A2. SOW 461513 Specs.pdf
PDF 54 KB Posted
- Solicitation number
- SS_461513
- Issued by
-
Department of Energy Office of Science
About this file
This Statement of Work (SOW) details the procurement of two specialized microscopy research instruments for Brookhaven National Laboratory. The first item is a Liquid Nitrogen (LN2) cooling Transmission Electron Microscope (TEM) holder with electrical feedthroughs, featuring precise heating and biasing control from -173°C to 800°C, double-tilt capability, compatibility with MEMS chips and TEM grids, and a removable 4-hour liquid nitrogen dewar. The second item is a Scanning Electron Microscopy (SEM) stage with high-frequency capabilities, designed with RF cable integration for in situ biasing experiments and a flippable head stage for switching between SEM and TKD configurations.
The deliverables have specific timeline requirements: the TEM holder is to be completed 150 days after contract award, while the SEM stage is to be delivered 360 days after contract award. This is a pre-solicitation Request for Information (RFI) issued by the Department of Energy Office of Science to conduct market research and identify potential sources capable of manufacturing these specialized scientific instruments. Interested parties are instructed to submit a capabilities statement and SAM.gov UEI number to Eric Nagel at enagel@bnl.gov, with the Buy American Act applicable to this requirement.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work
Brief description of the items:
The TEM sample holder is used to mount thin research samples for insertion into the transmission electron microscope (TEM). The item to be purchased has special capabilities that allows apply electric bias to the thin TEM samples at variable temperatures from LN2 (~ 100 K) to 1000 K (or 800 oC). The item should have MEMS chip that allows heating without significant sample drift. Additionally, this purchase includes the co-development of an SEM stage for high-frequency excitation of samples.
Specifications:
Item1: Liquid nitrogen (LN2) cooling TEM holder with electrical feedthroughs
-Simultaneous and precise heating and biasing control during cooling, with temperatures ranging from LN2 to 800°C.
-Double-tilt capability (range depends on pole-piece type).
-Compatibility with both MEMS chips and 3mm TEM grids.
-Removable, large liquid nitrogen dewar with at least 4 hours of operation time.
-Integrated heater and sensor inside the holder shaft for enhanced stability.
-Specialized MEMS chip design, including: 4-contact microheater for temperature stability in the mK regime, and capability to combine high E-field with 800°C without reaching breakdown voltage.
Item2: Scanning electron microscopy (SEM) stage with high-frequency capabilities
-Integration of RF cables for high-frequency in situ biasing experiments with temperature-controlled heating.
-Flippable and removable head stage for switching between in situ SEM (sample facing up) and in situ TKD (sample facing down).
-Compatibility with the same MEMS chip for both the Lightning Arctic TEM holder and SEM stage.
Deliverables:
Item 1: Liquid nitrogen (LN2) cooling TEM holder with electrical feedthroughs – 150 days after contract award.
Item 2: Scanning electron microscopy (SEM) stage with high-frequency capabilities – 360 days after contract award.
X Myung-Geun Han