Summary of Transport Ventilation Machine Specifications

This document contains technical specifications for two transport ventilation machine configurations intended for pediatric and adult patient care. Both systems are designed for patients weighing 8 kg and above, with portable capability for use in ground, aerial, and aquatic environments across indoor and outdoor settings. The machines feature shock-resistant, reinforced housing with vibration protection to withstand movement during patient transport. All specifications require 8.4-inch diagonal monitors with 640x480 pixel resolution and battery-powered operation.

The portable transport model includes dimensions of 320x220x270mm (unit only) or 630x630x1380mm (without handle) and 630x630x1433mm (with handle), weighing 6.5 kg with 4-8 hours of battery operating time. The stationary trolley-mounted model has compact unit dimensions of 310x210x245mm and trolley dimensions of 630x630x1380mm, weighing 4.9 kg (10.8 lbs) as a unit or 16.9 kg (37.3 lbs) with trolley, providing 4-6 hours of battery operation. Both systems must replicate bedside ventilation modes and settings. The trolley model includes an oxygen cylinder holder and provides expanded connectivity options (CO2, SpO2, Nurse Call, Humidifier integration, USB, RJ45 Ethernet), CPR ventilation capability, suctioning functionality, on-screen help, and support for High Flow O2, NIV S/T, nCPAP, and Neonatal ventilation modes.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications

Transport Ventilation Machine

1. Ventilator or producing parts must be used from a 8kg patient to an adult. Must be for peds and adult.

2. Must be able to lift easily and transport with the patient, mobility is a must.

3. Can be used interior and exterior of a building. Ground, in the air and water.

4. Shock-resistant and able to withstand movement while in use with patient in transport.

Reinforced housing with impact protection and vibration proof

5. Able to use the same mode and settings as one at the bedside.

6. Baterry powered when needed.

7. Dimensions (WxDxH) 320x220x270mm, with out a handle 630x630x1380mm, with handle

630x630x1433mm

8. Weight: 6.5kg

9. Monitor size and resolution: 8.4 in diagonal, 640x480 pixels

10. Battery operating time: 4-8 hours

1. Ventilator or producing parts must be used from a 8kg patient to an adult. Must be for peds and adult.

2. Has oxygen cylinder holder

3. Dimensions: (WxDxH) 310x210x245mm (unit), 630x630x1380mm (with trolley)

4. Weight: 4.9 kg/10.8 lb, 16.9kg/37.3 lb with trolley

5. Monitor size and resolution: 8.4 in diagonal, 640x480 pixels

6. Battery opeating time: 4-6 hours

7. Connectivity: CO2/Nurse Call/COM1, CO2/Sp02/COM1, CO2/Humidifer & COM1, USB Port, RJ-

45 Ethernet Port

8. CPR Ventilation

9. Suctioning tool

10. On-screen help

11. High Flow O2/NIV, NIV S/T, nCPAP, Neonatal

File details come from the government source that posted it. Updated .