Technical Specifications for Hydrostatic Test Stand.pdf

PDF 65 KB Posted

Attached to
Hydrostatic Test Stand Federal contract opportunity
Solicitation number
SP470624Q0029
Issued by
Defense Logistics Agency

About this file

This document contains the Technical Specifications for a Hydrostatic Test Stand required by the Defense Logistics Agency (DLA) Land and Maritime Product Test Center. The key requirements include:

The Hydrostatic Test Stand must be new, have high and low pressure pneumatic pumps, be capable of generating and holding pressures up to 20,000 psi, include 5 digital pressure gage ranges, have an accuracy of 0.5% or better, and have a protective see-through door/lid. The unit dimensions cannot exceed 110" wide, 48" deep, and 102" high. The contractor is responsible for initial setup, providing operation/maintenance manuals, and one day of on-site training. Initial calibration and certifications are to be included. The Test Stand will be used to conduct Proof, Leakage, Burst, and Electrical Conductivity testing of DLA customer supplied products like hoses and valves.

View the file

Other files for this federal contract opportunity

Other files attached to Hydrostatic Test Stand, newest first.
File Type Posted
SOW Hydrostatic Test Stand.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical Specifications for:

Hydrostatic Test Stand for pressure testing vessels

DLA Land and Maritime Engineering & Technical Support Directorate

Product Test Center Division, Mechanical Lab, VTM

1. Objective:

Our objective in acquiring a new Hydrostatic Test Stand consists of ensuring the new Test

Stand meets our strict requirements in pressure generation, test control, accuracy, ease of use, maintenance, simplified calibration and providing optimal safety when working with high pressures. As a result of purchasing a new unit the MPTC hopes to increase our scope of testing, provide a safe operating environment, reduce test cycle time and give DLA Customers a comprehensive and accurate test.

2. General Requirements:

1. Hydrostatic Test Unit must be new, not refurbished.

2. Unit must have High and Low pressure, pneumatically driven pumps.

3. Unit must have the ability to supply required pressures and hold indefinitely, as in no internal leakage or pressure loss to hoses, control valves, gages etc.

4. Contractor is responsible for initial setup of test unit.

5. Contractor is responsible for suppling operation and maintenance manuals.

6. System must be “turnkey” after setup. Contractor must supply all necessary connectors, hose, hardware, etc.

7. On-site training (1 day) for 5 associates at DSCC, DLA Land and Maritime Product Test

Center, Mechanical Lab, VTM, Columbus, OH to be included in purchase.

8. Initial calibration to be included in purchase and certifications provided.

9. Overall Pressure Measuring Parameters must be 0-20,000 psi.

10. Must include a minimum of 5 different pressure gage ranges. Anticipated gage ranges:

0-500, 0-1,000, 0-3,000, 0-10,000 and 0-20,000 pound per square inch (psi).

11. All gages must be digital if used as a standard or reference gage contributing to data collection.

12. All gages used in pressure generation or supply can be of analog type.

13. Accuracy of standard or reference gages must be 0.5% or better.

14. Overall unit dimensions cannot exceed 110” wide, 48” deep and 102” high.

15. Test chamber must have a protective and see through door/lid, illuminated and withstand testing pressures up to 20,000 psi.