About this opportunity
The FDA's Center for Devices and Radiological Health seeks a 6-axis collaborative robotic arm system with controller, teach pendant, and software to support regulatory science tools development, with quotes due September 11, 2026, at 4:00 PM EDT and questions due September 10, 2026, at 12:00 PM EDT. This is an unrestricted, full and open competition with no set-aside. Award will be made on a Lowest Price Technically Acceptable (LPTA) basis as a firm fixed-price purchase order without discussions.
Place of performance and delivery is Silver Spring, Maryland. The robotic arm must have 6 rotating joints, 5–10kg payload, 750–950mm reach, less than 0.035mm repeatability, maximum joint speed of 180 degrees per second, and weigh less than 30kg with IP40 or higher ingress protection. The controller requires 120V 50–60Hz AC power (220V acceptable), minimum 8 digital input/output ports, 2 USB ports, 1 Ethernet port, and 1 display port. The teach pendant must feature an 8-inch or larger touchscreen, emergency stop button, and wireless or 2-meter cable connectivity weighing less than 3.5kg. Software must include TCP calibration capability with permanent licensing, compatibility with Windows 10/11 or Linux, and SDK support for third-party integration (LabVIEW, MATLAB, Python, C++, C#, Visual Basic). All equipment must be new and non-refurbished with ISO 9283 certification, encoder-based position detection, and a minimum 1-year warranty. Delivery is required within 60 days of award. The NAICS code is 333998 with a 700-employee small business size standard. Quotes must include all shipping, handling, and technical support costs. Optional items include a gripper, additional I/O expansion boards, and adapter plates. The offer acceptance period is 90 calendar days from the proposal due date.
Notice text
Notice of Combined Synopsis/Solicitation. This is a combined synopsis/solicitation for commercial products or commercial services prepared in accordance with part 12. This announcement constitutes the only solicitation; quotes are being requested, and a separate written solicitation will not be issued.
Background:
The procurement is to support regulatory science tools development for the Interventional, Radiological Imaging and Visualization regulatory science research program. Developing measurement methodologies for technical performance is critical for consistent evaluation of head mounted displays and image registration in interventional procedures. Specifically, the procurement for a robotic arm to precisely control the position and orientation of surgical tools to quantitatively characterize the spatial and temporal stability and accuracy of image registration. In addition, the robotic arm will also allow for controlled movement of an augmented and virtual reality head mounted displays (HMDs) or camera system to evaluate registration performance visualized through the HMD including simulated user movement.
These types of technologies are increasingly being incorporated in medical devices across OPEQ. The range of motion of the robotic arm will allow for developing image registration regulatory science tools for a wider range of device types with more complex motions than are achievable with a linear translation stage. Finally, the robot will allow for greater flexibility in targets and cameras for characterizing registration using HMDs. This will align the Office of Science and Engineering Laboratories (OSEL) instrumentation with FDA recognized standards, such as the Instrumental Commission (IEC), for head mounted displays, which is essential for the OSEL mission of developing regulatory science tools. The collaborative robotic arm will be incorporated into an existing bench setup using custom developed software for controlling a collaborative robotic arm.
Minimum Performance Requirements:
Item 1:
Collaborative Robotic Arm
• Degrees of Freedom: 6 rotating joints
• Payload: 5-10kg
• Reach: 750-950 mm
• Tool Movement at Tool Center Point (TCP)
TCP location: Should support both TCP locations
100-200mm from center of face-plate on tool z-axis (along rotation axis of joint 6)
100-200mm from center of face-plate on the tool x-axis (orthogonal to rotation axis of joint 6).
Range of TCP movement (position and orientation):
X: at least 150mm
Y: at least 150mm
Z: at least 150mm
Yaw angle: at least +/- 45o
Pitch angle: at least +/- 45o
Roll angle: at least +/- 20o
• Maximum speed per joint: at least 180o/s
• Maximum TCP velocity: at least 0.5 m/s
• Joint movable range: Minimum rotation angle of each joint of all six joints must be more than 220o.
• Position detection: encoders
• Brakes mechanism that locks and unlock joint, including in an emergency and by user request.
• Repeatability: less than 0.035 MM
• Weight: less than 30kg
• Protection and Sealing of the robot: Protection from tools and small wires, such as IP40 or higher.
• Mounting: Standard required. Optional for side or inverted mounting.
• Cable length: at least 2 m
• Footprint Size: less than 175 mm x 175 mm
Item 2:
Controller
• Power: 120V 50-60Hz AC preferred, 220V acceptable.
• Maximum Controller size (W x H x D): 500 mm x 500 mm x 500 mm
• I/O ports:
Digital input ports: min 8 ports
Digital output ports: min 8 ports
Analog input ports: min 2 ports
• USB ports: min 2 ports (2.0 or 3.0)
• Ethernet port: min 1 port
• Display port for external screen: min 1 port.
Item 3:
A control box for programming the motions of a robot. Also called teach box or teach pendant.
• Manually move and reorient the robotic arm
• Ability to program using GUI through the teach pendant.
• Touch screen of a size 8 in or larger
• Emergency stop switch/button.
• Connectivity: Wireless or by a cable length of at least 2 m
• Weight: less than 3.5 kg
Item 4:
Software
• Tool Center Point (TCP) calibration procedure
• Permanent software license
• Support for controlling the robotic arm and camera systems with third party software, such as LabVIEW, MATLAB, Visual Basic, C++, C#, or Python. Support may include a software development kit (SDK) with all required application programing interface (API) and libraries for software development, drivers, or ActiveX support.
• Compatible with Windows 10, Windows 11, or Linux Operating System
Deliverables: The Contractor shall deliver the item and quantities ordered within 60 days of award.
All deliverables required under this contract shall be packaged, marked, and shipped in accordance with Government specifications. At a minimum, all deliverables shall be marked with the contract number and contractor’s name. The Contractor shall guarantee that all required materials shall be delivered in immediate new, usable, and acceptable condition.
Place of Supply Delivery and Place of Performance:
The equipment should be complete with all accessories and software necessary to work and to be delivered to:
US FDA CDRH DIDSR
Attention: Beams, Ryan
10903 New Hampshire Avenue
WO62 RM 3126
Silver Spring, Maryland 20993
PH: 240-402-3216
Fax: 301-796-9925
Email: Ryan.Beams@fda.hhs.gov
Period of Performance:
12 consecutive months commencing upon date of Award.
Please see the attached Statement of Work and corresponding ancillary documents for review and consideration.
Quotes are due by email on or before September 11, 2025, by 4:00 P.M. (Eastern Standard Time) and should be sent to brandon.rafus@fda.hhs.gov.