XR RAD1
VISN 4: VA Stars & Stripes Healthcare Network642 PHILADELPHIA PA642B69014XR US2
VISN 4: VA Stars & Stripes Healthcare Network642 PHILADELPHIA PA642B69016XR US1
VISN 4: VA Stars & Stripes Healthcare Network693 NEW ALLENTOWN PA693B63004XR MAMMO1
VISN 4: VA Stars & Stripes Healthcare Network693 WILKES BARRE PA693B63005XR RAD1
VISN 4: VA Stars & Stripes Healthcare Network693 WILKES BARRE PA693B63006XR US1
VISN 4: VA Stars & Stripes Healthcare Network693 WILKES BARRE PA693B63007XR RF1
VISN 4: VA Stars & Stripes Healthcare Network693 WILKES BARRE PA693B63008NM BONE DEN1
VISN 5: VA Capitol Health Care Network688 WASHINGTON DC688B61003XR US2
VISN 6: VA Mid-Atlantic Health Care Network558 DURHAM NC558B69009XR MAMMO1
VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NC565B69015XR US1
VISN 6: VA Mid-Atlantic Health Care Network637 ASHEVILLE NC637B69003NM PET CT1
VISN 6: VA Mid-Atlantic Health Care Network659 SALISBURY NC659B69001XR US1
VISN 7: VA Southeast Health Care Network508 DECATUR GA508B62052XR US1
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM AL521B62003XR US1
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM AL521B62004XR PORT RAD1
VISN 7: VA Southeast Health Care Network521 BIRMINGHAM AL521B62007XR PORT RAD1
VISN 7: VA Southeast Health Care Network534 CHARLESTON SC534B62034NM BONE DEN1
VISN 7: VA Southeast Health Care Network544 COLUMBIA SC544B62014XR MAMMO1
VISN 7: VA Southeast Health Care Network557 DUBLIN GA557B62003XR PORT RAD2
VISN 7: VA Southeast Health Care Network557 DUBLIN GA557B62004XR US1
VISN 8: VA Sunshine Healthcare Network516 BAY PINES FL516B60008XR US2
VISN 8: VA Sunshine Healthcare Network546 MIAMI FL546B60023XR US1
VISN 8: VA Sunshine Healthcare Network546 MIAMI FL546B60024XR US2
VISN 8: VA Sunshine Healthcare Network546 MIAMI FL546B60025XR US2
VISN 8: VA Sunshine Healthcare Network573 TALLAHASSEE FL573B65025XR US1
VISN 8: VA Sunshine Healthcare Network573 TALLAHASSEE FL573B65026NM BONE DEN1
VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FL573B65030XR US2
VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FL573B65031XR US1
VISN 8: VA Sunshine Healthcare Network573 LAKE CITY FL573B65032XR PORT C ARM1
VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FL573B65034XR US1
VISN 8: VA Sunshine Healthcare Network573 LAKE CITY FL573B65035XR US2
VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FL573B65036XR US1
VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FL573B65037XR US1
VISN 8: VA Sunshine Healthcare Network573 TALLAHASSEE FL573B65038XR US2
VISN 8: VA Sunshine Healthcare Network672 SAN JUAN PR672FC6008XR US1
VISN 8: VA Sunshine Healthcare Network672 SAN JUAN PR672FC6012XR HYBRID1
VISN 8: VA Sunshine Healthcare Network675 ORLANDO FL675B67006XR MRI1
VISN 8: VA Sunshine Healthcare Network675 ORLANDO FL675B67007XR IR1
VISN 8: VA Sunshine Healthcare Network675 ORLANDO FL675B67012XR US2
VISN 8: VA Sunshine Healthcare Network675 ORLANDO FL675B67014XR US2
VISN 8: VA Sunshine Healthcare Network675 ORLANDO FL675B67015XR US1
VISN 8: VA Sunshine Healthcare Network675 ORLANDO FL675B67016XR PORT C ARM1
VISN 8: VA Sunshine Healthcare Network675 ORLANDO FL675B67017XR MRI1
VISN 9: VA Mid South Healthcare Network621 MOUNTAIN HOME TN621B60002XR US2
VISN 9: VA Mid South Healthcare Network626 CHATTANOOGA TN626B64028XR US1

QA TEMPLATE

EQUIPMENT POVISNSTATIONEQUIPMENT TYPEVENDOR EQUIPMENT QUESTIONSGOVERNMENT EQUIPMENT RESPONSE
460B68009VISN 4: VA Stars & Stripes Healthcare Network460 WILMINGTON DEXR USRequest item 1.k. Maximum equipment weightbe increased from 87 lbs to 141 lbs.
Under Section 6. Analysis Section - do you require any automated software for IUD Placement?This Feature is not a requirement
460B68009VISN 4: VA Stars & Stripes Healthcare Network460 WILMINGTON DEXR US1.k. Can the maximum equipment weight be raised to 126 lbs to increasee competition?YES
460B68009VISN 4: VA Stars & Stripes Healthcare Network460 WILMINGTON DEXR US7.a. Can the PureWave feature be removed from the requirement as it appears to be brand-specific?YES
499B69000VISN 23: VA Midwest Health Care Network499 SIOUX FALLS SDXR USVISN 23: VA Sious Falls

What surgical procedures are to be performed now or in the future or is this soley for Transrectal and Transperineal Prostate procedures? If required for surgery, additional software applications and probes are required.

Image Fusion Software. Please confirm what procedures will be performed utilizing Fusion.

If Fusion is required, will it be needed for TP, TR, or both?

If Fusion is required, will it be required to be freehand or stabilized?

System solely used for TR and TP procedures.

No Image fusion software.

499B69004VISN 23: VA Midwest Health Care Network499 IOWA CITY IAXR UROLine 31K - Would changing the minimum Xray monitor size to 19 inches be considered a clinically equivalent or substantially complaint alternative to the specified 27 inch monitor size?Yes
499B69004VISN 23: VA Midwest Health Care Network499 IOWA CITY IAXR UROLine 32L-Would changing the minimum video monitor size to 19 inches be considered a clinically equivalent or substantially compliant alternative to the specified 27 inch monitor size?Yes
499B69004VISN 23: VA Midwest Health Care Network499 IOWA CITY IAXR UROline 28 h. Please verify the vertical travel spec requested. Can the vertical travel spec be changed to 20.7" for fair competition.Yes
499B69005VISN 23: VA Midwest Health Care Network499 IOWA CITY IAXR USLine 22C- Please advise if the control panel height range from the floor in 57.5-67.5 inches is correct requirement?Yes
499B69005VISN 23: VA Midwest Health Care Network499 IOWA CITY IAXR USLine 24E- Is the minimum storage of 2TB a combination of the operating software and patient information storage?Yes
499B69005VISN 23: VA Midwest Health Care Network499 IOWA CITY IAXR USLine 58I - Please explain Image editing feature and how it applies to the sonographer workflow?It allows us to optimize image after grabbed, improves sonographer productivity and exam quality
499B69005VISN 23: VA Midwest Health Care Network499 IOWA CITY IAXR USRequest item 1.g. Maximum Weight be increased to 265 lbs.Yes
506B60039VISN 10: VA Healthcare System of Ohio506 ANN ARBOR MIXR USRequest 1.a. Minimum screen size reduced to 14 in.
Request 2.m. Other scanning modes PIV moved to Added Value.1.a: Minimum screen size remains at 21 inches, as reducing it to 14 inches would significantly compromise the clinical users' needs. Clinical staff require a larger display to support effective treatment and diagnosis, which ultimately benefits patient care.
•2.m: The requirement for "Other scanning modes" is retained as a technical requirement, while reference to a specific scanning mode has been removed to allow vendors flexibility in meeting the clinical need. The underlying clinical justification is that access to multiple scanning modes is essential for staff to optimize image acquisition and adapt to diverse diagnostic and therapeutic scenarios, thereby supporting high standards of patient care and enabling effective diagnosis and treatment.
508B62052VISN 7: VA Southeast Health Care Network508 DECATUR GAXR US
Missing technical specifications and matrix formsTechnical specifications and compliance matrix added
508B62052VISN 7: VA Southeast Health Care Network508 DECATUR GAXR USBenchmark missingTechnical specifications and compliance matrix added
516B60008VISN 8: VA Sunshine Healthcare Network516 BAY PINES FLXR USRequest item 1.a. Minimum screen size be revised to 19 in.
Request item 9.g. Integration to Dose Tracking be removed.Tech Specs updated
521B62003VISN 7: VA Southeast Health Care Network521 BIRMINGHAM ALXR USA battery life of 60 minutes is required. The system will be used for liver resection and vascular surgical cases. Is a battery necessary since the system will not leave the surgical suite?Yes, the battery is necessary for transport from the room to the storage area where it will remain plugged in.
521B62003VISN 7: VA Southeast Health Care Network521 BIRMINGHAM ALXR USCan the minimum monitor size be updated to 19 inches to increase competition?

Can the minimum active screen be updated to 19 inches to increase competition?

Can the minimum transducer ports be updated to 4 to increase competition?

Can you please confirm functionality of the system as the description details liver resection and vascular procedures but lists obstetric, Mitral valve navigator, and Cardiac 3D Quantification.

Can the b. 4-way lap C9150LA frequency range be clarified.

Image Fusion Software. Please confirm what procedures will be performed utilizing Fusion.

If Fusion is required, will it be needed for TP, TR, or both?

If Fusion is required, will it be required to be freehand or stabilized? No, the minimum monitor size of 23" is requested by the clinical team for ease of viewing

No, the minimum active screen of 23" is requested by the clinical team for ease of viewing

No, the minimum transducer cannot be updated to 4

Ultrasound will be used for urology and general surgery

Can the b. 4-way lap C9150LA frequency range be clarified? Frequency range: 5.0 MHz (low) – 13.0 MHz (high). See updated Technical specifications and compliance matrix

Fusion software will not be used

521B62004VISN 7: VA Southeast Health Care Network521 BIRMINGHAM ALXR PORT RADDetector Pixel Size
•Would a detector with a 148 μm pixel size be considered a clinically equivalent or substantially compliant alternative to the specified 140 μm maximum pixel size requirement?

Display Monitor Size

• Would an integrated 17-inch display monitor be considered substantially compliant with the 21.5-inch minimum monitor size requirement if it provides equivalent workflow functionality and image review capabilities?

Detector Weight Capacity

• Would a detector with a 661 lb recumbent patient weight capacity be considered substantially compliant with the specified 881 lb distributed load capacity requirement, and if so, what clinical justification would be needed to support equivalency?

Generator kVp Range

• Would a system with a maximum generator output of 133 kVp be considered substantially compliant with the specified 150 kVp maximum generator requirement, provided all clinical imaging requirements can be met?

Display at Tube Head

• Is a dedicated monitor at the tube head a mandatory requirement, or would systems with a centralized touchscreen workstation providing equivalent positioning and workflow functionality be considered acceptable?

A detector with 148 um pixel size would be considered clinically equivalent to 140 um pixel size.

Integrated 17-inch display monitor would be considered compliant with the 21.5 if the workflow functionality and image review capabilities are comparable.

A detector with 661lbs recumbent patient weight capacity would be compliant to 881lbs distributed load capacity. The clinical need is for the bariatric patients.

Maximum generator of 133kVp would be considered compliant with the 150 kVp maximum generator requirement.

Display monitor on tube head would be a preferred option but not required. A centralized touchscreen workstation providing equivalent positioning and workflow would be considered an alternative option but not required. See updated Technical specifications and compliance matrix 521B62004 VISN 7: VA Southeast Health Care Network 521 BIRMINGHAM AL XR PORT RAD 1.a. Can you please change the generator power to 30kW for more open competition?

3.c. Please move the free release handles to value adds.

3.d. Please move the positioning light to value adds.

3.g. Please move ability to move portable from tube head to value adds.

3.p. Please move display monitor on tube head to value adds.

3.r. Please move SID positioning aides to value adds.

3.s. Please move tube and detector angle alignment software to value adds. Generator power was changed to 30kW on the updated technical specifications and compliance matrix

Free release handles were moved to Added Value on the Tech Specs and compliance matrix.

Positioning light is required.

The ability to move portable from tube head is required.

Display monitor on tube head would be a preferred option but not required.

SID positioning aides are required.

Tube and detector angle alignment software would be a preferred option but not required. This is moved to Added Value on the Tech Specs and compliance matrix.

521B62007VISN 7: VA Southeast Health Care Network521 BIRMINGHAM ALXR PORT RAD2.c. Please confirm need or remove need for navigation software. Usually this is used in IR or for procedures.
2.d. Please confirm need or remove need for image fusion software.Navigation software is needed because the IR physician use ultrasound for procedures such as injections and drainages.

Image fusion software can be an option but not required. This was moved to Added Value on the Tech Specs and compliance matrix.

526B60005 VISN 2: VA Healthcare Network Upstate New York 526 BRONX NY XR US Image Fusion Software. Please confirm what procedures will be performed utilizing Fusion.

If Fusion is required, will it be needed for TP, TR, or both?

If Fusion is required, will it be required to be freehand or stabilized?

Can 1. Clinical Training d. Technologists who complete the clinical applications training shall receive continuing education credits (CMEs) be added value?1. Fusion is required for both TP and TR. 2. Fusion should be freehand and stabilized. 3. Fine for it to be added value.
538B60001VISN 10: VA Healthcare System of Ohio538 CHILLICOTHE OHXR US2.b. Please confirm whether 3D freehand is needed. Request to have this moved to the value-add section or removed, as it is not needed.Yes, Please see revised Tech Specs and Compliance Matrix for changes.
538B60001VISN 10: VA Healthcare System of Ohio538 CHILLICOTHE OHXR US2.c. Please confirm if navigation software is needed. If not, please remove or move to value-add.Yes, Please see revised Tech Specs and Compliance Matrix for changes.
538B60001VISN 10: VA Healthcare System of Ohio538 CHILLICOTHE OHXR US2.d. Please confirm if image fusion software is needed. If not, please remove or move to value-add.Yes, Please see revised Tech Specs and Compliance Matrix for changes.
544B62014VISN 7: VA Southeast Health Care Network544 COLUMBIA SCXR MAMMOWill site also be trading in an Affirm Upright biopsy system?
If so, please provide the SN? It begins with 820We encourage and expect vendors to deinstall + trade-in old systems. Affirm Biopsy SN is 82011151298
544B62014VISN 7: VA Southeast Health Care Network544 COLUMBIA SCXR MAMMO1e. SID 70cm - to allow for fair competition please update to 65-70.

h. Max system weight 900lbs - to allow for fair competition please update to 925 to

1. What vacuum assisted device will you use for biopsy?

2. Can the window of the weight be widened to maximum weight to 925 to be more vendor neutral? Ok to set SID requirement to 65-70cm, updated on latest technical specifications and compliance matrix

OK to set weight requirement to 925lb, updated on latest technical specifications and compliance matrix

No, we are requiring a self-contained vacuum device/system. To provide otherwise will require site prep, which is not desired.

546B60023 VISN 8: VA Sunshine Healthcare Network 546 MIAMI FL XR US Line 27 - Requested height range only says 30". Is this from the floor? Our spec sheet does not specify height from floor.

Line 31 - Are they requesting a portable unit? These specifications are quite small.

Line 66 - CV systems don't offer Small Parts. Will this be necessary? Line 27- 1h. Removed specification. Original ask was total adjustable range for the control panel.

Line 31- 1l. Not requesting portable unit. Corrected equipment dimensions to 74in x 25 in x 36in

Line 66- 6d. Removed specification.

546B60024VISN 8: VA Sunshine Healthcare Network546 MIAMI FLXR USRequest item 1.a. and 1.b be reduced to 14 in.
Request item 1.i. Minimum battery life be reduced to 60 min.Tech Specs updated
546B60025VISN 8: VA Sunshine Healthcare Network546 MIAMI FLXR USImage Fusion Software. Please confirm what procedures will be performed utilizing Fusion.

If Fusion is required, will it be needed for TP, TR, or both?

If Fusion is required, will it be required to be freehand or stabilized?

This requirement is not on the tech specs. No changes made.
549B60505VISN 17: VA Heart of Texas Health Care Network549 DALLAS TXXR USLine 35B Please explain what procedures you will be using for Three-dimensional (3D) freehand scanning modeEvaluation of valvular stenosis and regurgitation, evaluation of vegetations, evaluation of Watchman and other structual devices
549B60505VISN 17: VA Heart of Texas Health Care Network549 DALLAS TXXR USLine 68J Please explain what procedures you will be using for shear wave and stain-based elastographyLV systolic function in patients on chemotherapy, infiltrative heart disease evalation
549B60505VISN 17: VA Heart of Texas Health Care Network549 DALLAS TXXR USLines 87A-91E Please clarify how you would like the quantities of the transducers reflected on the 9 unitsEach unit will have a 2D/3D transthoracic probe and pedhoff probe. 2D/3D transesophageal probes will be used on 4-5 machines.
549B60505VISN 17: VA Heart of Texas Health Care Network549 DALLAS TXXR USFor items 5q and 7.3. The solicitation calls for 9 CVUS systems, but only one ICE probe. Would you like for ICE software to be loaded on all 9 CVUS systems or just on 1 system??One system will need Versight capabilities, with a PIM cable to unlock these capabilities.
549B60505VISN 17: VA Heart of Texas Health Care Network549 DALLAS TXXR US5.j – Shear Wave and Strain-Based Elastography: Can you please clarify the intended cardiovascular clinical application for shear wave and strain-based elastography within this requirement? Is this referring to cardiac strain imaging, or is tissue elastography specifically required?

6.a – Abdominal Analysis Package: The intended clinical applications outlined in the solicitation are TTE, TEE, stress echocardiography, Doppler echocardiography, and other cardiovascular cases. Can you please clarify the intended cardiovascular clinical application for the abdominal analysis package and whether this functionality is required for the systems being procured?

7.a, 7.d & 7.e – Transducer Quantities: The solicitation requests nine ultrasound systems; however, the transthoracic echo and Doppler transducers are each listed at a quantity of 12. As well as the Intracardiac is only listed as 1. Can you please confirm the required quantities for these transducers to ensure the proposal includes the quantities intended by the VA?

5j. Yes , this is is supposed to be for LV/RV/LA strain.

6a. The abdominal package has been removed as a requirement.

7a.d.e. We need to pair each 2D/3D tranthoracic probe with each of the 9 machines. We will use the extra 3 2D/3D probes as backups if one of the 9 fail. We will pair each of the Doppler (pedoff) probes with each of the 9 machines. These probes aren't used as frequently, so we only need 9 of these rather than 12.

7.b.c [NOTE] For the TEE probes, it is possible to lower the requirement to 6 of the regular thickness probes (8-2 MHz) and 3 of the thinner probes (11-4 MHz) (rather than 8 and 4). The requirement has been adjusted.

552B50018VISN 10: VA Healthcare System of Ohio552 DAYTON OHXR US2.b. Please remove navigation software; this is not applicable for this care area.Yes, Please see revised Tech Specs and Compliance Matrix for changes.
552B50018VISN 10: VA Healthcare System of Ohio552 DAYTON OHXR US2.c. Please remove image fusion software; this is not applicable for this care area.Yes, Please see revised Tech Specs and Compliance Matrix for changes.
554B60021VISN 19: VA Rocky Mountain Network554 COLORADO SPRING COXR-PORT C-ARMline 26 f. Can orbital rotation spec be changed to 120 degrees for fair competition?Changing to 120 is acceptable
554B60021VISN 19: VA Rocky Mountain Network554 COLORADO SPRING COXR-PORT C-ARMline 27 g. Can pivot rotation spec be changed to 380 degrees for fair competition?Changing to 380 is acceptable
554B60021VISN 19: VA Rocky Mountain Network554 COLORADO SPRING COXR-PORT C-ARMline 31 k. Can minimum C-arm depth spec be changed to 18"for fair competition?Changing to 18" is acceptable
554B60021VISN 19: VA Rocky Mountain Network554 COLORADO SPRING COXR-PORT C-ARMline 34 n. Can image matrix size spec be changed to 1536x1532 pixels for fair competition?Changing the spec is acceptable
554B60021VISN 19: VA Rocky Mountain Network554 COLORADO SPRING COXR-PORT C-ARMline 47 j. Can the "rainbow" spec be changed to 120 degrees for fair competion?Current spec is 180 deg so this would be a 33% reduction and could make it more difficult to use. Do not concur.
554B60021VISN 19: VA Rocky Mountain Network554 COLORADO SPRING COXR-PORT C-ARMline 51 n. This is typically a full size c-arm feature, not available on mini, can this be moved to value add for fair competition?Removed
557B62003VISN 7: VA Southeast Health Care Network557 DUBLIN GAXR PORT RADDetector Pixel Size
•Would a detector with a 148 μm pixel size be considered a clinically equivalent or substantially compliant alternative to the specified 140 μm maximum pixel size requirement?

Display Monitor Size

• Would an integrated 17-inch display monitor be considered substantially compliant with the 21.5-inch minimum monitor size requirement if it provides equivalent workflow functionality and image review capabilities?

Detector Weight Capacity

• Would a detector with a 661 lb recumbent patient weight capacity be considered substantially compliant with the specified 881 lb distributed load capacity requirement, and if so, what clinical justification would be needed to support equivalency?

Generator kVp Range

• Would a system with a maximum generator output of 133 kVp be considered substantially compliant with the specified 150 kVp maximum generator requirement, provided all clinical imaging requirements can be met?

Display at Tube Head

• Is a dedicated monitor at the tube head a mandatory requirement, or would systems with a centralized touchscreen workstation providing equivalent positioning and workflow functionality be considered acceptable?

A detector with 148 um pixel size would be considered clinically equivalent to 140 um pixel size.

Integrated 17-inch display monitor would be considered compliant with the 21.5 if the workflow functionality and image review capabilities are comparable.

A detector with 661lbs recumbent patient weight capacity would be compliant to 881lbs distributed load capacity. Given our facility’s patient demographics, the highest patient weight encountered in the past three years has been less than 661 lbs. The portable detector would include safeguards and workflows to prevent use with patients above the manufacturer-rated capacity. The clinical risk is minimal and does not compromise patient safety or care. The 661 lb detector meets our operational needs, and alternative imaging arrangements are in place for higher-weight patients.

Maximum generator of 133kVp would be considered compliant with the 150 kVp maximum generator requirement. There is no anticipated impact on clinical care or diagnostic image quality. Exams requiring higher kVp are performed on fixed radiography systems. Thus, the proposed system with 133 kVp maximum output meets our operational and clinical requirements.

A dedicated display at the tube head is not a mandatory requirement. Systems that provide a centralized touchscreen workstation—offering equivalent functionality in terms of positioning guidance, workflow integration, and easy access to controls. A tube head display is not required, but having convenient access to critical workflow tools near the imaging area is very helpful for technologist ease of use, and the selection should consider how each detector system supports these needs. See updated technical specifications and compliance matrix 557B62003 VISN 7: VA Southeast Health Care Network 557 DUBLIN GA XR PORT RAD 1.a. Can you please change the generator power to 30kW for more open competition?

3.c. Please move the free release handles to value adds.

3.d. Please move the positioning light to value adds.

3.g. Please move ability to move portable from tube head to value adds.

3.p. Please move display monitor on tube head to value adds.

3.r. Please move SID positioning aides to value adds.

3.s. Please move tube and detector angle alignment software to value adds.Yes, please see updated technical specifications and compliance matrix.
557B62004VISN 7: VA Southeast Health Care Network557 DUBLIN GAXR US
Missing transducers on technical specification formYes, please see updated technical specifications and compliance matrix.
557B62004VISN 7: VA Southeast Health Care Network557 DUBLIN GAXR USLine 61, 70, and 73: These are not cardiovascular clinical options, should they be removed?

Line 81: Can this be changed to a Linear Array Probe (as 9L Array Probe is a single Vendor specific probe)?

Line 82: Can this be changed to a Convex Array Probe (as C1-6 Convex Array Probe is a single vendor specific probe)?

Line 83: Can this be changed to a Phased Array Probe (as 4VC-D Probe is a single Vendor specific probe)?Yes, please see updated technical specifications and compliance matrix.
558B69009VISN 6: VA Mid-Atlantic Health Care Network558 DURHAM NCXR MAMMOPlease provide the correct system trade-in SN. It will start with 810 or SDM.Updated
558B69009VISN 6: VA Mid-Atlantic Health Care Network558 DURHAM NCXR MAMMO1.q. Can the software license be changed from "C-View Software License" to synthetic software license to allow for vendor neutrality?

2.c. Curved/Flexed Paddle Options: Curved paddle is vendor specific. Please remove for open competition.

3.d. Comfort-focused paddles: This is vendor specific question. Request to be restated as "Patient comfort aides available at gantry.

4. Trade In Serial Number. Please validate serial number on Trade-in. It appears the serial number is incorrect "6525-433278" and should start with "8" versus "6".1q. Yes. Call it synthetic software license 2c. Removed from technical specification form 3d. Removed from Technical specification form and updated to Patient comfort aides available at gantry 4. SDM131700391
565B69015VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NCXR USRequest item 1.e. be reduced to 60 min.Reduced to 20 min
565B69015VISN 6: VA Mid-Atlantic Health Care Network565 FAYETTEVILLE NCXR US1.e. Please remove minimum battery life or change to 20min. No vendor can support 240min on a radiology unit.
5.n. Please remove ECG requirement as there is no stated need for a cardiac (phased array) probe.1.e reduced to 20 min

5.n. Requirement removed

573B65025VISN 8: VA Sunshine Healthcare Network573 TALLAHASSEE FLXR USThe technical specificaton & matrix forms is listed as Tallahasse, FL; however when you open them it is listed as Gainville, Fl. Can you confirm site of service address.Gainesville, FL.
573B65025VISN 8: VA Sunshine Healthcare Network573 TALLAHASSEE FLXR US1.i. Please remove minimum battery life or change to 20min.Tech Specs updated
573B65030VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USRequest item 1.f. Maximum Equipment weight be increased to 155 lbs.Tech Specs updated
573B65030VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR US2.c. Please confirm need or remove need for navigation software. Usually this is used in IR or for procedures- not seen in ICU or critical care.
2.d. Please confirm need or remove need for image fusion software.Tech Specs updated
573B65031VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USDo you need a D2 non imaging probe?Yes, D2 is required
573B65031VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USFor 2.c. Image Fusion - Can you verify if Imag Fusion Software - combining real-time Ultrasound Images with CT, MRI, or PET Scans - is required?Tech Specs updated
573B65032VISN 8: VA Sunshine Healthcare Network573 LAKE CITY FLXR PORT C ARMPlease clarify minimum horizontal travel, no c-arm meets 104cmTech Specs updated
573B65032VISN 8: VA Sunshine Healthcare Network573 LAKE CITY FLXR PORT C ARMPlease adjust minimum c-arm depth to 27" to allow for fair and open competitionTech Specs updated
573B65032VISN 8: VA Sunshine Healthcare Network573 LAKE CITY FLXR PORT C ARMPlease adjust maximum weight for fair and open competition to 920 lbsTech Specs updated
573B65034VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USRequest item 1.f. Control Panel Height be revised to 36 in.

Request item 1.i. Minimum battery life be revised to 30 min.

Request item 1.k. Maximum eqiupment weight be increased to 250.

Request item 8.a, 8.b, and 8.c. be removed as not typically done in Cardiology.Tech Specs updated
573B65034VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USLine 25: The height is listed as 575 inches. Based on the previous matrix, should this be corrected to 10 inches?

Lines 74, 75, and 76: These do not appear to be cardiovascular clinical options. Can these lines be removed from the matrix?

Line 91: This does not appear to be a cardiovascular option. Can this line be removed?

Line 94: This does not appear to be a cardiovascular option. Can this line be removed?Tech Specs updated
573B65035VISN 8: VA Sunshine Healthcare Network573 LAKE CITY FLXR USRequest item 1.g. Minimum image Storage of 6TB be reduced to 1TB.Tech Specs updated
573B65035VISN 8: VA Sunshine Healthcare Network573 LAKE CITY FLXR US1.h. Please remove minimum battery life or change to 20min to allow for full and fair competition.
1.g. Please change minimum image storage to 1TB to allow for full and fair competition.Tech Specs updated
573B65036VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USRequest item 1.h. Minimum battery life be revised to 30 min.
Request item 1.j. Maximum eqiupment weight be increased to 250.Tech Specs updated
573B65036VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USLine 25: The height is listed as 575 inches. Based on the previous matrix, should this be corrected to 10 inches?

Lines 74, 75, and 76: These do not appear to be cardiovascular clinical options. Can these lines be removed from the matrix?

Line 91: This does not appear to be a cardiovascular option. Can this line be removed?

Line 94: This does not appear to be a cardiovascular option. Can this line be removed? Line 25: Refers to minimum ports - 3 Lines 74,75, and 76: Removed Line 91: The original matrix indicates line 91 is OEM supported operating systems and must remain.

Line 94: The original matrix indicates line 94 is PACS compatbility and must remain.

573B65037VISN 8: VA Sunshine Healthcare Network573 GAINESVILLE FLXR USRequest item 1.a. and 1.b. be revised to 24 in.
Request item 8.b, and 8.c. be removed as not typically done in Gastroenterology.Tech Specs updated
573B65038VISN 8: VA Sunshine Healthcare Network573 TALLAHASSEE FLXR US1.i. Please remove minimum battery life or change to 20min.
5.m. Please remove phantom requirement or move to value-add section. This is usually a third party purchase.Tech Specs updated
578B68013VISN 12: VA Great Lakes Health Care System578 HINES ILXR USCan you verify that 2.c Navigation and 2.d. Image Fusion will be used in this department?
Transducers - can you verify if all of these transducers on the matrix are required?2c&d-No

Transducer-We will not need all of the transducers. We will just need the ML4-20-D probe and L6-24-D Array probe for mammography. We could also any standard probes that come with the machine.

578B68013VISN 12: VA Great Lakes Health Care System578 HINES ILXR USG. Is there a trade in?no
578B68013VISN 12: VA Great Lakes Health Care System578 HINES ILXR USConsidering the overview notes that the machine will be used exclusively for Breast ultrasound imaging, can you please clarify/confirm the following:
1.Do you still require Navigation Software and Image Fusion Software (under Section 2 - Scanning Modes)
2.Do you require Analysis packages besides breast if the system is used exclusively for breast ultrasound (i.e. Abdominal, Pelvic, Urology, Obstetric, Cardiovascular, etc.)
3.The transducer section lists 10 various transducers.
a.Considering the system will be used exclusively for breast ultrasound, do you want only transducers specific for breast imaging?
i.If yes, which transducers?
ii.If no, questions:
1.There is a quantity of two (2) 4-15 MHz transducers noted- Do you want a quantity of two?
2.There is a quantity of two (2) 2-9 MHz transducers noted - Do you want a quantity of two?
1. No

2. No 3i. Transducers for breast imaging ML4-20-D Probe L6-24-D Linear Array Probe Any other probes that may come with the machine.

3ii - In the Tech Specs, it looks like these are referring to ports and not transducers? Please clarify.

578B68014VISN 12: VA Great Lakes Health Care System578 HINES ILXR USRequest 4.a. a keyboard on the display be required in lieu of the Physical keyboard?Yes
578B68014VISN 12: VA Great Lakes Health Care System578 HINES ILXR US2.b. Please remove navigation software; this is not applicable for this care area.updated tech specs
578B68014VISN 12: VA Great Lakes Health Care System578 HINES ILXR US2.c. Please remove image fusion software; this is not applicable for this care area.updated tech specs
578B68014VISN 12: VA Great Lakes Health Care System578 HINES ILXR US5.m. Please remove the phantom requirement or move it to the value-add section. This is usually a third-party purchase.updated tech specs
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USRequest item 1.m. Maximum equipment weight be increased to 265 lbs.Maximum weight changed to 265 lbs.
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USPlease confirm the care area for this tech spec. Only two probes are selected (high-frequency small parts and high-frequency linear), and trade-ins are point-of-care systems. Is this for radiology or point-of-care (vascular access)?Radiology - Intervential Radiology - PICC Team
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US2.b. Please confirm whether 3D freehand is needed for use at the CBOC. Request to have this moved to the value-add section or removed, as it is not needed.This purchase is needed for the Main VA Hospital in Houston
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US2.c. Please confirm if navigation software is needed. If not, please remove or move to value-add.Not needed
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US2.d. Please confirm if image fusion software is needed. If not, please remove or move to value-add.Not needed
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US5.m. Please remove the phantom requirement or move it to the value-add section. This is usually a third-party purchase.Added to Added Value section
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US6. Only two selections have been made: (a) Procedure and (e) Vascular. Please confirm whether more packages need to be included.Confirmed. Only procedural and vasuclar are needed.
580B60005VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US7. Please confirm these are the only two probes required for radiology.Yes, only those 2 probes are needed.
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USThe technical specs read 580-B70027 but the Compliance Matrix reads 580-B0006, along with the spreadsheet. Please clarify which number is correct.Correct number is 580B60006 - Technical specifications and compliance matrix added
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USColumn A does not include this Equipment PO - 580B6006 in order to pre populate the VISN, Station and Equipment type. Missing specs benchmarkCorrect number is 580B60006 - Technical specifications and compliance matrix added
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USBenchmark MissingCorrect number is 580B60006 - Technical specifications and compliance matrix added
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USNo tech spec or matrix found.Correct number is 580B60006 - Technical specifications and compliance matrix added
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USTech spec provided but not on schedule of line items. This appears to be a 2027 item.Correct number is 580B60006 - Technical specifications and compliance matrix added
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US3.b. Please confirm whether 3D freehand is needed. Request to have this moved to the value-add section or removed, as it is not needed.Yes, it is needed
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US3.c. Please confirm if navigation software is needed. If not, please remove or move to value-add.Not needed
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US3.d. Please confirm if image fusion software is needed. If not, please remove or move to value-add.Not needed
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US6.j. Please confirm if shear-wave and/or strain-based elastography clinically needed. I don't believe any of the Houston sites are doing shear-wave. If not, please remove or move to the value-add section.Not needed
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US6.p. Please remove the phantom requirement or move it to the value-add section. This is usually a third-party purchase.Moved to added value section
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US8. Please provide more clarity on the transducers (linear, curved, microconvex); many of these have what appear to be brand features included.Curvilinear probe 5-1 MHZ

Linear probes with the following frequency range 12-3 MHz,18-4 MHz & 26-8 MHz Endocavitary probe 10-3 MHz

580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US9.a-g. Please confirm whether the fusion-targeted biopsy system, along with associated lines, is needed; if not, remove it or move it to the value-add section.Not needed
580B60006VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USg. Please confirm trade-in; three systems are listed to be traded in, but only two are requested.2 trade ins confirmed
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USRequest items 8.b., 8.c., 8.e, and 8.f. be removed as this is not typically used in this department.needed by clinicians
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US2.c. Please confirm if navigation software is needed. If not, please remove or move to value-add.needed by clinicians
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US2.d. Please confirm if image fusion software is needed. If not, please remove or move to value-add.It is needed.
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US5.l. Please remove the phantom requirement or move it to the value-add section. This is usually a third-party purchase.Moved to added value section
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US7. Please provide transducer clarification; the section appears to be focused on one vendor without probe details (linear, curved, etc), only specific item numbers for a specific vendor.Those respective probes are needed. Equivalents are also fine.
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US7.a. Please confirm ICE catheter is needed or clinically necessary; if not, move to the value-add section.It is needed.
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US8.a-h. Please confirm whether the fusion-targeted biopsy system, along with associated lines, is needed; if not, remove it or move it to the value-add section.It is needed.
580B60015VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USG. Is there a trade in?This item is an addition, no trade-ins.
580B60019VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USRequest Item 1.a. Minimum Screen Size reduced from 20 in to 19 in.

Request 6.i. Shear wave and strain-based elastography be removed.

Under 8. Transducers can you verify this is the total transducers across all seven systems? 1.a. Changed to 19in.

6.i. Moved to added value

8. Transducer list updated.

580B60019VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USPlease remove request for specific vendor and ensure this spec is vendor neutral.Specific ultrasound device requested, but no specific vendor requested
580B60019VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US2.c. Please confirm need for 3D -freehand for use in the vascular lab. Request to have this moved to value add section or removed it not needed.Moved to added value.

FYSA: These ultrasound devices serve multiple departments and are not limited to vascular surgery.

580B60019 VISN 16: South Central VA Health Care Network 580 HOUSTON TX XR US 2.d. Please confirm navigation software is needed. If not please remove or move to value add. This is typically a radiology procedure feature and not used in vascular surgery. Moved to added value.

FYSA: These ultrasound devices serve multiple departments and are not limited to vascular surgery.

580B60019 VISN 16: South Central VA Health Care Network 580 HOUSTON TX XR US 2.d. Please confirm the image fusion software. If not needed, please remove or move to value-add. This is typically a radiology procedure feature and typically not used in vascular surgery. Moved to added value.

FYSA: These ultrasound devices serve multiple departments and are not limited to vascular surgery.

580B60019 VISN 16: South Central VA Health Care Network 580 HOUSTON TX XR US 3.n Can the xPlane imaging feature be removed from the requirement as it appears to be brand-specific? Moved to added value.

FYSA: These ultrasound devices serve multiple departments and are not limited to vascular surgery.

580B60019 VISN 16: South Central VA Health Care Network 580 HOUSTON TX XR US 6.i. Is shear-wave and/or strain-based elastography clinically needed for vascular surgery? If not, please remove or move to the value-add section. Moved to added value.

FYSA: These ultrasound devices serve multiple departments and are not limited to vascular surgery.

580B60019VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US6.o. Please remove the phantom requirement or move it to the value-add section. This is usually a third-party purchase.Removed.
580B60019VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US7.g. Please confirm whether the mitral value navigator is clinically needed; if not, please remove it or move it to the value-add section. This is a cardiology feature, not vascular surgery.Transducer list updated.

FYSA: These ultrasound devices serve multiple departments and are not limited to vascular surgery.

580B60019 VISN 16: South Central VA Health Care Network 580 HOUSTON TX XR US 7.h. Please confirm whether the cardiact 3D quantification is clinically needed; if not, please remove it or move it to the value-add section. This is a cardiology feature, not vascular surgery. Removed.

FYSA: These ultrasound devices serve multiple departments and are not limited to vascular surgery.

580B60019VISN 16: South Central VA Health Care Network580 HOUSTON TXXR US7.k Please confirm whether the 3D auto left atrial appendage n is clinically needed; if not, please remove it or move it to the value-add section. This is a cardiology feature, not a vascular surgery feature.Removed.
580B60020VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USMissing specs for probesList of the seven required transducers defined in section 7.
580B60020VISN 16: South Central VA Health Care Network580 HOUSTON TXXR USRequest item 1.m. Maximum equipment weight be increase from 250 to 265.
Can you verify if 8.b. 3D-generated model of the prostate is required?1.m Increased

8.b. Moved to value add