Statement of Work Summary

This Statement of Work defines the requirement for delivery of two (2) chiropractic examination tables to the VA SLC Utah Health Care System located at 500 Foothill Drive, Salt Lake City, Utah 84148. The contractor must deliver all equipment within 30 days of contract award in manufacturer's original sealed containers marked with the purchase order and contract number. Delivery and installation will be coordinated through designated on-site points of contact: Sara Prosser and Rodolfo Cox (biomedical), and Jeff Maez (warehouse).

The chiropractic tables must meet specific technical specifications including: adjustable height between 21"-29" (optional 24.5"-32") via electric foot pedal, maximum length of 77", standard width of 27", tilting and flexion headpieces with manual forward and straight drop capability, touchscreen control panel, lateral flexion capability with locking up to 21 degrees, maximum weight of 300 pounds, minimum lifting capacity of 450 pounds, armrests with patient gripper bars, air-activated thoracic breakaway and drop functions, lumbar and pelvic drops, slide-out ankle support with harnesses, and compatibility with 110V 60Hz or 220V 50Hz electrical requirements. The contractor shall provide one (1) year warranty on all parts and labor, including travel and shipping costs for warranty repairs, along with one binder and one digital copy of manufacturer operation and maintenance manuals. Security and C&A requirements do not apply.

View the file

Other files

File Type Posted
Questions and Answers.docx DOCX document
Solicitation 36C215926Q0811 - Revisions.docx DOCX document
852.219-76 VA Notice of Limitations on Subcontracting—Certificate of Compliance for Supplies and Products.docx DOCX document
36C25926Q0811 0001.docx DOCX document
Wage Determination 2015-5489.pdf PDF
Buy-American-Certificate.pdf PDF
36C25926Q0811.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work Chiropractic Exam Table

PURPOSE

The purpose of this requirement is to deliver two (2) Chiropractic Table to VA SLC Utah Health Care System, located at 500 Foothill Drive Salt Lake City, Utah 84148.

SCOPE
The Contractor shall deliver all items listed
All items must meet all salient characteristics defined
The Contractor shall furnish all personnel, supplies, and equipment required for delivery.
ITEM SPECIFICATIONS AND SALIENT CHARACTERISTICS
Detailed Item Description Example: Chiropractic Table
Quantity: Two (2) each

Technical Specifications/Salient Characteristics

· Height will range between 21 ½”-29” (Optional 24.5” – 32”) controlled by electric foot pedal

· Length must not exceed 77”

· Width must be the standard 27”

· Must include Tilting Headpiece and Flexion Headpiece

· Must have manual forward and straight drop headpiece

· Will include control panel touchscreen

· Lateral Flexion must move and lock up to 21°

· Weight of table not to exceed 300lbs

· Must support lifting capacity up to 450lbs

· Armrests will have patient gripper bars

· Must have air activated thoracic breakaway, thoracic drop with release, lumbar drop, pelvic drop, and lateral flexion with locking mechanism

· Must have slide out ankle support with harnesses

· Must be compatible with electric requirement of 110V 60Hz-7A-5A or 220V-50Hz-4A

· Minimum one (1) year warranty on all parts

DELIVERY AND INSTALLATION
Delivery Timeframe: within 30 days of award unless specified differently within scope of contract.

Delivery Location: Contractor shall deliver all equipment to the VA SLC Utah Health Care System, located at 500 Foothill Drive Salt Lake City, Utah 84148.

Delivery Markings: Contractor shall deliver items in manufacturer's original sealed containers with manufactures’s name marked thereon. Deliveries shall be marked with the PO and contract number.

Delivery and Installation will be coordinated through the On-Site Point of Contact (POC). POCs for delivery and install are:

Sara Prosser – sara.prosser@va.gov (biomed) Rodolfo Cox - rodolfo.cox@va.gov (biomed) Jeff Maez - russell.maez@va.gov (warehouse)

OPERATION AND MAINTENANCE MANUALS

Binders - Quantity (1) each for items identified Digital Copies - Quantity (1) each for items identified Deliver compilation of all manufacturer recommended maintenance schedule and operation materials packaged in binder(s) to On-Site Point of Contact upon completion of installation.

SECURITY REQUIREMENTS

The C&A requirements do not apply. A Security Accreditation Package is not required.

WARRANTY

The contractor shall provide all manufacturer(s) warranties on all parts and labor at time of delivery.

The warranties shall include all travel and shipping costs associated with any warranty repair.