This document is a Presolicitation Notice for a federal contract opportunity related to a laboratory water purification system. The VA Bedford Healthcare System is seeking a wall-mounted, ultrapure and pure water filtration system to provide point-of-use water for specific laboratory applications such as immunocytochemistry, immunohistochemistry, DNA/RNA assays, and mass spectrometry.

The required system must utilize reverse osmosis and electrodeionization technologies to produce pure (Type 2) and ultrapure (Type 1) water meeting strict specifications for resistivity, microorganisms, particulates, pyrogens, RNase, DNase, and total organic carbon. The system must have a small footprint not exceeding 35"H x 50"W x 30"D and utilize a low-voltage DC pump generating less than 52 decibels of sound. Electrical requirements are 120V, and the minimum inlet water pressure is 1 to 6 bar. The system must incorporate multiple UV lamps and filters for bacteria control and contaminant removal. The solicitation number is 36C24124Q0861, and the response is due by August 23, 2024. The contracting office is the Department of Veterans Affairs VA New England HCS.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Presolicitation Notice Presolicitation Notice

SUBJECT*
Lab Water Purification System BRAND NAME OR EQUAL to Millipore Milli-Q IQ 7005

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
01053-9764
SOLICITATION NUMBER*
36C24124Q0861
RESPONSE DATE/TIME/ZONE
08-23-2024
ARCHIVE
60 DAYS AFTER THE RESPONSE DATE

SET-ASIDE

PRODUCT SERVICE CODE*
6640
NAICS CODE*
334516
PLACE OF PERFORMANCE
VA Bedford Medical Center

Research Dept. Blg 18, Rm 213

200 Springs Rd

Bedford MA

POSTAL CODE
01730
COUNTRY
USA

CONTACT INFORMATION

CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

VA NEW ENGLAND HCS

Acquisitions-90 421 North Main Street Leeds MA 01053-9764

Johanne.Menard@va.gov

DESCRIPTION

P09 – PERFORMANCE WORK STATEMENT & SALIENT CHARACTERISTICS

Salient Characteristics and PWS

BRAND NAME OR EQUAL TO

Millipore Milli-Q IQ 7005

1. VA Bedford Healthcare System requires a wall-mounted, ultrapure and pure water filtration system to provide point-of-use water for specific laboratory demands including immunocytochemistry, immunohistochemistry, biochemical assays for DNA and RNA, and mass spectroscopy.

2. The commodity/equipment shall include a PURE and ULTRA-PURE water dispensing unit that produces filtered water directly from a tap water source with a typical water production rate of 5L per hour and variable flow water dispensing. The volumetric water dispensing should deliver a specific water volume, from 20mL to 100L automatically (with 1 mL incremental). The ultrapure water system shall utilize Reverse Osmosis (RO) technology where pump pressure is temperature and resistivity compensated to keep water production rate constant, and the system shall ultilize Electrodeionisation (EDI) technology to ensure constant pure and deionized water which is required for our HPLC, immunocytochemistry and DNA/RNA assays.

3. Space Specifications:

Due to laboratory space and bench to ceiling constraints, the footprint of all components must not exceed 35”H x 50”W x 30”D.

4. The ultrapure water system will utilize a low-voltage DC pump which will generate less than 52 decibels of sound at one meter from the system for virtually silent operation.

5. Electrical and Water Specifications

ELECTRICAL REQUIREMENTS:….120V

MINIMUM INLET PRESSURE:……………………………………….1 to 6 bar

6. Other Purification Technology requirements:

Water system has a UV lamp with emission at 265 nm to inactivate bacteria. For safety reasons and hazardous waste management this is preferably a mercury-free lamp.

Water system also uses a UV lamp with emission at 172 nm to reduce remaining dissolved organic contamination and destroy bacteria, thus reducing background noise in all assays. For safety reasons and hazardous waste management this is preferably a mercury-free lamp.

Additionally, storage tank has a UV lamp with 265 nm emission to inactivate bacteria and prevent biofilm formation within the system at all times. For safety reasons and hazardous waste management this is preferably a mercury-free lamp.

Storage tank has an electronic overflow sensor and pressure sensor to allow the system to run afterhours as needed when water is depleted at the end of the day without risk of flooding.

System will have 3 point-of-use filters for additional bacteria control and removal of larger biological structures, pyrogens, endotoxins, RNases, DNases and proteases.

7. Pure (Type 2) water production requirements:

Point-of-Use Water Delivery Flow Rate up to 2L/min Pure Water Resistivity (MΩ.cm at 25°C) > 5 Microorganisms (cfu/mL) – < 0.01 Particulates < 0.22 μm (/mL) – < 1 Pyrogen Levels (EU/mL) – < 0.001 RNase Level (ng/mL) – < 0.01 DNase Level (pg/μL) – < 4 TOC (ppb) < 30

8. Ultapure (Type 1) water production requirements for DNA/RNA/ELISA assays:

Point-of-Use Water Delivery Flow Rate 0.05 to 2L/min UltraPure Water Resistivity (MΩ.cm at 25°C) 18.2 Microorganisms (cfu/mL) – < 0.01 Particulates < 0.22 μm (/mL) – < 1 Pyrogen Levels (EU/mL) – < 0.001 RNase Level (ng/mL) – < 0.01 DNase Level (pg/μL) – < 4 Proteases (ug/ml) - <0.15 TOC (ppb) < 5

9. The commodity/equipment is compliant with the USP regulations, and the water purification system can report the resistivity and conductivity in both temperatures compensated and non-temperature compensated formats.

10. The contractor providing the commodity will include the purification filters/lamps/tablets and mounting brackets for the system and storage tank for initiation.

11. The contractor providing the requirement/commodity shall also provide installation, on-site training on set-up of protocols and analysis, unlimited instrument and assay and remote technical support.

12. The contractor providing the requirement/commodity shall provide 12-months maintenance and service following the first-year warranty, one preventative maintenance visit per year and discount on replacement parts for emergency repair.

13. Deliver within 8 weeks after receipt of order (ARO).

14. Include all shipping charges.