Bioprotect E3 max (class II biological safety cabinet for lab instrumentation) (ambis 2225990)

Closed Pre-Solicitation Posted

Solicitation number
NOI-NIAID-24-2225990
Agency
Responses due
Set-aside
No set-aside

Opportunity facts

Place of performance
Bethesda, Maryland 20892, United States

Notice details come from SAM.gov. Updated .

About this opportunity

Summary of Key Details:

The National Institute of Allergy and Infectious Diseases (NIAID) of the National Institutes of Health (NIH) intends to negotiate a sole-source contract with The Baker Company, Inc. to procure a BioPROTECT e3 MAX (Class II Biological Safety Cabinet) for their Research Technologies Section Flow Cytometry Unit. This cabinet will replace a broken biosafety cabinet used to safely process BSL-3 samples for NIAID researchers. The cabinet must meet specific requirements, including size, HEPA filtration, and the ability to accommodate various laboratory instruments. All responsible sources that could provide comparable services may submit capability statements by 4:00pm EST on July 15, 2024 through the NIAID electronic Simplified Acquisition Submission System (eSASS) website.

This is not a set-aside opportunity. The Baker Company, Inc. is the only known responsible source, as the cabinet must match the existing footprint and equipment requirements. No award value or budget range is provided. The quantity is one cabinet to be installed at the NIH facility in Bethesda, MD.

Notice text

This solicitation NOI-NIAID-24-2225990 is a Notice of Intent, not a request for proposal.  The National Institute of Allergy and Infectious Diseases (NIAID) of the National Institutes of Health (NIH) intends to negotiate on an other than full and open competition basis with The Baker Company, Inc,, 175 Gate House Road, Sanford, ME 04073-2482, United States, to procure a BioPROTECT e3 MAX (Class II Biological Safety cabinet for Lab Instrumentation).

The National Institutes of Health (NIH), National Institute of Allergy and Infectious Diseases (NIAID), conducts and supports basic and applied research to better understand, treat, and ultimately prevent infectious, immunologic, and allergic diseases.  Within the NIAID, the Research Technologies Section (RTS) provides intramural scientists with state-of-the-art facilities and expertise in flow cytometry, microscopy and genomics. The RTS Flow Cytometry Unit regularly performs state-of-the-art research for clinicians and researchers within the DIR towards the NIAID goal of discovery of new diagnostics, therapeutics and vaccines against disease.

Project Description:

The RTS Flow Cytometry unit currently uses biosafety cabinet to safely conduct experiments in the BSL – 3 on behalf of all the NIAID DIR scientists. This acquisition is to replace a broken biosafety cabinet that holds sorting instruments and is used to process BSL-3 samples. The Baker cabinet currently matches the exterior footprint of the old cabinet, so no modifications or alterations to the space are needed. This cabinet also has the interior space to hold the required equipment, and it has the custom HEPA filters needed to maintain the safety of the space. Using this cabinet will protect the laboratory's current investment in both the equipment and the research in progress.

The current requirements are as follows:

  1. BioPROTECT e3 MAX, vented to room, quantity 1
  2. Installation – CPD Authorized Baker Installer

Features:

  • "Walk-in" biosafety cabinet, featuring Class II performance for containment and protection
  • The sidewall and rear wall plenum are under negative pressure
  • Available in 3 standard widths with additional options available for extra deep or extra tall configurations
  • Instrument table is adjustable, designed to accommodate multiple heights for user access points and varying sizes of instruments
  • Allows roll in cart, for instrument loading and service
  • Space in the base for fluidics and other ancillary equipment
  • Designed for high-volume robotic and automated equipment applications
  • Accommodates high-throughput robotic systems (including ancillary devices), ultra-centrifuges, flow cytometers, liquid handlers and other large pieces of laboratory equipment
  • Ideal for high-throughput screening, combinatorial chemistry, drug discovery, immunology, tissue culture, clinical research, molecular biology and quality control assays
  • Product protection: HEPA-filtered, particulate-free airflow throughout the work area provides an ISO Class 5 environment and minimizes cross contamination
  • Personnel protection: Intake velocity of 100 fpm [0.5 m/s] into front suctions grilles located in the access doors preventing escape of potentially hazardous aerosols and particulate
  • Environmental protection: The cabinet’s internal fan provides HEPA filtered exhaust flow from the cabinet
  • The exhaust can be vented back into the lab space or connected to the building exhaust system with a canopy exhaust connection should volatile toxic compounds be present
  • Containment and Protection – External Dimensions:
  • Width Standard 116.88"
  • Depth 61.63"
  • Height 102.5"

Place of Performance: NIH, Building 33, 33 North Drive, Bethesda, MD 20892, United States.  FOB: Destination

The statutory authority for this sole source requirement is 41 U.S.C. 253 (c) (1) as implemented by FAR 13.106-1 only one responsible source and no other supply of service will satisfy agency requirements. THIS IS NOT A REQUEST FOR PROPOSAL.  All responsible sources that could provide comparable services may submit a capability statement by 4:00pm, eastern stand time July 15, 2024 with referencing #NOI-NIAID-24-2225990.  

All Capability Statements must be submitted via the NIAID electronic Simplified Acquisition Submission System (eSASS) website at https://esass.nih.gov. All vendors must register in the eSASS system to submit a quotation. Instructions on how to register / submit quotes are included on the website. Late submissions shall be treated in accordance with the solicitation provision at FAR 52.212-1(f). All responsible sources may submit an offer that will be considered by this Agency.  A determination not to compete this requirement, based upon responses to this notice, is solely within the discretion of the Government. 

Similar open opportunities

Opportunity Agency Set-aside Responses due
1-Year OEM Service Contract - (2) Revvity Celigo S Imaging Instruments (VRC) Bethesda, Maryland 20892, United States · NAICS 811210 National Institute of Allergy and Infectious Diseases No set-aside
7125--Sterile Storage Cabinets VA Greater Los Angeles (GLA) Healthcare System Los Angeles CA 90073, United States · NAICS 339113 Veterans Integrated Service Network 22 No set-aside
Preventative Maintenance and FullCare Service Support for Cytiva Sepax C-PRO Cell Separation Systems Bethesda, Maryland 20892, United States · NAICS 811210 Office of Logisitics and Acquisition Operations No set-aside
Annual Preventative Maintenance and Service Support for Seven (7) Miltenyi CliniMACS Prodigy Instruments Bethesda, Maryland 20892, United States · NAICS 811210 Office of Logisitics and Acquisition Operations No set-aside
NIA:Non-Competitive Combined Synopsis/Solicitation NAICS 561990 Office of Logisitics and Acquisition Operations No set-aside