SPS Water Purification System Salient Characteristics.docx
DOCX document 29 KB Posted
- Solicitation number
- 36C25525Q0089
About this file
This document contains salient characteristics for SPS water purification systems to be installed at two VA Medical Center locations - Leavenworth and Topeka. The systems are designed to meet ST108 water requirements using house soft water as input.
Both locations require similar core components including PE Carbon filters, UV sanitizers, reverse osmosis systems (2.5 GPM/3,600 GPD for Leavenworth; 5 GPM/7,200 GPD for Topeka), endotoxin post filters, 100-gallon cone-bottom storage tanks, recirculation pumps, dual PE Mixed Bed DI tanks, and resistivity monitoring sensors. The installations include removal of old equipment, new equipment installation, and temporary DI systems during installation. Both locations require identical maintenance services including monthly water quality testing, filter replacements, loop disinfection, quarterly carbon tank exchanges, biannual endotoxin filter changes, annual UV bulb replacement, and regular water testing at multiple locations (critical monthly testing and quarterly testing for utility hot/cold water and steam).
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Topeka Site Map.pdf | ||
| S01_36C25525Q0089_EKHCS_SPS Water Purification Systems.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
?.## SPS WATER PURIFICATION SYSTEM(S) SALIENT CHARACTERISTICS
This is the salient characteristics to meet ST108 water requirements. House soft water will feed the system to meet the ST108 requirement for EKHCS VAMC at both Leavenworth and Topeka campus.
SPS System – Leavenworth
· PE Carbon Ultraviolet Sanitizer
· Reverse Osmosis o 2.5 GPM / 3,600 GPD
· (2) Ø4” x 40” Membranes
· 120V x 1-Ph Up to 30 GPM Flow
· Endotoxin Post Filter
· Stainless Steel housing
· Holds (4) 20” filter cartridges
· 100-Gal Storage Tank
· Cone-Bottom
· ∅50” x 92”H
· Includes hydrophobic vent, level sensors, pressure gauges, sample port, loop return spray ball, etc
· Recirculation Pump
· 0-60 GPM Flow
· 120V x 1-Ph
· (2) PE Mixed Bed DI Tanks
· Includes Quality Indicator Lights, 1-MegOhm (115V x2)
· 12 GPM Max Flow
· 1.5 ft³ Mixed Bed Resin Media 20 GPM Flow Capacity
· Resistivity Monitoring Sensor
· Monitor conductivity/resistivity and flow.
· Includes remote monitoring.
· Accuracy ±3% of reading / ±1% of repeatability
· LCD display
· 115V (x1)
· Installation
· Includes removal of old equipment, installation of new equipment, and temporary DI during install
· Temporary DI system will include: 12” Carbon, 12” Tri-bed, 12” Mixed Bed, and Endotoxin Post Filter
This is the same system as Leavenworth, except the RO is sized larger to accommodate feeding the steam boiler.
SPS System – Topeka
· PE Carbon Ultraviolet Sanitizer
· Reverse Osmosis
· 5 GPM / 7,200 GPD
· (4) Ø4” x 40” Membranes
· 120V x 1-Ph Up to 30 GPM Flow
· Endotoxin Post Filter
· Stainless Steel housing
· Holds (4) 20” filter cartridges
· 100-Gal Storage Tank
· Cone-Bottom
· ∅50” x 92”H
· Includes hydrophobic vent, level sensors, pressure gauges, sample port, loop return spray ball, etc
· Recirculation Pump
· 0-60 GPM Flow
· 120V x 1-Ph
· (2) PE Mixed Bed DI Tanks
· Includes Quality Indicator Lights, 1-MegOhm (115V x2)
· 12 GPM Max Flow
· 1.5 ft³ Mixed Bed Resin Media 20 GPM Flow Capacity
· Resistivity Monitoring Sensor
· Monitor conductivity/resistivity and flow
· Includes remote monitoring
· Accuracy ±3% of reading / ±1% of repeatability
· LCD display
· 115V (x1)
· Installation
· Includes removal of old equipment, installation of new equipment, and temporary DI during install
· Temporary DI system will include: 12” Carbon, 12” Tri-bed, 12” Mixed Bed, and Endotoxin Post Filter
SERVICES & CONSUMABLES – TOPEKA
| WATER SYSTEM QUALITY TESTING |
| MONTHLY |
| 12 |
| 10" 25-MICRON PRE-FILTER |
| MONTHY |
| 12 |
| LOOP DISINFECT |
| MONTHLY |
| 12 |
| 9" MIXED BED DI |
| QUARTERLY |
| 4 |
| 12" CATALYTIC CARBON TANK EXCHANGE |
| QUARTERLY |
| 4 |
| 20" ENDOTOXIN FILTER |
| BIANNUALLY |
| 2 |
| UV BULB & SLEEVE |
| ANNUALLY |
| 1 |
| WATER TESTING - CRITICAL |
| MONTHLY |
| 12 |
IN, OUT, WASHER, ULTRA SONIC, & SINK (5-LOCATIONS)
| WATER TESTING - UTILITY HOT |
| QUARTERLY |
| 4 |
IN, OUT, WASHER, ULTRA SONIC, & SINK (5-LOCATIONS)
| WATER TESTING - UTILITY COLD |
| QUARTERLY |
| 4 |
IN, OUT, WASHER, ULTRA SONIC, & SINK (5-LOCATIONS)
| WATER TESTING - STEAM (1-LOCATION/SAMPLE) |
| QUARTERLY |
| 4 |
*FACILITY WILL NEED TO PROVIDE SAMPLE PORTS TO PERFORM TESTING*
SERVICES & CONSUMABLES - LEAVENWORTH
| WATER SYSTEM QUALITY TESTING |
| MONTHLY |
| 12 |
| 10" 25-MICRON PRE-FILTER |
| MONTHY |
| 12 |
| LOOP DISINFECT |
| MONTHLY |
| 12 |
| 9" MIXED BED DI |
| QUARTERLY |
| 4 |
| 12" CATALYTIC CARBON TANK EXCHANGE |
| QUARTERLY |
| 4 |
| 20" ENDOTOXIN FILTER |
| BIANNUALLY |
| 2 |
| UV BULB & SLEEVE |
| ANNUALLY |
| 1 |
| WATER TESTING - CRITICAL |
| MONTHLY |
| 12 |
IN, OUT, WASHER, ULTRA SONIC, & SINK (5-LOCATIONS)
| WATER TESTING - UTILITY HOT |
| QUARTERLY |
| 4 |
IN, OUT, WASHER, ULTRA SONIC, & SINK (5-LOCATIONS)
| WATER TESTING - UTILITY COLD |
| QUARTERLY |
| 4 |
IN, OUT, WASHER, ULTRA SONIC, & SINK (5-LOCATIONS)
| WATER TESTING - STEAM (1-LOCATION/SAMPLE) |
| QUARTERLY |
| 4 |
File details come from the government source that posted it. Updated .