Specifications.pdf

PDF 157 KB Posted

Attached to
Thermogravimetric Analysis Instrument Federal contract opportunity
Solicitation number
12505B24Q0177
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Midwest Area

About this file

This document contains the specifications for a Thermogravimetric Analysis (TGA) Instrument required under federal contract solicitation number 12505B24Q0177. The TGA Instrument is needed to analyze and characterize changes in material weight as a function of temperature or time. Key requirements include: automated furnace control up to 1000°C, linear heating rates of 0.1 to 100°C/min, sample weight capacity of 1,000 mg or less, weight baseline drift of 25 μg or lower, temperature precision of ±0.1°C or better, integrated programmable autosampler, and capability for future expansion to include spectroscopic instruments. The procurement is a 100% Total Small Business Set-Aside with the associated NAICS code of 334516 (Analytical Laboratory Instrument Manufacturing). Quotes are due by 10:00 AM CDT on May 24, 2024 and must include technical specifications and pricing. Award will be made on a Lowest Price Technically Acceptable (LPTA) basis to the responsible offeror submitting the lowest evaluated price.

View the file

Other files for this federal contract opportunity

Other files attached to Thermogravimetric Analysis Instrument, newest first.
File Type Posted
Clauses and Provisions Amend 01.pdf PDF
SF-30 - Amend 01.pdf PDF
Clauses and Provisions.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Thermogravimetric Analysis (TGA) Instrument

SPECIFICATIONS

Scope: Delivery, installa�on, and familiariza�on of a fully automated Thermogravimetric Analysis (TGA) Instrument used for analyzing and characterizing the change in material weight as a func�on of temperature or �me.

A TGA monitors the increase or decrease in the mass of a sample as a func�on of temperature or �me as the sample is subjected to a controlled temperature program under a controlled atmosphere. The instrument consists of a sample pan or crucible that is weighed on a precision balance. The pan is placed into a furnace where it is heated in a controlled atmosphere and the weight of the sample is monitored during the experiment. The atmosphere gas can be inert or reac�ve and flows over the sample and is exited through an exhaust port. Materials for characteriza�on include foods, carbohydrates, proteins, fats and lipids, and petroleum products among other types of samples.

The TGA instrument will be capable of quan�fying water, solvent, and plas�cizer loss, decarboxyla�on, pyrolysis, oxida�on, decomposi�on, weight percent filler, amount of metallic cataly�c residue remaining on carbon-based sample, and weight ash percentage. The TGA will have an integrated programmable autosampler for placing and removing sample pans from the measurement cell. Ideally, the autosampler will also have the ability to pierce or remove pan lids use to keep moisture out of a sample un�l the sample can be run by the TGA. Many experiments will be conducted from room temperature to 1000 °C and the hea�ng furnace of the TGA should be capable of reaching an upper opera�ng temperature of at least 1000 °C and capable of maintaining an isothermal temperature and using a controlled hea�ng ramp rates as needed. Instrument so�ware provides programmed control of hea�ng and cooling temperature ramps, isothermal scans, an autosampler for placing and removing sample pans and mass flow controllers for switching furnace atmosphere gases within the furnace during an experiment. The so�ware records data and analyzes the results, graphs the data and exports tabulated data and graphs to electronic format for interpre�ng the results. Preferably, the TGA instrument should not require a mechanical cooler (refrigerated cooling system-not dependent on liquid nitrogen) to cool the furnace a�er an experiment.

Features, Components, Accessories, Modules, So�ware and/or Services to be included:

1. Automated Thermogravimetric Analysis (TGA) Instrument with built in furnace and integrated precision balance for analyses.

2. Automated furnace control with a temperature range from ambient to 1000 °C.

3. Linear controlled hea�ng rates of 0.1 to 100 °C/min or beter.

4. Sample weight capacity 1,000 mg or less.

5. Weight baseline dri� 25 µg or lower.

6. Resolu�on 0.1 µg or lower.

7. Temperature precision = ± 0.1 °C or beter.

8. Thermally isolated precision balance with low dri� and high sensi�vity to deliver the most accurate real-�me data.

9. Touch screen user interface and/or computer interface and analyses so�ware.

10. Computer, monitor, keyboard, and so�ware for interface, instrument control, and graphing, measurement, analysis, and export of TGA experiments and data.

Thermogravimetric Analysis (TGA) Instrument

11. Integrated programmable autosampler for placing and removing sample pans from the measurement cell.

12. Sealed pan op�on is desirable to maintain the atmosphere of the sample un�l the experiment can be run. The ability to pierce or remove pan lids by instrument.

13. Mass flow controllers for automated switching of furnace atmosphere/purge gases.

14. Power requirements: U.S.-compa�ble (120 VAC, 50/60 Hz, single phase).

15. Capability to u�lize pla�num pans or ceramic crucibles capable of withstanding 1000 °C, and aluminum pans (600 °C temperature limit).

16. Capable of future expansion for spectroscopic instruments such as Fourier Transform Infrared

Spectrometer (FTIR) or Mass spectrometer (MS).

17. Start-up opera�ng tools and supplies.

18. Onsite installa�on and training.

19. Minimum one-year parts and labor warranty on instrument

Delivery: Delivery of this item will require a li� gate.