Y24-612-KS.pdf
PDF 162 KB Posted
- Attached to
- Inductively Coupled Plasma Mass Spectrometer
- Solicitation number
- Y24-612-KS
- Issued by
- Orange County, Orlando City, Florida
About this file
This document is a detailed specifications document for an Inductively Coupled Plasma Mass Spectrometer (ICP-MS) solicitation by Orange County Utilities Central Laboratory in Florida. The county is seeking to replace its existing Agilent 7700 series ICP-MS with a state-of-the-art benchtop instrument for analyzing trace metals in environmental samples, including potable waters, environmental waters, and bio-solids. The solicitation allows for three pre-approved instrument models (Perkin Elmer NexION 2200, ThermoScientific RQPlus, and Agilent 7900 ICP-MS) and is open for bids until June 13, 2024, with electronic submissions required through the ProcureNow OpenGov e-Procurement Platform.
The specifications outline comprehensive technical requirements for the instrument, including compliance with USEPA methods, specific performance metrics, computer hardware and software specifications, and additional support provisions. The county seeks a fully automated system with advanced features such as mass spectrometry capabilities, comprehensive software control, and compatibility with existing laboratory systems. Optional annual service and software support agreements are available, and the vendor will be required to provide installation, testing, and training at the Orange County Utilities Central Laboratory, including establishing instrument and method detection limits and performing various calibration procedures.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Inductively_Coupled_Plasma_Mass_Spectrometer.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SPECIFICATIONS
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETER
ICP-MS)
The Orange County Utilities Central Laboratory is soliciting bids for one (1) state-of-the-art Inductively Coupled Mass Spectrometer (ICP-MS) computer-controlled bench top instrument that offers maximum performance and versatility. This instrument will be used for analysis of trace metals in environmental samples, including potable waters, environmental waters, and bio-solids. Due to space limitations, the County requires a benchtop instrument. The ICP-MS instrument with a dedicated data system shall meet or exceed the specifications listed below.
The Perkin Elmer NexION 2200, the ThermoScientific RQPlus and the Agilent 7900 ICP-MS have been approved by the County as acceptable. This bid is for the replacement of the County’s existing Agilent 7700 series ICP-MS installed in 2015. Trade-in discount may be considered pending approval by the Orange County BCC. All other equivalent bids shall meet or exceed the minimum salient specifications listed below:
A. INSTRUMENT SPECIFICATIONS
1. Fully compliant with USEPA 200.8 & SW846 6020 methods and able to meet all
QA/QC EPA method requirements for 200.8 and SW846 6020. Detection limits should be at or below the suggested MDLs indicated in each method.
2. Bench-top design with simultaneous solid-state detector and a mainframe footprint not to exceed 35"W x 35"D (this only encompasses the instrument, excluding data station, autosampler and chiller).
3. Instrument much be equipped with a quadrupole (single or triple ICP-MS).
4. Mass spectrometer should have ~2 to 240 unit mass range and ~1.0-unit mass resolution or comparable mass range and resolution.
5. Mass spectrometer should have abundance sensitivity (at Cs) at
~10-7 or comparable abundance sensitivity.
6. The dynamic range of the detector should be ~109 or greater.
The instrument will use helium or similar collision gas for operation.
7. An extra set of consumables must be included.
8. The instrument must have an uninterrupted power supply included.
9. Instrument should operate in an ambient lab temperature environment
(approximately 59 to 95oF).
10. Software must fully control all ICP-MS components.
11. Computers must be provided with the required software to run instrument(s), process the acquired data, generate a customer report and be compatible with the in-house
LIMS.
12. Instrument software must contain data acquisition, qualitative, and quantitative analyses.
13. Instrument software will allow for stand-alone data processing.
14. Must provide a library/database of the mass spectra in the field of environmental science.
15. Detector to allow for full frame imaging.
16. Mass-flow controllers for gas flows.
17. The RF generator shall have an output power range of 750-1600 watts.
18. The generator shall have a 75% minimum power transfer efficiency with a 0.l minimum power output stability.
19. Shielding from stray UV and RF emissions
20. Safety interlocks shall be monitored continuously and if any interlock is breached, the plasma will shut down.
21. Interlocks that prevent damage to the RF Generator and instrument in the event of power failure.
22. Sample tray capacity shall be >80 samples.
23. Closed coupled, three-channel integral peristaltic pump, computer controlled, must be integrated within the ICP-MS system.
24. The system must be interface able with ESI Auto samplers, prepFast and Fast systems
(quoted). Specifications below.
ESI prepFAST
24.1 Auto calibration from a single stock solution
24.2 Variable in-line dilution performed with a mixing valve up to 400x.
24.3 QC Auto dilution of calibration over range or internal standard out of range.
24.4 The FAST system offers many advantages for analyzing a variety of sample types and ICPMS including:
24.4.1 Lowers reagent cost by using lower concentrations of internal standards.
24.4.2 Better precision.
24.4.3 Lower cost of consumables and routine maintenance: torch, nebulizer, cones, pump oil.
24.4.4 Reduced deposition on cones, etc.- due to smaller sample sizes minimizes drift.
24.4.5 Lower facilities cost with shorter run times for less consumable usage-peripump tubing, Ar gas, etc.
24.4.6 Improves sensitivity.
24.4.7 Lowers oxides
24.4.8 Cleaner, faster washout, reduces sample carry-over.
24.4.9 Eliminates peristaltic pump memory effects by sample never touching peripump tubing.
24.4.10 Dual rinse station improves rinse out and reduces blank contamination.
24.4.11 Fewer QC failures.
24.4.12 Faster stabilization time.
24.4.13 Optional online constant dilution with micro-peripump.
24.4.14 Can add volatile analytes such as Hg to ICPMS methods.
24.4.15 High speed sample loading minimizes sample uptake time.
24.4.16 Supports online internal standard addition.
24.4.17 Pre-loaded FAST and ICPMS methods for rapid implementation included.
24.4.18 Modular design enables multi-application flexibility (with plug and play ease-of-set-up).
24.5 The prepFAST offers many advantages as well to sample preparation and analysis including:
24.5.1 Software control and functionality: the prepFAST software allows for flexibility in method development due to the unique functions available in the software.
24.5.2 Multiple syringe system: The prepFAST utilizes the S400V which contains multiple syringe modules with various barrel volumes for precise dilution up to 400X.
24.5.3 Priority dilution: After a run of samples, the user can simply select the over-range samples from the list and then rerun without moving samples, rewriting shorter methods, etc.
24.5.4 Auto calibration: The prepFAST allows for calibration from a high standard with the user simply selecting the calibration points they desire and the prepFast does the rest.
24.5.5 Inline dilution: The prepFAST mixes/dilutes the sample inline thereby eliminating the need to take an aliquot of the sample and diluents and deposits it into another tube that could be the source of con-tamination.
24.5.6 The prepFAST automatically mixes the sample with the internal standard accurately.
25. Recirculator shall not be mounted directly on the instrument chassis.
26. Capability to interface to laser ablation or speciation instrumentation at a future date.
No external instrument is to be bid on as part of this IFB.
B. COMPUTER HARDWARE & SOFTWARE SPECIFICATIONS
1. Software shall be based on the Windows 10 operating system and shall provide full control of all instrument functions including gas flows, viewing position, plasma ignition, and monitoring of safety interlocks. The spectrometer shall be controlled by an industry-standard computer complete with a 19" monitor.
2. Fully automated system start-up and shutdown via instrument control window.
3. Full instrument control with the ability to display parameters and instrument response in real time.
4. Software shall provide a complete comprehensive mass library with preferred mass for element and have the capability of automatic spectral interference identification for selected masses.
5. Software shall have the capability to store, recall and display all spectra for all samples analyzed.
6. Software shall have the capability of reprocessing data post-run.
7. Software shall be multi-tasking.
8. Software shall be able to meet all QA/QC method requirements for 200.8 and 6020.
9. Software shall comply with all applicable NELAP/NELAC requirements.
10. Network compatible Laser jet printer with 500 sheet paper tray.
C. SUPPORT
1. Provide technical telephone support Monday through Friday, 8:00 A.M. through 5:00
P.M. at no additional charge to the County.
2. Provide on-site service call requests by a factory-trained service engineer on site within 72 hours when trouble shooting over the telephone is ineffective at no additional charge to the County.
D. INSTALLATION/TESTING/TRAINING
1. Installation shall be performed by a factory-trained service engineer.
Immediately upon initial installation, at no additional cost, and prior to final acceptance by the County, the Vendor shall simultaneously provide a minimum of three (3) days, eight (8) hours per day, of on-site instrument and software operation training at Orange County Utilities Central Laboratory and instrument testing which shall include, but not be limited to, completing testing per Method 200.8, Rev. 5.4 as follows and SW 846-6020:
Establish Instrument Detection Limits Establish Method Detection Limits Determine Linear Dynamic Range Linear least squares Weighted linear least squares.
Linear forced-through-zero Method of standard additions Additions calibration
E. OPTIONS
1. ANNUAL SERVICE AGREEMENT OPTION - to be effective upon expiration of original warranty. The Annual Service Agreement shall fully cover all areas covered in the original warranty including telephone technical support and 72-hour response on-site repair at no additional cost. If applicable, Contractor shall include copies of proposed Annual Service Agreement with bid submittal.
The County reserves the right to accept, reject or alter any proposed agreement.
2. ANNUAL SOFTWARE SUPPORT AGREEMENT OPTION - to be effective upon expiration of original warranty. The Annual Software Support Agreement shall fully cover all updates and telephone support Monday through Friday, 8:00 A.M. through 5:00 P.M. If applicable, the Contractor shall include copies of proposed Annual Software Agreement with bid submittal. The County reserves the right to accept, reject or alter any proposed agreement.
3. Lift gate required upon delivery.
4. Inside delivery required.
File details come from the government source that posted it. Updated .