Notice of Intent to Sole Source USAFA Autoclave System and Reverse Osmosis System
Closed Special Notice Posted
- Solicitation number
- USAFAAutoclaveSystemandReverseOsmosisSystem
- Agency
-
Air Force Academy
Headquarters Air Force, Department of Defense
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 339113 Surgical Appliance and Supplies Manufacturing
- PSC
- 6640 Laboratory Equipment And Supplies
- Place of performance
- Usaf Academy, Colorado 80840, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The United States Air Force Academy intends to award a sole-source, Firm-Fixed-Price contract to Getinge USA Sales, LLC for lifecycle replacement of one Autoclave System and one Reverse Osmosis System, with a response deadline of August 7, 2026. The Dean of Faculty Department of Biology requires these systems to serve over 800 cadets annually for sterilizing materials and processing biohazardous waste. The Autoclave System must feature a chamber volume of at least 275 liters, capability for gravity, vacuum, and liquid cycles across 104–135°C temperature ranges, freestanding configuration, integrated electric steam generator with 480V AC three-phase input, touchscreen control interface with PACS 3500 programmable logic controller, and safety interlocks and steam safety valves. The Reverse Osmosis System must be wall-mounted with associated bladder tank, 115V single-phase input, 400 gallons-per-day production capacity, three pre-filters and single membrane configuration, manual bypass, and polyphosphate feeder cartridge. The procurement includes deinstallation, removal, and haul-away of existing equipment. The contract will be awarded under FAR Part 12 simplified procedures for commercial products under NAICS 339113.
Notice text
The United States Air Force Academy (USAFA) intends to award a sole-source, Firm-Fixed-Price contract to Getinge USA Sales, LLC (Getinge) under the simplified procedures for commercial products in accordance with RFO Part 12. The proposed North American Industry Classification System (NAICS) Code is 339113 (Surgical Appliance and Supplies Manufacturing).
Requirement:
The USAFA Dean of Faculty Department of Biology (DFBI) requires the lifecycle replacement of one Autoclave System and one Reverse Osmosis (RO) System for use by over 800 Cadets per year for critical daily operations, including sterilizing materials for classes and research and processing biohazardous waste.
The systems must meet the following salient characteristics:
Steam Sterilizer System (Autoclave)
- Quantity: One (1) new system
- Chamber volume: At least 275L
- Gravity Cycles: Capable of 110C – 135C temperatures
- Vacuum Cycles: Capable of 110C – 135C temperatures
- Liquid Cycles: Capable of 104C – 135C temperatures
- Exterior Dimensions (W x H x D): Close to 30” x 74” x 48”
- Configuration: Freestanding unit
- Utilities: Integrated electric steam generator with 480V AC, 60Hz, 3-phase electrical requirements; sterilizer control with 115V AC, 60Hz, 1-phase electrical requirements
- Control Interface: Touchscreen interface and a PACS 3500 Programmable Logic Controller with preprogrammed cycles
- Safety Features: Chamber safety interlock and steam safety valves to protect against over pressurization
Reverse Osmosis (RO) Water Treatment System
- Quantity: One (1) new system
- Configuration: Wall-mounted system with associated bladder tank
- Utilities: 115V, 60Hz, 1-phase electrical requirements
- Capacity: Capability to produce up to 400 gallons per day of RO water
- Rough Dimensions (L x W x D): Close to 38” x 26” x 7”
- Filter Elements: 3 pre-filters and single membrane
- Bypass: Manual bypass capability
- Feeder: Polyphosphate feeder cartridge
Removal and Installation Services
- Deinstallation, removal, and complete haul-away of the Department's existing Autoclave and RO systems.
- Turnkey installation, final utility connections, calibration, and startup testing of the new system.
Justification:
In accordance with RFO 12.102(a), the Government has determined that only one source is reasonably available to fulfill this requirement.
Market research and responses to a prior Request for Information (RFI) confirmed that alternative manufacturer models that meet the minimum 275L chamber volume are physically too large to fit into the laboratory room’s highly restricted, pre-built spatial footprint. Getinge is the only known manufacturer whose 275L unit complies with the strict physical room dimensions.
Additionally, DFBI currently has an active preventative maintenance and service contract with Getinge for other departmental laboratory autoclaves. Standardizing on Getinge equipment allows the Government to integrate this new system under the existing active maintenance contract immediately. This standardizes training for laboratory technicians and biological safety personnel, ensures immediate compatibility of replacement parts, and reduces administrative and logistical overhead by avoiding the management of multiple service contracts.
To prevent unacceptable supply chain, installation, and biological safety certification risks, the Government requires this procurement to be executed directly with Getinge USA Sales, LLC to guarantee OEM-certified utility connections, check-out, and warranty integration.
Disclaimer:
This notice of intent is not a request for competitive quotations, and a formal solicitation will not be issued. However, any responsible source that believes it can meet these requirements may submit a capability statement. Information received will be considered solely for the purpose of determining whether to conduct a competitive procurement. A determination not to compete this action is solely within the discretion of the Government.
Responses:
Submit capability statements electronically to Nicole Cibula at nicole.cibula@us.af.mil and Amy Krause at amy.krause.1@us.af.mil by 11:00 AM (MDT) on 7 August 2026.
Notice history
| Notice | Type | Posted |
|---|---|---|
| This notice | Special Notice | |
| USAFA Autoclave System and Reverse Osmosis System RFI | Pre-Solicitation |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Getinge Steam Sterilizer Repair | Naval Supply Systems Command | No set-aside | |
| 6515--Sterilizer | Veterans Integrated Service Network 17 | No set-aside | |
| Notice of Award Maintenance for Government owned BacT/ALERT 3D Equipment. | Medical Command | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity