This document is a specifications sheet for a temperature/humidity chamber required by the Illinois Emergency Management Agency (EMA) for their Nuclear Calibration Laboratory. The chamber must meet specific requirements for testing electrical leakage in direct reading electrostatic/electroscope type dosimeters, including size, weight, temperature and humidity ranges, ramp times, and utilities. The EMA is seeking bids for 1 unit of the Z-Plus model ZP(H)-16 chamber, which has dimensions of 30" x 30" x 30". The vendor will be responsible for delivery, installation, start-up, and training services.

The related state and local contract opportunity solicits bids for the 1 temperature/humidity chamber, with the contract term anticipated to be a one-time purchase. The solicitation is open to all interested vendors, with no mention of any set-asides or incumbent contractors. Pricing and funding sources are not detailed, but the agency indicates a fork lift and utilities will be available on-site for the vendor's use.

View the file

Other files

File Type Posted
Standard Terms and Conditions v.20.1~70.pdf PDF
Brand Name Only Specification 22200089.pdf PDF
Standard Certifications V.18.1~55.doc DOC document
Forms B Section~16.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPECIFICATIONS FOR Nuclear Calibration Laboratory 1301 Knotts Street | Springfield, IL 62703 EMA.CalibrationLab@illinois.gov 217.786.7221

TEMPERATURE/HUMIDITY CHAMBER

REQUIREMENTS

Purpose:
Electrical leakage testing as described in ANSI N322
Test Item Description:
Direct reading electrostatic/electroscope type dosimeters (loaded in test racks)

Rack diameter: 24” Rack height: 6” Weight (per rack): ~17 lbs

Chamber Size:
Z-Plus Model ZP(H)-16: 30” x 30” x 30”
Test Conditions:
-10 to +50 °C

Ramp time undefined (as quick as possible) Ambient conditions: 72 °F, 20 – 60% RH

Utilities:
208V 1PH 60Hz
Transportation:
Fork lift available, no dock
Accessories:
Casters

Recirculating water Humidity water filtration system Stainless steel shelves Running time meter Power cord

INSTALLATION AND ONSITE SERVICES

· Vendor will deliver systems to the agency, include freight quote

· Onsite installation/start-up and training services should be provided by vendor image1.jpeg