The U.S. Army Contracting Command–Redstone Arsenal issued a Request for Information on July 9, 2026, with a response deadline of July 16, 2026, to identify sources capable of supplying two Geiger-Mueller detector systems for the Army Primary Standards Laboratory at Redstone Arsenal, Alabama. This is market research only and does not constitute a commitment to contract or solicitation for proposal.
The requirement covers one Model 1056A-200 Ratemeter Digital GM detector and one Model 897A-210 GM Detector System. The 1056A-200 must deliver a 5-decade range between 10² and 10⁵ mR/h, 15% energy response from 80 keV to 1.5 MeV, integral analog preamplifier, remote ratemeter capability up to 2,000 feet via single cable, no external power required at detector location, an 8 Ci ³⁶Cl check source, and life expectancy up to 10⁵ rads, with LCD graphical display and IP65 wallmount rating. The 897A-210 must be compatible with the 955A GM Area Monitor, offer three selectable 5-decade ranges from 10⁻² to 10⁵ R/h, feature integral preamp, seismic qualification, approximately 10⁵ rads life expectancy, remote ratemeter capability up to 1,700 feet, single cable connection, and 8 Ci ³⁶Cl check source.
Interested parties must respond via SAM.gov with company name, business size, socio-economic status, points of contact, CAGE Code, and Unique Entity Identifier. The procurement is classified under NAICS 334515 and PSC 6625. Set-aside status is to be determined. Contact Portia R. Sampson, Contracting Officer, at portia.r.sampson.civ@army.mil or 256-955-9135.
Notice text
The U.S. Army TMDE Activity (USATA) has a requirement to implement a brand name purchase of one (1) Model 1056A-200 Ratemeter Digital and one (1) Model 897A-210, GM Detector Systems that will be used for customer support. These items shall be delivered to the Army Primary Standards Laboratory (APSL), Building 5435 Fowler Road, Redstone Arsenal, AL. The units shall include the following specifications or equivalent (if not brand name):
1.0 Requirements:
Model 1056A-200 Ratemeter Digital Geiger-Mueller (GM) Qty:1
• Range: 5 decades between 10^2 and 10^5 mR/h
• Energy response: 15 % from 80 keV to 1.5 MeV
• No external power needed at detector location
• Analog preamplifier integral with detector
• Ratemeters may be remotely located up to 2000 feet (using 50-100 cable or equivalent)
• Single cable between ratemeter and detector
• 8 µCi 36 Cl check source
• Life expectancy: up to 10^5 rads (for detector)
• Seismically tested with high reliability options available
• LCD graphical display showing radiation value
• Wallmount version rated IP65
Model 897A-210, Geiger-Mueller (GM) Detector Systems Qty:1
• For use with the 955A GM Area Monitor
• 3 ranges available. Each cover 5 decades from 10-2to 105R/h
• Energy response: +15 % for 80 keV to 1.5 MeV
• Preamp integral with detector
• Seismically qualified
• Life expectancy: approximately 105 rads
• All power supplied from ratemeter
• Ratemeter may be remotely located up to 1700 feet
• Single cable between ratemeter/detector
• 8 µCi36Cl check source
2.0 Delivery Requirements
Shipping Address: U.S. Army TMDE Activity, Attn: Patrick Marine, Building 5435, Redstone Arsenal, AL 35898
Files
| File | Type | Posted |
|---|---|---|
| Purchase Description.pdf | ||
| Request for Information Detector Hardware System (SAM).pdf |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Detector, radio frequen | Land and Maritime | Total Small Business | |
| Handheld X-Ray Fluorescence for ERDC GSL | Engineer Research and Development Center | No set-aside | |
| Detector, Gas | Land and Maritime | Total Small Business | |
| SAP - Purchase of 4 Canberra gamma radiation detec | National Vehicle and Fuel Emissions Laboratory | No set-aside | |
| Case, radiac set | Land and Maritime | Total Small Business |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity