Annual Preventive Maintenance and Certification for Biological Safety Cabinets (Class II, Type A2), Chemical Fume Hood, and PCR Cabinets/Clean Benches
Closed Pre-Solicitation Posted
- Solicitation number
- N6311626Q0516
- Agency
-
Naval Supply Systems Command
Department of the Navy, Department of Defense
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 541380 Testing Laboratories and Services
- Place of performance
- San Diego, California 92152, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The Department of the Navy Naval Supply Systems Command is seeking a responsible vendor to provide annual preventive maintenance and certification services for laboratory equipment, including Biological Safety Cabinets (Class II, Type A2), a Chemical Fume Hood, and PCR Cabinets/Clean Benches. The contractor must possess the capability to perform comprehensive testing and certification in accordance with specific industry standards, including NSF/ANSI 49 for biological safety cabinets, OSHA 1910.1450 and CAL OSHA Title 8 5154.1 for chemical fume hoods, and ISO 5 (Class 100) standards for PCR cabinets and clean benches. Services must encompass downflow velocity profiling, HEPA filter integrity testing, smoke pattern testing, alarm function verification, and detailed certification reporting for each piece of equipment. The pre-solicitation notice was posted on February 9, 2026, with a due date of February 17, 2026, providing interested vendors with an eight-day window to submit their responses or questions.
The contract carries no set-aside designation and is open to all responsible sources. The scope of work encompasses twenty-two pieces of laboratory equipment located at the Naval Health Research Center: eighteen Biological Safety Cabinets (Class II, Type A2), one Chemical Fume Hood, and three PCR Cabinets/Clean Benches. The contractor must supply all necessary labor, tools, materials, and equipment to complete the annual preventive maintenance and certification services, with deliverables including detailed service reports for each piece of equipment that document all tests performed, applicable standards, and pass/fail results. The notice does not specify a total contract value or award amount; however, the scope indicates this is an annual maintenance and certification contract requiring specialized expertise in laboratory equipment testing and compliance.
Notice text
The Government is seeking responsible sources capable of performing annual preventive maintenance and certification for Biological Safety Cabinets (Class II, Type A2), a Chemical Fume Hood, and PCR Cabinets/Clean Benches. See below scope of work.
SCOPE
The vendor shall provide all necessary labor, tools, materials, and equipment to perform annual preventive maintenance (PM) and certification for a total of twenty-two (22) pieces of laboratory equipment located at the Naval Health Research Center. The equipment is as follows:
- Biological Safety Cabinets (Class II, Type A2) – Qty: 18
- Chemical Fume Hood – Qty: 1
- PCR Cabinets/Clean Benches – Qty: 3
The contractor's performance shall include, but not be limited to, the following services for each piece of equipment as applicable:
Biological Safety Cabinets (Class II, Type A2):
- Testing and certification in accordance with NSF/ANSI 49 standards.
- Services must include downflow velocity profiling, inflow velocity measurement, HEPA filter integrity testing (leak scan), smoke pattern testing for airflow visualization, and site installation assessment.
- Alarm function verification (e.g., airflow and window position alarms).
- Decontamination of the work area before and after service.
Chemical Fume Hood:
- Testing and certification in accordance with OSHA 1910.1450 and CAL OSHA Title 8 5154.1.
- Services must include face velocity measurement, smoke visualization testing to verify containment, and alarm function verification.
PCR Cabinets / Clean Benches:
- Testing and certification to ensure ISO 5 (Class 100) air quality conditions are met.
- Services must include HEPA filter integrity testing and non-viable particle counting.
- UV light function verification.
Upon completion of service, the vendor shall provide the following:
- Certification Reports: A detailed service report must be provided for each individual piece of equipment. Each report must clearly document the tests performed, the standards used, the results of each test (pass/fail), and any adjustments made.
- Certification/Testing Labels: A physical label must be provided and affixed to each individual piece of equipment. The label must indicate date of testing/certification and the next scheduled certification date.
The location of this requirement is San Diego, CA 92152.
Interested sources may submit capabilities information by email to the point of contact listed below. Reference N6311626Q0516 on submissions and include Company Name, Full Address, CAGE Code and Point of Contact information including Name, Phone Number and E-Mail address.
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Comprehensive full-service maintenance/repair plan for Vitek 2 Compact 60 | Naval Supply Systems Command | No set-aside | |
| H941--Chemical Hoods & HEPA PM | Veterans Integrated Service Network 10 | Service-Disabled Veteran-Owned Small Business | |
| 7125--Sterile Storage Cabinets | Veterans Integrated Service Network 22 | No set-aside | |
| Comprehensive full-service maintenance/repair plan for Illumina Nova/Seq 6000 | Naval Supply Systems Command | No set-aside | |
| Air sampler | Naval Supply Systems Command | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity