MAR 2025 EQ Consolidation - Schedule of Line Items w Vendor Questions - Revised.xlsx

XLSX spreadsheet 404 KB Posted

Attached to
6525--MAR 2025 Equipment Only Consolidation
Solicitation number
36A79725Q0006
Issued by

View the file

Other files

File Type Posted
36A79725Q0006 0002.docx DOCX document
Amendment 0002 - Continuation Page.docx DOCX document
553B50041_XR US VAMC DETROIT MI_compliance matrix updated.xlsx XLSX spreadsheet
Amendment 0001 - Continuation Page.docx DOCX document
402B51207 - 5085B5013.zip ZIP file
36A79725Q0006 0001.docx DOCX document
MAR 2025 EQ Consolidation - Schedule of Line Items w Vendor Questions.xlsx XLSX spreadsheet
MAR 2025 EQ Consolidation - RFO.pdf PDF
36A79725Q0006_1.docx DOCX document
5085B5010 - 5085B5013.zip ZIP file
603B54008 - 642B59012.zip ZIP file
589B50009 - 595B59008.zip ZIP file
Memorandum for Training.pdf PDF
MAR 2025 Equipment Only Consolidation Schedule.pdf PDF
VA Directive 6550 - New Appendix A - published Jun. 2023.pdf PDF
580B50010 - 589B50008.zip ZIP file
Vendor Workup Spreadsheet - 36A79725Q0006.xlsx XLSX spreadsheet
402B51207 - 519B50015.zip ZIP file
Vendor Folders.zip ZIP file
528B47008 - 553B50041.zip ZIP file
646B52005 - 673B50041.zip ZIP file
673B50042 - 691B50048.zip ZIP file
553B50042 - 580B50009.zip ZIP file
MAR 2025 EQ Consolidation - Schedule of Line Items.xlsx XLSX spreadsheet
Show all 24

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Schedule of Line Items

HTME VISNStationEQ POEquipment TypeQTY
VISN 1: VA New England Healthcare System402 AUGUSTA, ME402B51207NM BONE DEN1
VISN 23: VA Midwest Health Care Network618 MINNEAPOLIS, MN499B59001XR MAMMO1
VISN 4: VA Stars & Stripes Healthcare Network503 ALTOONA, PA503B50004XR PORT C ARM1
VISN 17: VA Heart of Texas Health Care Network504 AMARILLO, TX504B57001XR PORT C ARM2
VISN 17: VA Heart of Texas Health Care Network504 AMARILLO, TX504B57002XR PORT RAD3
VISN 7: VA Southeast Health Care Network508 ATLANTA, GA5085B5010XR US1
VISN 7: VA Southeast Health Care Network508 ATLANTA, GA5085B5011XR US2
VISN 7: VA Southeast Health Care Network508 DECATUR, GA5085B5012XR US1
VISN 7: VA Southeast Health Care Network508 DECATUR, GA5085B5013XR US1
VISN 10: VA Healthcare System of Ohio515 BATTLE CREEK, MI515B50008XR RF1
VISN 8: VA Sunshine Healthcare Network516 BAY PINES, FL516B40030XR URO1
VISN 8: VA Sunshine Healthcare Network516 BAY PINES, FL516B40031XR URO1
VISN 17: VA Heart of Texas Health Care Network519 BIG SPRING, TX519B50015XR US1
VISN 2: VA Healthcare Network Upstate New York528 BUFFALO, NY528B47008XR RAD1
VISN 2: VA Healthcare Network Upstate New York528 ROCHESTER, NY528B57005XR US1
VISN 4: VA Stars & Stripes Healthcare Network529 BUTLER, PA529B52003XR PORT RAD1
VISN 10: VA Healthcare System of Ohio539 CINCINNATI, OH539B40035XR US2
VISN 10: VA Healthcare System of Ohio539 CINCINNATI, OH539B50003XR US1
VISN 10: VA Healthcare System of Ohio539 CINCINNATI, OH539B50012XR RF1
VISN 10: VA Healthcare System of Ohio539 CINCINNATI, OH539B50013XR US1
VISN 10: VA Healthcare System of Ohio539 CINCINNATI, OH539B50014XR US1
VISN 10: VA Healthcare System of Ohio552 DAYTON, OH552B50002XR MRI1
VISN 10: VA Healthcare System of Ohio553 DETROIT, MI553B50041XR US1
VISN 10: VA Healthcare System of Ohio553 DETROIT, MI553B50042XR PORT C ARM1
VISN 19: VA Rocky Mountain Network554 AURORA, CO554B53013XR US1
VISN 19: VA Rocky Mountain Network554 AURORA, CO554B53014XR RAD3
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B51009XR PORT RAD2
VISN 12: VA Great Lakes Health Care System556 NORTH CHICAGO, IL556B51010XR PORT C ARM1
VISN 2: VA Healthcare Network Upstate New YorkEAST ORANGE, NJ561B50004XR US3
VISN 21: VA Sierra Pacific Network570 FRESNO, CA570B50006XR US1
VISN 8: VA Sunshine Healthcare Network573 LAKE CITY, FL573B55010XR PORT C ARM1
VISN 16: South Central VA Health Care Network580 HOUSTON, TX580B50009NM BONE DEN1
VISN 16: South Central VA Health Care Network580 HOUSTON, TX580B50010NM BONE DEN1
VISN 16: South Central VA Health Care Network580 HOUSTON, TX580B50011XR PORT C ARM1
VISN 5: VA Capitol Health Care Network581 HUNTINGTON, WV581B52033XR RAD1
VISN 10: VA Healthcare System of Ohio583 INDIANAPOLIS, IN583B50008XR US1
VISN 10: VA Healthcare System of Ohio583 INDIANAPOLIS, IN583B50010XR US3
VISN 15: VA Heartland Network589 KANSAS CITY, MO589B50005NM SPECT1
VISN 15: VA Heartland Network589 KANSAS CITY, MO589B50006XR US1
VISN 15: VA Heartland Network589 KANSAS CITY, MO589B50007XR IR1
VISN 15: VA Heartland Network589 KANSAS CITY, MO589B50008XR IR1
VISN 15: VA Heartland Network589 KANSAS CITY, MO589B50009XR CATH LAB1
VISN 15: VA Heartland Network589 KANSAS CITY, MO589B50013XR CATH LAB1
VISN 15: VA Heartland Network589 WICHITA, KS589B51005NM PET CT1
VISN 15: VA Heartland Network589 COLUMBIA, MO589B53002XR CATH LAB1
VISN 15: VA Heartland Network589 COLUMBIA, MO589B53003XR CATH LAB1
VISN 15: VA Heartland Network589 COLUMBIA, MO589B53004XR IR1
VISN 4: VA Stars & Stripes Healthcare Network595 LEBANON, PA595B59006NM SPECT CT1
VISN 4: VA Stars & Stripes Healthcare Network595 LEBANON, PA595B59007XR US2
VISN 4: VA Stars & Stripes Healthcare Network595 LEBANON, PA595B59008XR US1
VISN 9: VA Mid South Healthcare Network603 LOUISVILLE, KY603B54008XR US4
VISN 9: VA Mid South Healthcare Network603 LOUISVILLE, KY603B54013XR US1
VISN 21: VA Sierra Pacific Network612 MARTINEZ, CA612B59005XR PORT C ARM1
VISN 21: VA Sierra Pacific Network612 STOCKTON, CA612B59006XR US1
VISN 21: VA Sierra Pacific Network612 STOCKTON, CA612B59007XR PORT C ARM1
VISN 19: VA Rocky Mountain Network623 MUSKOGEE, OK623B54007NM BONE DEN1
VISN 9: VA Mid South Healthcare Network626 NASHVILLE, TN626B54012XR US1
VISN 4: VA Stars & Stripes Healthcare Network642 PHILADELPHIA, PA642B59010XR US2
VISN 4: VA Stars & Stripes Healthcare Network642 PHILADELPHIA, PA642B59012RT LINAC1
VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B52005XR PORT RAD1
VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B52007XR US2
VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B52008XR US1
VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B52010XR US2
VISN 4: VA Stars & Stripes Healthcare Network646 PITTSBURGH, PA646B52011XR US4
VISN 21: VA Sierra Pacific Network654 RENO, NV654B50034XR US1
VISN 6: VA Mid-Atlantic Health Care Network658 SALEM, VA658B59005US2
VISN 21: VA Sierra Pacific Network662 SAN FRANCISCO, CA662B50003XR HYBRID1
VISN 8: VA Sunshine Healthcare Network673 TAMPA, FL673B50041XR US1
VISN 8: VA Sunshine Healthcare Network673 TAMPA, FL673B50042XR CT1
VISN 8: VA Sunshine Healthcare Network673 TAMPA, FL673B50058XR US1
VISN 8: VA Sunshine Healthcare Network673 TAMPA, FL673B50059XR US4
VISN 8: VA Sunshine Healthcare Network673 TAMPA, FL673B50060XR US3
VISN 17: VA Heart of Texas Health Care Network674 TEMPLE, TX674B50226XR PORT C ARM2
VISN 22: VA Desert Pacific Healthcare Network678 TUCSON, AZ678B52007XR IR1
VISN 5: VA Capitol Health Care Network688 WASHINGTON, DC688B51002XR US2
VISN 5: VA Capitol Health Care Network688 WASHINGTON, DC688B51004XR US1
VISN 22: VA Desert Pacific Healthcare Network691 LOS ANGELES, CA691B50048XR PORT C ARM1

Q&A

VISNSTATIONEQUIPMENT REQUISITION NUMBERVENDOR EQUIPMENT QUESTIONSGOVERNMENT EQUIPMENT RESPONSE
1AUGUSTA, ME402B512071. 1a-d Should this be asking for "max" scan time?

2. DEXA is not done as a %CV. This isn't applicable.

3. 1g change to 18" to allow all vendors to bid (This is a lock-out bid).

4. 1j remove to allow all vendors to bid (This is a lock-out bid specific only to Hologic). # of detectors is irrelevant as detector technology isn't the same. 1. 1a-d Minimum changed to Maximum in updated technical specifications

2. No change to technical specifications.

3. 1g changed to 18”

4. 1j. Removed

2ROCHESTER, NY528B570052. Minimum Battery Life: Can the Scan on Battery time be reduced from 50 mins to 45 mins to increase competition?Yes it can be
2ROCHESTER, NY528B570052. Can the Max Height be increased from 51" to 55" to increase competition?Yes it can be
2ROCHESTER, NY528B570052. Can the Max Depth be increased from 35.5" to 38" to increase competition?Yes it can be
2ROCHESTER, NY528B570052. Can the Max Weight be increased from 278 lbs to 290 lbs to increase competition?Yes it can be
2ROCHESTER, NY528B570053c. Is Image Fusion Software clinically necessary? If it is not, can this be moved to the Value Add section?No its not. Yes it can be
2BUFFALO, NY528B470082. Unit Physical Specifications:

d. Maximum radiographic mA exposure @ 100 kVp: 10-1000 → With an 80 kW HVG, a maximum of 800 mA can be supported at 100 kVp (same for all manufacturers).

Please remove the requirement condition "@ 100 kVp".it can be removed
2BUFFALO, NY528B470083. Additional Specifications:

k. Motor-driven tabletop movement in the X and Y axis at stated maximum weight capacity

→ Movement of the tabletop in the X and Y axis is only possible manually (same for all manufacturers) - please remove requirement .it can be removed
4BUTLER, PA529B520031. Unit Physical Specifications

d. Generator mAs range: 0.2-630 Please adjust higher range to mAs 500 to ensure full competition.

Tech Spec changed
4BUTLER, PA529B52003l. Minimum tube reach: 28
→ Please adjust minimum tube rich to 31.2 inch to ensure competition.Tech Spec changed
4LEBANON, PA595B590061. Line 24: Please change this to 440mA to allow for full and fair competition

2. Line 26: please increase to 10,000 lbs to allow for full system weight (NM gantry, CT gantry, table, etc…) as spec states

3. Line 36: Does this mean you require CZT detectors? Other specs in the Matrix do seem to point to CZT, please clarify and update spec as it is not specifically called out.

4. Lines 30-31: If CZT is required then these 2 collimators are not available on a CZT based SPECT/CT scanner form any vendor and can be removed from the matrix

5. Line 60: Please remove as only vendor can meet this requirement

6. Line 61: Please remove as only vendor can meet this requirement 1. Line 24 changed to 440mA.

2. Line 26 changed to 10,000 lbs

3. Line 36, Requiring CZT detectors

4. Lines 30-31 removed

5. Lines 60-61 moved to added value.

4LEBANON, PA595B590071. Could the minimum screen size be changed to 19 inches to increase competition?
2. Is shear wave a hard requirement?1. Changed to 18 in

2. Yes, shear wave is hard requirement

4LEBANON, PA595B590078a. Can this specification be moved to Value Add to increase competition?Clinical requires for their usage of the ultrasound.
4LEBANON, PA595B590081. Could the minimum screen size be changed to 19 inches to increase competition?
2. Is shear wave a hard requirement?1. Changed to 18 in

2. Yes, shear wave is hard requirement.

4LEBANON, PA595B590088a. Can this specification be moved to Value Add to increase competition?Clinical requires for their usage of the ultrasound
4PHILADELPHIA, PA642B590101. Request 1.a. Minimum screen size of 21 in be reduced to 19 in.

2. Request 1.k. Maximum equipment weight 125 lbs be increased to 140 lbs.

3. Can you verify items 6.j. Mitral Valve Navigator is required? 1. Request 1.a. Minimum screen size of 21 in be reduced to 19 in.

No. As Philadelphia VAMC is a teaching hospital, the bigger screen size will assist the Anesthesiology staff in providing a more condusive learning environment.

2. Request 1.k. Maximum equipment weight 125 lbs be increased to 140 lbs.

No. Unit will be transported throughout different areas of the facility. Weight requirement supports staff's ability to constantly move the machine throughout the facility.

3. Can you verify items 6.j. Mitral Valve Navigator is required?

Yes. This is a functional requirement.

4 PHILADELPHIA, PA 642B59010 1. Can the minimum screen size be changed to 19 to increase competition?

2. Can the minimum battery life be changed to 120 minutes to increase competition?

3. Could the minimum weight be changed 200 lbs to increase competition?

4. Is Phantom for quality control a hard requirement?

5. Can the frequency range on 7.c. be changed to 18 MHz to increase competition? 1. Can the minimum screen size be changed to 19 to increase competition?

No. As Philadelphia VAMC is a teaching hospital, the bigger screen size will assist the Anesthesiology staff in providing a more condusive learning environment.

2. Can the minimum battery life be changed to 120 minutes to increase competition?

No. Battery life needs to be sufficient to support an efficient workflow for the Anesthesiology department. The device will be used to evaluate multiple patients throughout various loacations of the hospital, so the device needs to withstand heavy usage throughout the day.

3. Could the minimum weight be changed 200 lbs to increase competition?

No. Unit will be transported throughout different areas of the facility. Weight requirement supports staff's ability to constantly move the machine throughout the facility.

4. Can the frequency range on 7.c. be changed to 18 MHz to increase competition?

No. The range was listed due to a functional need.

4PITTSBURGH, PA646B52011Line 12 reads VA Pittsburgh’s emergency medicine department is requesting two new ultrasound systems and multiple transducers. This should read VA Pittsburgh’s Critical Care department is requesting two new ultrasound systems and multiple transducersedited
4PITTSBURGH, PA646B52011Line 73 reads Curved Transducer – maximum scan depth of 35cm. This should read Phased transducer - maximum scan depth 35cm.changed to Phased
4PITTSBURGH, PA646B520071. Can you verify items 6.k. Mitral Valve Navigator is required?
2. Can you verify items 8.a. Capable of accepting MRI (both 1.5T and 3.0T) and ultrasound images from other modalities as input and displaying these images on a screen is required?not required, removed
4PITTSBURGH, PA646B520081. Request 1.b. Can you verify the Minimum active screen size of 32 - should this of been 22 in.
2. Request 1.m. Maximum equipment weight 180 lbs be increased to 190 lbs.changed to 22in, changed to 190
4PITTSBURGH, PA646B520081.b. Can the active screen size be reduced from 32" to 14" to increase competition?changed to 14
4PITTSBURGH, PA646B520081.l. Can the Max dimenstions be increased from 52x20x36 to 54x22x38 to increase competition?The max dimensions have been increased
4PITTSBURGH, PA646B520101. Can you verify items 6.j. Mitral Valve Navigator and 6.k. Cardiac 3D Quantification are required as there is no cardiac probe required?
2. Can you verify items 8.a. Capable of accepting MRI (both 1.5T and 3.0T) and ultrasound images from other modalities as input and displaying these images on a screen is required?not required, removed
4PITTSBURGH, PA646B52010Line 73 reads Curved Transducer – maximum scan depth of 35cm. This should read Phased transducer - maximum scan depth 35cm.changed to Phased
4PITTSBURGH, PA646B520111. Can you verify items 6.j. Mitral Valve Navigator and 6.k. Cardiac 3D Quantification are required as there is no cardiac probe required?
2. Can you verify items 8.a. Capable of accepting MRI (both 1.5T and 3.0T) and ultrasound images from other modalities as input and displaying these images on a screen is required?not required, removed
5WASHINGTON, DC688B51002Will the selected ultrasound system be required to be compatible with Fujifilm Synchronicity? If so, do you only need to share images via DICOM?Specifications are correct

Ultrasound system will need to be compatible with point-of-care image repository. Images are sent via DICOM to destination

5WASHINGTON, DC688B51004Request 1.f. Maximum equipment weight 160 lbs be changed to 190 lbs.Specifications updated.
5WASHINGTON, DC688B510041. Can the minimum screen size be changed to 19 to increase competition?

2. What are the procedures being performed for the Breast Application?

Specifications updated.

This equipment will also be used for ultrasound guided cyst aspirations/abscess drainages, ultrasound guided core biopsies and ultrasound guided wire localizations.

5WASHINGTON, DC688B510041.f. Can the Max Weight be increased from 160 to 290 to increase competition?Specifications updated.
5HUNTINGTON, WV581B520332. Unit Physical Specification:

g. Minimum spatial resolution [lp/mm]: 5 → (It is assumed that "Minimum" should be "Maximum" instead.)

Please change to "3.5 or higher".Specifications updated.
5HUNTINGTON, WV581B52033h. Maximum pixel pitch [μm]: 100
100 um will not be the best pixel size for chest x-rays. Please change to 140.Specifications updated.
5HUNTINGTON, WV581B520333. Additional Specifications:

q. Charging station for wireless detector (as required by vendor)

Some vendors do not have a separate charging station. Please change to 'detector battery must be able to be charged in the table/wall bucky as well as to a wall outlet'.Specifications updated.
6SALEM, VA658B59005Can you verify the Trade-in systems as they are the same from obligation 658B49002?Corrected
6SALEM, VA658B590051.c. Is the Min Battery Life of 160 mins referring to Standby time or active Scan Time? If it is active Scan Time, can this be reduced to 75 mins to increase competition?Changed to 75 Minutes
6SALEM, VA658B590058 Is the Fusion-Targeted Biopsy System clinically necessary? If so, can this be moved to Value-Add section to increase competition?YES – Clinically necessary
7ATLANTA, GA5085B50103.d. Is Image Fusion Software clinically necessary? If it is not, can this be moved to the Value Add section?No, it is not clinically necessary, and yes, it can be added to the Value Add section. Tech Specs and Compliance Matrix were updated.
7ATLANTA, GA5085B5011The requirement overview says that they are replacing 2 ultrasounds that have reached end-of-life, but on the trade-in section, it says that they have no trade-in units to offer. Please confirm if trade ins are to be quoted and provide details to determine value if so.One trade-in would be considered. The trade-in is a Philips Affiniti 70 with SN US516F0897. Tech Specs and Compliance Matrix were updated.
7ATLANTA, GA5085B50111.b. Can the Minimum active screen size be reduced from 21.5 to 14 to increase competition?No. Several vendors offer 21.5 inch or larger monitors, as this is the standard size for a cardiac ultrasound machine used in most hospitals. Decreasing the size to 14 inches would not be acceptable.
7ATLANTA, GA5085B50116.e Is this specification clinically necessary? If not, can it be moved to the Value Add section?Yes, 3D mitral valve functional assessments are clinically necessary as these ultrasounds will used in cardiology.
7ATLANTA, GA5085B50116.f Is this specification clinically necessary? If not, can it be moved to the Value Add section?Yes, cardiac 3D quantification is clinically necessary.
7DECATUR, GA5085B5012Request 1.g. Minimum Battery Life 90 mins be reduced to 60 minutes.Yes, the minimum battery life was reduced to 40 mins. Tech Specs and Compliance Matrix were updated.
7DECATUR, GA5085B50121.b. Can the Minimum active screen size be reduced from 21.26 to 14 to increase competition?No, a larger screen size is required for more accurate anatomy viewing. However, minimum screen sizes were reduced to 20 inches to increase competition. Tech Specs and Compliance Matrix was updated.
7DECATUR, GA5085B50121.g Is the minimum battery life of 90 mins referring to Standby time or active Scan Time? If it is active Scan Time, can this be reduced to 40 mins to increase competition?Yes, the active Scan Time can be reduced to 40 minutes. Tech Specs and Compliance Matrix were updated.
7DECATUR, GA5085B50122.d Is Image Fusion Software clinically necessary? If it is not, can this be moved to the Value Add section?No, it is not clinically necessary, and yes, it can be added to the Value Add section. Tech Specs and Compliance Matrix was updated.
7DECATUR, GA5085B50126.e Is this specification clinically necessary? If not, can it be moved to the Value Add section?No, it is not clinically necessary, and yes, it can be added to the Value Add section. Tech Specs and Compliance Matrix was updated.
7DECATUR, GA5085B50126.f Is this specification clinically necessary? If not, can it be moved to the Value Add section?No, it is not clinically necessary, and yes, it can be added to the Value Add section. Tech Specs and Compliance Matrix was updated.
7DECATUR, GA5085B5013Request 1.l. Maximum equipment weight 230 lbs be changed to 265 lbs.The maximum equipment weight can be increased from 230 lbs to 265 lbs. Tech Specs and Compliance Matrix was updated.
8TAMPA, FL673B50042The Word document (Tech Spec) and Excel document (Compliance Matrix) do not match up. Please correct.Correct tech spec and compliance matrix uploaded.
8LAKE CITY, FL573B550101. Please provide the Biomedical Engineering Point of Contact. 2. Is a printer needed?1. Coleman Blincoe, HTM Supervisor 2. Yes, printer is needed.
8TAMPA, FL673B50058Line 32 m. Please clarify if the maximum weight requirement can expanded to over 250 lbs.No, the maximum weight specified remains 140 lbs.
8TAMPA, FL673B50058Line 30 k. Please clarify if a battery is required for quick boot-up or is standby and hibernate also an option? Standby and hibernate do not require a battery.Battery is needed for quick boot-up
8TAMPA, FL673B50060Line 27 h. Please clarify if a battery is required for quick boot-up or is standby and hibernate also an option? Standby and hibernate do not require a battery.Battery is needed for quick boot-up
8TAMPA, FL673B50060Line 29 j. Please clarify if the maximum weight can expanded to over 250 lbs.No, the maximum weight specified remains 140 lbs.
8TAMPA, FL673B50042The tech spec document and matrix have two different values. We are thinking the excel Matrix is correct, pelase confirm.Correct tech spec and compliance matrix uploaded.
8TAMPA, FL673B50058Please confirm if Small Parts, Pelvic, Urology, Obstetric, Breast Cardiovascular Clinical Application are required as requirement overview states 'guide interventional jointDoes not need to be able to do Pelvis, Urology, OB, or Cardiovascular. It is being used for Musculoskeletal use only but small parts are needed at times (for small joints and nerves etc…)
8TAMPA, FL673B500581.k. Is the minimum battery life of 60 mins referring to Standby time or active Scan Time? If it is active Scan Time, can this be reduced to 40 mins to increase competition?Active scan time. No, the minimum battery life specified remains 60 minutes to optimize scan time on patients, especially back to back, reducing downtime.
8TAMPA, FL673B500581.l. Can the Max Depth dimension be increased from 26 to 38 to increase competition?No, the maximum equipment dimensions will remain as specified to accommodate the size of exam rooms.
8TAMPA, FL673B500581.m. Can the Max Weight be increased from 140 to 176 to increase competition?No, the maximum weight specified remains 140 lbs.
9LOUISVILLE, KY603B54008Can you verify item 9.g. Integration to Dose Tracking System - Radimetrics is required?This requirement has been removed.
9LOUISVILLE, KY603B540081.h Is the minimum battery life of 60 mins referring to Standby time or active Scan Time? If it is active Scan Time, can this be reduced to 40 mins to increase competition?The clinicians travel frequently with their systems and need 60 minutes of runtime. It cannot be reduced.
9LOUISVILLE, KY603B540131. Request 1.c. Minimum Battery Life 120 min changed to 60 min.

2. Request 1.e. Maximum equipment weight 160 lbs be changed to 185 lbs.

1.c. The requirement has been updated to reflect 60-minutes.

1.e. Max weight has been adjusted to 185 lbs.

9LOUISVILLE, KY603B540131.e. Can the Max Weight be increased from 160 to 176 to increase competition?Max weight has been adjusted to 185 lbs.
9LOUISVILLE, KY603B540131.c. Is the Min Battery Life of 120 mins referring to Standby time or active Scan Time? If it is active Scan Time, can this be reduced to 40 mins to increase competition?The clinicians travel frequently with their systems and need no less than 60 minutes of runtime. It cannot be reduced below 60-minutes.
9NASHVILLE, TN626B54012Line 30k. Please clarify if the maximum weight can expanded to over 100 lbs.Yes it can be expanded to 110 lb
9NASHVILLE, TN626B54012Line 28 i. Please clarify if a Is a battery required for quick boot-up or is standby and hibernate also an option? Standby and hibernate do not require a battery.Yes, a battery is required
9NASHVILLE, TN626B540121. Can you verify if 2.b. Three-dimensional (3D) - freehand is required?
2. Can you verify if 2.c. Navigation Software is required?1. No

2. No 9 NASHVILLE, TN 626B54012 1. What are the procedures being performed for the Cardiovascular Clinical Application?

2. Could the minimum frequency range for 7. a.be expanded to 2-5 MHz to increase competition?

3. Could the minimum frequency range for 7. b.be expanded to 4-13 MHz to increase competition?

4. Could the minimum frequency range for 7. c.be expanded to 2-6 MHz to increase competition? 1. Surface Echo, line placement, diagnosis of pulmonary embolism and pleural effusion

2. No, the provided frequency range is required for procedures performed

3. No, the provided frequency range is required for procedures performed

4. No, the provided frequency range is required for procedures performed

10BATTLE CREEK, MI515B50008Is this a RF system or an Ultrasound system requirement? The requirement is labeled as a RF system, but in the documentation it is listing ultrasound system requirements.This is for an Ultrasound system.
10BATTLE CREEK, MI515B500082.b. Is "three-dimensional (3D)--freehand clinical necessary? Can this be moved to the "value add" section to increase competition?Tech spec moved to value add
10BATTLE CREEK, MI515B500082. c. Is "navigation software" clinically necessary? As this is an interventional procedure done in combination with CT/MRI. Can this be moved to "value add" section?Tech spec moved to value add
10BATTLE CREEK, MI515B500082. Is 3D/4D software required, there is a 3D endocavity probe included but no ask for 3D/4D software. You can only achieve 3D imaging with a 3D probe if the 3D/4D software is included. Freehand 3D is not the same as 3D/4D software.Tech spec updated - transducer removed
10BATTLE CREEK, MI515B500085.o Is a physical phantom required? These are typically not provided by the manufacturers, Biomed typically has these already or they can purchased from outside companiesPhantom will be procured via third party (not NAC)
10CINCINNATI, OH539B500031. Request 1.g. Minimum Battery Life 15 mins moved to Value Add Section.
2. Request 5.f. Ability to enter standby mode or sleep mode be moved to Value Add Section.1. Request denied; equipment requires 15mins battery-only operation to fill intended purpose as a mobile POC device

2. Request granted; Sleep/standby requirement moved to Value Added Section

10CINCINNATI, OH539B500121. 3b. Please change requirement to floor or ceiling mounted to allow for full and fair competition.
2. 3c. Please change requirement to floor or ceiling mounted to allow for full and fair competition.1. Requirement for floor mounted system moved to added value section.

2. Requirement for floor mounted system moved to added value section.

10CINCINNATI, OH539B500131. Request 1.h. Minimum Battery Life 15 mins moved to Value Add Section.
2. Request 5.j. Ability to enter standby mode or sleep mode be moved to Value Add Section.1. Request granted; 15min battery life requirement moved to Value Added Section

2. Request granted; Sleep/standby requirement moved to Value Added Section

10CINCINNATI, OH539B50014Request 1.a. Minimum screen size of 20 in be reduced to 19 in.1. Request granted; Minimum screen size reduced to 19in
10DAYTON, OH552B50002Please review the following specs as they are seen as vendor specific lockout specs that only one vendor can meet. The ask is to open these for fair competition:

1.f.

2.c 2.g.

4.a.

5.a.

5.d.

5.g. 1f – Supports whole body imaging, multiple high-element count coils connected simultaneously; minimum of 128 receive channels activated concurrently.

2c – 450-550 lbs.

2g – the table must lower enough to accommodate safe and easy access for geriatric and disabled individuals.

-moved to added value: table lowers to 56cm from floor -moved to added value: large castors with “easier maneuverability for the technologist”

- removed: single touch patient positioning.

4a – 24” -moved to added value: dual 24” 5a – Torso/body array – High-element count, compatible with acceleration techniques in any orientation.

-moved to added value: System must have capability to connect multiple coils at the same time, to do imaging of multiple body parts without moving the patient.

-moved to added value: Second torso/body array.

5d – Keep Head/neck/spine array -moved to added value: table integrated with respiratory sensors 5g – High-element coil compatible with acceleration techniques -removed: cable-less with direct table connect

10DETROIT, MI553B500411. Request 1.a. Minimum screen size of 20 in be reduced to 15.6 in.
2. Request 1.i. Minimum number of transducer ports be reduced from 4 to 3.1. Minimum screen size to remain as is due to clinical need. 2. Tech specs updated to 3.
10DETROIT, MI553B500421. Please provide the Biomedical Engineering Point of Contact. 2. Is a printer needed?POC: John Velazquez-Garcia. A printer is needed for backup.
10INDIANAPOLIS, IN583B500102. Can "navigation software" be removed, as this is a feature used primarily for abdominal biopsies In combination with CT/MRI not musculoskeletal examsYes, this can be removed.
12NORTH CHICAGO, IL556B51009Q1.AA Change Weight Limit to 990lbs. Q3.E Make Optional Q3.Y Make Optional or Remove for fair competition.Technical specifications updated.
15WICHITA, KS589B51005Line 17 a. Please clarify if this requirement is for acquired or reconstruction slices?Tech spec updated providing clarity
15WICHITA, KS589B51005Line 20 d. Please change the minimum axial scan field of view to 26cm to allow for fair and open competition.Tech spec updated
15WICHITA, KS589B51005Line 31 O. Please change maximum system weight to 8,300 lbs to allow competitionTech spec updated
15WICHITA, KS589B51005Line 35 c. Please clarify is deviceless PET gating requested or PET/CT respiratory gating? If the latter, is an Anzai or RGSC trigger device required?Deviceless PET gating is required. Anzai or a trigger device is not acceptable. "Must be deviceless" has been added to tech specs to clarify.
15WICHITA, KS589B51005Line 38 f. Please clarify if the required injector is a ceiling or floor mounted?Floor mounted
15WICHITA, KS589B51005Line 48 b. Please clarify if you require automated quality control that does not require an external radioactive source?External or internal sources are acceptable
15WICHITA, KS589B51005Line 22 f. Please change table height range from floor to 21 inches to allow competitionTech spec updated
15WICHITA, KS589B510051. 1b. min 70 for fair competition 1k. increase to 385 for fair competition

2. 2a. time of flight system with list mode reconstruction or equivalent

3. 2l. Open to BGO crystals with analog or BGO paired with ultra high sensitivity or resolution SIPM to allow fair competition.

4. 4c. is this required for cardiac, oncology, or both?

5. 2d. what is the maximum reconstruction time required for cardiac PET CT data sets?

6. 1i. recommend changing minimum number of crystals 29,000 to allow fair competition 1b: The technical spec remains unchanged because a larger aperature is needed for patient comfort, bariatric and claustrophobic patients.

1k. The technical spec remains as is because we need increased resolution.

2a. Tech spec has been updated to include "or equivalent" to the requirement.

2l. No, prefer the Lu-type or LSO crystals for faster scan time, improved image quality, and lower dose. Tech specs are now more specific to reflect this requirement.

4c. Remote aligning/positioning of patient to center without repeating image. Tech specs have been updated to reflect this clarification.

2d. There is no maximum reconstruction time. ECG cardiac gating for PET or CT for single or multiphase. This clarification has been added to the tech specs.

1i. Number of crystals remains as 32,000 because we prefer a higher number of crystals for higher spatial resolution, better sensitivity, enhanced signal-to-noise ratio leading to more detailed images.

15COLUMBIA, MO589B53002Line 52h. Please clarify the size of required Digital detector.Tech spec updated
15COLUMBIA, MO589B53002Please clarify if there is a need for any Digital Subtraction as this is standard in a Cath Lab configuration?Tech spec updated
15COLUMBIA, MO589B53002Please clarify if there is a need for IVUS in the Cath Lab as an ultrasound is typically part of a standard configuration? If there is a need, what type of IVUS system should be included?Tech spec updated
15COLUMBIA, MO589B53002Please clarify if there is a need for software that provides Peripheral Run-offs is typically part of the standard configuration?Tech spec updated
15COLUMBIA, MO589B53002Please clarify if there is a need for Stent enhancement software is typically part of the standard configuration?Tech spec updated
15COLUMBIA, MO589B53002Will any structural heart cases be performed in this room?

1. Will this room be a hybrid – and need an OR certified table?

2. Please confirm that room dimensions are for the procedure room only? Yes hearts

1. Not a Hybrid, no OR certified table

2. Room dimensions are for the procedure room only

15COLUMBIA, MO589B53003Line 51h. Please clarify the size of the required Digital Detector.Tech spec updated
15COLUMBIA, MO589B53003Please clarify if there is a need for any Digital Subtraction as this standard in a Cath Lab configuration?Tech spec updated
15COLUMBIA, MO589B53003Please clarify if there is a need for IVUS in the Cath Lab as an ultrasound is typically part of the standard configuration? If there is a need what type of IVUS system should be included ?Tech spec updated
15COLUMBIA, MO589B53003Please clarify if there is a need for software that provides Peripheral Run-offs is typically part of the standard configuration?Tech spec updated
15COLUMBIA, MO589B53003Please clarify if there is a need for Stent enhancement software is typically part of the standard configuration?Tech spec updated
15COLUMBIA, MO589B53003Is CARTO interface required?Tech spec updated
15COLUMBIA, MO589B53004Line 39 l. Please clarify the type of IVUS system should be included in the configuration?Tech spec updated
15COLUMBIA, MO589B53004Line 40 m. Please clarify if the Hemodynamic monitoring system (hardware and software) will be provided for this lab by the vendor?Yes
15COLUMBIA, MO589B53004Please clarify is there a need for software that provides Peripheral Run-offs is typically software required for the Biplane system?Tech spec updated
15COLUMBIA, MO589B530041. What cases will be performed specifically…do not leave anything out in response. Detailed cases needed.

2. Will they do any neuro in this room? Stroke?

3. What’s the purpose / need of a bi-plane?

4. Do they need advanced apps to help with difficult cases – overlay guidance?

5. Please confirm that room dimensions are for the procedure room only?

6. Please confirm if the width of the procedure room can be increased to 14’5” or 17’1”? 1. Updated requirement overview

2. Will do stroke, no neuro

3. Kyphoplasty

4. Yes

5. Room dimensions are for the procedure room only

6. Cannot be increased

16HOUSTON, TX580B50011Please provide the Viomedical Engineering Point of Contact.Andrew Chi, andrew.chi3@va.gov
16HOUSTON, TX580B50011Is a printer and/or wifi capability needed?Optional. Added value, not required.
16HOUSTON, TX580B500091. 1a-d Should this be asking for "max" scan time?

2. DEXA is not done as a %CV. This isn't applicable.

3. 1g change to 18" to allow all vendors to bid (This is a lock-out bid).

4. 1j remove to allow all vendors to bid (This is a lock-out bid specific only to Hologic). # of detectors is irrelevant as detector technology isn't the same. 1. Removed 1a-e

2. Removed 1a-e

3. 1g is MAX height, 18in is within 25 so do not see the issue. All vendors with table height 25 and under are allowed to bid.

4. 1j not sure what is being referred. 1j is regarding HD storage. # of detectors is not listed anywhere.

16 HOUSTON, TX 580B50010 1. 1a-d Should this be asking for "max" scan time?

2. DEXA is not done as a %CV. This isn't applicable.

3. 1g change to 18" to allow all vendors to bid (This is a lock-out bid).

4. 1j remove to allow all vendors to bid (This is a lock-out bid specific only to Hologic). # of detectors is irrelevant as detector technology isn't the same. 1. Removed 1a-e

2. Removed 1a-e

3. 1g is MAX height, 18in is within 25 so do not see the issue. All vendors with table height 25 and under are allowed to bid.

4. 1j not sure what is being referred. 1j is regarding HD storage. # of detectors is not listed anywhere.

17AMARILLO, TX504B570011.c. Respectfully request minimum detector size (cm) be changed from 31cm to 30cm.Done. The requirement has been changed
17AMARILLO, TX504B570011.i. Respectfully request explanation of what is meant by "forward tube source".The requirement has been removed
17AMARILLO, TX504B570012.y. Respectfully request removal of UPS on board requirement. This is a lock-out feature and limits competition.Please quote your UPS requiremennt
17AMARILLO, TX504B570012.dd. Respectfully request explanation of "Variable Landmarking" reference for use and impact in study.Tracking instruments and anatomical structures
17AMARILLO, TX504B570012.ff. Respectfully request explanation of "Peak opaciification" reference for use and impact in study.Vascular flow visualisation
17AMARILLO, TX504B570013.c. Respectfully request Max 30 fps be adjusted to 25 fps to increase competition.The requirement has been changed to 25fps
17AMARILLO, TX504B57001Can we please verify the SN of the OEC 9900 for trade in. You have D326082. The serial number should start with an E and look like EX-XXX.The serial number is accurate
17AMARILLO, TX504B57001Largest size in market is 30.8cm, is a value 30.x acceptable to meet the 31cm requirement?The requirement has been changed to 30cm
17AMARILLO, TX504B57001Request states "UPS on board" - can we provide UPS in the equote in place of it being a part of the C-Arm?Yes you can
17AMARILLO, TX504B57001Request states 32" monitor; would a smaller monitor or a secondary monitor be acceptable (making the view larger)?Monitor size dropped to 27in minimum
17AMARILLO, TX504B57001Line 24e. Please change the minimum range of motion horizontal travel (cm) to 15cm-26cm to allow competitionThe requirement has been changed to 15cm
17AMARILLO, TX504B57002Please confirm if this is for C-ARM or PORT RAD.This is for PORT RAD
17AMARILLO, TX504B57002Line 13 Please correct the requirement overview to reflect Portable XrayDone. The requirement has been adjusted
17AMARILLO, TX504B57002Line 25f. Please clarify if you are asking for a wireless detector 14 x 17?Yes, wireless
17AMARILLO, TX504B57002Line 33d. Please clarify the detector distributed weight limit in lbs?8lbs
17AMARILLO, TX504B57002Line 42g. Please move badge reader to optional for to allow for more competitionThe requirement will remain as is
17AMARILLO, TX504B57002Line 47 l. Please explain "Trickle Charge" method for detector battery ?The requirement will remain as is
17AMARILLO, TX504B57002Line 50 O. Please clarify if navigate remote the same as exposure remote?Yes, correct
17AMARILLO, TX504B57002Line 58 w. Please clarify if the display monitor on the tube head can be optional?Preferred
17BIG SPRING, TX519B500151. What are the procedures being performed for the Urology Application?

2. What are the procedures being performed for the Cardiovascular Clinical Application?

3. Could the maximum frequency range for 8.b High Frequency be changed to 18 MHz to increase competition?

4. Could the minimum frequency range for 8.e ultra broad band be expanded to 5 MHz to increase competition?

5. Do you want a biplane or triplane transducer?

6. Do you want Transrectal (TR) or Transperineal (TP)?

7. Do you want stabilized or freehand?

8. Is Phantom for quality control a hard requirement? 1. Future utilisation expansions considered

2. Future utilisation expansions considered

3. The requirement will remain as is

4. The requirement will remain as is

5. Biplane

6. Neither

7. Freehand

8. No. The requirement has been moved to the added value section

17BIG SPRING, TX519B500153.b is "three-dimensional (3D)-freehand" clinically necessary, can this be moved to "value add" section?The requirement has been moved to the added value section
17BIG SPRING, TX519B500153.c Is "navigation software" clinically necessary? As this is an interventional procedure done in combination with CT/MRI. Can this be moved to "value add" section?This requirement will remain as is
17BIG SPRING, TX519B500153.d Is "Image fusion software" clinically necessary? Similar interventional procedure to navigation software. This is done in combination with CT/MRI. Can this be moved to "value add" section?This requirement will remain as is
17BIG SPRING, TX519B500156.p. Is a physical phantom required? These are typically not provided by the manufacturers, Biomed typically has these already or they can purchased from outside companiesThe requirement has been moved to the added value section
17BIG SPRING, TX519B500156.j can you please expand on needs for strain-based elastrography, is this for breast, thryoid or another need?For liver elastography
17BIG SPRING, TX519B500156.s. Live xPlane is Philips lockout spec, can this be removed or moved to the "value add" section to increase competition?The requirement has been reworded
17BIG SPRING, TX519B500159 a-f Can these features be moved to value add section if 3.c-d are moved as these are all features of fusion and Image fusion?Those requirements will remain as is
17BIG SPRING, TX519B500159.a. Is "prostate gland volume and edge boundaries computation capabilities" clinically necessary, or can this be moved to the "value add"section?This requirement will remain as is
17BIG SPRING, TX519B500159.b. Is "tracking and recording of needle movement during biopsy in 3D coordinates in real-time clinical necessary or can this be moved "value add" section?This requirement will remain as is
17BIG SPRING, TX519B500159.c. Is "the availability of projected, future needle movement during biopsy in 3D coordinate" cinically necessary or can this be moved to "value add" section?This requirement will remain as is
17BIG SPRING, TX519B500159.d. Is" Ability to to overlay biopsy images on top of real time prostate images" clinically necessary or can this be moved to "value add" section?This requirement will remain as is
17BIG SPRING, TX519B500159. e. Is the 3D-generated model of prostate that can be exported to report, adjustable by additional non-imaging information, such as pathology-related data clinically necessary or can this moved to "value add section"?This requirement will remain as is
17BIG SPRING, TX519B500159. f Is standard planning templates for prostate biopsies clinical neccesary or can this be moved to "value add" section?This requirement will remain as is
17TEMPLE, TX674B50226#28 horizontal travel SPEC 150 OEC has 20.0 or 8". I'm not aware of a system that has this much horizontal travel, can we please verify this specifiction.The requirement has been modified
17TEMPLE, TX674B50226Number 29 vertical travel spec asked for 50. We're at 18 inches or 46 cm. I believe the specification may be off, can we please verify.The requirement has been modified
17TEMPLE, TX674B50226Maximum system dimensions 100 x 75? Please provide clarification of which measurement is being discussed here.The requirment has been removed
17TEMPLE, TX674B50226Maximum system weight spec asked for 300 pounds. Can this be confirmed?The requirement has been modified
17TEMPLE, TX674B50226Line 33 L. Please change the minimum generator output power kW to 2.3kW to allow competition.The requirement has been modified
17TEMPLE, TX674B502261.p. Respectfully request that the Minimum Matrix Size be changed from 3840 x 2160 to 1956x1956 to expand competitionThe requirement will remain as is
17TEMPLE, TX674B502261.r. Respectfully request Maximum System Weight (lbs) be increased from 300lbs to 700lbs to increase competiton unless this solicitation is for a Mini C-armThe requirement has been modified
17TEMPLE, TX674B502262.h. Respectfully request explanation of what is meant by "forward tube source".The requirment has been removed
17TEMPLE, TX674B502262.y. Respectfully request removal of UPS on board requirement. This is a lock-out feature and limits competition.Please quote your UPS requiremennt
17TEMPLE, TX674B502262.bb. Respectfully request clarification as to what is meant by "Remote control for entire unit". Is this a reference to wireless foot pedal?Operate C arm from a distance
19AURORA, CO554B53014Are there trade-in systems for this consideration line/RFO? The tech spec sheet provided does not specify.Yes, 3 GenRad GE Discovery 656 available for trade-in (installed 2014). Tech spec has been updated to reflect this.
19AURORA, CO554B53014Based on the overall configuration, if the detector is 17”x17” at the wall stand, would a static wall stand be acceptable? Wireless 17”x17” would still be available for the table.No change made: Fixed/Static wall-stand not acceptable.
19MUSKOGEE, OK623B540071. 1a-d Should this be asking for "max" scan time?

2. DEXA is not done as a %CV. This isn't applicable.

3. 1g change to 18" to allow all vendors to bid (This is a lock-out bid).

4. 1j remove to allow all vendors to bid (This is a lock-out bid specific only to Hologic). # of detectors is irrelevant as detector technology isn't the same. 1. 1a-e has been changed

2. In the context of bone mineral density (BMD) measurements, CV stands for coefficient of variation. It's a measure of precision and reproducibility of BMD scans, specifically how well the scan can produce consistent results if repeated. A lower CV indicates better precision and reproducibility. Please highlight your company's version of this terminology in your submission.

3. 1g has been changed to a maximum of 28" from the ground. The request to change the height to 18 would drastically reduce the competition.

4. 1j is referencing hard drive storage, not detectors. No matter the detector technology it should have a minimum hard drive storage of 150 GB

19AURORA, CO554B530131. Request 1.i. Minimum Battery Life 45 mins moved to Value Add Section.
2. Request 5.k. Ability to enter standby mode or sleep mode be moved to Value Add Section.1. Moved to added value

2. Moved to added value

19AURORA, CO554B530131. Can the minimum screen size be changed to 19 to increase competition?
2. What are the procedures being performed for the Breast Application?1. No change made

2. There are no breast applications needed, this requirement is for a vascular/cardiovascular ultrasound

21MARTINEZ, CA612B590051.c. Respectfully request minimum detector size (cm) be changed from 31cm to 30cm.Minimum detector size updated to 30cm
21MARTINEZ, CA612B590051.h. Respectfully request explanation of what is meant by "forward tube source".Removed requirement
21MARTINEZ, CA612B590052.x. Respectfully request removal of Battery Backup. This is a lock-out feature and limits competition.Battery backup removed from technical requirements section and added to Added Value Section
21MARTINEZ, CA612B590052.cc. Respectfully request explanation of "Peak opaciification" reference for use and impact in study.Peak opacification is for observing contrast flow in subtracted and un-subtracted runs and used for/along with Road mapping
21MARTINEZ, CA612B590052.dd. Respectfully request explanation of "Variable Landmarking" reference for use and impact in study.Landmarking is for seeing the position of vessels in relation to certain background anatomies during subtraction and road mapping. Landmarking enables ability to vary the percentage of background anatomy that displays in a subtracted image
21MARTINEZ, CA612B590052.ee. Respectfully request change from max of 30fps to max of 25fps for Recording and Playback.Requirement is not asking for a max fps. The requirement is for a variation of recording/playback rates
21MARTINEZ, CA612B590053.c. Resepectfully request Road Mapping be changed from Max 30fps be adjusted to Max 25fps to increase competition.Updated to 25fps
21STOCKTON, CA612B590061. Could the minimum screen size be changed to 19 inches to increase competition?

2. Could the minimum active screen size be changed to 19 inches to increase competition?

3. Could the minimum tilting be reduced 45 deg to increase competition?

4. Could the minimum image storage be reduced to 500gb to increase competition?

5. Is shear wave a hard requirement?

6. What function will the multi-function pedal be used for?

7. What are the procedures being performed for the Cardiovascular Clinical and Mitral Valva Navigator Applications?

8. Can the maximum frequency of 7. c. be changed to 18MHz to increase competition? 1. Changed to 19 inches to allow competition

2. Changed to 19 inches to allow competition

3. Changed to 90 deg to allow competition

4. Other manufacturers are able to meet the requirement for 1 TB storage. Will not change.

5. Confirmed with Vascular Service the requirement is not needed. Removed line item.

6. To allow functions to be programmed into each pedal for streamlined workflow for clinical staff. Moved to Added Value section.

7. No procedures are being performed at CBOC. Removed line items.

8. Changed to 18MHz to allow competition

21STOCKTON, CA612B590065.q. Is this feature found on other vendors' systems or is it a lock of spec? If it is a lockout can it be removed or moved to "value add" section to increase competitionMoved to Added Value section to allow competition
21STOCKTON, CA612B590066.c. Is"mitral value navigator" clinically necessary? This is a 3D Transesophageal feature, and there is not a TEE probe included in the spec. Can this be moved to "value add" section to increase compeition?Removed Mitral Valve Navigator line item.
21STOCKTON, CA612B59007Please provide the Biomedical Engineering Point of Contact.Not available.
21STOCKTON, CA612B59007The specs list both an image intensifier & a flat panel detector. Please confirm that you are requesting a flat panel detector.Yes, we are requesting a flat panel detector. Removed image intesifier line item.
21RENO, NV654B500341. Request 1.k. Minimum Battery Life 3 hours changed to 60 min.

2. Request 1.m. Maximum equipment weight 153 lbs be changed to 190 lbs.

1. Yes

2. Changed to 200 lbs.

21 RENO, NV 654B50034 1. Can the minimum screen size be changed to 19 to increase competition?

2. Can the minimum…