Glovebox

Closed Solicitation Posted

Solicitation number
SLAC_345266_glovebox
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Menlo Park, California 94025, United States
Points of contact

Notice details come from SAM.gov. Updated .

About this opportunity

The SLAC National Accelerator Laboratory, operated by Stanford University for the Department of Energy, is seeking proposals for a 3-chamber argon purged glovebox with a purification unit. The glovebox must maintain oxygen and water concentrations below 1ppm, and include 2 antechambers linking the working chambers and 4 external antechambers. Technical evaluation criteria include requirements for the purification units, regeneration units, glove port feedthroughs, vacuum pumps, HEPA filters, sensors, purging systems, electronic controls, and leak rate. Quotes are due by COB on May 10, 2024 and should be submitted to kcoh@SLAC.stanford.edu. Source selection will be based on the Lowest Priced Technically Acceptable method.

There is no set-aside designation for this solicitation. The resulting purchase order will be issued on a fixed-price basis, subject to SLAC's standard terms and conditions. The Buy American Act certification applies for awards over $10,000, and all subcontractors must be registered in the System for Award Management (SAM.gov). The place of performance is West Menlo Park, CA 94025, USA. Vendors are requested to provide pricing, lead time, and any potential discounts, as well as a link to their published price list if available.

Notice text

3 versions

Update #3 · Latest ·

SLAC National Accelerator Laboratory is operated by Stanford University for the Department of Energy.  As a management and operating contractor, subcontracts awarded by the Laboratory are not Federal procurements, and are not directly subject to the Federal Acquisition Regulations in 48 CFR. Nonetheless, certain Federal laws, Executive Orders, and regulations may affect our DOE approved purchasing system, as required by statute, regulation, or contract terms and conditions.

Amendment 02

HI,

The following Addendum 02 is being issued for the subject Request for Quotation: 345266

Answers to questions:

  1. Delivery date? I could not find it. : Answer: preferable 3 months ARO
  2. What "Panal" means? Panel or touchscreen? Answer: spelling error, panal refers to touch screen panel
  1. Each chamber has box purging system operated by panal, flow rate max 200L/min

This Addendum modifies the Request for Quotation and must be acknowledged. Failure to acknowledge all addenda could make your proposal non-responsive.  Please confirm that the submitted proposal is not changed because of this addendum. If proposal sent is revised because of this, please submit the revised proposal.

The following changes are hereby made to the appropriate documents contained the above-mentioned Request for Quotation (RFQ). All other terms and conditions of the RFQ remain the same.

Thank you very much.

Katherine Jill A. Coh

SCM Commodities Team Lead

SLAC National Accelerator Laboratory

2575 Sand Hill Road

Menlo Park, CA 94025

Tel: 650-926-2770

Email: kcoh@slac.stanford.edu

Amendment 01 to RFP 345266:

The following Addendum is being issued for the subject Request for Quotation: 345266 Glovebox

Previous Technical Criteria is being replaced by its entirety. NEW TECHNICAL CRITERIA:

  1. 3 chamber glove boxes linked with T shaped with antechambers. Access external and in between chambers. Hinged covers.  2 T antechambers total. Manual valve control. End left and right chambers will have mini antechambers with hinged covers. Manual 3 way valve. See drawing Side Left and right ) chambers are 1500 mm width.  Center chamber is 1250 mm width.
  2. Each chamber has: Purification unit to decrease oxygen and water concentrations below 1ppm
  3. Each chamber has: Regeneration unit for oxygen and water filters
  4. Each chamber has: water analyzer 0-500 ppm
  5. Each chamber has: Oxygen analyzer 1-1000 ppm
  6. Each chamber has: Circulation blower
  7. Each chamber has: automatic pressure control -15 + 15 mbar
  8. Each chamber has: touch screen control unit
  9. Each chamber has: Electronic control system for operation pressure
  10. Each chamber has: Foot pedal control system for operation pressure
  11. Each chamber has: stainless steel circulation pipework, 1 piece stand 1 meter height, bolted on side panels and moving castors, front window polycarbonate feedthroughs, lighting and main antechamber doors
  12. 1500 mm width chambers have 3 port glove feedthroughs, 1250 mm width chamber has 2 glove port feedthroughs
  13. Each chamber has rotary vane pump speed up 10 10CFM including oil mist filter and oil return kit
  14. Each chamber has: 3 height adjustable shelves
  15. Each Chamber has 1 electrical feedthrough and 4 flanges DN40

Each chamber has 2 pc HEPA dust filter (inlet/outlet)

  1. Each chamber has box purging system operated by panal, flow rate max 200L/min
  2. Flange mounting KF40 Recesessed conductors 25 pin 500V 5 amp rating per pin mounting on QF 50 flange
  3. ALL chambers: Leak rate below 0.05 Vol%/h
  4. Crating and shipping included
  5. Installation included

Please take note that statement from solicitation: Quote is for product only; no on-site services permitted” is now removed. ONSITE INSTALLATION IS INCLUDED AS CALLED FOR IN TECHNICAL EVALUATION CRITERIA.

This Addendum modifies the Request for Quotation and must be acknowledged. Failure to acknowledge all addenda could make your proposal non-responsive.  Please confirm that the submitted proposal is not changed because of this addendum. If proposal sent is revised because of this, please submit the revised proposal.

The following changes are hereby made to the appropriate documents contained the above-mentioned Request for Quotation (RFQ). All other terms and conditions of the RFQ remain the same.

Thank you very much.

Katherine Jill A. Coh

SCM Commodities Team Lead

SLAC National Accelerator Laboratory

2575 Sand Hill Road

Menlo Park, CA 94025

Tel: 650-926-2770

Email: kcoh@slac.stanford.edu

Required item(s) listed below:

1.) 3 chamber argon purged glove box with purification unit. Oxygen and water concentrations under 1ppm. 2 Antechambers linking each working chamber together and 4 antechambers (2 large, 2 small) external

Technical evaluation criteria are as follows:

  1. 3 chamber glove boxes linked with antechambers
  2. Purification units to decrease oxygen and water concentrations below 1ppm
  3. Regeneration unit for oxygen and water filters
  4. 3- or 2-piece glove port feedthroughs per box
  5. Vacuum pumps for antechambers
  6. HEPA filters for gloveboxes
  7. Oxygen (3) and wate (3) sensors for glove box
  8. Glove box purging system
  9. Electronic control system for operation pressure
  10. Foot pedal control system for operation pressure
  11. Leak rate below 0.05 Vol%/h
  12. Flanges for electrical / gas feedthroughs (at least 4)
  13. Crating and shipping included
  14. Installation included

Quote is for product only; no on-site services permitted.

Source selection will be based on Lowest Priced Technically Acceptable Method.

Please provide a price and lead time quotation by COB  May 10, 2024. 

Send quote to kcoh@SLAC.stanford.edu.

Please be advised that the requirement was also posted in SAM.gov.

Is there any room for additional discount such as for Academic, ICPT, GSA, or Government Institutions?

Buy American Act Certification (BAA) applies to awards over $10,000.

Representations & Certifications Rider is required. Once completed the Reps and Certs is valid for one year until renewal. If you have already completed one within the year please disregard. Otherwise, please complete and return.

All subcontractors are required to register in the System for Award Management (www.sam.gov).

If possible, please provide a link to your webpage for published price list to substantiate for the prices quoted.

The resulting purchase order will be issued on a fixed price basis. SLAC National Accelerator Laboratory Terms and Conditions for Fixed Price Commercial Supplies and Services (Rev. August 2023) will be made a part of the resulting order. And SLAC Onsite Non-Green Rider Terms and Conditions. Please mark up terms and conditions if you wish to negotiate. No mark ups mean that you accept SLAC Terms and Conditions.

Please also see attached Certificate of Insurance and Illness and Injury Prevention Plan and return filled out document together with your proposal.

Payment Terms will be NET-30

Freight terms will be FOB Destination Prepaid and Add

You are requested to mark and identify proprietary/confidential information provided in response to the solicitation.

Contact me directly with any questions.

Sincerely,

Katherine Jill A. Coh

Procurement Commodities Team Lead

SLAC National Accelerator Laboratory

2575 Sand Hill Road

Menlo Park, CA 94025

Phone: 650-926-2770

Email: kcoh@slac.stanford.edu

Update #2 ·

SLAC National Accelerator Laboratory is operated by Stanford University for the Department of Energy.  As a management and operating contractor, subcontracts awarded by the Laboratory are not Federal procurements, and are not directly subject to the Federal Acquisition Regulations in 48 CFR. Nonetheless, certain Federal laws, Executive Orders, and regulations may affect our DOE approved purchasing system, as required by statute, regulation, or contract terms and conditions.

Amendment 01 to RFP 345266:

The following Addendum is being issued for the subject Request for Quotation: 345266 Glovebox

Previous Technical Criteria is being replaced by its entirety. NEW TECHNICAL CRITERIA:

  1. 3 chamber glove boxes linked with T shaped with antechambers. Access external and in between chambers. Hinged covers.  2 T antechambers total. Manual valve control. End left and right chambers will have mini antechambers with hinged covers. Manual 3 way valve. See drawing Side Left and right ) chambers are 1500 mm width.  Center chamber is 1250 mm width.
  2. Each chamber has: Purification unit to decrease oxygen and water concentrations below 1ppm
  3. Each chamber has: Regeneration unit for oxygen and water filters
  4. Each chamber has: water analyzer 0-500 ppm
  5. Each chamber has: Oxygen analyzer 1-1000 ppm
  6. Each chamber has: Circulation blower
  7. Each chamber has: automatic pressure control -15 + 15 mbar
  8. Each chamber has: touch screen control unit
  9. Each chamber has: Electronic control system for operation pressure
  10. Each chamber has: Foot pedal control system for operation pressure
  11. Each chamber has: stainless steel circulation pipework, 1 piece stand 1 meter height, bolted on side panels and moving castors, front window polycarbonate feedthroughs, lighting and main antechamber doors
  12. 1500 mm width chambers have 3 port glove feedthroughs, 1250 mm width chamber has 2 glove port feedthroughs
  13. Each chamber has rotary vane pump speed up 10 10CFM including oil mist filter and oil return kit
  14. Each chamber has: 3 height adjustable shelves
  15. Each Chamber has 1 electrical feedthrough and 4 flanges DN40

Each chamber has 2 pc HEPA dust filter (inlet/outlet)

  1. Each chamber has box purging system operated by panal, flow rate max 200L/min
  2. Flange mounting KF40 Recesessed conductors 25 pin 500V 5 amp rating per pin mounting on QF 50 flange
  3. ALL chambers: Leak rate below 0.05 Vol%/h
  4. Crating and shipping included
  5. Installation included

Please take note that statement from solicitation: Quote is for product only; no on-site services permitted” is now removed. ONSITE INSTALLATION IS INCLUDED AS CALLED FOR IN TECHNICAL EVALUATION CRITERIA.

This Addendum modifies the Request for Quotation and must be acknowledged. Failure to acknowledge all addenda could make your proposal non-responsive.  Please confirm that the submitted proposal is not changed because of this addendum. If proposal sent is revised because of this, please submit the revised proposal.

The following changes are hereby made to the appropriate documents contained the above-mentioned Request for Quotation (RFQ). All other terms and conditions of the RFQ remain the same.

Thank you very much.

Katherine Jill A. Coh

SCM Commodities Team Lead

SLAC National Accelerator Laboratory

2575 Sand Hill Road

Menlo Park, CA 94025

Tel: 650-926-2770

Email: kcoh@slac.stanford.edu

Required item(s) listed below:

1.) 3 chamber argon purged glove box with purification unit. Oxygen and water concentrations under 1ppm. 2 Antechambers linking each working chamber together and 4 antechambers (2 large, 2 small) external

Technical evaluation criteria are as follows:

  1. 3 chamber glove boxes linked with antechambers
  2. Purification units to decrease oxygen and water concentrations below 1ppm
  3. Regeneration unit for oxygen and water filters
  4. 3- or 2-piece glove port feedthroughs per box
  5. Vacuum pumps for antechambers
  6. HEPA filters for gloveboxes
  7. Oxygen (3) and wate (3) sensors for glove box
  8. Glove box purging system
  9. Electronic control system for operation pressure
  10. Foot pedal control system for operation pressure
  11. Leak rate below 0.05 Vol%/h
  12. Flanges for electrical / gas feedthroughs (at least 4)
  13. Crating and shipping included
  14. Installation included

Quote is for product only; no on-site services permitted.

Source selection will be based on Lowest Priced Technically Acceptable Method.

Please provide a price and lead time quotation by COB  May 10, 2024. 

Send quote to kcoh@SLAC.stanford.edu.

Please be advised that the requirement was also posted in SAM.gov.

Is there any room for additional discount such as for Academic, ICPT, GSA, or Government Institutions?

Buy American Act Certification (BAA) applies to awards over $10,000.

Representations & Certifications Rider is required. Once completed the Reps and Certs is valid for one year until renewal. If you have already completed one within the year please disregard. Otherwise, please complete and return.

All subcontractors are required to register in the System for Award Management (www.sam.gov).

If possible, please provide a link to your webpage for published price list to substantiate for the prices quoted.

The resulting purchase order will be issued on a fixed price basis. SLAC National Accelerator Laboratory Terms and Conditions for Fixed Price Commercial Supplies and Services (Rev. August 2023) will be made a part of the resulting order. And SLAC Onsite Non-Green Rider Terms and Conditions. Please mark up terms and conditions if you wish to negotiate. No mark ups mean that you accept SLAC Terms and Conditions.

Please also see attached Certificate of Insurance and Illness and Injury Prevention Plan and return filled out document together with your proposal.

Payment Terms will be NET-30

Freight terms will be FOB Destination Prepaid and Add

You are requested to mark and identify proprietary/confidential information provided in response to the solicitation.

Contact me directly with any questions.

Sincerely,

Katherine Jill A. Coh

Procurement Commodities Team Lead

SLAC National Accelerator Laboratory

2575 Sand Hill Road

Menlo Park, CA 94025

Phone: 650-926-2770

Email: kcoh@slac.stanford.edu

Update #1 ·

SLAC National Accelerator Laboratory is operated by Stanford University for the Department of Energy.  As a management and operating contractor, subcontracts awarded by the Laboratory are not Federal procurements, and are not directly subject to the Federal Acquisition Regulations in 48 CFR. Nonetheless, certain Federal laws, Executive Orders, and regulations may affect our DOE approved purchasing system, as required by statute, regulation, or contract terms and conditions.

Required item(s) listed below:

1.) 3 chamber argon purged glove box with purification unit. Oxygen and water concentrations under 1ppm. 2 Antechambers linking each working chamber together and 4 antechambers (2 large, 2 small) external

Technical evaluation criteria are as follows:

  1. 3 chamber glove boxes linked with antechambers
  2. Purification units to decrease oxygen and water concentrations below 1ppm
  3. Regeneration unit for oxygen and water filters
  4. 3- or 2-piece glove port feedthroughs per box
  5. Vacuum pumps for antechambers
  6. HEPA filters for gloveboxes
  7. Oxygen (3) and wate (3) sensors for glove box
  8. Glove box purging system
  9. Electronic control system for operation pressure
  10. Foot pedal control system for operation pressure
  11. Leak rate below 0.05 Vol%/h
  12. Flanges for electrical / gas feedthroughs (at least 4)
  13. Crating and shipping included
  14. Installation included

Quote is for product only; no on-site services permitted.

Source selection will be based on Lowest Priced Technically Acceptable Method.

Please provide a price and lead time quotation by COB  May 10, 2024. 

Send quote to kcoh@SLAC.stanford.edu.

Please be advised that the requirement was also posted in SAM.gov.

Is there any room for additional discount such as for Academic, ICPT, GSA, or Government Institutions?

Buy American Act Certification (BAA) applies to awards over $10,000.

Representations & Certifications Rider is required. Once completed the Reps and Certs is valid for one year until renewal. If you have already completed one within the year please disregard. Otherwise, please complete and return.

All subcontractors are required to register in the System for Award Management (www.sam.gov).

If possible, please provide a link to your webpage for published price list to substantiate for the prices quoted.

The resulting purchase order will be issued on a fixed price basis. SLAC National Accelerator Laboratory Terms and Conditions for Fixed Price Commercial Supplies and Services (Rev. August 2023) will be made a part of the resulting order. And SLAC Onsite Non-Green Rider Terms and Conditions. Please mark up terms and conditions if you wish to negotiate. No mark ups mean that you accept SLAC Terms and Conditions.

Please also see attached Certificate of Insurance and Illness and Injury Prevention Plan and return filled out document together with your proposal.

Payment Terms will be NET-30

Freight terms will be FOB Destination Prepaid and Add

You are requested to mark and identify proprietary/confidential information provided in response to the solicitation.

Contact me directly with any questions.

Sincerely,

Katherine Jill A. Coh

Procurement Commodities Team Lead

SLAC National Accelerator Laboratory

2575 Sand Hill Road

Menlo Park, CA 94025

Phone: 650-926-2770

Email: kcoh@slac.stanford.edu

Files

Similar open opportunities

Opportunity Agency Set-aside Responses due
BL18 In-Vacuum Undulator Menlo Park, California 94025, United States · NAICS 333248 Office of Science No set-aside
PIP-II ssrs and LB650 vacuum gauge and readout Illinois 60510, United States · NAICS 334513 Fermi National Accelerator Laboratory No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity