Portable reverse osmosis for fargo ND

Closed Solicitation Posted

This opportunity was awarded. See the award notice from .

Solicitation number
36C26326Q0493
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Fargo, North Dakota 58102, United States

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs, Veterans Integrated Service Network 23, is soliciting quotations for a portable reverse osmosis (RO) water purification system intended for the Dialysis unit at the Fargo VA Medical Center. The solicitation (36C26326Q0493) is issued as a Request for Quotation (RFQ) under FAR 12 Commercial Procedures for the Evoqua W3T591409 Kit, RO Eon Rover ALX SCB system or an equivalent brand-name product that meets specified performance standards. Key contract requirements include FDA approval as a 510(K) medical device, certification to UL and IEC standards, compliance with AAMI-Grade water standards per ANSI/AAMI 13959, a fully integrated endotoxin removal filter capable of removing contaminants greater than 0.005 micron, multi-stage filtration systems, heat-based disinfection, and multiple leak detection options. Evaluation will be based on offerors' ability to demonstrate capability and understanding of the Statement of Work requirements through a technical narrative and firm fixed pricing that covers all labor, overhead, profit, materials, and costs. Questions are due by 10:00 AM CST on March 31, 2026, with quotations due by 10:00 AM CST on April 3, 2026, with award anticipated on April 3, 2026.

This is an unrestricted procurement with no set-aside designation, and the government currently maintains a preventative maintenance contract with Evoqua for dialysis equipment units. The solicitation requires two Evoqua W3T591409 RO Eon Rover ALX SCB kits and two Evoqua W3T578738 Booster kits, along with orientation and labor services, to be delivered as new, OEM-supplied equipment only. Specific award value or budget range information is not disclosed in the solicitation documents. The contractor must complete delivery to the Fargo Health Care System Warehouse at 2101 N Elm Street, Fargo, North Dakota 58102, within 60 days after receipt of order during weekday business hours (7:30 a.m. to 3:30 p.m.) on an F.O.B. destination basis. Offerors must submit a single proposal document not exceeding 30 pages, including a signed SF 1449, technical narrative, firm fixed pricing, and a completed Buy American Act Certificate certifying domestic content of products being offered.

Notice text

COMBINED SYNOPSIS SOLICITATION

VA RFQ 36C26326Q0493

Portable Reverse Osmosis-Brand Name or Equal, Unrestricted

This is a combined synopsis/solicitation for commercial products or commercial services prepared in accordance with the format in Revolutionary FAR Overhaul, FAR Subpart 12, as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; quotes are being requested and a separate solicitation will not be issued.

SOLICITATION NUMBER: 36C26326Q0493

This solicitation is issued as a Request for Quotation (RFQ).

The solicitation document and incorporated provisions and clauses are those in effect through Federal Acquisition Circular FAC 2025-06 10/01/2025, and Revolutionary FAR Overhaul FAR 12 as of 02/25/2026.

This solicitation is Brand Name or Equal, Unrestricted.

The associated NAICS code and small business size standard:

                NAICS code:        339112

Size Standard:    1000 employees

Description of Requirements:

Salient Characteristics

Brand Name or Equal:  Evoqua W3T591409 Kit, RO Eon Rover ALX SCB

       Fully integrated endotoxin removal filter Ultrafilter assists in preparing water or bicarbonate concentrate to
       meet  the requirements for the production of ultrapure dialysate according to the ANSI/AAMI 11663 standard.         The filter’s unique proprietary single-stage design removes bacteria, viruses, endotoxins and other
contaminants greater than 0.005 micron (5 nanometers)

Multi-Stage Filtration System, including carbon filtration, reverse osmosis (RO), and ultra-filtration in one unit

Features easy-to-use controls and instant access to key parameters

Prevents RO shutdown when the inlet water has low PSI

Uses heat to disinfect rather than chemicals

       Continuous flow of hot water with no time constraints

Provides single heat, storage heat

Multiple Leak Detector options, internal or remote sensor with user alert or shutdown

Color touch screen LCD control panel, clear display of operational functions

Recirculation product line

Filter and membrane access – removable back panel, machine disassembly not required

Certified to UL 61010-1, IEC 61010-1, CAN/CSA C22.2 No: 61010-1, IEC 61326-1

FDA approved 510(K) Medical Device

Must be designed to meet AAMI-Grade water standards according to ANSI/AAMI 13959 standards

Leak Detection System

Maximum Product Flow Rate at 77°F       Up to 0.4 gpm (1.5 lpm)

Inlet Temperature                                            40-104°F (4-40°C)

Inlet Pressure (Dynamic)                       Minimum: 5 psi (0.34 bars)          Maximum: 120 psi (8.3 bars)

Inlet Water Flow/Water Feed                      Minimum: 1.5 gpm (5.7 lpm)

Drain Capacity                                                   Minimum 2.0 gal/min

Total Hardness   < 10 grains (pH dependent)

Silt Density Index (SDI)   < 5

Total Chlorine     < 0.1 ppm

CART:

Rust-proof and corrosion resistant. Aluminum frame, including built-in rubber bumper guard

Large 6” dual swivel casters with locking features

       Rear directional locking casters with integrated braking mechanism

Centrally located monitoring area that features pressure and temperature gauges, and sample ports with drip tray

       Includes booster pump for low-pressure inlet situations

Open design to allow for easy access to filter and other areas of maintenance

Complies with ETL Tip and Threshold Testing Standards, and meets and adheres to the latest ANSI, AAMI, ISO Standards

provides hose and cord management options

All-in-one water treatment delivery system

Brand Name or Equal:  Evoqua W3T578738 Kit, Booster, MHX/WRO, Rover Family

Includes booster pump for low-pressure inlet situations

        System Design

      Small footprint for ease of use in vicinity of patients without interrupting patient care due to limited space in      patient area

       Includes installation and start up

       System orientation provided

INSPECTION AND ACCEPTANCE

-The contractor shall conduct a joint inspection with the COR upon delivery

equipment.

-The contractor shall provide the VA COR a list of missing and /or damaged items

found upon inspection and have those items replaced at no cost to the Government.

-The COR shall ensure all work is completed satisfactorily prior to acceptance.

Disputes shall be resolved by the VA Contracting Officer

This service does not involve storage, generating, transmitting, or exchanging VA sensitive information.

“Privacy training is required if the contractor(s) will be working in areas with sensitive Veteran information with an increased risk of inadvertent disclosures.  These are areas where only authorized personnel are allowed (nurse stations, PACT team areas, Laboratory, Pharmacy, etc.) and where visitors are typically be escorted and are not allowed to roam around freely (4B, ER, 3M, ICU, CLC, etc.).  The contractor(s) must read and sign TMS 20939, VA Privacy Training for Personnel without Access to VA Computer Systems or Direct to or the Use of VA Sensitive Information, prior to beginning work in these areas.

Delivery.

             Fargo Health Care System Warehouse

                2101 N Elm St.

                Fargo, ND 58102

                Delivery Times:  M-F 7:30 am to 3:30 pm

  1. To see the solicitation PROVISIONS, CLAUSES, SCHEDULE and all other terms that apply to this acquisition, see the attached RFQ 1449 36C26326Q0493.

10.  SUBMITTAL OF OFFERS:

All offers shall be received by the CO no later than 10:00 AM CT Friday April 3, 2026.

All offers shall be emailed to Becky.delossantos@va.gov

Emails shall include this RFQ number in the Subject Line: 36C26326Q0493     

               

INSTRUCTIONS TO OFFERORS:

All Instructions to Offerors added by the CO in addition to the standard 52.212-1 INSTRUCTIONS TO OFFERORS are here on this page.  Read this entire page.

This solicitation is issued under Revolutionary FAR Overhaul, FAR 12 Simplified Acquisitions.

The Government intends to award without Discussions.

Partial Offers are not being accepted.

The Contract will be Single Award, Firm Fixed Price, Purchase Order.

Offerors are required to be registered in SAM when submitting a quote and at time of award.

Late offers will be treated according to 52.212-1 Instructions to Offerors.

The Government intends to evaluate all line items on a group basis and make a single award for BAA/TAA applicability.

OFFERS MUST INCLUDE:

52.225-2 Buy American Certificate is attached.  Offeror shall indicate the Country of Origin for each line item and fill in and sign the Certificate.  See Yellow Highlights on Certificate.

SUBMISSION OF OFFERS:

All offers shall be received by the CO no later than 10:00 AM CT Friday April 3, 2026.

All offers shall be emailed to Becky.Delossantos@va.gov.

Emails shall include this RFQ number in the Subject Line: 36C26326Q0493.

DELIVERY:

Delivery is FOB Destination.

Delivery: 60 Days ARO

Files

File Type Posted
52.225-2 Buy American Act Certificate-COTS.pdf PDF
SPECIFICATIONS 36C26326Q0493.docx DOCX document
36C26326Q0493.pdf PDF

Notice history

Notice Type Posted
Portable reverse osmosis fargo Award Award Notice
This notice Latest solicitation Solicitation

Similar open opportunities

Opportunity Agency Set-aside Responses due
J046--Reverse Osmosis PM NAICS 811210 Veterans Integrated Service Network 21 Total Small Business
J065--Reverse Osmosis Systems Maintenance Services Ann Arbor VA Medical Center 2215 Fuller Road, United States · NAICS 541380 Veterans Integrated Service Network 10 No set-aside
F109--Water Management Program and Chemical Testing VAMC Boston Healthcare System Department of Veterans Affairs Vamc Boston Healthcare System Boston MA 02130, United States · NAICS 561210 Veterans Integrated Service Network 1 Service-Disabled Veteran-Owned Small Business
Z1DA--CON-NRM-668-22-106 Correct Domestic Water System Spokane, Washington 99205, United States · NAICS 238220 Veterans Integrated Service Network 20 Service-Disabled Veteran-Owned Small Business
Demineralizer, water, IO NAICS 333310 Construction and Equipment Total Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity