This statement of work outlines the technical requirements for delivery, installation, and training of a thermogravimetric analyzer - mass spectrometer system. Key requirements include a TGA with autosampler and furnace capable of evolved gas analysis, a heated transfer line, mass spectrometer with range of 1-300 amu, software compatible with Windows 10 without internet, and a computing device. Installation and training by a service technician is required at the specified delivery location within 180 days of award. The related federal contract opportunity is a notice of intent to sole source from the US Department of Agriculture Agricultural Research Service for the thermogravimetric analyzer - mass spectrometer described in the statement of work.

View the file

Other files

File Type Posted
Sole Source Justification.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work Thermogravimetric Analyzer – Mass Spectrometer (TGA-MS)

Scope:

Delivery, installation, and familiarization of a thermogravimetric analyzer – mass spectrometer (TGA-MS) system with PC or laptop, control, all data acquisition software for both traditional TGA data and hyphenated TGA data, and warranty included.

Features, Components, Accessories, Modules, Software and/or Services to be included:

1. Thermogravimetric Analyzer (TGA).

2. Autosampler.

3. Furnace capable of Evolved Gas Analysis (EGA).

4. Heated Transfer Line.

5. Mass Spectrometer (MS).

6. Consumables in quantities required for training and evaluation.

7. Tools and or accessories for routine maintenance and sample preparation

8. PC/monitor or laptop with software that operates the instrument and analyzes the data.

9. Power cords or other cables and plumbing required for proper instrument operation.

10. Installation and training by a service technician.

Technical Requirements/Tasks

1. Sample weight range = 0 – 1,000 mg.

2. Temperature range = Ambient to 1,000°C.

3. Heating rates = 0.1 to 500°C/min.

4. Modulated temperature control for determination of activation energy.

5. High resolution for separation of overlapping weight losses.

6. Gas flow control for balance and sample purge.

7. Autosampler - accommodates 16 or more samples.

8. Gas transfer lines capable of delivery of evolved gasses to MS or IR in parallel.

9. Mass spectrometer with easily maintained ion source.

10. Mass spectrometer with a minimum range of 1 – 300 amu.

11. Consumables in quantities capable of being used during training and evaluation during installation.

12. Software that can be activated and used without internet connectivity, compatible with

Windows 10 and greater for control as well as for analysis and reporting of both TGA and MS instruments.

13. A general computing device with mouse and keyboard input with visual monitor and audio output having specifications and peripheral boards required to connect the instruments and run the required software.

14. Installation and training

Delivery:

Government will be present during delivery and installation of equipment. Equipment will be delivered to 1815 North University Street, Peoria, IL 61604 within 180 days of award. Delivery will require the use of a liftgate.