Neptune 3 Waste Management System
Cancelled Solicitation Posted
- Solicitation number
- 36C24W25Q0274
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 339112 Surgical and Medical Instrument Manufacturing
- Place of performance
- Oklahoma City, Oklahoma 73105, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The Department of Veterans Affairs Oklahoma City Veterans Affairs Health Care System is seeking to procure one Stryker Neptune 3 Rover waste management system for their new Community Based Outpatient Clinic to support general surgery and orthopedic examinations and procedures. This procurement follows a "Brand Name or Equal" approach, requiring equipment that meets specific technical specifications including multiple suction range indicators, integrated smoke evacuator, mobile unit capabilities, and a height-adjustable powered IV pole with precise operational parameters. The contractor must provide a comprehensive solution including all necessary components for functionality such as installation, any required VA staff training, equipment mounts, cables, and coordination with the VA's Biomedical Engineering Department for maintenance. The solicitation is classified under NAICS code 339112 (Surgical and Medical Instrument Manufacturing) and PSC 6515 (Medical And Surgical Instruments, Equipment, And Supplies). Responses to this solicitation are due by July 16, 2025, at 12:00 PM Central Time and must be submitted via email to the designated contracting officer.
This procurement does not utilize any set-aside designations and is open to all qualified vendors capable of meeting the technical requirements. No incumbent contractor or current service provider is identified in the solicitation documentation. While no specific contract value or budget range is disclosed, the procurement involves a single unit of specialized medical equipment with detailed specifications including 120V electrical requirements, 24-liter waste capacity, and operational parameters within specific environmental conditions. The equipment will be delivered to the Oklahoma City Veterans Affairs Health Care System at 50 N.E. 42nd Street, Oklahoma City, Oklahoma 73105, with delivery restricted to Monday through Friday between 8:00 AM and 4:00 PM. The contractor will have no access to patient records, and comprehensive VA staff training is not required as part of this procurement.
Notice text
STATEMENT OF WORK
Stryker Neptune 3 Rover
Veterans Affairs Health Care System
Oklahoma City, Oklahoma
Brand Name or Equal
- General: Currently, the Oklahoma City Veterans Affairs Health Care System (herein after referred to as OKC VAHCS) is looking to purchase one (1) Stryker Neptune 3 Rover for our new Community Based Outpatient Clinic (CBOC) for General Surgery and Orthopedic examinations and procedures.
- Delivery Location: Oklahoma City Veterans Affairs Health Care System, 50 N.E. 42nd St. Oklahoma City, OK 73105
- Delivery Times:Delivery times shall be Monday through Friday from 08:00 to 16:00 (8:00AM to 4:00PM).
- Definitions/Acronyms
-
- OKC VAHCS: Oklahoma City Veterans Affairs Health Care System
- Scope:
The Contractor will provide the Brand Name or Equal to Stryker Instruments necessary equipment and accompanying software, if applicable, to the OKC VAHCS for one (1) Stryker Neptune 3 Rover. The requirement is for a comprehensive and whole solution to the requirement. The Contractor is required to be an expert in the field of examination chairs and equipment and requirements listed in this Statement of Work (SOW). As such the Contractor is required to include any items not specifically listed in the pricing schedule but is necessary to enable the equipment or accompanying software to function. Examples include installation expenses, VA staff training, equipment mounts, display charts, cables, connectors, plugs, and special equipment wiring requirements, if applicable. The company shall coordinate with OKC VAHCS Biomedical Engineering Department to coordinate any preventive maintenance or replacement of any machine.
- Training for VA Employees: Comprehensive training for VA staff is not required.
- Specifications:
Below are the salient characteristics for the Brand Name or Equal requirement:
Manufacturer: Stryker Instruments
Model: Neptune 3 Waste Management System
Item #
Description/Part Number*
Qty
0703-001-000
Neptune 3 Rover (120V)
1
Brand Name or Equal
-
-
- Multiple suction range indicators with high suction audio alerts
- Integrated smoke evacuator
- Technology that locks away suctioned biohazards
- Mobile unit
- Height-adjustable, powered IV pole
- Two manifold receptacles
- Specifications
- Electrical power requirements:
- 120 V, 60 Hertz, 12.0 Amps, single phase
- Dimensions
- Width: 48.3 cm (19 inch)
- Height: 259 cm (102 inch) with powered IV pole at maximum height; 177.8 cm (70 inch) with powered IV pole at minimum height
- Depth: 58.4 cm (23 inch)
- Mode of operation: continuous
- Sound pressure: 57-62 dB
- Adjustable suction limit: 50 to 520 mm-Hg; measured with all ports closed
- Vacuum measurement accuracy: ± 5% of full scale
- Suction limit accuracy: ≤ 26 mm-Hg or 10% of setting
- Mass: 131 kg (288 lb) – collection of canisters empty (prefill only); 155 kg (341 lb) – collection canisters full
- Volume: 24-liter capacity (combination of 4-liter and 20-liter canisters)
- Volume measurement accuracy: 4-liter canister, ± 50 mL; 20-liter canister, ± 150 mL
- IV pole capacity: 12000 mL or 3000 mL per IV pole hook
- Inclined plane of operation: ± 2.5 degrees
- Environmental conditions:
- Temperature limitation:
- Operation: 14º C to 26º C
- Storage and transportation (before initial use): -20º C to 40º C
- Storage and transportation (after initial use): 10º C to 40º C
- Humidity limitation:
- Operation: 20% to 50%
- Storage and transportation (before initial use): 10% to 75%
- Storage and transportation (after initial use): 10% to 75%
- Atmospheric pressure limitation:
- Operation: 70 kPa to 106 kPa
- Storage and transportation (before initial use): 50 kPa to 106 kPa
- Storage and transportation (after initial use): 50 kPa to 106 kPa
- Temperature limitation:
- Power cord general specifications:
- Current rating: 15 A
- Voltage rating: 125 VAC
- Frequency rating: 60 Hz
- Conductor size: 14 AWG
- Connector type: IEC 60320 C13
- Cord lengths: 5.5 m
- Temperature rating: 105-degrees C minimum
- Electrical power requirements:
-
- Security Requirements
The contractor will have no access to patient records. The A&A requirements do not apply, and a Security Accreditation Package is not required.
Files
| File | Type | Posted |
|---|---|---|
| S02 36C24W25Q0274 Sources Sought Opportunities.pdf |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| 6515--NX EQ Image Guided Surgical Navigation System (VA-26-00083972) | VA Strategic Acquisition Center Frederick | Service-Disabled Veteran-Owned Small Business | |
| NX EQ surgical tissue management system | VA Strategic Acquisition Center Frederick | No set-aside | |
| Surgical Operational Efficiency Specialist Services | Veterans Integrated Service Network 16 | Service-Disabled Veteran-Owned Small Business | |
| 6540--NX EQ Scanning Systems Ultrasonic Ophthalmic AB Scan (VA-27-00003110) | VA Strategic Acquisition Center Frederick | Service-Disabled Veteran-Owned Small Business | |
| 6640--Perfusion Cell Saver Lease and Service | Veterans Integrated Service Network 21 | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity