36C26126Q1003.docx

DOCX document 19 KB Posted

Attached to
4110--Helmer iPR245-GX i. Series Pharmaceutical Refrigerator (Brand Name or Equal)
Solicitation number
36C26126Q1003
Issued by

About this file

This is a Sources Sought Notice issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 21 for information and planning purposes only. The notice seeks potential sources capable of providing two (2) units of Helmer iPR245-GX i.SeriesA Pharmacy Refrigerators with a capacity of 44.9 cubic feet (1,271 liters) for the VA Manila Outpatient Clinic. The solicitation allows for brand name or equal equipment. Responses to this notice are for planning purposes only and will not result in reimbursement for any costs associated with providing information.

The essential specifications for the required refrigerators include a temperature range of +2°C to +10°C with a +5°C set point, 115V 60Hz or 220-240V 50-60Hz electrical power, and high-performance storage capabilities for medical and scientific products. The equipment must be QPS certified to UL and CSA standards, comply with IEC/UL61010-2-011:2016 and IEC 61010-1:2010 AMD1:2016, and be ETL certified to NSF/ANSI 456 Vaccine Storage Standard. Additional requirements include ENERGY STAR certification, a VCC compressor-based refrigeration system with variable speed capability (1,300-4,000 rpm), automatic defrost, R600a refrigerant, a 7-inch color touchscreen user interface with microprocessor control, lockable doors with password protection, and stainless steel RTD sensor control with an ambient operating temperature range of +15°C to +32°C.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sources Sought Notice Sources Sought Notice

SUBJECT*
Helmer iPR245-GX i.Series Pharmaceutical Refrigerator (Brand Name or Equal)

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
95652-2613
SOLICITATION NUMBER*
36C26126Q1003
RESPONSE DATE/TIME/ZONE
08-19-2026 4:00pm PACIFIC TIME, LOS ANGELES, USA
ARCHIVE
15 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
4110
NAICS CODE*
333415
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

VA Sierra Pacific Network (VISN 21) VA Northern California HealthCare System 3237 Peacekeeper Way Bldg 200 Suite 204 McClellan CA 95652-2613

POINT OF CONTACT*

Contracting Officer Cristina Mae Lansangan cristina.lansangan@va.gov

PLACE OF PERFORMANCE

ADDRESS
Department of Veterans Affairs

Manila Outpatient Clinic

1501 Roxas Boulevard

Pasay City

POSTAL CODE
1302
COUNTRY
Philippines

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS

EMAIL DESCRIPTION

DESCRIPTION

SOURCES SOUGHT NOTICE – BRAND NAME OR EQUAL - Helmer iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft. for VA Manila Outpatient Clinic

This is a Sources Sought Notice issued solely for information and planning purposes ONLY. It does not constitute a solicitation. Responses to this notice will assist the Government in identifying potential sources capable of providing the required equipment. No reimbursement will be made for any costs associated with providing information in response to this notice.

The Department of Veterans Affairs, Manila Outpatient Clinic is seeking potential sources capable of providing two (2) units Helmer iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft, (Brand name or Equal).

The essential/significant physical, functional or performance characteristics of the requirement are:

Application, Rating, & Electrical Data

Application
High performance storage of medical and scientific products
Storage Volume
44.9 cu ft | 1271 L
Temperature Range
+2°C to +10°C
Set Point
+5°C
Electrical Power
115V 60Hz | 220-240V 50-60Hz
Certification/Agency Listing
QPS (Certified to UL and CSA Standards)

IEC/UL61010-2-011: 2016 | IEC 61010-1:2010, AMD1:2016 ETL certified to NSF/ANSI 456 Vaccine Storage Standard

Energy Star
Yes, ENERGY STAR® Certified High Performance Refrigerator
Ambient Operating Temperature
+15°C to +32°C | +59°F to +90°F

Refrigeration

Refrigeration System
VCC compressor-based, forced-air hydrocarbon
Compressor
Hermetic, variable speed (VCC). Rated speed range: 1300-4000 rpm
Condenser Type
Air-cooled fin and tube, sealed
Defrost Method
Automatic
Refrigerant
R600a, Isobutane, EPA SNAP compliant

Controller / Configuration Settings

User Interface
i.C3® color touchscreen, door mounted, 7in | 177mm, °C or °F
Power Switch
On/Off
Controller Type
Microprocessor with touchscreen input & display
Security
Lockable doors, password protected settings
Control Sensor
RTD, stainless steel

iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft (1271 Liters) (with NSF/ANSI 456 Vaccine Storage Certified Labeling and Certificate of Calibration

Communication Ports
USB, Remote alarm contacts, RJ45 (Ethernet). API available for Ethernet connectivity development
Power Failure Alarm
Yes
High/Low Alarms - Tests
Yes, fully adjustable, automatic (Peltier) testing
Door Ajar Alarm
Yes, fully adjustable
Condenser Temperature Alarm
Yes, fully adjustable
Min/Max Temperature
Yes, display and reset
Download
Yes, via USB. CSV and PDF
Event Log
Yes, all alarms, door openings, download via USB
Temperature Log
Yes, download via USB
Interactive Temperature Graph
Yes
Temperature Monitoring Ballast
Yes, two 4 oz bottles with screw caps, solid ballasts optional
Battery Back-up
Yes, 20 hr. display & alarm back up for i.C3® touchscreen | Rechargeable 12V, lead acid
Interior Storage (w x h x d)
2 ventilated, powder coated shelves, adjustable

22.5 x 21.9 in | 572 x 557 mm, 100 lb | 46 kg max capacity/shelf 12 ventilated, powder coated, pull-out drawers, adjustable

21.5 x 3.9 x 22 in | 546 x 100 x 560 mm, 100 lb | 46 kg max capacity/drawer iPR245-GX i.Series® Pharmacy Refrigerator, 44.9 cu ft (1271 Liters) (with NSF/ANSI 456 Vaccine Storage Certified Labeling and Certificate of Calibration

Options | Accessories
Stainless steel interior, Chart recorder, Chart paper, Leveling feet, Solid ballast, Remote alarms, Floor and Wall bracket kit, Seismic anchoring kit, Storage trays, Screen protector, Locking drawer, Solid or heated glass doors, Top cover, Side-wall access ports, Temperature validation, Validation guide (IQ/OQ), Extended warranty
Certificate of Calibration
Yes, generated with ISO 17025 certified, NIST traceable thermometer
Warranty
Rel.i™ Plus Warranty - 7 years compressor, 2 years parts, 1 year labor

Outside US and Canada - Contact your distributor for warranty information

The anticipated NAICS code for this requirement is 333415 – Air-Conditioning and warm Air Heating Equipment and Commercial and Industrial Refrigeration Equipment Manufacturing with Size Standard of 1,250 employees.

No set-aside determination has been made at this time. The Government is seeking information to support acquisition planning and any applicable socioeconomic determination.

Only firms interested and capable should send their information and any questions by email to Cristina.lansangan@va.gov no later than 4:00 PM Pacific Standard Time, August 19, 2026.

Please provide the following information in your response:

1. Company Name and UEI

2. Point of Contact: Name, Phone Number and E-mail address

3. SBA Socio-economic Status

4. Capability Statement

5. Indicate whether this equipment is available under GSA/FSS/NAC contract, including contract number and expiry date

6. Authorized Distributor Letter (if applicable)

7. Estimated Price

DISCLAIMER

This Sources Sought notice is issued solely for information and planning purposes only and does not constitute a solicitation. All information received in response to this request that is marked proprietary will be handled accordingly. The Government will not compensate an offeror for providing information in response to this notice. Furthermore, the Government reserves the right to cancel this requirement at its discretion.

*= Required Field
Sources Sought Notice

Sources Sought Notice

File details come from the government source that posted it. Updated .