Questions_and_Answers_Solicitation_W911QX-16-T-0004.docx

DOCX document 12 KB Posted

Attached to
One (1) Silicon Carbide Horizontal Hot-Wall Chemical Vapor Deposition System (HWCVD) in accordance with the attached salient document.
Solicitation number
W911QX-16-T-0004
Issued by

About this file

Questions and answers for W911QX-16-T-0004

View the file

Other files

File Type Posted
A07_-_Solicitation_(Amendments).docx DOCX document
A07_-_Solicitation_Q As.xlsx XLSX spreadsheet
A01_-_Salient_Characteristics.doc DOC document
A07_-_Executive_Summary_Solicitation.docx DOCX document
A07_-_Solicitation_(Amendments).docx DOCX document
A01_-_DD1423.doc DOC document
A01_-_Salient_Characteristics.doc DOC document
A07_-_Evaluation_Criteria_(LPTA).doc DOC document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Questions and Answers Regarding Solicitation W911QX-16-T-0004

Question 1: Paragraph 5: The supply pressure given for different gases, like argon and carbon gas, is given as 30 psi. Is this psig (gauge pressure) or psia? (absolute pressure)?

Answer: This is psig.

Question 2: Paragraph 13: A reference is given to "input lines as detailed in C.1.5". To which document is this reference made?

Answer: C.1.5 was changed to paragraph 5.

Question 3: CLIN 0004 Option: Emergency power (UPS). Is the intention that the complete system, including the heating system, should be supplied during a powerfailure?

Answer: It is the intention that the UPS powers the control system so that it can return to a safe state. It is not the intention to power the heating system.