This document is a Revised Salient Characteristics specification for the Anritsu MS2085A Site Master Cable and Antenna Analyzer, a highly portable, next-generation handheld device designed for field installation and maintenance of RF communication systems. The required configuration must include the MS2085A base unit with Option 21 (2-port Transmission Measurement) and Option 804 (CAA Frequency Range 5 kHz to 4 GHz), along with specified accessories: the MS2085A-0804-0099 Premium Calibration option, TOSLN50A-8 Coaxial Calibration Kit, 15RNFN50-1.5-R Test Port Extension Cable, and power cord. The specification explicitly clarifies that Option 15 (Vector Voltmeter) must not be included, Frequency Options 904 and 906 are incorrect, and the 6 GHz frequency option (Option 806) is not required despite being mentioned in the related solicitation.

The MS2085A features a large 10-inch high-resolution capacitive multi-touch screen with IK08 impact protection, weighs 3.8 kg with Option 21 installed, and supports reflection and transmission measurements with a frequency range of 5 kHz to 4 GHz, frequency accuracy of 2.5 ppm, and dynamic range exceeding 100 dB up to 3 GHz. Key capabilities include 1-Port ReadyCal calibration support, Return Loss, VSWR, Cable Loss, Smith Chart analysis, TDR, Distance-to-Fault measurements, 2-port transmission measurements (S21), rugged Type-N RF ports (50 Ω impedance), Gigabit Ethernet and USB connectivity, approximately 5-6 hours battery life, and environmental compliance with MIL-PRF-28800F Class 2 operating standards and IP52 ingress protection. The related federal contract opportunity, Solicitation W9124B27QA001 issued by the Department of the Army Materiel Command Mission and Installation Contracting Command Fort Hood, has response due dates listed as November 30, 2027 at 12 pm PST and December 8, 2026 at 12 pm PST.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications or Salient Characteristics for an Anritsu MS2085A Site Master Cable and Antenna Analyzer The Anritsu MS2085A is a highly portable, next-generation handheld Cable and Antenna Analyzer (CAA) designed for field installation and maintenance of RF communication systems. The required configuration covers a frequency range of 5 kHz to 4 GHz and includes 2-port Vector Network Analyzer (VNA) transmission capability.

Required Configuration Quotes shall be based on the following exact configuration (matching the validated Government requirement):

1. MS2085A – Site Master Cable and Antenna Analyzer (base unit)

1. MS2085A-0021 (Option 21) – 2-port Transmission Measurement – REQUIRED

1. MS2085A-0804 (Option 804) – CAA Frequency Range 5 kHz to 4 GHz – REQUIRED

1. MS2085A-0804-0099 (Option 99) – Premium Calibration

1. TOSLN50A-8 – Coaxial Calibration Kit, Type N(m), DC to 8 GHz, 50 Ω

1. 15RNFN50-1.5-R – Test Port Extension Cable, Armored, 1.5 meters, N(m)–N(f), 6 GHz, 50 Ω

1. Power cord Important clarifications:

1. Option 15 (Vector Voltmeter) is NOT required and shall not be included.

1. Frequency Options 904 and 906 are incorrect and shall not be used. The required frequency option is Option 804 only.

1. The 6 GHz frequency option (Option 806) is not required.

Salient Characteristics & Features

Characteristic
Details
Port Configuration
2-Port, 1-Path setup allowing standard reflection and transmission measurements (requires Option 21).
Display
Large 10-inch high-resolution capacitive multi-touch screen with IK08 impact protection.
Weight
3.8 kg (8.39 lb) with required Option 21 installed (manufacturer published specification).
Portability
Optimized for field technicians; highly portable handheld form factor.
Sweep Speed & Resolution
Fast sweep rate of 350 µs per data point with up to 10,049 maximum data points.
Split Screen View
Supports simultaneous dual-trace viewing (e.g., Return Loss and Distance-to-Fault).
Reporting
Built-in PDF/HTML report generator for on-site test result compilation.

Key Technical Specifications Frequency & Power

1. Frequency Range: 5 kHz to 4 GHz (Option 804 – required).

1. Frequency Accuracy: ≤ ±2.5 ppm (typical) with aging of ±1 ppm per year.

1. Frequency Resolution: 1 Hz.

1. Output Port Power: Settable in 0.1 dB steps.

VNA & Transmission Measurements (requires Option 21)

1. Dynamic Range (Transmission): >100 dB up to 3 GHz (typical >105 dB).

1. IF Bandwidth selections: 10 Hz to 100 kHz.

1. Calibration: Supports 1-Port ReadyCal, OSL, and FlexCal capabilities.

Supported Measurement Types

1. Reflection: Return Loss, VSWR, Cable Loss, Smith Chart, 1-Port Phase, TDR, and Distance-to-Fault (DTF).

1. Transmission (Option 21 – Required): 2-port transmission measurements (S21).

Connectivity & Environmental

1. RF Ports: Rugged Type-N (female), 50 Ω impedance.

1. Data Interfaces: RJ45 Gigabit Ethernet, USB 3 Type-A, and USB 3 Type-C (USBTMC).

1. Battery Life: Approximately 5–6 hours of continuous operation.

1. Environmental: Meets MIL-PRF-28800F Class 2 operating standards (−10 °C to 55 °C) and IP52 ingress protection when housed in soft case.