36C24225Q0828.docx

DOCX document 23 KB Posted

Attached to
6515--36C242-25-AP-3692 | 528-25-3-8085-0085 Microwave Ablation System (VA-25-00090638)
Solicitation number
36C24225Q0828
Issued by

About this file

This document is a Sources Sought Notice for a Microwave Ablation System solicitation by the Department of Veterans Affairs (VA) at the Samuel S. Stratton VA Medical Center in Albany, NY. The procurement seeks to upgrade medical technology by replacing radiofrequency ablation with microwave ablation technology for Interventional Radiology. The specific requirement is for a standalone microwave ablation unit with technical specifications including a 2.45 GHz frequency, 15.6-inch foldable touchscreen, two integrated peristaltic pumps, and output power ranging from 0-90W for a single channel and 0-160W for two channels.

Key solicitation details include Solicitation Number 36C24225Q0828, with responses due by 07-16-2025 at 12:00 pm Eastern Time. The product service code is 6515, and the NAICS code is 334517. The place of performance is the Albany VAMC, located at 113 Holland Avenue in Albany, New York. The contracting officer is Corey Swinton, who can be contacted at Corey.Swinton@va.gov. The solicitation will be archived for 99 days after the response date, and the procurement does not specify any specific set-aside status.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SUBJECT*

36C242-25-AP-3692 | 528-25-3-8085-0085 Microwave Ablation System (VA-25-00090638)

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
12208-3410
SOLICITATION NUMBER*
36C24225Q0828
RESPONSE DATE/TIME/ZONE
07-16-2025 12:00 pm EASTERN TIME, NEW YORK, USA
ARCHIVE
99 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
6515
NAICS CODE*
334517

CONTRACTING OFFICE ADDRESS

Department of Veterans Affairs Samuel S . Stratton VA Medical Center 113 Holland Avenue (2-90NCA) Albany NY 12208-3410

POINT OF CONTACT*

Contracting Officer Corey Swinton Corey.Swinton@va.gov

PLACE OF PERFORMANCE

ADDRESS
Albany VAMC

113 Holland Ave

Albany New York

POSTAL CODE
12208
COUNTRY
USA

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

Sources Sought Notice Sources Sought Notice

DESCRIPTION

*= Required Field
Sources Sought Notice

Sources Sought Notice

Statement of Work Microwave Ablation unit For Interventional Radiology Albany VA Medical Center 6/11/2025

GENERAL INFORMATION

The general scope of this procurement is to purchase a Microwave Ablation unit

Equipment Purchase

SCOPE

Technology upgrade; replacing radiofrequency ablation with microwave ablation technology.

REQUIREMENT/SALIENT CHARACTERISTICS/SPECIFICATIONS

Common Nomenclature: Microwave Ablation

Kind of material: Plastic housing, Single stand along unit, non-critical RME.

Electrical Data: 140 -160W microwave power amplifiers operating at 2.45 GHz

Dimensions, size, capacity: N/A

Principles of operation: Microwave energy refers to a specific subset of the radio frequency (RF) spectrum operating between 300 MHz to 300 GHz. This device uses probes to direct microwave pulses to energize targeted tissues to treat them, especially cancer cells. The ablation unit to be procured in this requirement should have an output power of 0-80W for a single channel and 0-160W for a 2 channel.

Technical Requirements:

Component
Specification
Pumps
Two (2) fully integrated peristaltic pumps
Touchscreen Monitor
15.6-inch foldable touchscreen
Swivels
180-degrees CW/CCW
Operable with gloved fingers
Yes
Operable with sterile drape
Yes
Input / Output
One (1) USB port and one (1) Ethernet port
Tract Ablation
Cauterizes needle tract
Needle Lock
Probe migration solution
Coolant
1000mL or greater

Sterile water or 0.9% saline (user provided)

IntelliBlate Microwave Ablation System Instructions for Use 19

Performance Characteristics

MWA Generator Frequency
2.45 GHz ± 50 MHz
Applied Part Type
Type BF – Dual probe support and independent channel controls. Two remote temperature probe support.
Output power (tip referenced)
Single Channel: 0-90W
Two (2) Channel:
0-80W (each), 0-160W (linked)
Microwave Output Power Accuracy
Within 10 seconds of activation of output energy on the touchscreen
Power (tip referenced) <25W:
± 20%
Ximitry Probe Antenna Temperature Measurement
10-60°C (±3°C accuracy), 60-85°C (±6°C accuracy), 85-150°C (±10°C accuracy)
Temperature Probe Measurement
10-60°C (±2°C accuracy), 60-85°C (±5°C accuracy), 85-150°C (±10°C accuracy)
Weight (without Cart)
38 lbs. (17.2 kg)
Weight (with Cart)
95 lbs. (43 kg)
Dimensions (without Cart)
17.5 in. x 14.8 in. x 8.2 in. (445mm x 375 mm x 208mm)
Dimensions (with Cart)
W – 24 in. (609.6 mm), D – 22 in. (558.8 mm), H – 43.3 in. (1100 mm) Touchscreen closed

H – 52.75 in. (1340 mm) Touchscreen open (max)

Electrical Power Connection
100VAC-240VAC, 50Hz/60Hz, 900 VA
Power Cord Length
25 ft. (7.62 m)

Intended use: Ablation Procedures with Interventional Radiology and Surgery

Equipment with which the item is used, Job related tools: single use disposable probes

End item application: Ablation during procedures

Original equipment manufacturer part number: IB-CON

Other pertinent information that describes the item, material or service required:

The system must include the following components for acceptance:

1. BioMed

OTHER UNIQUE REQUIREMENTS

Warranty: 1 year from date of purchase Training: On-site orientation Safety/ISO/IT/BioMed/SPS & Other Departmental Compliance: Connects to CT and PACS

DELIVERY

· Contractor must deliver all components of the system as stated in this solicitation within 30 days of award notice.

· Contractor must deliver all items in this solicitation to -

· Upon receipt, the Government will inspect items for condition and working operation and acknowledge condition within 1 week of receipt.

· Standard FOB Destination is required

Page 1 of

Page 1 of

File details come from the government source that posted it. Updated .