This equipment drawing document details the STERIS Reliance® 1227 Cart and Utensil Washer/Disinfector, a vented, pit-mounted unit designed for medical or laboratory environments. The comprehensive technical specifications cover utility requirements, physical specifications, installation parameters, and dimensional details for a complex industrial washing system. Key specifications include a chamber with internal dimensions of 34 x 80 x 106 inches, an internal volume of 166.85 cubic feet, and specific utility requirements for hot water, cold water, compressed air, steam, electrical connections, and ventilation. The unit has precise installation requirements, including floor load specifications, pit dimensions, and environmental operating conditions such as temperature range of 41-104°F and altitude up to 6,560 feet.

The washer's utility requirements are extensive, including electrical connections for 200-208V or 460-480V three-phase power, compressed air with specific purity standards, steam inlet with pressure range of 30-80 psig, and ventilation exhaust at 500 SCFM for vapor and drying. The system includes chemical dosing capabilities for detergent, rinse aid, and descaler, with precise consumption measurements per cycle. Physical requirements detail shipping weights and dimensions across multiple crates, with the heaviest uncrated piece (mechanical core) weighing 890 pounds. The documentation emphasizes precise installation requirements, including floor load specifications, pit dimensions, and recommendations for facility preparation, making it a highly specialized piece of industrial equipment.

View the file

Other files

File Type Posted
Screenshot Bldg 1 RM 258 Utilities.png PNG image
36C24425B0023 0002.pdf PDF
36C24425B0023 0001.pdf PDF
TIL Requirements SPD Cart Washer.pdf PDF
36C24425B0023_1.pdf PDF
SPD Cart Washer Photo.jpeg JPEG file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

80 [2032]

LOADING

CLEARANCE

HEIGHT

112-3/4 [2864]

CLEARANCE

FOR RELIEF

VALVE

OPERATION

102 [2591]

UNIT

HEIGHT

114-1/2 [2908]

72 [1829]

UNIT WIDTH

12-3/8 [314] 34 [864]

LOADING

CLEARANCE

WIDTH

13-1/8 [333]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

TYPICAL INSTALLATION (1/2)

NOTES:

(A) UNLESS OTHERWISE SPECIFIED, ALL DIMENSIONS ARE IN INCHES [mm].

(B) CUSTOMER MUST BE SURE THAT MACHINE STANDS ON A NON-COMBUSTIBLE FLOOR.

(C) ALLOW SUFFICIENT SPACE IN FRONT OF LOADING AREA TO LOAD / UNLOAD CARTS

ACCORDING TO CUSTOMER CARTS AND DIMENSION AND STERIS WASHING ACCESSORIES:

CONTAINER RACK 63 1/2 [1612] LG

UTENSIL RACK 65 1/2 [1663] LG

(D) ENCLOSURES COULD BE BUILDING WALLS OR SERVICE AND NON-SERVICE S/S PANELS

ACCESSORIES (SEE PAGE 3)

(E) FOR UNIT WITHOUT SERVICE SIDE ACCESS PANELS, SERVICE AREA ACCESS

DOOR IS INSTALLED ON LOAD SIDE OF THE WASHER (NOT BY STERIS).

LOAD SIDE CAN BE EITHER ON LEFT OR RIGHT SIDE OF THE MECHANICAL CORE.

(F) IF RECOMMENDED FLOOR CLEARANCE FOR INSTALLATION IS NOT AVAILABLE

ADDITIONAL MANPOWER AND EQUIPMENT WILL BE REQUIRED.

(G) SERVICE AREA REQUIRED IF STANDARD UNIT.

(H) SERVICE AREA REQUIRED IF NON-VENTED OPTION (NOT SHOWN) IS CHOSEN.

FRONT VIEW

Page 1 of 16 920515107.D

Released

120-1/2 [3061]

UNIT LENGTH

74 [1880]

WALL

OPENING

112 [2845]

68 [1727]

20 [508]

24 [610]

92-5/8 [2353]

SEE PAGE 1 *(NOTE H)

82-5/8 [2099]

SEE PAGE 1 *(NOTE G)

126 [3200] 126 [3200]

120 [3048]

SERVICE CLEARANCE

ABOVE FINISHED

FLOOR

98 [2489] AT 104 [2642]

CLEARANCE MIN.

4-1/4 [108]

1 [25]

105 [2667]

WALL OPENING

3-1/2 [89]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

TOP VIEW

SERVICE

AREA

FLOOR CLEARANCE

FOR INSTALLATION

SEE PAGE 1 *(NOTE F) SEE PAGE 1 *(NOTE F)

FOR INSTALLATION

FLOOR CLEARANCE

SERVICE DOOR

SEE PAGE 1 *(NOTE E)

SEE PAGE 1 *(NOTE D)

SEE PAGE 1 *(NOTE D)

SIDE VIEW

CEILING DECK (IF PRESENT)

TYPICAL INSTALLATION (2/2)

WITHOUT SIDE ACCESS PANELS (ACCESSORY)

Page 2 of 16 920515107.D

106 [2692]

UNIT WIDTH

[660]

108 [2743]

WALL OPENING

82-1/2 [2096]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

TYPICAL INSTALLATION

SIDE ACCESS PANELS (ACCESSORY)

NON-SERVICE

SIDE WALL

S E

R V

IC

E

A R

E A

Page 3 of 16 920515107.D

A

BB

112 [2845] 3 [76]

118 [2997]

[1829]

2X 21-1/4 [540]

2X 12 [305]

4X 15 [381]

67-1/2 [1715]

[1880]

2 [51]

*(NOTE E)

[559]

4-1/8 [105]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

STANDARD CONSTRUCTION

(WITHOUT SEISMIC ANCHORAGE)

1/2

FLOOR SLOPE

TOWARD FLOOR

DRAIN DEGREE: 1

FLOOR DRAIN

LIMIT OF PIT

*(NOTE D)

SERVICE

AREA

NOTES:

(A) ALL MATERIALS (INCLUDING CONCRETE, LEVELING PADS, STAINLESS STEEL ANGLES, FLOOR SINK...) ARE

BY OTHERS THAN STERIS. PIT SLOPE TOWARDS DRAIN. STAINLESS STEEL ANGLES ARE

RECOMMENDED BY STERIS BUT ARE NOT NECESSARY FOR WASHER INSTALLATION.

(B) REFER TO PAGE 13 FOR FLOOR DRAIN FLOW RATE CAPACITY REQUIREMENTS.

(C) STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.

(D) THIS CART WASHER CAN FIT INTO AN EXISTING 76 in [1930 mm] PIT, BUT WITH RESTRICTED ACCESS TO

COMPONENTS LOCATED IN THE PIT. THE DRAIN LOCATION MAY NEED TO BE RELOCATED.

(E) FOR EXISTING 76 in [1930 mm] PIT WASHER, THE CLEARANCE WILL BE 2 in [51 mm] INSTEAD OF 3 in [76 mm].

(F) STERIS RECOMMENDS A WATER TIGHT PIT.

SUMP

4X CORNER

LEVELING PAD

Page 4 of 16 920515107.D

FLOOR SINK

8 [203] MAX.

FLOOR DRAIN

4 [102] MIN

2X 12 [305] 2X 21-1/4 [540]

4X 15 [381]

6 [152] MIN

6-3/4 [171] MAX

67-1/2 [1715]

6-3/4 [171]

MIN

22 [559]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

STANDARD CONSTRUCTION

(WITHOUT SEISMIC ANCHORAGE)

2/2

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

CORNER

LEVELING PAD

SEE PAGE 4 *(NOTE C)

FINISHED FLOOR

(RIGHT END)

WASHER DRAIN OUTLET

SEE PAGE 4 *(NOTE B AND C)

B-BSECTION

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

CORNER

LEVELING PAD

WASHER DRAIN OUTLET

SEE PAGE 4 *(NOTE B AND C)

FINISHED FLOOR

(SERVICE AREA)

A-ASECTION

Page 5 of 16 920515107.D

BB

112 [2845]

[1829]

3 [76]

118 [2997]

4X 10 [254]

4X 8 [203]

60-3/4 [1543]

22 [559]

67-1/2 [1715] 74

[1880]

4-1/8 [105]

2 [51]

*(NOTE E)

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

CONSTRUCTION WITH SEISMIC ANCHORAGE

(SEE REPORT #10183017 AGAINST CBC-2013, REPORT #11863511

AGAINST CBC-2019 AND REPORT #11053380 AGAINST CBC-2022)

1/2

SUMP

FLOOR DRAIN

FLOOR SLOPE

TOWARD FLOOR

DRAIN DEGREE: 1

LIMIT OF PIT

*(NOTE D)

SERVICE

AREA

NOTES:

(A) ALL MATERIALS (INCLUDING CONCRETE, LEVELING PADS, STAINLESS STEEL ANGLES, FLOOR SINK...)

ARE BY OTHERS THAN STERIS. PIT SLOPE TOWARDS DRAIN. STAINLESS STEEL ANGLES ARE

RECOMMENDED BY STERIS BUT ARE NOT NECESSARY FOR WASHER INSTALLATION.

(B) REFER TO PAGE 13 FOR FLOOR DRAIN FLOW RATE CAPACITY REQUIREMENTS.

(C) STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.

(D) THIS CART WASHER CAN FIT INTO AN EXISTING 76 in [1930 mm] PIT, BUT WITH RESTRICTED ACCESS TO

COMPONENTS LOCATED IN THE PIT. THE DRAIN LOCATION MAY NEED TO BE RELOCATED.

(E) FOR EXISTING 76 in [1930 mm] PIT WASHER, THE CLEARANCE WILL BE 2 in [51 mm] INSTEAD OF 3 in [76 mm].

(F) STERIS RECOMMENDS A WATER TIGHT PIT.

(G) WELDING OF THE SUMP ATTACHMENT TO FOUNDATION PLATE IS PERFORMED BY OTHERS.

(H) REFER TO SEISMIC REPORT FOR FOUNDATION PLATES AND CONCRETE CONSTRUCTION REQUIREMENTS.

STERIS OFFERS SEISMIC PRE-INSTALL ACCESSORY (FD33-5). WHICH INCLUDES THE FOUNDATION PLATES.

(I) ALL FOUR CORNER FOUNDATION PLATES UPPER SURFACE SHALL BE LEVELLED. THE PLATE STANDING ON HIGHEST PART OF THE CONCRETE FLOOR SHALL MEET THE SHOWN POSITION. STERIS PROVIDES SHIMS TO LEVEL THE OTHER CORNERS TO A MAXIMUM OF 1/2 in [12.7 mm] FROM NOMINAL POSITION.

4X CORNER

FOUNDATION PLATES

*(NOTE H AND I)

Page 6 of 16 920515107.D

7+1/4 +6 -0 [178-0 ]

SEE PAGE 6 *(NOTE I)

FLOOR SINK

8 [203] MAX.

FLOOR DRAIN

4 [102] MIN

4X 8 [203]

8-1/8 [206]

REF

60-3/4 [1543]

4X 10 [254]

10 [254]

MIN

67-1/2 [1715]

22 [559]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

A-ASECTION

4X CORNER

FOUNDATION PLATES

SEE PAGE 6

*(NOTE H AND I)

FINISHED FLOOR

(RIGHT END)

WASHER DRAIN OUTLET

SEE PAGE 6

*(NOTE B AND C)

B-BSECTION

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

SEE PAGE 6 *(NOTE C)

FINISHED FLOOR

(SERVICE AREA)

WASHER DRAIN OUTLET

SEE PAGE 6 *(NOTE B AND C)

RECOMMENDED

2 X 2 STAINLESS-STEEL

ANGLES AROUND

PERIMETER OF PIT

CONSTRUCTION WITH SEISMIC ANCHORAGE

(SEE REPORT #10183017 AGAINST CBC-2013, REPORT #11863511

AGAINST CBC-2019 AND REPORT #11053380 AGAINST CBC-2022)

2/2

Page 7 of 16 920515107.D

78-3/4 [2000]

73-3/8 [1864]

91 [2311]

61-3/4 [1568]

11-3/4 [298]

48-1/4 [1226]

93 [2362]

67-1/2 [1715]

91-1/8 [2315]

2X 102-1/4 [2597]

107-1/2 [2731] 108-3/8

[2753]

68-1/4 [1734]

87-3/8 [2219]

79-7/8 [2029]

64-1/4 [1632]

HW HOT WATER

CW COLD WATER

A COMPRESSED AIR

D FLOOR DRAIN LOCATION *(NOTE F)

S STEAM *(NOTE C)

CR CONDENSATE RETURN

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

RELAY FUNCTION SEE PAGE 9 *(NOTE B)

CR10 HOT WATER REQUEST

CR12 COLD WATER REQUEST

UTILITY (PLUMBING)

CONNECTION POSITION

FRONT VIEW

HW

CW

A

D

CR S

NOTES:

(A) ALL CONNECTIONS SHOULD BE IN ACCORDANCE

WITH LOCAL CODES.

(B) EXTERNAL ISOLATION VALVES, UNIONS, VACUUM

BREAKERS ON UTILITY LINES TO WASHER ARE

NOT PROVIDED BY STERIS.

(C) ALL OVER PRESSURE DEVICES TO CONTROL

PRESSURE OF STEAM, WATER AND AIR ARE NOT

INCLUDED. IT IS RESPONSIBILITY OF CUSTOMER

TO INSTALL THOSE EQUIPMENT ON MAIN UTILITIES.

(D) DO NOT PLACE PIPING IN SERVICE ACCESS AREA.

(E) SIGNAL AVAILABLE FOR WATER REQUEST INSIDE

ELECTRICAL BOX.

(F) WASHER DRAIN OUTLET MAY NOT ALIGN PERFECTLY

WITH CENTER OF FLOOR SINK. EDGES OF FLOOR SINK

MUST BE AT LEAST 1 in [25 mm] FROM THE WASHER DRAIN.

(G) AIR LINE, ELECTRICAL CONDUIT, DUCT WORK, STEAM PIPING,

CONDENSATE RETURN PIPING AND WATER PIPING SHOULD

BE ROUTED ABOVE WASH CHAMBER TO AVOID INTERFERING

WITH SERVICE OF COMPONENTS ON WASHER ROOF.

SIDE VIEW

HW

CW

A

D

CR

S

*(NOTE E)

TOP VIEW

HW

CW

A

D

CR

S

SERVICE

AREA

*(NOTE D)

6-3/4 MIN

[171]

Page 8 of 16 920515107.D

16-1/2 [419]

61-3/4 [1568]

48 [1219]

E ELECTRICAL

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

RELAY FUNCTION *(NOTE B)

CR5 CYCLE IN OPERATION

CR6 LOW VENTILATION REQUEST

CR7 HIGH VENTILATION REQUEST

CR8 ALARM

CR9 CYCLE COMPLETE

CR10 HOT WATER REQUEST

CR12 COLD WATER REQUEST

UTILITY (ELECTRICITY)

CONNECTION POSITION

NOTES:

(A) ALL CONNECTIONS SHOULD BE IN ACCORDANCE WITH

LOCAL CODES.

(B) AUXILIARY DRY CONTACTS FOR CUSTOMER UTILITY CONTROL

AVAILABLE INSIDE CONTROL BOX.

SEE ELECTRICAL SCHEMATIC: #10132900

(C) FOR STANDARD THROUGH-WALL UNIT, CLEARANCE BETWEEN

CLOSED DOOR OF ELECTRIC CABINET AND FACILITY WALL, OR

ANY OPPOSING FIXTURE, MUST MEET LOCAL CODES.

(D) STERIS RECOMMENDS A LIGHT IN SERVICE

AREA ALONG WITH A PROVISION OF AN

ELECTRICAL OUTLET FOR MAINTENANCE.

FRONT VIEW

E

SIDE VIEW

E

*(NOTE C)

TOP VIEW

E SERVICE

AREA

Page 9 of 16 920515107.D

118-7/16 [3008]

58 [1473]

75 [1905]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

RELAY FUNCTION *(NOTE F)

CR5 CYCLE IN OPERATION

CR6 LOW VENTILATION REQUEST (250 SCFM)

CR7 HIGH VENTILATION REQUEST (500 SCFM)

UTILITY (VENTILATION)

CONNECTION POSITION

NOTES:

(A) INSTALLATION OF EXTERNAL DUCTING FROM WASHER EXHAUST FAN TO THE EXTERIOR OF THE BUILDING (NOT BY STERIS) MUST NOT EXCEED 1 in H2O IN PRESSURE LOSS TO ENSURE OPTIMAL DRYING PERFORMANCE OF THE WASHER AND FAN MOTOR EFFICIENCY.

IF PRESSURE LOSS OF EXTERNAL DUCTING EXCEEDS 1 in H2O, AN EXTERNAL FAN MUST BE INSTALLED TO COMPENSATE (NOT BY STERIS).

(B) BUILDING DUCT MUST BE SCALED UP TO 10 in [254 mm] O.D. MINIMUM AS CLOSE AS POSSIBLE TO WASHER DUCT CONNECTION (ADAPTOR FROM 6 in [152 mm] TO BUILDING DUCT O.D. NOT PROVIDED BY STERIS).

(C) FACILITY ROOM VENTILATION MUST COMPENSATE FOR WASHER EXHAUST FLOW RATE.

(D) STERIS DOES NOT RECOMMEND INSTALLATION OF SPRINKLER OVER THE UNIT, BUT IF REQUIRED BY LOCAL CODE IT MUST BE FUSED TO 260 F [126 C] OR ABOVE. ANY SPRINKLER HEADS IN FRONT OF THE LOAD/UNLOAD DOORS MUST ALSO BE FUSED TO 260 F [126 C] OR ABOVE.

(E) DO NOT PLACE DUCT WORK TO INTERFERE WITH SERVICE ACCES AREA.

(F) SIGNAL AVAILABLE FOR EXTERNAL VENTILATION CONTROL INSIDE ELECTRICAL BOX.

FRONT VIEW

EX

SIDE VIEW

EX

*(NOTE F)

TOP VIEW

EX

SERVICE

AREA

*(NOTE E)

EX VENTILATION

Page 10 of 16 920515107.D

60 [1524]

REF

8-3/8 [213]

132 [3353]

MAX

6-1/2 [165]

CONSUMPTION / CYCLE STANDARD DECONTAM

CHEMICAL

(CH)

DETERGENT 2.5 oz @ 1/10 oz/gal [76 ml @ 0.8 ml/L] N/A

RINSE AID 0.5 oz @ 1/50 oz/gal [15 ml @ 0.16 ml/L] N/A

DESCALER N/A 50 oz @ 2 oz/gal [1520 ml @ 16 ml/L]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

DOSING STATION

*NOTES:

(A) MAXIMUM CONTAINER HEIGHT FROM FLOOR IS 24 in (610 mm).

AMBIENT TEMPERATURE SHALL NOT EXCEED 40°C (104°F)

(B) UP TO 150 ft AVAILABLE ON SPECIAL ORDER

FRONT VIEW

CEILING

CHEMICAL

CONTAINERS

*(NOTE A)

12 ft [3.6 m] OF

INJECTION LINE

(FROM CONTAINERS

TO DOSING STATION)

DOSING STATION

WEIGHT : 7 lbs [3.2 kg]

UP TO 2 TUBES

*(NOTE B)

100 ft (30 m) OF INJECTION LINE

AND SIGNAL CABLE *(NOTE B)

DETERGENT LINES SHALL NOT BE IN CONTACT WITH HOT SURFACES.

STERIS RECOMMENDS CUSTOMER TO ROUTE CHEMICAL LINES AND

SIGNAL CABLE THROUGH A 3 in [76 mm] ID CONDUIT.

EXCESS LENGTH MUST BE CUT BEFORE CONNECTING TO WASHER/STATION

TO AVOID FORMING BUNDLES THAT COULD AFFECT THE CHEMICAL INJECTION

SYSTEM'S PERFORMANCE (DO NOT CUT SIGNAL CABLE).

Page 11 of 16 920515107.D

UTILITY REQUIREMENTS CHART (PLUMBING)

UTILITY PARAMETER REQUIREMENT NOTES

HOT WATER

(HW)

CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT

1. INSTALLATION OF A WATER HAMMER ARRESTOR

(NOT PROVIDED BY STERIS) IS RECOMMENDED.

2. CUSTOMER MUST PROVIDE UTILITY

CONNECTIONS WITH SHUTOFFS DISCONNECTS

WITHIN 2 FEET OF THE PERIMETER OF THE

EQUIPMENT AND BELOW THE CEILING DECK.

3. OPTIMAL WASHER PERFORMANCES OBTAINED

WHEN UTILITIES ARE PROVIDED AT MAXIMUM

VALUE OF RANGES.

4. OPTIMAL CLEANING EFFICIENCY WHEN HOT

WATER IS AT A MAXIMUM OF 135 F [57.2 C].

5. IF CYCLE WASH PHASE SET POINT

TEMPERATURE IS SET HIGHER THAN 140 F

[60 C], CONSUMPTION FOR DRAIN COOLING

IS DOUBLE.

6. IF DRAIN COOLING SYSTEM IS ENABLED,

COLD WATER MAXIMUM FLOW RATE IS

INCREASED TO : 60 US gpm [227 L/MIN].

PRESSURE RANGE

15 - 50 psig DYNAMIC

[103-345 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE

19 US GPM [72 L/min] AT 15 psig [103 kPa] TO

35 US GPM [132 L/min] AT 50 psig [345 kPa]

TOTAL CONSUMPTION

FOR STANDARD CYCLE

50 US GAL [189 L]

(25 US GAL [95 L] PER PHASE)

TEMPERATURE RANGE 110°F [43°C] TO 150°F [65.5°C]

WATER QUALITY HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

COLD WATER

(CW)

*(NOTE 6)

CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT

PRESSURE RANGE

25 - 40 psig DYNAMIC

[172-276 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING FLOW RATE

19 US GPM [72 L/min] AT 25 psig [172 kPa] TO

23 US GPM [23 L/min] AT 40 psig [276 kPa]

TOTAL CONSUMPTION

FOR STANDARD CYCLE

DRAIN COOLING:

19 US GAL [72 L] AT 25 psig [172 kPa]

TEMPERATURE RANGE 40 - 70°F [4.4 - 21°C] MAXIMUM

WATER QUALITY HARDNESS (CaCO3) : 50-80 ppm (MAX. 120 ppm)

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 12 of 16 920515107.D

UTILITY REQUIREMENTS CHART (PLUMBING)

UTILITY PARAMETER REQUIREMENT NOTES

COMPRESSED

AIR

(A)

CONNECTION SIZE / TYPE 1/2 in / FEMALE NPT 1/2 in / MALE BSPT

1. AIR PURITY PER ISO-8573-1:2010 [7:4:4]

MAX. PARTICULATE DENSITY: 10 mg/m³

MAX. DEW POINT FOR WATER CONTENT:

3°C [37°F]

MAX. OIL CONCENTRATION FOR THE OIL

CONTENT: 5 mg/m³

2. STERIS RECOMMENDS A REFRIGERATED AIR

DRYER WHERE ENVIRONMENTAL DEW POINT

CONDITIONS ARE HIGHER THAN

RECOMMENDED (CONTACT STERIS SALES

REPRESENTATIVE).

3. STERIS RECOMMENDS A MAXIMUM 8 in [203 mm] ROUND OR SQUARE FLOOR SINK BY 6 in [152 mm] DEEP.

4. IF DRAIN COOLING SYSTEM IS ENABLED,

DRAIN MAXIMUM FLOW RATE IS INCREASED TO:

120 US GPM [454 L/min]

5. MAXIMUM FLOW RATE OCCURS AT RESERVOIR

COMPLETE DRAINING (EX: POWER OFF). IT WILL

BE MAINTAINED FOR A MAXIMUM OF:

50 SECONDS FOR UNITS WITH 2 RESERVOIRS.

6. STERIS REQUIRES AT LEAST 97% SATURATED

DRY PLANT STEAM, FREE OF DIRT, RESIDUES

AND CONTAMINANTS.

7. PROPER OPERATION OF THE WASHER

REQUIRES A MINIMUM DYNAMIC PRESSURE

DIFFERENTIAL OF 21.5 psig [148 kPa] BETWEEN

STEAM INLET AND CONDENSATE RETURN.

8. PEAK OPERATING FLOW RATES OCCUR WHEN

RECIRCULATION HEATER AND RESERVOIR

HEATER ARE ACTIVATED AT THE SAME TIME.

IT WILL BE MAINTAINED FOR A MAXIMUM OF

90 SECONDS.

9. STERIS REQUIRES A LOW PRESSSURE

CONDENSATE RETURN SYSTEM.

PRESSURE RANGE MIN. 80 psig DYNAMIC [552 kPa] MAX. 125 psig STATIC [860 kPa]

OPERATING FLOW RATE 6 SCFM [170 L/min]

DRAIN

(D)

CONNECTION SIZE / TYPE

*(NOTE 3) 4 in [101 mm] OD

OPERATING FLOW RATE

*(NOTE 4 - 5) MAX. 60 US GPM [227 L/min]

STEAM INLET

(S)

CONNECTION SIZE / TYPE 1 1/2 in / FEMALE NPT 1 1/2 in / MALE BSPT

PRESSURE RANGE

30 - 80 psig DYNAMIC

[206-552 kPa] MAX. 90 psig STATIC [620 kPa]

OPERATING

MASS FLOW RATE

PEAK 900 lb/h [409 kg/h] AT 30 psig [207 kPa]

PEAK 1400 lb/h [636 kg/h AT 80 psig [550 kPa]

TOTAL CONSUMPTION

FOR STANDARD CYCLES 53 lb [24.0 kg]

CONDENSATE

RETURN

(CR)

CONNECTION SIZE / TYPE 1 in / FEMALE NPT 1 in / MALE BSPT

PRESSURE RANGE MAX. 10 psig [68.9 kPa] STATIC

OPERATING FLOW RATE PEAK : 2.9 US GPM [11 L/min]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 13 of 16 920515107.D

UTILITY REQUIREMENTS CHART (ELECTRIC)

UTILITY

PARAMETER AND REQUIREMENT

NOTES

CONNECTION

SIZE/TYPE

NOMINAL

AMPERAGE

RECOMMENDED

CIRCUIT

ELECTRICITY

(EL)

200-208V 60 Hz, 3-Ph

1 - 1 1/4 in [35 - 44 mm]

CONDUIT

25.5A 40A

1. ALL CONNECTIONS SHOULD BE IN

ACCORDANCE WITH LOCAL CODES.

2. NEUTRAL WIRE NOT REQUIRED.

3. ALWAYS FOLLOW LOCAL ELECTRICAL CODES

AND SAFETY-RELATED WORK PRATICES TO

DETERMINE THE VALUE OF PROTECTION.

4. WIRING ON THE EQUIPMENT TERMINATES AT

A JUNCTION BOX. WIRING BETWEEN JUNCTION

BOX AND BUILDING SERVICE LINES ARE NOT

SUPPLIED BY STERIS.

5. DISCONNECT SWITCHES WITH OFF POSITION

LOCKOUT (SUPPLIED BY OTHERS) MUST BE

INSTALLED IN ELECTRIC SUPPLY LINES NEAR THE

EQUIPMENT.

6. BREAKERS NOT PROVIDED BY STERIS

7. MAIN SUPPLY TERMINAL BLOCK WIRE

SIZE: 6 AWG MAX.

8. PROTECTIVE EARTH TERMINAL BLOCK WIRE

SIZE: 4 AWG MAX.

460-480V 60 Hz, 3-Ph

1 - 1 1/4 in [35 - 44 mm]

CONDUIT

13.5A 20A

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

Page 14 of 16 920515107.D

PHYSICAL SPECIFICATIONS

PARAMETER SPECIFICATION NOTES

OPENING AREA FROM CLEAN SIDE

TO SERVICE AREA LESS THAN 30 in

[19354 mm

CHAMBER INTERNAL DIMENSIONS

W x H x L

34 in X 80 in X 106 in [864 mm X 2032 mm X 2692 mm]

CHAMBER INTERNAL VOLUME 166.85 ft [4.73 m

NOISE LEVEL: A-WEIGHTED EQUIVALENT SURFACE

SOUND PRESSURE LEVEL 65 dB (A) AS PER ISO 3746

PERMISSIBLE ENVIRONMENT

INDOOR USE ONLY:

- ALTITUDE OF OPERATION UP TO 6560 ft (2000 m);

- TEMPERATURE OF 41 TO 104 F (5 TO 40 C);

- MAXIMUM RELATIVE HUMIDITY IS 80% FOR

TEMPERATURES UP TO 88 F (31 C) DECREASING

LINEARLY TO 50% RELATIVE HUMIDITY AT 104 F (40 C);

- POLLUTION DEGREE 2: EQUIPMENT MUST BE

INSTALLED IN AN ENVIRONMENT WHERE NORMALLY

ONLY NON-CONDUCTIVE POLLUTION OCCURS BUT

WHERE OCCASIONAL, TEMPORARY CONDUCTIVITY

CAUSED BY CONDENSATION CAN BE EXPECTED.

AS DETERMINED

ACCORDING TO

INTERNATIONAL STANDARD

EN/IEC 61010-1, SECOND

EDITION

HEAT LOSS

AVERAGE STANDARD VALIDATED CYCLE WITH ROOM @ 75 F [24 C]

TOTAL HEAT LOSS

FREE-STANDING INSTALLATION (ONE WALL) THROUGH-WALLS INSTALLATION (TWO WALLS)

FRONT OF WALL BACK OF WALL AT EACH END BETWEEN WALLS

14600 Btu/h [4278 W]

4400 Btu/h [1289 W]

10200 Btu/h [2989 W]

4400 Btu/h [1289 W]

5800 Btu/h [1700 W]

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

UTILITY REQUIREMENTS CHART

UTILITY PARAMETER REQUIREMENT NOTES

VENTILATION

EXHAUST

(EX)

CONNECTION SIZE / TYPE

*(NOTE 1)

6 in [152 mm]

OUTSIDE DIAMETER

1. BUILDING DUCT MUST BE SCALED UP TO 10 in [254 mm] O.D.

MINIMUM AS CLOSE AS POSSIBLE TO WASHER DUCT

CONNECTION (ADAPTOR FROM 6 in [152 mm] TO BUILDING

DUCT O.D. NOT PROVIDED BY STERIS)

2. FOR ALL VENTILATION DUCTING FROM WASHER, STERIS

RECOMMENDS THE INSTALLATION OF A DEDICATED

CORROSION-PROOF AND WATERTIGHT DUCT RATED

TO AN OPERATING TEMPERATURE OF 210 F [99 C] OR

MORE TO THE EXTERIOR OF THE BUILDING, SLOPED

TOWARD THE WASHER AND FREE OF DEAD LEGS.

3. INTERNAL DRYING AND EXHAUST FANS PRODUCE 475 SCFM

AND BUILDING EXHAUST NEEDS TO EXHAUST AT THE RATE

OF 500 SCFM.

PRESSURE DROP OF

BUILDING VENTILATION

DUCT WORK

< 1 in H2O @ 1000 scfm

OPERATING FLOW RATE

*(NOTE 3)

VAPOR EXHAUST AND DRYING:

500 scfm [14.15 m

/min]

WASH AND RINSE:

250 scfm [7.07 m

/min]

TEMPERATURE RANGE MAX. 212°F [100°C]

Page 15 of 16 920515107.D

T I T L E

EQUIPMENT DRAWING OF

RELIANCE® 1227 CART AND UTENSIL WASHER/DISINFECTOR

VENTED, PIT MOUNTED UNIT

Corporation

SIZE: SCALE: DRAWN BY: DRAWING NUMBER: PART NUMBER

& REVISIONA N/A Michel Blouin 920515107

This document contains confidential and proprietary information of STERIS Corporation. Neither this document nor the information herein are to be reproduced, either in part or in whole, except as specifically authorized by STERIS Corporation.

PHYSICAL REQUIREMENTS

ITEM REQUIREMENTS SPECIFICATIONS OR REFERENCE

TRANSPORT

VERIFY THAT HALLS, DOORS, ELEVATORS AND ALL OTHER ENTRANCES MEET WEIGHT AND DIMENSIONAL CLEARANCE

SPECIFICATIONS:

ACCESS TO INSTALLATION SITE ACCESSIBLE BY PALLET JACK AND/OR FORK LIFT.

SHIPPING WEIGHT AND

DIMENSIONS (W x H x L)

CRATE A: 1720 lb [780 kg] 49 x 82 x 121 in [1245 x 2083 x 3074 mm]

CRATE B: 1710 lb [776 kg] 49 x 83 x 121 in [1245 x 2108 x 3074 mm]

CRATE C: 1205 lb [547 kg] 49 x 54 x 121 in [1245 x 1372 x 3074 mm]

CRATE D: 825 lb [374 kg] 39 x 54 x 97 in [991 x 1372 x 2464 mm]

CRATE E: 1330 lb [603 kg] 37 x 90 x 81 in [940 x 2286 x 2057 mm]

LARGEST UNCRATED PIECE:

WEIGHT AND DIMENSIONS

(W x H x L)

SUMP (CRATE A):

WEIGHT: 388 lb [176 kg]

65 x 7 x 113 in [1651 x 178 x 2870 mm]

HEAVIEST UNCRATED PIECE:

WEIGHT AND DIMENSIONS

(W x H x L)

MECHANICAL CORE (CRATE E):

MAXIMUM WEIGHT: 890 lb [404 kg]

26 x 79 x 80 in [660 x 2006 x 2032 mm]

WALLS AND

CEILING

VERIFY THAT WALLS AND CEILING MEET DIMENSIONAL AND POSITION REQUIREMENTS:

WALL OPENING DIMENSIONS

AND CEILING HEIGHT SEE PAGE 1,2 AND 3

SERVICE AREA CLEARANCE SEE PAGE 1, 2 AND 3

FIRE PREVENTION DEVICES

STERIS DOES NOT RECOMMEND INSTALLATION OF SPRINKLER OVER THE

UNIT BUT IF REQUIRED BY LOCAL CODE IT MUST BE FUSE TO 260 F (126 C)

OR ABOVE. ANY SPRINKLER HEADS IN FRONT OF THE LOAD/UNLOAD DOORS

MUST ALSO BE FUSE TO 260 F (126 C) OR ABOVE.

FLOOR

VERIFY THAT FLOOR AND PIT MEET DIMENSIONAL AND CONSTRUCTION REQUIREMENTS:

MAXIMUM OPERATING WEIGHT

(WATER AND TWO HEAVY CASE CARTS)

LOADED FREE-STANDING UNIT:

5960 lb [2703 kg]

LOADED THROUGH-WALLS UNIT:

5150 lb [2336 kg]

AVERAGE FLOOR LOAD

OVER WASHER FOOTPRINT 0.46 psi [3.23 kPa]

MAX. FLOOR LOAD

AT WASHER FOOT PADS

(1 ½ in [38 mm] DIAMETER PADS)

SUMP, NON-SERVICE SIDE (2): 490 psi [3378 kPa] SUMP, SERVICE SIDE (2): 1275 psi [8791 kPa]

MECHANICAL CORE (4): 280 psi [1930 kPa]

PIT DIMENSIONS

SEE PAGE 4 & 5 FOR STANDARD PIT MOUNTED UNIT

SEE PAGE 6 & 7 FOR PIT MOUNTED UNIT WITH SEISMIC

CORNER LEVELING PADS/

FOUNDATION PLATES POSITION

FLOOR DRAIN DIMENSIONS

AND POSITION

FLOOR CLEARANCE FOR INSTALLATION SEE PAGE 1-2

Page 16 of 16 920515107.D

Feuille 1
Vues
new_view_35
Zones
A1
Feuille 2
Vues
new_view_37
new_view_57
Zones
A1
Feuille 3
Vues
new_view_87
new_view_93
Zones
A1
Feuille 4
Vues
new_view_80
Zones
A1
Feuille 5
Vues
right_81
bottom_82
Zones
A1
Feuille 6
Vues
new_view_55
Zones
A1
Feuille 7
Vues
A
right_70
Zones
A1
Feuille 8
Vues
new_view_38
new_view_39
new_view_40
Zones
A1
Feuille 9
Vues
new_view_44
new_view_45
new_view_46
Zones
A1
Feuille 10
Vues
new_view_41
new_view_42
new_view_43
Zones
A1
Feuille 11
Vues
new_view_47
Zones
A1
Feuille 12
Zones
A1
Feuille 13
Zones
A1
Feuille 14
Zones
A1
Feuille 15
Zones
A1
Feuille 16
Zones
A1

File details come from the government source that posted it. Updated .