AMD 0001 Attachment 001 Performance Work Statement and QASP.pdf

PDF 178 KB Posted

Attached to
Laboratory Sterilizer Maintenance Service - Manhattan KS Federal contract opportunity
Solicitation number
1232SA25Q0926
Issued by
Department of Agriculture Agricultural Research Service

About this file

This Performance Work Statement (PWS) details laboratory sterilizer maintenance service requirements for the Center for Grain and Animal Health (CGAHR) Research. The contract involves calibrating temperature and pressure sensors on two Getinge Model 533LS autoclaves located in Room 117, with specific performance requirements including: conducting preventative maintenance once per base and option periods, providing 4-hour telephone response and 48-business hour onsite response, and submitting comprehensive calibration reports. Key compliance requirements mandate that the contractor provide a repair service technician authorized by the original equipment manufacturer, and all personnel must be escorted by a USDA representative while on location. The Quality Assurance Surveillance Plan (QASP) establishes a 100% acceptable quality level for all technical requirements outlined in Section 2 of the document.

View the file

Other files for this federal contract opportunity

Other files attached to Laboratory Sterilizer Maintenance Service - Manhattan KS, newest first.
File Type Posted
Attachment 003 Transmittal Summary.pdf PDF
Attachment 002 Wage Determination 2015_5337_R27.pdf PDF
RFQ 1232SA25Q0926 Clauses and Provisions.pdf PDF
Attachment 001 Performance Work Statement and QASP.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

AMD 0001 ATTACHMENT 001 PERFORMANCE WORK STATEMENT AND QASP

Laboratory Sterilizer Maintenance Service

1. General

1.1. Background: The Center for Grain and Animal Health (CGAHR) Research is seeking a company with extensive knowledge and competence calibrating temperature and pressure sensor on Getinge sterilizers and have Getinge autoclave passwords. Downtime of the sterilizers during calibration services shall be kept to a minimum. The facility will be occupied during this work. The CGAHR campus address is 1515 College, Manhattan, KS.

1.2. Scope : The contractor shall calibrate temperature and pressure sensors for sterilizers located at CGAHR. Additionally, the contractor shall provide documentation (calibration certificate, supporting records, and measurement uncertainties) that the equipment being used to calibrate the sensors have been calibrated within one year. The contractor shall submit reports of the calibration work.

1.3. Equipment List:

Description Location Getinge Model Serial Number

AutoClave Room 117 Model 533LS 08H09028

Bench Top Autoclave Room 117 Model 533LS 09M09650

2. Contractor Performance Requirements:

2.1. Conduct preventative maintenance travel and labor once within the base period and once per any awarded option periods.

2.2. Repair travel and labor.

2.3. Telephone response within 4 hours.

2.4. Onsite response 48 (business hours).

2.5. Provide calibration reports, including supporting records and measurement uncertainties, for all serviced equipment.

2.6. Building will be occupied while service is ongoing. Contractor shall schedule work with location to allow location to meet current scientific and research needs.

2.7. Contractor’s Key Personnel: The contractor shall provide a repair service technician that is authorized by the OEM manufacturer of the serial numbered equipment detailed in Section 1.3 above to perform and document the service requirements.

2.8. Security Requirements: All contractor personnel shall always be in the presence of a

USDA escort while on location.

3. Quality Assurance Surveillance Plan (QASP): The Government shall establish a QASP as a means of evaluating whether the Contractor is meeting the performance standards/quality levels identified in the PWS to ensure that the government pays only for the level of services received.

Tasks Standards Acceptable Quality

Level (AQL)

Technical Requirements and

Tasks identified in Section 2

Standards as established in Technical requirements and tasks identified in

Section 2

100%