About this opportunity
The VA Bedford Healthcare System is seeking to procure a wall-mounted, ultrapure and pure water filtration system to provide point-of-use water for specific laboratory applications such as immunocytochemistry, immunohistochemistry, DNA/RNA assays, and mass spectrometry. The required system must utilize reverse osmosis and electrodeionization technologies to produce pure (Type 2) and ultrapure (Type 1) water meeting strict specifications for resistivity, microorganisms, particulates, pyrogens, RNase, DNase, and total organic carbon. The system must have a small footprint and utilize a low-voltage DC pump generating less than 52 decibels of sound. The solicitation number is 36C24124Q0861, and responses are due by August 29, 2024.
This opportunity has no set-aside designation. The contract does not identify any potential incumbents or current contractors. The notice does not provide any information on the anticipated award value or budget range. The required system must have the capacity to produce up to 5 liters per hour of pure and ultrapure water for laboratory use.
Notice text
P09 PERFORMANCE WORK STATEMENT & SALIENT CHARACTERISTICS
Salient Characteristics and PWS
BRAND NAME OR EQUAL TO
Millipore Milli-Q IQ 7005
VA Bedford Healthcare System requires a wall-mounted, ultrapure and pure water filtration system to provide point-of-use water for specific laboratory demands including immunocytochemistry, immunohistochemistry, biochemical assays for DNA and RNA, and mass spectroscopy.
The commodity/equipment shall include a PURE and ULTRA-PURE water dispensing unit that produces filtered water directly from a tap water source with a typical water production rate of 5L per hour and variable flow water dispensing. The volumetric water dispensing should deliver a specific water volume, from 20mL to 100L automatically (with 1 mL incremental). The ultrapure water system shall utilize Reverse Osmosis (RO) technology where pump pressure is temperature and resistivity compensated to keep water production rate constant, and the system shall ultilize Electrodeionisation (EDI) technology to ensure constant pure and deionized water which is required for our HPLC, immunocytochemistry and DNA/RNA assays.
Space Specifications:
Due to laboratory space and bench to ceiling constraints, the footprint of all components must not exceed 35 H x 50 W x 30 D.
The ultrapure water system will utilize a low-voltage DC pump which will generate less than 52 decibels of sound at one meter from the system for virtually silent operation.
Electrical and Water Specifications
ELECTRICAL REQUIREMENTS: .120V
MINIMUM INLET PRESSURE: .1 to 6 bar
Other Purification Technology requirements:
Water system has a UV lamp with emission at 265 nm to inactivate bacteria. For safety reasons and hazardous waste management this is preferably a mercury-free lamp.
Water system also uses a UV lamp with emission at 172 nm to reduce remaining dissolved organic contamination and destroy bacteria, thus reducing background noise in all assays. For safety reasons and hazardous waste management this is preferably a mercury-free lamp.
Additionally, storage tank has a UV lamp with 265 nm emission to inactivate bacteria and prevent biofilm formation within the system at all times. For safety reasons and hazardous waste management this is preferably a mercury-free lamp.
Storage tank has an electronic overflow sensor and pressure sensor to allow the system to run afterhours as needed when water is depleted at the end of the day without risk of flooding.
System will have 3 point-of-use filters for additional bacteria control and removal of larger biological structures, pyrogens, endotoxins, RNases, DNases and proteases.
Pure (Type 2) water production requirements:
Point-of-Use Water Delivery Flow Rate up to 2L/min
Pure Water Resistivity (M .cm at 25°C) > 5
Microorganisms (cfu/mL) < 0.01
Particulates < 0.22 m (/mL) < 1
Pyrogen Levels (EU/mL) < 0.001
RNase Level (ng/mL) < 0.01
DNase Level (pg/ L) < 4
TOC (ppb) < 30
Ultapure (Type 1) water production requirements for DNA/RNA/ELISA assays:
Point-of-Use Water Delivery Flow Rate 0.05 to 2L/min
UltraPure Water Resistivity (M .cm at 25°C) 18.2
Microorganisms (cfu/mL) < 0.01
Particulates < 0.22 m (/mL) < 1
Pyrogen Levels (EU/mL) < 0.001
RNase Level (ng/mL) < 0.01
DNase Level (pg/ L) < 4
Proteases (ug/ml) -
Files
| File | Type | Posted |
|---|---|---|
| 36C24124Q0861_2.docx | DOCX document |
Notice history
| Notice | Type | Posted |
|---|---|---|
| This notice | Pre-Solicitation | |
| 6640-- Lab Water Purification System BRAND NAME OR EQUAL to Millipore Milli-Q IQ 7005 | Solicitation |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| H246--Reverse Osmosis, Deionization Equipment and PE Water Softene 2026-2027 | Veterans Integrated Service Network 22 | No set-aside | |
| Capital Equipment: Purchase of (1) Millipore Sigma | EPA Region 2 | No set-aside | |
| AFIT/ENG-Cleanroom DI-RO Type I Ultra-Pure Water System | AFLCMC Wright Patterson AFB | No set-aside | |
| F103--D and D laboratory steam & water testing | Veterans Integrated Service Network 6 | No set-aside | |
| Sources Sought: Alternative Sources of Water for Small Units. (ACC-DTA CSO Amend 42) | Department of the Army | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity