About this file
This specification document supports a micro purchase for Automated External Defibrillator (AED) and Personal Protective Equipment (PPE) kits issued by the City of Pennsylvania. The procurement requests quotes for all items included in the specification to ensure bidder eligibility for award consideration. The micro purchase will be awarded as a whole or by individual line items, whichever serves the City's best interests, to the lowest responsive and responsible bidder. Vendors must not deliver products or provide services until receiving formal notification from the department. This procurement will result in a requirements contract upon award.
The specification establishes pricing terms whereby all bidders must submit quotes for the complete scope of items to qualify for consideration. The City reserves discretion in award methodology to optimize value and operational efficiency. No specific funding amounts, renewal options, set-asides, or incumbent information is detailed within this reference specification document. Bidders should anticipate that award will establish an ongoing requirements-based relationship with the City for the specified AED and PPE kit products.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Physical
Item Specification
Size (H × W × D) 5.25 × 9.50 × 11.50 in (13.3 × 24.1 × 29.2 cm)
Weight 6.7 lb (3.1 kg), including batteries
Ingress protection IP55 (dust-protected and protected against water jets)
Display Optional LCD with moving bar, 2.6 × 1.3 in (6.6 × 3.3 cm); optional 2.6 s ECG window at 25 mm/s
Defibrillator
Waveform ZOLL Rectilinear Biphasic
Energy (factory default)
Adult: 120 J, 150 J, 200 J. Pediatric (Pedi-padz II): 50 J, 70 J, 85 J. Energy follows the electrode type connected. Configurable within rules (each shock same or higher than previous).
Charge time Less than 10 seconds with new batteries
Charge hold time 30 seconds
Patient impedance range
0–300 ohms
Patient safety All patient connections electrically isolated; defibrillation-protected ECG circuitry
Shockable rhythms (typical)
VF with average amplitude >100 µV; wide-complex VT >150 BPM (adult) or >200 BPM (pediatric)
Power
Batteries 10 user-replaceable Type 123A lithium manganese dioxide photo-flash cells (off-the-shelf, e.g. Duracell)
Capacity (typical new, 20 °C)
About 5 years standby, or ~225 shocks, or ~13 hours continuous monitoring
End-of-life indication Red X in status window (typically ~9 shocks remaining)
Electrodes
• CPR-D-padz: one-piece adult electrodes with integrated CPR sensor; ~5-year shelf life; cable ~48 in (1.2 m)
• Pedi-padz II: pediatric (generally under 8 years or <25 kg / 55 lb)
• Stat-padz II also compatible
Real CPR Help (depth/rate feedback) requires CPR-D-padz and is intended for adult victims.
CPR feedback (Real CPR Help)
• Adaptive metronome, typically 60–100 CPM (designed to support AHA/ERC target rates)
• Compression depth sensing range commonly listed as 1.9–8.9 cm (¾–3.5 in)
• Voice and visual prompts such as “PUSH HARDER” / “GOOD COMPRESSIONS”
Self-test and readiness
• Automatic self-test configurable every 1–7 days (default 7 days)
• Monthly full-energy (200 J) test
• Checks battery, electrode connection, ECG and charge/discharge circuits, microprocessor, CPR circuitry/sensor, and audio Environment
Item Typical specification (PS / standard models)
Operating temperature 32° to 122°F (0° to 50°C)
Storage temperature About −22° to 140°F (−30° to 60°C); some aviation/PS listings go to 158°F / 70°C
Humidity 10–95% RH, non-condensing
Shock IEC 68-2-27; 100 G
Vibration MIL-STD-810F, minimum helicopter test
Altitude −300 to 15,000 ft (−91 to 4,573 m)
Data and connectivity
• Records ECG and CPR data (capacity depends on configuration; examples include several hours of ECG/CPR, or ~20 minutes when audio recording is enabled)
• IrDA port for configuration and event download
• PC software historically specified for Windows (not Mac)
Standards (typical listings) Meets applicable requirements of IEC 60601-1, IEC 60601-2-4, AAMI/ANSI DF-39, and related EMC/safety standards. Classification is commonly Class II, internally
JL Industries
Model # :WBB765034
MPN #: C1413F12
JL Industries AED Cabinet Surface Mount, 85 db Audible Alarm, Steel https://www.globalindustrial.com/shopbybranditems?cp=0&p=~brand=JL%20Industries
Write a Review
Questions & Answers (0) Shop All Defibrillator/AED Cabinets & Accessories
Product Description JL AED cabinets fit most AED's on the market. These durable powder coat painted cabinets are equipped with audible 85 db alarm and available in fully recessed, semi recessed or surface mount.
Features:
• CRS with white powder coat finish
• 1-3/4" Trim on face and door
• Mounted in a 5/8" door stop and attached by a continuous hinge
See Less Deta
Contents Description Piece Count
AED Instruction Card 1 Bandage Shears 1
BZK Antiseptic Towelettes 2 CPR Face Shield 1 Disposable Razor 1 Emergency Blanket 1 https://www.globalindustrial.com/c/medical-lab/medical_devices_instruments/defibrillator_accessories?pGid=51185&cid=72045#PG-72045-51185
Nitrile Exam Gloves 2