About this opportunity

The Department of Veterans Affairs, Regional Procurement Office East, issued a notice of intent to award a sole-source, firm-fixed-price contract to Fidelis Sustainability Distribution, LLC for a Ravin Cam Gamma Camera with wall mount, with capability submissions due by 11:00 AM ET on Wednesday, August 26, 2026.

The VA intends to acquire this equipment for VA Pittsburgh Healthcare System's Quality & Patient Safety Department to support real-time radiation detection, contamination assessment, isotope identification, and dose rate measurement during surgical procedures. The gamma camera must detect multiple simultaneous isotopes with 1.1% energy resolution at 662 keV, spectroscopy range of 30–662 keV, optical and radiation field of view specifications, weigh 5.2 lbs, and provide battery life exceeding 6 hours with a detachable tablet for real-time data display. Delivery and installation occur at VA Pittsburgh Healthcare System (University Drive C, Pittsburgh, PA 15240); the vendor must provide a temporary camera during calibration.

Responsible sources believing they can meet requirements may submit capability information demonstrating ability to fulfill specifications. Submissions must include company name, SAM UEI number, address, point of contact details, and business size, supported by relevant corporate experience, applicable contract history with dollar amounts and performance periods, product details, and customer references with phone and email. The Government will evaluate responses solely to determine whether to conduct competitive procurement and reserves discretion to award competitively. No Government cost responsibility applies to response preparation.

Notice text

Page 1 of

Page 5 of 5

Page 1 of

NOTICE OF INTENT TO SOLE SOURCE:

IAW RFO FAR 5.101 this notice is to advise the public that The Department of Veterans Affairs, Regional Procurement Office East (RPOE), Network Contracting Office 4, intends to award a sole source, Firm-Fixed-Price contract pursuant to RFO FAR 12.102 to Fidelis Sustainability Distribution, LLC located at 1281 Terminal Way, Suite 202, Reno, Nevada, for Ravin Cam Gamma Camera and wall mount.

THIS NOTICE OF INTENT IS NOT A REQUEST FOR COMPETITIVE PROPOSALS or QUOTATIONS.

Responsible sources that believe they can meet the requirements may give written notification by submitting capability information that clearly demonstrates the capability to provide the stated products. Capability information is required prior to the response due date and time, to establish the ability of the source to meet this requirement. Supporting evidence must be furnished in sufficient detail to demonstrate the ability to fulfill the requirements and must include relevant corporate experience along with the applicable contract number, total contract dollar amount, contract period of performance, details of the relevant products and customer point of contact with corresponding telephone number and email address.

Responses must include following information:

1. Company Name

2. SAM UEI number

3. Company s address

4. Point of Contact information (title, phone number and email address)

5. Business size.

A determination by the Government not to compete this proposed contract based upon responses to this notice is solely within the discretion of the Government.

Information received will be considered solely for the purpose of determining whether to conduct a competitive procurement. The Government will not be responsible for any costs associated with the preparation of responses to this notification.

Responses are due by 11:00 am ET, Wednesday, August 26, 2026. No submissions will be accepted after this date and time. Responses shall be submitted via email to ryan.hatfield@va.gov. Telephone inquiries will not be accepted.

The NAICS code for this acquisition is 334516 and the size standard is 1,200 Employees.

STATEMENT OF NEED

The Quality & Patient Safey Department at the VA Pittsburgh Healthcare System (VAPHS) is requesting the purchase of a Ravin CAM gamma camera with wall mount.

General Requirements:

Specifications as set forth in this proposal are minimum specifications and shall not be construed as limiting the overall quality, quantity, or performance of the instrument.

The offeror shall meet or exceed the minimum requirements and shall be held responsible for the performance and overall quality of the requested instrument during the instrument s warranty period.

The instruments, based upon the specifications requested, shall be the newest and the most current model.

Company will offer a temporary Ravin CAM, while instrument is in for calibration.

Equipment required:

Ravin CAM Gamma Camera (1):DRAFT

Must be able to:

Detect multiple radiation sources at one time

Show real-time contamination

Show real-time isotope identification, dose rate, and count rate

Tablet to show real-time data

Lightweight and easy to carry

Measure and calculate dose delivered to the tumor and surrounding organs

Wall Mount (1):

Mount to wall of surgery room to show real-time data during procedures

Salient Characteristics:

Detects and shows multiple simultaneous isotopes

Energy Resolution: 1.1% FWHM @ 662 keV

Spectroscopy Range: 30-662 keV

Radiation Detected: Gamma Rays

Optical Field of View: 90o x 65o

Radiation Field of View (coded aperture) : 90o x 90o

Weight: 5.2 lbs

Dimensions: 22cm (H) x 20 cm (W) x 25.5 cm (L)

Battery Life: 6+ hours

Power: 75.5 Whr

Detachable Tablet

Installation, including Verification/Validation, Training:

Installation of wall mount can be done by in-house resources. No verification/validation, training needed.

DELIVERY:

All equipment deliveries must be coordinated with the Biomedical department prior to delivery. Failure to do so may result in the return of the equipment at vendor cost.

Location:

VA Pittsburgh Healthcare System

University Drive C

Pittsburgh, PA 15240

Repair and Maintenance:

Manufacturer will provide troubleshooting manual.

The VAPHS Bio-Medical Department shall manage the service and repair contracts on the equipment.

The manufacturer should define maintenance requirements and indicate what maintenance is performed by the user.

Company will send a temporary Ravin CAM for use during calibration.

DRAFT

Files

File Type Posted
36C24426Q0903.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
6505--FY27 Low Energy General Radiopharmaceuticals 36C25027Q0052 See Section 11 of the Cleveland and Kent, Ohio, United States · NAICS 325412 Veterans Integrated Service Network 10 No set-aside
H365--460-27-1-132-0001 | SPE 49273 | VISN 4 Radiation Safety Officer (RSO) | NCO 4 Services 3 (VA-27-00003148) 9 VA Medical Facilities In the NCO 4 Service Area · NAICS 541380 Veterans Integrated Service Network 4 Service-Disabled Veteran-Owned Small Business
J065--Philips Imaging Equipment preventative maintenance and repair services at the Northport VA Medical Center NAICS 811210 Veterans Integrated Service Network 2 No set-aside
6515--NX EQ Image Guided Surgical Navigation System (VA-26-00083972) Nationwide · NAICS 334510 VA Strategic Acquisition Center Frederick Service-Disabled Veteran-Owned Small Business
J066--Azenta NitroStore System PM and Support Services NAICS 811210 Veterans Integrated Service Network 4 No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity