J091--Bldg. 129 Medical Gas Delivery System Deficiencies Repair(s) from 2024 Annual Inspection

Closed Special Notice Posted

Solicitation number
36C24725Q0375
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Department of Veterans Affairs Central Alabama Healthcare Systems Tuskegee AL 36083

Notice details come from SAM.gov. Updated .

About this opportunity

The Department of Veterans Affairs (VA), specifically the Central Alabama Healthcare Systems (CAVHCS) Tuskegee facility, has issued a Special Notice announcing its intent to sole source a procurement for medical gas delivery system repairs at Building 129. The notice pertains to a one-time service contract for addressing deficiencies identified during the 2024 Annual Inspection of the medical gas infrastructure. The VA intends to award a firm-fixed-price contract to Environmental & Medical Gas Services (EMGS), a small business located in Villa Rica, GA, under FAR Part 13. The procurement will utilize NAICS code 622110 for General Medical and Surgical Hospitals, with a small business size standard of $47.0 million. The planned contract will cover specific repair and maintenance tasks for the medical gas system at the Tuskegee VA Health Care System.

The primary objective of this Special Notice is to repair critical medical gas delivery system deficiencies, including replacing a dew point monitor, two desiccant medical air dryers, repairing a high temperature alarm, and installing proper piping labels. Contractors interested in potentially competing for this opportunity must submit capability statements by February 14, 2025, at 10:00 AM CST to Contracting Officer Terri Hudson via email. While the VA intends to sole source the contract to EMGS, they will consider submitted capability statements to determine whether to maintain the sole source approach or open the procurement to competition. Eligible contractors must be registered in the System for Award Management (SAM) and provide clear documentation demonstrating their technical capabilities to satisfy the specific repair requirements. The VA emphasizes that this notice is not a formal request for competitive quotes, and any costs incurred by interested parties in responding will not be reimbursed. The estimated contract award date is February 18, 2025, with the work to be performed at the Tuskegee VA facility.

Notice text

3 versions

Update #3 · Latest ·

STATEMENT OF WORK

2024 Medical Gas Delivery System Deficiencies Repair(s)

OBJECTIVE:

The objective of this agreement is to establish a contract inclusive of all personnel, management, materials, parts, equipment, tools, supplies, labor, and transportation require to perform recommended repairs on the Medical Gas Delivery System at the Central Alabama Veterans Health Care Systems (CAVHCS).

SITE VISIT:

Bidders shall visit the site to confirm locations, dimensions, associated equipment, and equipment conditions relative to the recommended repair(s) at Tuskegee VA Bldg. 129 Medical Gas Delivery System Deficiencies findings doing the 2024 inspection. The site visit shall be for the purpose of observing/verifying any conditions, concerns, dimensions, etc. relevant to the fulfillment of this contract. No consideration will be given to claims based on a lack of knowledge of the existing conditions, except where contract documents make definite provisions for adjustment in cost or extension due to existing conditions not readily foreseeable. The site addresses are:

V.A. Medical Center - East

2400 Hospital Rd.

Tuskegee, AL 36083

Bldg.129

CONTRACTOR RESPONSIBILTY:

Contractor shall provide all personnel, management, materials, parts, equipment, tools, supplies, labor, and transportation require to perform the recommended repairs findings on the 2024 Annual Inspection on Medical Gas Delivery System in Bldg. 129 Tuskegee. See Annual Medical Gas Testing and PM reports for recommended repair(s) details.

Issue #25086 Replace (1) dew point monitor.

Issue #25086 Replace (2) desiccant medical air dryer(s) w/integrated controller (includes all filters needed)

Issue #25085 Repair medical air high temperature alarm (not working at Master Alarm Panels)

Issue #25102 Install proper label(s) on Med-Air Intake and Med-Vacuum Exhaust piping.

Contractor shall test all alarms, equipment, and associated equipment for normal and proper operating once work has been completed.

Contractor shall ensure area is clean of all debris material before departing site.

BLDG. 129 MEDICAL GAS SYSTEM EQUIPMENT LIST

Bulk Oxygen Tank w/Reserve Tank

Praxair

Medical Air System

Duplex Compressor(s) Amico_A-SCD-P-080P-SS-N-075-46063-T

Duplex Dryer(s) Trident Pneumatics_Dryspell

Dew Point Monitor Amico

CO Monitor GFG_ABL-4021

Medical Vacuum System

Duplex Pump(s) Patton s_72-21-033C

Mater Alarm Panels

Amico Alert-2 Master

Areal Alarm Panels

Amico Alert-3 LCD

CONTRACTOR QUALIFICATIONS:

Contractor shall preform all work in accordance with VHA Directive 7515(1), the contractor shall be certified in American Society of Sanitary Engineering (ASSE) 6010 for system installations, ASSE 6020 for system inspections, ASSE 6030 for systems verifier, and ASSE 6040 systems maintenance whichever is applicable to the work contracted to perform.

Contractor shall be licensed and certified to provide the testing and inspection services specified in this contract.

Contractor shall provide service tickets, at the time services are rendered, for all services performed. Service tickets shall be signed by the COR in order to confirm services were rendered.

Contractor shall hold a current and valid ASSE Medical Gas Verifier 6030 certification, ensuring all qualifications meet the requirements as stated in the latest version of NFPA 99.

Any personnel representing the contractor and performing the actual inspections shall also be qualified to at least the ASSE Medical Gas Verifier 6030 certification level.

The Contractor shall provide copies of the certifications upon acceptance of any contract award, or no work shall be performed, nor accepted, until proof of licensing is verified.

Contractor shall use testing equipment that is calibrated and adheres to the standards and criteria set forth by the National Institute of Standards and Technology (NIST), as recommended by the ISO-9001 requirements. As proof of current calibration, the Contractor shall provide a Traceable Calibration Certificate to the COR or their designee, presented upon initial contract issuance and each time certificates are updated. Evidence of all required licenses, as well as documentation of the qualifications of personnel, shall be provided to the CO and COR prior to award of the contract.

Contractor personnel, while on VA property, shall adhere to all requirements and regulations that govern the VA Health Care System and its property. Among all other requirements, the VA Health Care System and its property is now entirely smoke-free. No weapons are permitted on the premises. Medical masking in all required areas.

GENERAL REQUIREMENTS:

All work shall be conducted in accordance with National Fire Protection Association (NFPA 99), Compress Gas Association (CGA), Occupational Safety and Health Administration (OSHA), Joint Commission on Accreditation of Healthcare Organizations (JCAHO), ASSE 6000: Professional Qualifications Standard, VA standards, and latest edition or other applicable current recommendations.

The Individuals working/testing/inspecting on Medical Gas Systems must show proof of the following certification. ASSE 6010, ASSE 6020, ASSE 6030, ASSE 6040

All certifications must be up to date. Any Contractors/Vendor not showing proof of Certifications shall be disqualified.

Contractor shall be registered in SAM.GOV with an active Dunn & Bradstreet number.

All shutdowns shall be coordinated with COR. Shutdown might occur at hours outside of traditional 1st shift, on Holidays, or on weekends. Include such implications in quote.

Contractor shall clean-up work sites DAILY. NO debris can be left at work sites.

Contractor shall visit the work sites and verify all existing conditions and include all such implications in the quote.

The work shall include notification to the department point of contact immediately if life threatening condition exist during the inspections, testing and repairs.

INTERRUPTION OF MEDICAL GAS SERVICES

The interruption of medical gas services (i.e., closure of valves, raising and lowering pressures etc.) shall be prohibited without the approval of the contracts point of contact. Interruption of medical gas services shall be considered a medical gas shutdown, and the Facility shutdown procedure shall be followed. Interruption of medical gas service shall be requested using the following procedure.

Notification of the contracts point of contact not less than (3) days in advance of the proposed interruption of medical gas services.

Do not proceed with interruption of medical gas service without the contracts written permission from the Engineering department.

HOURS OF WORK:

The hours of work shall take place during the normal business hours of 8:00 AM and 4:30 PM, Monday through Friday, Federal Holidays excluded. Request for weekend work must be submitted in writing. Inspections will require a prior five-day notification submitted to the Contract Officers Representative (COR) or government representative.

Federal Holidays and National Holidays observed by the VA Hospital are:

New Years Day Labor Day

Martin Luther King Day Columbus Day

President s Day Veteran s Day

Memorial Day Thanksgiving Day

Juneteenth Christmas Day

Independence Day

PERIOD OF PERFORMANCE:

The Period of Performance is three months ARO.

DOWNTIME:

All downtime shall be scheduled 72 hours in advance to minimize impact on Hospital thru COR

NOTES:

All modifications shall be reviewed and approved prior to execution.

Contractor shall specify cost of modifications IAW GSA pricing schedule.

Testing should be during normal hours 8:00am 4:30pm

WARRANTY:

Contractor shall provide warranty on the integrity of services, parts and equipment installed supporting the CAVHCS Medical Gas Delivery System. The warranty period shall be for one (1) year after completion of work.

Copy of Annual Medical Gas Testing and PM Report

RECORDS MANAGEMENT AND CONTRACT SECURITY

A. Contractor shall comply with VA Handbook 6300.1 Records Management Procedures and VA Handbook 6500.6 Contract Security. Both can be found at:

https://www.va.gov/vapubs/search_action.cfm?dType=2

**Note: This link is the complete list of VA Handbooks and are in sequential order.

**Instructions for accessing VA Handbooks at link listed above:

Click on link above and once it is opened, please scroll down on the left side until you get to VA Handbook 6300.1.

Once there, go to the PDF Format Column and click on PDF.

VA Handbook 6300.1 will then ask you to open or save the document.

B. Please follow the same instructions listed above for accessing VA Handbook 6500.6.

All contractor employees working on this requirement are required to complete mandatory annual training entitled, VA Privacy and Information Security Awareness and HIPPA Training prior to providing service for each year of the contract with copies sent to the COR for each and available for audit as requested. The course numbers are VA10203 and VA 10176. You must select that you are a contractor. Contractor employees are to self-enroll into TMS via the link: https://www.tms.va.gov/learning/user/SelfRegistrationUserSelection.do

A VA employee will escort the contractor when contractor is required to enter/work in sensitive areas.

IDENTIFICATION, SMOKING AND VA REGULATIONS

The Contractor's FSE shall wear visible identification always while on the premises of the Augusta VA Medical Center. Contractor shall contact the Contracting Officer s Representative (COR) upon contract award, to receive instructions on how to obtain a VA Identification Badge during the performance of this contract. Augusta VA Medical Center is a smoke free campus which includes both tobacco and e-cigarettes; tickets can be issued by Police Services. Possession of weapons is prohibited. Enclosed containers, including tool kits, shall be subject to search. Violations of VA regulations may result in citation answerable in the United States (Federal) District Court, not a local district, state, or municipal court. Photography or Video on medical center premises is strictly prohibited. Additionally, pursuant to 38 CFR 1.218(b) (23) Use of recording devices of any kind to record patients or employees is strictly prohibited.

Update #2 ·

***INTENT TO SOLE SOURCE NOTICE-*** Estimated Date of Award: February 18, 2025 Synopsis: The Department of Veterans Affairs, Central Alabama Healthcare Systems (CAVHCS) Tuskegee intends to award a sole source procurement for the following line items. This requirement is for a one-time service to complete virtual training via Environmental & Medical Gas Services (EMGS) for Engineering Services for the Tuskegee VA Health Care System (CVAHCS), 2400 Hospital Road Tuskegee, AL 36083. LINE ITEM 0001: CONTRACTOR RESPONSIBILTY: (1) Contractor shall provide all personnel, management, materials, parts, equipment, tools, supplies, labor, and transportation require to perform the recommended repairs findings on the 2024 Annual Inspection on Medical Gas Delivery System in Bldg. 129 Tuskegee. See Annual Medical Gas Testing and PM reports for recommended repairs details. 1- Issue #25086 - Replace (1) dew point monitor. 2- Issue #25086 - Replace (2) desiccant medical air dryer(s) w/integrated controller (includes all filters needed) 3- Issue #25085 - Repair medical air high temperature alarm (not working at Master Alarm Panels) 4- Issue #25102 - Install proper label(s) on Med-Air Intake and Med-Vacuum Exhaust piping. (2) Contractor shall test all alarms, equipment, and associated equipment for normal and proper operating once work has been completed. (3) Contractor shall ensure area is clean of all debris material before departing site. LINE ITEM 0002: Shipping & Handling The Department of Veterans Affairs, Network Contracting Office 7 located in Tuscaloosa, AL intends to issue a sole source firm fixed priced contract to EMGS, a small business, located in Villa Rica, GA 30180. The contract is expected to be awarded under FAR Part 13. A Limited Sources Justification has been prepared. This procurement is being conducted in accordance with FAR 13.106-1(b) (1)(ii), Soliciting from a Single Source. The NAICS code utilized for this procurement is 622110 General Medical and Surgical Hospitals with a Small Business Size Standard of $47.0 Million. THIS NOTICE OF INTENT IS NOT A REQUEST FOR COMPETITIVE QUOTES. However, Contractors may submit a capability statement, proposal, or quotation which shall be considered by the contracting officer whether to continue with the sole source procurement or to provide for competition. Such responses MUST be received no later than February 14, 2022 at 10AM CST. Forward responses by e-mail to Contracting Officer, Terri Hudson. Responses received will be evaluated; however, a determination by the Government not to compete the proposed procurement based upon responses received to this notice is solely within the discretion of the Government. The Department of Veterans Affairs, NCO 7 will NOT be responsible for any costs incurred by interested parties in responding to this notice of intent. Only written responses will be considered. All responses from responsible sources will be fully considered. As a result of analyzing responses to this notice of intent, the Government shall determine if a solicitation will be issued. Any prospective contractor must be registered in the System for Award Management (SAM) to be eligible for award. Interested parties are encouraged to furnish information by email only with Response to Intent to Sole Source for Medical Gas Delivery System Repair in the subject line. All documents submitted will not be returned. All interested parties shall submit clear and convincing documentation demonstrating their capabilities to satisfy the requirements listed to Contracting Officer, Terri Hudson by email at terri.hudson@VA.Gov. The capability documentation should include general information and technical background describing in detail the prospective contractor's capabilities and experience that will demonstrate the ability to support this requirement.

Update #1 ·

***INTENT TO SOLE SOURCE NOTICE-***

Estimated Date of Award: February 18, 2025

Synopsis:

The Department of Veterans Affairs, Central Alabama Healthcare Systems (CAVHCS) Tuskegee intends to award a sole source procurement for the following line items.

This requirement is for a one-time service to complete virtual training via Environmental & Medical Gas Services (EMGS) for Engineering Services for the Tuskegee VA Health Care System (CVAHCS), 2400 Hospital Road Tuskegee, AL 36083.

LINE ITEM 0001: CONTRACTOR RESPONSIBILTY: (1) Contractor shall provide all personnel, management, materials, parts, equipment, tools, supplies, labor, and transportation require to perform the recommended repairs findings on the 2024 Annual Inspection on Medical Gas Delivery System in Bldg. 129 Tuskegee. See Annual Medical Gas Testing and PM reports for recommended repairs details. 1- Issue #25086 - Replace (1) dew point monitor. 2- Issue #25086 - Replace (2) desiccant medical air dryer(s) w/integrated controller (includes all filters needed) 3- Issue #25085 - Repair medical air high temperature alarm (not working at Master Alarm Panels) 4- Issue #25102 - Install proper label(s) on Med-Air Intake and Med-Vacuum Exhaust piping. (2) Contractor shall test all alarms, equipment, and associated equipment for normal and proper operating once work has been completed. (3) Contractor shall ensure area is clean of all debris material before departing site.

LINE ITEM 0002: Shipping & Handling

The Department of Veterans Affairs, Network Contracting Office 7 located in Tuscaloosa, AL intends to issue a sole source firm fixed priced contract to EMGS, a small business, located in Villa Rica, GA 30180. The contract is expected to be awarded under FAR Part 13. A Limited Sources Justification has been prepared. This procurement is being conducted in accordance with FAR 13.106-1(b) (1)(ii), Soliciting from a Single Source.

The NAICS code utilized for this procurement is 622110 General Medical and Surgical Hospitals with a Small Business Size Standard of $47.0 Million. THIS NOTICE OF INTENT IS NOT A REQUEST FOR COMPETITIVE QUOTES. However, Contractors may submit a capability statement, proposal, or quotation which shall be considered by the contracting officer whether to continue with the sole source procurement or to provide for competition. Such responses MUST be received no later than February 14, 2022 at 10AM CST. Forward responses by e-mail to Contracting Officer, Terri Hudson. Responses received will be evaluated; however, a determination by the Government not to compete the proposed procurement based upon responses received to this notice is solely within the discretion of the Government. The Department of Veterans Affairs, NCO 7 will NOT be responsible for any costs incurred by interested parties in responding to this notice of intent. Only written responses will be considered. All responses from responsible sources will be fully considered. As a result of analyzing responses to this notice of intent, the Government shall determine if a solicitation will be issued. Any prospective contractor must be registered in the System for Award Management (SAM) to be eligible for award. Interested parties are encouraged to furnish information by email only with Response to Intent to Sole Source for Medical Gas Delivery System Repair in the subject line. All documents submitted will not be returned. All interested parties shall submit clear and convincing documentation demonstrating their capabilities to satisfy the requirements listed to Contracting Officer, Terri Hudson by email at terri.hudson@VA.Gov. The capability documentation should include general information and technical background describing in detail the prospective contractor's capabilities and experience that will demonstrate the ability to support this requirement.

Files

File Type Posted
36C24725Q0375_1.docx DOCX document
36C24725Q0375.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
J065--Medical Gas Prevenative Maintenance and Service Albany VAMC Department of Veteran Affairs Albany Vamc Albany, New York 12208, United States · NAICS 811210 Veterans Integrated Service Network 2 No set-aside
H365--annual medical gas testing POP: 12/01/2026 - 09/30/2031 VA Loma Linda Healthcare System 11201 Benton Street Loma Linda CA 92357, United States · NAICS 541380 Veterans Integrated Service Network 22 Service-Disabled Veteran-Owned Small Business
Medical Gas System Activation and Bulk Oxygen Tank Services Louisville, Kentucky 40222, United States · NAICS 325120 Veterans Integrated Service Network 9 Service-Disabled Veteran-Owned Small Business
6835--Med Gas Base Plus Four Department of Veterans Affairs Tucson VA Medical Center Tucson, Arizona 85723, United States · NAICS 325120 Veterans Integrated Service Network 22 Service-Disabled Veteran-Owned Small Business
FY27: Replacement Dental Vacuum Pump System United States · NAICS 811210 Veterans Integrated Service Network 7 Service-Disabled Veteran-Owned Small Business

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors