Technical_Specifications-Automated_HEM_SGT-HEM_Extraction_System_R1.pdf
PDF 181 KB Posted
- Attached to
- Automated HEM SGT-HEM Extraction System
- Solicitation number
- 26-P-00172
- Issued by
- Hillsborough County, Florida
About this file
This is a technical specifications document for the City of Tampa, acting through Hillsborough County, Florida, for the procurement of one Automated HEM/SGT-HEM Extraction System (AES) for the Howard F. Curren Advanced Wastewater Treatment Plant Analytical Laboratory. The system must be capable of accurately determining hexane extractable materials (HEM) and silica gel treated hexane extractable materials (SGT-HEM) concentrations in compliance with USEPA Method 1664A across various aqueous sample matrices, including untreated wastewaters, highly treated wastewater effluents, and industrial wastewaters containing high concentrations of suspended particulates and colloidal solids. The system must simultaneously process three 1000mL samples onto 100mm solid phase extraction disks and include six SPE disk holders, nine sample extract receiving vessels, a controller-automated system, an n-hexane evaporator system, US-standard electrical cables and plugs, and all necessary tubing, connectors, clamps, adapters, hoses, cables, solvent bottles, and waste containers. Installation must occur at the Administration Building, Howard F. Curren Advanced Wastewater Treatment Plant, located at 2700 Maritime Boulevard, Tampa, Florida, and must be completed and fully operational within 60 days of purchase order issuance. All installation costs, initial system testing, and familiarization training must be included in the bid price, with the City of Tampa permitted to video record the training. The system must pass a 30-day conformance period following training; failure to conform results in full refund obligations.
The document specifies that the AES must operate in tandem with the City's existing Horizon SPE-DEX 3000 system and be compatible with Pacific Premium Oil & Grease Disks (100mm), Pacific O&G Fast Flow Pre-Filters (90mm), and Aluminum 125mL Weigh Pans (105mm). Bidders with documented histories of unsatisfactory service with the City of Tampa will be deemed non-compliant. Successful bidders must provide one year of no-charge repairs, parts, labor, travel, and living expenses following installation and acceptance, and must make available optional one-year service maintenance contracts for the subsequent 10 years. Bidders must submit documented proof of an established service and technical support network and provide references from at least three TNI-accredited environmental or wastewater laboratories currently operating the specified system. All required documentation, including system manuals, training manuals, and parts catalogs, must be provided in both print and digital formats. The document does not specify pricing terms, set-asides, incumbent vendors, budget amounts, or contract award dates.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
TECHNICAL SPECIFICATIONS
AUTOMATED HEM/SGT-HEM EXTRACTION SYSTEM
1. General Description The Howard F. Curren Advanced Wastewater Treatment Plant Analytical Laboratory is seeking an automated hexane extractable materials (hereinafter "HEM") / silica gel treated hexane extractable materials (hereinafter "SGT-HEM") extraction system (hereinafter "AES") capable of accurately and precisely determining the concentrations of HEM and SGT-HEM as per United States Environmental Protection Agency (hereinafter "USEPA'') Method 1664A in a variety of aqueous sample matrices including, but not limited to, untreated wastewaters, highly treated wastewater effluents, and industrial wastewaters containing high concentrations of suspended particulates and colloidal solids.
For each of the specifications listed in items 2 through 8, the Bidder must clearly state whether the specification is met. For those specifications requiring information to be provided, Bidder must submit the specified information with the bid response.
2. Regulatory Compliance Specifications:
The system supplied must meet the following regulatory compliance specifications:
2.1. The system must be able to meet all method requirements as specified for USEPA SM 1664A.
2.2. The system supplied must be able to meet the relevant quality control standards adopted by
The NELAC Institute (hereinafter “TNI”) as specified under Rule 64E-1.0015 Florida Administrative Code (hereinafter "FAC'') available at https://www.flrules.org/gateway/ruleNo.asp?id=64E-1.0015.
3. Hardware Specifications:
The system hardware supplied must consist of at least the following components:
3.1. Completeness of system
3.1.1. The system must come with US standard electrical cables and plugs compatible with standard US receptacles.
3.1.2. All connection hardware, including any interface or connection cables, must be provided at the time of installation.
3.1.3. The system provided must include all necessary components to perform the automated extraction and evaporation of samples.
3.2. Extraction System
3.2.1. AES shall be capable of simultaneously processing three 1000mL samples onto three 100mm solid phase extraction (hereinafter “SPE”) disks.
3.2.2. AES shall include a controller capable of automating the performance of HEM and SGT-HEM analyses in a manner fully compliant with the analytical and quality control requirements specified in USEPA Method 1664A.
3.2.3. AES shall include all controller actuated valves as required to automate the transfer and processing of samples, solvents, and waste streams.
3.2.4. AES shall include a total of six SPE disk holders appropriate for use with 100mm SPE disks to allow sample rotations and to speed processing efficiency.
3.2.5. AES shall include a total of size sample container cap connectors suitable fo ruse with 1L wide mouth sample bottles using 53mm-400 screw caps.
3.2.6. AES shall include a total of nine appropriate sample extract receiving vessels to allow three vessels to be in use with the remaining vessels available to facilitate the rotation of samples for processing.
3.2.7. AES shall include any vacuum pump necessary to operate to meet the specifications stated herein.
3.2.8. AES shall include one appropriately designed solvent (n-hexane) evaporator system capable of being operated independently of each other and the sample extraction https://www.flrules.org/gateway/ruleNo.asp?id=64E-1.0015 equipment. These solvent recovery systems shall be capable of positioning large (105mm) aluminum weighing pans.
3.2.9. AES shall include all required tubing, connectors, clamps, adapters, hoses, cables, solvent bottles, and waste containers necessary to operate to meet the specifications herein.
3.3. Compatibility with Existing System
3.3.1. AES shall be operated in tandem with our existing Horizon SPE-DEX 3000 system and must be compatible with the existing consumables in use, specifically:
3.3.1.1. Pacific Premium Oil & Grease Disk, 100 mm
3.3.1.2. Pacific O&G Fast Flow Pre-Filter, 90 mm
3.3.1.3. Aluminum 125mL Weigh Pan, 105 mm
4. Installation Specifications:
4.1. Bidder shall install the system in the Administration Building at the Howard F. Curren Advanced Wastewater Treatment Plant located at 2700 Maritime Blvd., Tampa, Florida.
4.2. All costs of installation and initial system testing for conformance to specifications at the time of installation shall be included in the bid price.
4.3. Bidder shall provide familiarization training at the time of system installation.
4.4. Bidder agrees to allow City of Tampa personnel to video record this training for internal purposes.
4.5. The system installation must be complete, and instrument made fully operable by Bidder no later than 60 days after issuance of purchase order.
5. Conformance Specifications:
5.1. The system shall be operated for a period of 30 days following training and must be found to conform to the technical specifications above.
5.2. Bidder agrees to accept the return of the system should the system fail to conform to any required specification within this 30-day period.
5.3. Should the system be returned to the Bidder, the Bidder agrees to provide a full refund of the bid price of the system to the City of Tampa.
6. Document Specifications:
The following documents shall be provided in both print and digital format with the system:
6.1. System manuals for all hardware components.
6.2. Training manuals.
6.3. Parts, supplies, and accessories catalog.
7. Service Specifications:
7.1. Any Bidder with a documented history of unsatisfactory service with the City of Tampa relating to the types of services and/or products described in this bid will be deemed non-compliant.
7.2. For a period of one year following installation and acceptance, all repairs, parts, labor, travel, and living expenses shall be provided by the Bidder on a no charge basis to the City of Tampa as needed to maintain the system to the specifications listed in this bid and to the Bidder's product specifications.
7.3. For a period of 10 years commencing with the end of the one-year period following installation and acceptance, the Bidder shall make available one year service maintenance contract for preventive maintenance and emergency service on both the system hardware and software on a reoccurring annual basis if requested by the City.
7.4. Bidder shall submit with its bid documented proof of an established network of service and technical applications support personnel that shall respond promptly and effectively to any problems associated with the operation of the system bid.
7.5. References of at least 3 TNI accredited environmental / wastewater laboratories which are currently operating the system as specified in this bid must be provided to allow evaluation of Bidder's service and technical support network. References shall be included in the bid submittal.
| 1. General Description |
| 2. Regulatory Compliance Specifications: |
| 2.1. The system must be able to meet all method requirements as specified for USEPA SM 1664A. |
| 2.2. The system supplied must be able to meet the relevant quality control standards adopted by The NELAC Institute (hereinafter “TNI”) as specified under Rule 64E-1.0015 Florida Administrative Code (hereinafter "FAC'') available at https://www.flrule... |
| 3. Hardware Specifications: |
| 3.1. Completeness of system |
| 3.1.1. The system must come with US standard electrical cables and plugs compatible with standard US receptacles. |
| 3.1.2. All connection hardware, including any interface or connection cables, must be provided at the time of installation. |
| 3.1.3. The system provided must include all necessary components to perform the automated extraction and evaporation of samples. |
| 3.2. Extraction System |
| 3.2.1. AES shall be capable of simultaneously processing three 1000mL samples onto three 100mm solid phase extraction (hereinafter “SPE”) disks. |
| 3.2.2. AES shall include a controller capable of automating the performance of HEM and SGT-HEM analyses in a manner fully compliant with the analytical and quality control requirements specified in USEPA Method 1664A. |
| 3.2.3. AES shall include all controller actuated valves as required to automate the transfer and processing of samples, solvents, and waste streams. |
| 3.2.4. AES shall include a total of six SPE disk holders appropriate for use with 100mm SPE disks to allow sample rotations and to speed processing efficiency. |
| 3.2.5. AES shall include a total of size sample container cap connectors suitable fo ruse with 1L wide mouth sample bottles using 53mm-400 screw caps. |
| 3.2.6. AES shall include a total of nine appropriate sample extract receiving vessels to allow three vessels to be in use with the remaining vessels available to facilitate the rotation of samples for processing. |
| 3.2.7. AES shall include any vacuum pump necessary to operate to meet the specifications stated herein. |
| 3.2.8. AES shall include one appropriately designed solvent (n-hexane) evaporator system capable of being operated independently of each other and the sample extraction equipment. These solvent recovery systems shall be capable of positioning large (1... |
| 3.2.9. AES shall include all required tubing, connectors, clamps, adapters, hoses, cables, solvent bottles, and waste containers necessary to operate to meet the specifications herein. |
| 3.3. Compatibility with Existing System |
| 3.3.1. AES shall be operated in tandem with our existing Horizon SPE-DEX 3000 system and must be compatible with the existing consumables in use, specifically: |
| 3.3.1.1. Pacific Premium Oil & Grease Disk, 100 mm |
| 3.3.1.2. Pacific O&G Fast Flow Pre-Filter, 90 mm |
| 3.3.1.3. Aluminum 125mL Weigh Pan, 105 mm |
| 4. Installation Specifications: |
| 4.1. Bidder shall install the system in the Administration Building at the Howard F. Curren Advanced Wastewater Treatment Plant located at 2700 Maritime Blvd., Tampa, Florida. |
| 4.2. All costs of installation and initial system testing for conformance to specifications at the time of installation shall be included in the bid price. |
| 4.3. Bidder shall provide familiarization training at the time of system installation. |
| 4.4. Bidder agrees to allow City of Tampa personnel to video record this training for internal purposes. |
| 4.5. The system installation must be complete, and instrument made fully operable by Bidder no later than 60 days after issuance of purchase order. |
| 5. Conformance Specifications: |
| 5.1. The system shall be operated for a period of 30 days following training and must be found to conform to the technical specifications above. |
| 5.2. Bidder agrees to accept the return of the system should the system fail to conform to any required specification within this 30-day period. |
| 5.3. Should the system be returned to the Bidder, the Bidder agrees to provide a full refund of the bid price of the system to the City of Tampa. |
| 6. Document Specifications: |
| 6.1. System manuals for all hardware components. |
| 6.2. Training manuals. |
| 6.3. Parts, supplies, and accessories catalog. |
| 7. Service Specifications: |
| 7.1. Any Bidder with a documented history of unsatisfactory service with the City of Tampa relating to the types of services and/or products described in this bid will be deemed non-compliant. |
| 7.2. For a period of one year following installation and acceptance, all repairs, parts, labor, travel, and living expenses shall be provided by the Bidder on a no charge basis to the City of Tampa as needed to maintain the system to the specification... |
| 7.3. For a period of 10 years commencing with the end of the one-year period following installation and acceptance, the Bidder shall make available one year service maintenance contract for preventive maintenance and emergency service on both the syst... |
| 7.4. Bidder shall submit with its bid documented proof of an established network of service and technical applications support personnel that shall respond promptly and effectively to any problems associated with the operation of the system bid. |
| 7.5. References of at least 3 TNI accredited environmental / wastewater laboratories which are currently operating the system as specified in this bid must be provided to allow evaluation of Bidder's service and technical support network. References s... |