The Defense Health Agency issued a request for information (RFI) on July 22, 2026, to conduct market research on a Cardiac Output Monitoring System, with responses due by July 27, 2026. This RFI does not constitute a solicitation, a request for quotes, or an obligation to award a contract; the Government will not pay for information submitted and will not respond to requests for solicitation.
The Defense Health Agency Contracting Activity, located at Fort Detrick, Maryland, seeks to identify qualified contractors capable of supplying a cardiac output monitoring device for the United States Army Institute of Surgical Research (USAISR). The device must monitor cardiac output and derive hemodynamic parameters in a hospital setting using a Swan-Ganz catheter. A critical requirement is that the device's proprietary coded output must be compatible with USAISR's unique internal data logging systems to maintain continuity of historic and current protocol data and workflow. Loss of compatibility would prevent comparison of current and future studies to USAISR historical data, resulting in data loss and wasted effort.
Respondents must submit written responses demonstrating technical expertise and functional area experience sufficient to fulfill the requirement in detail. The notice applies to manufacturers of surgical appliances and supplies (NAICS 339113) and providers of medical and surgical instruments, equipment, and supplies (PSC 6515).
Notice text
This is a REQUEST FOR INFORMATION (RFI) ANNOUNCEMENT ONLY. In accordance with Revolutionary Federal Acquisition Regulation Overhaul 15.101(c), the Government does not intend to award a contract based on this RFI or otherwise pay for information requested. The information will be treated as information only and not as a proposal. This is not a solicitation announcement nor a request for quotes and does not obligate the Government to award a contract. Requests for a solicitation will not receive a response. Responses to this RFI must be in writing.
The purpose of this RFI announcement is to conduct market research to make appropriate acquisition decisions and to gain knowledge of potential qualified offerors interested and capable of providing the “Cardiac Output Monitoring System” described below. Documentation of technical expertise must be presented in sufficient detail for the Government to determine that your company possesses the necessary functional area of expertise and experience to fulfill the requirement.
The Defense Health Agency Contracting Activity, 808 Schreider Street, Fort Detrick, MD 21702, is seeking to understand the capabilities of potential qualified Contractors who can provide the supplies and/or services for the following:
Minimum Essential Characteristics: Cardiac Monitoring Platform
The United States Army Institute of Surgical Research (USAISR) seeks a device that can monitor cardiac output (CO) and derive hemodynamic parameters in a hospital setting. The USAISR requires a CO monitoring device that utilizes a Swan-Ganz catheter to monitor hemodynamic parameters. The device’s proprietary coded output must be compatible with USAISR’s unique internal data logging systems. This is required to maintain historic and current protocol data continuity and workflow. Inability to meet these requirements will inhibit comparisons of current and future studies to USAISR historical data resulting in loss of data, and efforts.
The CO Monitoring Platform shall have the following specifications:
- Ability to measure the following: mixed venous oxygen saturation (SvO2), CO, stroke volume, systemic vascular resistance, right ventricular ejection fraction, right ventricular end diastolic volume, pulmonary artery pressure, and pulmonary artery occlusion pressure.
- Utilization of the Edwards Lifesciences 7.5 FR Swan-Ganz CCOmbo V set or a Swan-Ganz of equal functionality and performance that provides continuous monitoring of SvO2 and measures continuous calculation of cardiac output by using thermodilution principles.
- Equipment dimensions equal to or smaller than 342 x 388 x 208 mm (height x width x depth)
- Equipment shall weigh less than 22 lbs
- Equipment ingress protection of at least IPX1
- Display size of at least 15 inches.
- At least two (2) COM1 serial port connectors, RS-232
- Operational environmental specifications:
- Temperature range: 0 to 37℃
- Relative humidity: 10-90% non-condensing
- Altitude: 0 to 3,000 meters
- Transportation environmental specifications:
- Temperature range: -18 to 45℃
- Relative humidity: 20 to 90% non-condensing
- Altitude: Maximum 6,096 m for up to 8 hours
- Monitor battery must have these technical characteristics:
- Output voltage (nominal) 14.4 V
- Maximum discharge current 4.096 A (8.5 A at 25℃)
- Cells 8 x Li-Ion (lithium ion)
- Monitoring platform technical specifications:
- Input/Output
- Projective capacitive touch screen
- At least two (2) RS-232 serial ports: Maximum data rate = 57.6 kilo baud
- At least three (3) USB ports: 2 x USB 3.0; 1 x USB 2.0
- At least one (1) RJ-45 ethernet port
- At least one (1) HDMI port
- Heart Rate average display
- Response time: 40 seconds for step increase from 80 to 120 beats per minute (BPM), 29 seconds for step decrease from 80 to 40 BPM.
- CO monitoring on: Averaging time: Time between CO measurements (3 to 21 minutes); Update rate: Approximately 1 minute; Response time: 175 seconds for step increase from 80 to 120 BPM, 176 seconds for step decrease from 80 to 40 BPM.
- Electrical
- Rated supply voltage: 100 to 240 Vac; 50/60 Hz
- Rated input: 1.5 to 2.0 Amps
- Fuses: T 2.5AH, 250V; High breaking capacity; Ceramic
- Alarm
- Sound pressure level: 45 to 85 dB(A)
- Input/Output
- Utilization of a Swan-Ganz patient cable with the following parameter measurement specifications:
- Continuous Cardiac Output
- Range: 1 – 20 Liter/minute
- Reproducibility: ±6% or 0.1 Liter/minute, whichever is greater
- Maximum thermal filament surface temperature 48℃
- Intermittent (bolus) Cardiac Output
- Range: 1 – 20 Liter/minute
- Reproducibility: ±3% or 0.1 Liter/minute, whichever is greater
- Blood temperature
- Range: 15 to 45℃
- Accuracy: ±3℃
- Continuous Cardiac Output
- Roll Stand:
- Compatible rolling stand with bracket and plates for cardiac output monitoring system
- Base Dimensions: 12-15 inches wide x 10-12 inches deep to allow
- Height: Non-specific but need to be adjustable
- Compatible rolling stand with bracket and plates for cardiac output monitoring system
- Warranty and Product Support:
- 12-month standard warranty
- 24-month product support and protection plan
The North American Classification System code for this acquisition is 339113 –
Surgical Appliance and Supplies Manufacturing (Size Standard: 800 Employees).
Notice to potential offerors: All offerors who provide goods or services to the United States Federal Government must be registered in the System for Award Management (SAM) at www.sam.gov and complete online representations and certifications in SAM.
All interested parties shall submit information by email to both Tammy McRae and Justin Pickett at tammy.n.mcrae.civ@health.mil and justin.h.pickett.civ@health.mil, no later than 5:00 PM Eastern Time on 27 July 2026.
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Operational Medicine Test and Evaluation Cell Salvage Supplemental - (AFMEDCOM) | Air Force Installation Contracting Agency | No set-aside | |
| Microbial Identification System (MIS) | Medical Command | No set-aside | |
| Medtronic cardiovascular "procedural packages" | Medical Supply Chain | No set-aside | |
| HD endoscope_gastroscope solution | Veterans Integrated Service Network 5 | No set-aside | |
| Request For Information M67854-26-I-2023: Own Force Monitoring at Echelon (OFM-E) Electromagnetic Spectrum Sensing | Marine Corps | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity