The Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 21 is issuing a sources sought notice for two Helmer iPR245-GX i.SeriesA Pharmacy Refrigerators (44.9 cubic feet) for the VA Manila Outpatient Clinic, with responses due by August 19, 2026. This is a planning notice only; responses do not constitute a solicitation and will not generate reimbursement.

Equipment must meet specified performance and certification requirements: temperature range of +2°C to +10°C with +5°C set point; electrical supply options of 115V 60Hz or 220-240V 50-60Hz; QPS certification to UL and CSA standards; compliance with IEC/UL61010-2-011:2016 and IEC 61010-1:2010 AMD1:2016; ETL certification to NSF/ANSI 456 Vaccine Storage Standard; and ENERGY STAR certification. The refrigeration system must use a VCC compressor-based, variable-speed design (1,300–4,000 rpm) with automatic defrost, R600a refrigerant (EPA SNAP compliant), air-cooled condenser, and operate reliably in ambient conditions of +15°C to +32°C. Control systems include a 7-inch color touchscreen interface with microprocessor management, lockable doors with password protection, and RTD stainless steel sensor control. Brand name or equal offerings are acceptable. The notice seeks to identify potential sources capable of meeting these requirements for procurement planning purposes.

Notice text

SOURCES SOUGHT NOTICE BRAND NAME OR EQUAL - Helmer iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft. for VA Manila Outpatient Clinic

This is a Sources Sought Notice issued solely for information and planning purposes ONLY. It does not constitute a solicitation. Responses to this notice will assist the Government in identifying potential sources capable of providing the required equipment. No reimbursement will be made for any costs associated with providing information in response to this notice.

The Department of Veterans Affairs, Manila Outpatient Clinic is seeking potential sources capable of providing two (2) units Helmer iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft, (Brand name or Equal).

The essential/significant physical, functional or performance characteristics of the requirement are:

Application, Rating, & Electrical Data

Application

High performance storage of medical and scientific products

Storage Volume

44.9 cu ft | 1271 L

Temperature Range

+2°C to +10°C

Set Point

+5°C

Electrical Power

115V 60Hz | 220-240V 50-60Hz

Certification/Agency Listing

QPS (Certified to UL and CSA Standards)

IEC/UL61010-2-011: 2016 | IEC 61010-1:2010, AMD1:2016 ETL certified to NSF/ANSI 456 Vaccine Storage Standard

Energy Star

Yes, ENERGY STAR® Certified High Performance Refrigerator

Ambient Operating Temperature

+15°C to +32°C | +59°F to +90°F

Refrigeration

Refrigeration System

VCC compressor-based, forced-air hydrocarbon

Compressor

Hermetic, variable speed (VCC). Rated speed range: 1300-4000 rpm

Condenser Type

Air-cooled fin and tube, sealed

Defrost Method

Automatic

Refrigerant

R600a, Isobutane, EPA SNAP compliant

Controller / Configuration Settings

User Interface

i.C3® color touchscreen, door mounted, 7in | 177mm, °C or °F

Power Switch

On/Off

Controller Type

Microprocessor with touchscreen input & display

Security

Lockable doors, password protected settings

Control Sensor

RTD, stainless steel

iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft (1271 Liters) (with NSF/ANSI 456 Vaccine Storage Certified Labeling and Certificate of Calibration

Communication Ports

USB, Remote alarm contacts, RJ45 (Ethernet). API available for Ethernet connectivity development

Power Failure Alarm

Yes

High/Low Alarms - Tests

Yes, fully adjustable, automatic (Peltier) testing

Door Ajar Alarm

Yes, fully adjustable

Condenser Temperature Alarm

Yes, fully adjustable

Min/Max Temperature

Yes, display and reset

Download

Yes, via USB. CSV and PDF

Event Log

Yes, all alarms, door openings, download via USB

Temperature Log

Yes, download via USB

Interactive Temperature Graph

Yes

Temperature Monitoring Ballast

Yes, two 4 oz bottles with screw caps, solid ballasts optional

Battery Back-up

Yes, 20 hr. display & alarm back up for i.C3® touchscreen | Rechargeable 12V, lead acid

Interior Storage (w x h x d)

2 ventilated, powder coated shelves, adjustable

22.5 x 21.9 in | 572 x 557 mm, 100 lb | 46 kg max capacity/shelf

12 ventilated, powder coated, pull-out drawers, adjustable

21.5 x 3.9 x 22 in | 546 x 100 x 560 mm, 100 lb | 46 kg max capacity/drawer

iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft (1271 Liters) (with NSF/ANSI 456 Vaccine Storage Certified Labeling and Certificate of Calibration

Options | Accessories

Stainless steel interior, Chart recorder, Chart paper, Leveling feet, Solid ballast, Remote alarms, Floor and Wall bracket kit, Seismic anchoring kit, Storage trays, Screen protector, Locking drawer, Solid or heated glass doors, Top cover, Side-wall access ports, Temperature validation, Validation guide (IQ/OQ), Extended warranty

Certificate of Calibration

Yes, generated with ISO 17025 certified, NIST traceable thermometer

Warranty

Rel.i Plus Warranty - 7 years compressor, 2 years parts, 1 year labor

Outside US and Canada - Contact your distributor for warranty information

The anticipated NAICS code for this requirement is 333415 Air-Conditioning and warm Air Heating Equipment and Commercial and Industrial Refrigeration Equipment Manufacturing with Size Standard of 1,250 employees.

No set-aside determination has been made at this time. The Government is seeking information to support acquisition planning and any applicable socioeconomic determination.

Only firms interested and capable should send their information and any questions by email to Cristina.lansangan@va.gov no later than 4:00 PM Pacific Standard Time, August 19, 2026.

Please provide the following information in your response:

Company Name and UEI

Point of Contact: Name, Phone Number and E-mail address

SBA Socio-economic Status

Capability Statement

Indicate whether this equipment is available under GSA/FSS/NAC contract, including contract number and expiry date

Authorized Distributor Letter (if applicable)

Estimated Price

DISCLAIMER

This Sources Sought notice is issued solely for information and planning purposes only and does not constitute a solicitation. All information received in response to this request that is marked proprietary will be handled accordingly. The Government will not compensate an offeror for providing information in response to this notice. Furthermore, the Government reserves the right to cancel this requirement at its discretion.

Files

File Type Posted
36C26126Q1003.docx DOCX document

Similar open opportunities

Opportunity Agency Set-aside Responses due
REFRIGERATOR-FREEZER, and 6 more items Medical Supply Chain Total Small Business
36C25227Q0048 - coolers and freezer North Chicago, Illinois 60064, United States · NAICS 333415 Veterans Integrated Service Network 12 Service-Disabled Veteran-Owned Small Business Sole Source
7125--Sterile Storage Cabinets VA Greater Los Angeles (GLA) Healthcare System Los Angeles CA 90073, United States · NAICS 339113 Veterans Integrated Service Network 22 No set-aside
Refrigeration unit NAICS 333998 Naval Supply Systems Command No set-aside
Condenser, refrigeratio NAICS 333415 Construction and Equipment No set-aside

On GovTribe

Work this opportunity

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity

GovTribe for Federal Contractors