6515--36C242-25-AP-3692 | 528-25-3-8085-0085 Microwave Ablation System (VA-25-00090638)
Closed Pre-Solicitation Posted
Not the latest version. See the latest solicitation from .
- Solicitation number
- 36C24225Q0828
- Agency
-
Veterans Integrated Service Network 2
Veterans Health Administration, Department of Veterans Affairs
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 334517 Irradiation Apparatus Manufacturing
- Place of performance
- Albany Vamc 113 Holland Ave Albany, New York 12208, United States
- Points of contact
Notice details come from SAM.gov. Updated .
About this opportunity
The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 2 is seeking to procure a Microwave Ablation System for the Albany VA Medical Center's Interventional Radiology department. This procurement represents a technology upgrade from radiofrequency ablation to microwave ablation technology, requiring a standalone unit operating at 2.45 GHz frequency with 140-160W microwave power amplifiers. The system must include specific technical requirements such as two fully integrated peristaltic pumps, a 15.6-inch foldable touchscreen that operates with gloved fingers and sterile drapes, and dual probe support with independent channel controls. The unit must provide output power ranging from 0-90W for single channel operation and 0-160W for two-channel operation, with additional features including tract ablation capabilities, needle lock probe migration solution, and support for remote temperature probes. Responses to this Sources Sought Notice are due by July 16, 2025, at 12:00 pm Eastern Time under solicitation number 36C24225Q0828, with Corey Swinton serving as the contracting officer.
This procurement does not specify any set-aside designation and falls under NAICS code 334517 for Irradiation Apparatus Manufacturing and PSC code 6515 for Medical and Surgical Instruments, Equipment, and Supplies. The solicitation does not identify any incumbent contractors or provide specific award values or budget ranges for this equipment purchase. The procurement focuses on acquiring a single microwave ablation unit that will be installed at the Albany VAMC facility located at 113 Holland Avenue in Albany, New York. The device specifications indicate it should accommodate coolant capacity of 1000mL or greater using sterile water or 0.9% saline solution, and the system is classified as non-critical RME (Reusable Medical Equipment) with plastic housing construction. The solicitation will remain archived for 99 days following the response deadline, and interested parties can contact the contracting officer at Corey.Swinton@va.gov for additional information.
Notice text
2
1
Statement of Work
Microwave Ablation unit
For Interventional Radiology Albany VA Medical Center
6/11/2025
GENERAL INFORMATION
The general scope of this procurement is to purchase a Microwave Ablation unit
Equipment Purchase
SCOPE
Technology upgrade; replacing radiofrequency ablation with microwave ablation technology.
REQUIREMENT/SALIENT CHARACTERISTICS/SPECIFICATIONS
Common Nomenclature: Microwave Ablation
Kind of material: Plastic housing, Single stand along unit, non-critical RME.
Electrical Data: 140 -160W microwave power amplifiers operating at 2.45 GHz
Dimensions, size, capacity: N/A
Principles of operation: Microwave energy refers to a specific subset of the radio frequency (RF) spectrum operating between 300 MHz to 300 GHz. This device uses probes to direct microwave pulses to energize targeted tissues to treat them, especially cancer cells. The ablation unit to be procured in this requirement should have an output power of 0-80W for a single channel and 0-160W for a 2 channel.
Technical Requirements:
Component
Specification
Pumps
Two (2) fully integrated peristaltic pumps
Touchscreen Monitor
15.6-inch foldable touchscreen
Swivels
180-degrees CW/CCW
Operable with gloved fingers
Yes
Operable with sterile drape
Yes
Input / Output
One (1) USB port and one (1) Ethernet port
Tract Ablation
Cauterizes needle tract
Needle Lock
Probe migration solution
Coolant
1000mL or greater
Sterile water or 0.9% saline (user provided)
IntelliBlate Microwave Ablation System Instructions for Use 19
Performance Characteristics
MWA Generator Frequency
2.45 GHz ± 50 MHz
Applied Part Type
Type BF Dual probe support and independent channel controls. Two remote temperature probe support.
Output power (tip referenced)
Single Channel: 0-90W
Two (2) Channel:
0-80W (each), 0-160W (linked)
Microwave Output Power Accuracy
Within 10 seconds of activation of output energy on the touchscreen
Power (tip referenced)
Files
| File | Type | Posted |
|---|---|---|
| 36C24225Q0828.docx | DOCX document |
Notice history
| Notice | Type | Posted |
|---|---|---|
| 6515--36C242-25-AP-3692 | 528-25-3-8085-0085 Microwave Ablation System | Solicitation | |
| 6515--36C242-25-AP-3692 | 528-25-3-8085-0085 Microwave Ablation System | Pre-Solicitation | |
| This notice | Pre-Solicitation |
Similar open opportunities
| Opportunity | Agency | Set-aside | Responses due |
|---|---|---|---|
| Q525--LASER SERVICES - BUFFALO UROLOGY Holmium (30 and 80 Watt) Lasers CO2 Laser Lithotripsy (ESWL) Laser | Veterans Integrated Service Network 2 | No set-aside | |
| 6515--Medtronic IPC Instruments | Veterans Integrated Service Network 16 | No set-aside | |
| Neuro Micro Instruments | Veterans Integrated Service Network 16 | No set-aside |
On GovTribe
Work this opportunity
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity