Sheet1

New NDT Cabinet X-Ray System
Salient characteristicsVendor AVendor BVendor CVendor DVendor E
System Capabilities17 axis
2Digital radiography computed tomography system
3Ability to scan a minimum of 60 inches in diameter and 60 inch length
4Fully shielded
5Interlocked cabinet style
6Must comply with applicable FDA 21CRF1020.40
7Flat panel detector
8225 kV Micro focus Tube
9450 HP Tube
10DR Imaging processing acquisition workstation & software
11Separate CT reconstruction workstation including software
System Specifics12Advanced 2D X-ray Inspection
132D Slice reconstruction
14CT volume reconstruction for 3D inspection
153D internal & external surface scanning
16Measurement and analysis tool for factors 12-15
X-ray Tube17Micro focus 225 kV directional tube voltage
1825-225 kV mA Range 0-3 mA Power 320 W max
19Focal spot size < 6 microns
20Detail recognition < 2 microns
2110 meter high tension cable
22Closed loop water cooler
23X-ray software must be at least Windows 10 based
24Automatic tube warm up
25Automatic filament current adjustment
26Automatic focusing
27Automatic target loading control
28Intensity control
29Spare parts & service kit included
450 kV HP Tube30450/11 metal ceramic tube head
31EN 12543 Spec. 0.4mm & 1.0mm dual focal spots
32Continuous rating: 700W/1.5mA; 1500W/3.3mA
335mm Beryllium (Be) window
34Target angle 11 degrees
35Radiation coverage 30 degrees x 40 degrees
36CF-450 High stability generators 1800W
37Digital controller with enclosure features 6 operating modes: Warm up, Radiographic, Fluoroscopic, Auto, Watt & Remote Control via RS232
38Up to 600 stored techniques
39I/O power distribution panel
402EA 10 meter high tension cables
41Oil to air cooler with flow & temperature interlocks
42All other associated/applicable cables, cooling lines & documentation
X-ray Tube Shutter 450 kV43Four leaf, two axis tube shutter shielded for 450 kV
44Independent axis control
45Laser alignment system
46Spare pack of 225kV filaments
47Spare 225kV targets
X-ray Detector484343N Gadox 17 inches x 17 inches (432mm x 432mm) image area
49DRZ+ (Gadox) Scintillator
50150 micron pixel pitch
512880 x 2880 pixel matrix
5216 bit A/D conversion
5315 fps (1 x 1); 30 fps (2x2 pixel binning)
54Premium CT grade tungsten shielded to 225kV (increased lifetime)
55All other associated/applicable power & Ethernet cables
Manipulator56Movement of the inspection object is limited to x, z, tilt & rotation, along with detector & tube vertical movement
57All drives are motorized with variable speed multi directional joy stick operation
58Each axis control is dependent allowing simultaneous movement of axis
59Tethered desktop control console
60Focal distance: 450kV tube to DDA -- 92 inches (239cm)
61Focal distance :450kV tube to LDA -- 95 inches (241 cm)
62225kV Micro focus directional tube to DDA -- 80 inches (203cm)
63Vertical axis detector -- 63 inches (160cm)
64Vertical Axis Tube --69 inches (175cm)
65Part manipulator -- X axis/front to back translation -- 36 inches (91cm)
66Part manipulator -- Z axis/tube to detector translation -- 65 inches (165cm)
67Part manipulator -- Z axis/tube to detector translation -- 65 inches (165cm)
68Tilt Range +20/-20 degrees
69CT precision rotation stage 360 degrees continuous Resolution 0.001 degrees
70DDA lateral shift 32 inches (81cm)
71Minimum 500lb (187kg) static capacity for part manipulator (not including tilt)
7218 inch nickel plated aluminum turn table platter, with integrated electronics
73Heavy duty welded & precision machined frame construction
74Precision linear bearing rail
75Vibration isolation system
Cabinet76FDA 21CFR 1020.40 compliance required
77Radiation shielding to energies of 450 kV constant potential
78Approximate dimensions: 111 inches deep x 137 inches tall x 161 inches wide (282cm deep x 347cm tall x 409cm wide)
79Steel/lead/steel construction
80Cable access port with cover
81Large motorized sliding access door
82Primary & secondary X-ray safety interlocks
83Interior lighting
84X-ray on warning light and audio alarm (minimum of 5 secs to alert)
85Designed to meet all PA State and Federal X-ray safety regulations, including FDA 21 CRF 1020.40 compliance
86Motion control module shall provide operator friendly teach/learn interface to be easily customized
87Additional 31 inch (787mm) 4K monitor (image review quality)
884 Camera Color surveillance system: Internal camera surveillance system with split screen monitors and 4 individual cameras (image review quality)
Computer Hardware & Software89Computer (stand alone, no network requirement) to operate the system
DR Work Station90Windows 10 Operating system
91Eight-Core Intel Xeon Processor
928 GB RAM
93512 GB Operating System Drive
943 TB Data storage in RAID 0 configuration
95DVD +/- RW drive
9610/100/1000 network interface
97Nvidia High Performance Display adaptor
9831.5 inch (800mm) 4K color display (image review quality)
991 EA wired keyboard & 1 EA wired mouse
DR Software100High performance image analysis system with a comprehensive set of image analysis tools -- Minimum ASTM E2737 & ASTM E2698 standards
101At least Windows 10 based
102NON PROPRIETARY image storage format
103Multiple windows interface
104High performance image processing & measurement functions utilizing GPU technology
105Automatic parallelization for systems with multiple CPUs and GPUs
106Live imaging -- up to 30 frames per second
107Live averaging
108Live histogram
109Live line profile
110Live rotation between portrait and landscape mode
111Live measurements with 1 pixel resolution
112Live image gain/offset calibration & defective pixel correction
113Automatic creation of technique sheet for operator record
114Capture video into AVI files
115Supports digital flat panel detectors LDAs & digital/analog cameras at 8, 10, 12 & 16 bits
116Read & store images in TIFF 32 bit/16 bit/ 8 bit BMP, JPEG, RAW, & Diconde
DR CT Acquisition Module117Automated CT acquisition capabilities
118Capable of CT acquisition up to 10,000 angle steps
119Automatic creation of technique sheet
120Monitor X-ray performance & will pause the CT scan if X-ray interruption is detected
121Acquisition of spiral CT spans
122Ring reduction acquisition capabilities
123Continuous fast CT acquisition module projections for fast CT scans from 5 seconds to minutes for thousands of projections
Reconstruction Work Station124Windows 10 Operating system
125Dual 10-core processors (total 20 cores)
126384 GB RAM
1271 TB SSD Operating system drive
12824 TB Hard drive with high speed raid 5 capabilities
1294 Nvidia High performance display adaptors for display and reconstruction
13015,360 total processing cores
13196 GB GPU memory
1323D rendering capabilities
13331.5 inch 4K display (image review quality)
1341 EA wired keyboard & 1 EA wired mouse
CT Computed Tomography Software135CT calibration, Cone beam reconstruction, 3D real time visualization & analysis tools in the same interface
136No limiting on the size of the reconstruction
137Combines 2 or more datasets taken at different current settings
138Significantly improves CT quality of widely varying densities
139Increases the dynamic range of the CT dataset
140Reduces noise & photon starvation in the CT reconstruction
141Capable to perform local CT reconstructions of a small area of interest inside a larger object
142Simplified, user friendly mode that allows the operator quickly navigate the full CT scanning process
Automatic Calibration143Optimize cone-beam CT reconstruction process & simplifies CT & µCT equipment
Cone Beam Reconstruction144Capabilities to perform reconstruction on GPU hardware
1453D volume rendering
146Real time volume rendering
147Export slices at any angle
148Create movie animations
149Perform localized reconstruction
150Ring reduction, which includes a combination of hardware & software to virtually eliminate ring artifacts caused by non-uniform pixel responses
3D Surface Rendering1513D surface rendering which includes surface extraction to CAD compatible format
152Export to STL (stereo lithography) cloud points OBJ, DXF, WRL, & PLY
153Visualization in wireframe mode
154Map textures with reflection
155Export an executable viewer module with a dataset to allow others to view, slice, rotate, window & manipulate the dataset on a traditional laptop or PC
156Viewer capabilities or equivalent to the post-processing capabilities of the CT work station
High Accuracy Geometry Tools157High accuracy tools similar to those used with metrology applications which calculates the geometry more accurately of the scan.
158Serialized small, medium & large tools
159Certificate of conformance of the tools
160High accuracy CMM inspection capabilities
161Precision engineered mounting fixtures
162Heavy duty carrying case
Spiral Acquisition163Perform automated spiral acquisition with a digital flat-panel detector for scanning elongated objects that cannot fit in a traditional single exposure
164Volume stitching & the object length is only limited by the system scanning area size
165Reduce or eliminate the cone angle artifacts caused by flat, planar features common in non-spiral CT techniques
Image Stitching166Capabilities to scan large parts in separate scans and then stitch multiple images into one seamless image of the part or assembly
Sub-Pixel Resolution167Capabilities which include the designed motion of the manipulator & the software to analyze & display the scan data utilizing sub-pixel analysis routines that double the detector capabilities
Time & Motion Analysis1684D computed tomography module
169Creates and analyzes moving or changing parts as a dynamic, time based dataset
170Acquires time-based data in either continuous or stepped mode
171Generates time-sequenced images (videos) showing the motion or change in the part using continuous or stepped scans
Expanded Capabilities172Programmable CNC motion control utilizing programmable CNC motion control
173Detector lateral shift of at least 60 inches
174CT continuous scan mode for faster CT scans in seconds or minutes
175Local CT reconstruction capabilities (reconstruct a small area of interest inside a larger object)
176NO limiting on the size of the reconstruction -- 10000 x 10000 x 10000 voxels reconstruction capabilities
Detector Qualification Package177Control software in an interface that allows operators to routinely monitor digital detectors
178Capabilities to allow for the qualification of DDA pixel mapping & verification of the detector quality, spatial resolution & contrast sensitivity in accordance with ASTM E2737 & ASTM E2698
179SMPTE Display Test pattern capabilities used for checking the quality of the display monitor for Pixel mapping qualification, Detector quality check, Spatial resolution, Contrast sensitivity check, & SMPTE display test pattern.
180EN-462 D13 Duplex wire gauge
181Photo light meter -- Spectroline AccuMax XRP-3000 Raido meter/photo meter kit with spectroline XS-555/L adaptor -- Kit with hard case
Secondary Hard Drive Set1822EA Hot swappable hard drives in both the DR & CT work stations (acquistion & review)
183Capability to separate sensitive and non-sensitive scan data
184Protective storage/carrying case
DR Hardware & Software Package185Review capabilities with high performance image analysis systems & a comprehensive set or image & analysis tools
186Windows 10 based operating system
187NON PROPREITARY image storage format
188High performance image processing & measurement functions utilizing GPU technology
189Automatic parallelization for systems with multiple CPUs and GPUs
190Histogram
191Line profile
192Rotation between portrait & landscape modes
193Measurement and analysis tool for factors (Lines 185-192)
194Read & store images in TIFF 32 bit/16 bit/ 8 bit BMP, JPEG & DICONDE
195Seamless integration with DR software
196DR review work station with Windows 10 operating system
197Eight-Core Intel Xeon Processor
1988 GB RAM
199512 GB Operating System Drive
2003 TB Data storage in RAID 0 configuration
201DVD +/- RW drive
20210/100/1000 network interface
203Nvidia High Performance Display adaptor
20431.5 (800mm) 4Kcolor display (image review quality)
2051 EA wireless keyboard & 1 EA wireless mouse
Additional Components206Copper filter kit: set of copper filters for use with micro focus tubes -- 0.005 inch, 0.010 inch, 0.020 inch, 0.040 inch, 0.060 inch, 0.080 inch thick filters
207High energy copper filter kit -- set of copper filters for use with minifocus tubes -- selection of 0.093 inch, 0.250 inch, 0.375 inch thick filters
Documents208Technical manual -- in English -- covers system operation, routine maintenance & basic trouble shooting information -- standard commercial user/technical manual for the hardware& software requirements
Installation209Install the system in designated building at LEMC by the COR
Training210DR/CT System Training as designated by COR, covering digital radiography, computed tomography & operation/maintenance/safety of the system
2116 seats to digital radiography & intro to CT training (40 hour course) as designated by the COR
2123 seats to Advanced CT Training (40 hour course) as designated by the COR
Warranty213Standard manufacturers warranty
214Extended warranty for 2 additional years
Maintenance215Year 1 -- hardware and software warranty/service -- 1 EA PM visit
216Year 2 -- hardware and software warranty/service -- 1 EA PM visit
217Year 3 -- hardware and software warranty/service -- 1 EA PM visit