Programmable AC/DC Power Supply System
Solicitation number
SP4706-27-Q-0002
Issued by

About this file

This is a Technical Specifications document dated November 2026 for a Programmable AC/DC Power Supply System intended for laboratory testing, avionics/electronics evaluation, power conversion testing, and hardware validation. The system must operate from standard 110-120 VAC, 60 Hz single-phase input via NEMA 1-15, NEMA 5-15 (15A), or NEMA 5-20 (20A) plugs. Due to facility electrical infrastructure constraints, multiple power connections to separate standard 110-120 VAC receptacles are explicitly acceptable, and synchronized modular units (such as three or four 1.5-2.0 kVA benchtop/rack units operating in master-slave parallel or polyphase synchronization) are fully acceptable and encouraged. The system must incorporate active power factor correction exceeding 0.95 at full load.

AC output specifications require variable voltage from 12-360 VAC (line-to-neutral), frequency range from 16-800 Hz (covering railway, commercial, and avionics frequencies), user-selectable single-phase or three-phase (Wye) output, 0-15 A RMS single-phase or 0-5 A RMS per phase three-phase capability, total harmonic distortion less than 1.5% from 16-400 Hz and 3.0% up to 800 Hz, and line/load regulation of ±0.1%/±0.2% full scale respectively. DC output specifications require variable voltage from 0 to ±250 VDC (supporting positive and negative polarity), 0-20 A DC with constant voltage and constant current modes, and output ripple/noise less than 0.5% RMS. The system must provide real-time telemetry display via either front-panel high-contrast display or Windows 10/11 compatible computer control software with USB, Ethernet/LAN, RS-232, or GPIB connectivity; providing both interfaces is preferred. Built-in protections must include over-voltage, over-current (with programmable limit), over-power, over-temperature, and continuous output short-circuit protection. The system requires internal thermostatically controlled forced-air fan cooling, compliance with UL 61010-1/IEC 61010-1 safety and FCC Part 15 Class A/EN 61326-1 EMC standards, operation in ambient temperatures 0°C to +40°C with 10-80% non-condensing relative humidity, minimum one-year manufacturer warranty, and NIST-traceable calibration certificate upon delivery including user/programming manuals and synchronization/interconnect cables.

View the file

Other files

File Type Posted
SF1449_SP470627Q0002.pdf PDF
Power Supply PWS Redacted.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

TECHNICAL SPECIFICATIONS

Programmable AC/DC Power Supply System

Date: November 2026 Class: Unclassified / Procurement

1. SCOPE AND PURPOSE

This Technical Specifications Document defines the minimum mandatory electrical, mechanical, operational, and interface requirements for a Programmable Alternating Current / Direct Current (AC/DC) Power Supply System. The system will be utilized for laboratory testing, avionics/electronics evaluation, power conversion testing, and hardware validation requiring variable voltage, current, and frequency across single-phase AC, three-phase AC, and bipolar/unipolar DC operational modes while operating from standard North American 110-120 VAC wall receptacles.

2. INPUT POWER REQUIREMENTS & ARCHITECTURE

Due to facility electrical infrastructure constraints, the power supply system must operate from standard commercial/laboratory branch circuits without requiring high-voltage three-phase facility hardwiring.

• Input Voltage & Frequency: Must accept standard 110-120 VAC, 60 Hz single-phase input power.

• Plug / Connector Compatibility: Must interface directly via standard NEMA 1-15, NEMA 5-15 (15A), or NEMA 5-20 (20A) plugs.

• Multi-Cord / Multi-Unit Topologies Permitted:

■ Because the maximum output power requirements (up to 5.4 kVA AC or 5.0 kW DC) exceed the continuous power limit of a single NEMA 5-15 (1.44 kW continuous) or NEMA 5-20 (1.92 kW continuous) receptacle, multiple power connections to separate standard 110-120 VAC receptacles are explicitly acceptable.

■ Synchronized Modular Units: A system comprised of multiple individual programmable power supply units synchronized together (e.g., three or four 1.5 kVA - 2.0 kVA benchtop/rack units operating in master-slave parallel and/or polyphase synchronization via high-speed sync cables) is fully acceptable and encouraged to meet the aggregate single-phase, three-phase, and DC output requirements.

• Power Factor Correction (PFC): Active PFC (> 0.95 typical at full load) to maximize usable power draw per 120 VAC branch circuit.

Page 1 of 3UNCLASSIFIED / FOR OFFICIAL PROCUREMENT USE ONLY

3. MINIMUM OUTPUT POWER REQUIREMENTS

3.1 Alternating Current (AC) Output Specifications

Parameter Minimum Specification Requirement Notes / Configuration

Output Voltage Range Variable from 12 VAC to 360 VAC (Line-to-Neutral)

Continuous or dual-range electronic switching

Output Frequency Range Variable from 16 Hz to 800 Hz Covers 16.7 Hz railway, 50/60 Hz commercial, and 400-800 Hz avionics/defense

Output Phases Selectable between 1-Phase (Single-Phase) and 3-Phase (Wye)

User-selectable via front panel or control software

Output Current (1-Phase Mode)

Variable from 0 to 15 A RMS Achieved via single chassis or paralleled master-slave units

Output Current (3-Phase Mode)

Variable from 0 to 5 A RMS per phase Balanced 3-phase output (120° phase separation ±0.5°)

Total Harmonic Distortion

(THD)

< 1.5% from 16-400 Hz; < 3.0% up to 800 Hz

Into linear resistive load

Line & Load Regulation ±0.1% of full scale (Line); ±0.2% of full scale (Load)

Remote sense capability included

3.2 Direct Current (DC) Output Specifications

Parameter Minimum Specification Requirement Notes / Configuration

Output Voltage Range Variable from 0 to ±250 VDC Supports positive and negative polarity (via bipolar electronic output or automated/manual polarity reversal / dual-channel configuration)

Output Current Range Variable from 0 to 20 A DC Constant Voltage (CV) and Constant Current (CC) modes

Output Ripple & Noise < 0.5% RMS of full scale Measured at 20 Hz to 20 MHz bandwidth

4. TELEMETRY, DISPLAY, AND CONTROL INTERFACE REQUIREMENTS

• Real-Time Telemetry Display: The system shall incorporate a means of real-time display of output voltage (V RMS / V DC), output current (A RMS / A DC), and output frequency (Hz).

• Implementation Options: Real-time metering must be provided via either:

1. An incorporated high-contrast front-panel display (LCD/LED/OLED digital readout per phase/unit), OR

2. Accompanying computer control software (Windows 10/11 compatible GUI via USB, Ethernet/LAN, RS-232, or GPIB) providing real-time digital readouts, logging, and setpoint control. (Providing both

TECHNICAL SPECIFICATIONS: PROGRAMMABLE AC/DC POWER SUPPLY

Page 2 of 3UNCLASSIFIED / FOR OFFICIAL PROCUREMENT USE ONLY front-panel and PC software interfaces is preferred).

5. PROTECTION, SAFETY, AND ENVIRONMENTAL COMPLIANCE

• Built-In Protections: Over-Voltage Protection (OVP), Over-Current Protection (OCP) with programmable current limit, Over-Power Protection (OPP), Over-Temperature Protection (OTP), and continuous Output Short-Circuit Protection.

• Cooling: Internal thermostatically controlled forced-air fan cooling exhausting to rear or side panels.

• Safety & EMC Standards: UL 61010-1 / IEC 61010-1 safety compliance; FCC Part 15 Class A / EN 61326-1 EMC compliance.

• Operating Environment: Ambient temperature 0°C to +40°C; relative humidity 10% to 80% non-condensing.

• Warranty & Documentation: Minimum one (1) year manufacturer warranty, NIST-traceable calibration certificate upon delivery, user/programming manuals, and synchronization/interconnect cables.

TECHNICAL SPECIFICATIONS: PROGRAMMABLE AC/DC POWER SUPPLY