This is a Statement of Work (SOW) for the procurement of calibration standards and ancillary equipment to support Aviation Fleet Intermediate Calibration Activities and Navy Calibration Laboratories. The requirement specifies the purchase of three (3) MAHR 130B-16 Gage Block Comparators designed for measuring long gage blocks up to 24 inches or 600mm. Each unit must include a one-year manufacturer warranty with warranty certificate, calibration certificate, calibration sticker, serialization, manufacturer's standard operating and servicing manuals, and shipping to the designated destination. The equipment must meet specific technical specifications including a measuring range of 1.100-24.000 inches or 50-600mm, resolution selectable between 1 microinch or 0.1 microinch/0.1mm or 1.01mm, repeatability of 6 sigma ±1 (25nm), and linearity within 1 microinch (0.025mm) in the central 50 inches (1.25mm) and elsewhere. The system base weight is 400 pounds (180kg) with dimensions of 15 inches length by 15 inches width by 30 inches height, and requires laboratory temperature maintenance in an ideal range of minus 68 to 0.5°F or 20 to 0.25°C with a maximum range of minus 68 to 1.0°F or 20 to 0.5°C.
All procurement must follow an approval process including fleet submission, fleet TYCOM review and prioritization, NAVAIR METCAL technical review for workload justification and standardization, and overall procurement strategy assessment. The related solicitation (N00421-26-Q-1093) indicates this is a sole source procurement with Mahr, Inc., issued by the Department of the Navy Naval Air Systems Command, with quotes required by March 20, 2026 at 12:00pm Eastern Standard Time. All offerors must be registered in SAM.GOV prior to submission, and the Government will acquire this on a firm fixed-price basis with new products only, with no substitutions permitted.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Solicitation - N0042126Q1093.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work A. Background The use of NAVAIR METCAL equipment funding is to procure Calibration Standards (CALSTDs) and ancillary equipment for Aviation Fleet Intermediate Calibration Activities, Fleet Training Activities and Aviation Navy Calibration Laboratories (NCLs) to support Aviation Organizational and Intermediate (I) level maintenance assets. All Cal Standards listed for procurement must first go through an approval process. This process includes initial submission by the fleet; review and prioritization by the Fleet TYCOM’s; laboratory workload justification and standardization review by a NAVAIR METCAL technical review team; overall procurement strategy review to assess overall program objectives, priorities and funding.
REQUIREMENTS:
1. Mfr. one-year warranty with a warranty certificate.
2. Calibration Certificate and calibration sticker (as applicable for each unit)
3. Units to be serialized.
4. Include shipping to destination below.
5. Manufacturer’s Standard Operating and servicing manuals (CD version acceptable) to be provided with each unit. The DFARS clause listed below must be referenced in all contracts and the manufacturer made aware that the use/reproduction of tech data/manuals (for government use only) will follow these use guidelines.
DFARS SUBPART 227.71 - Rights in Technical Data
B. Tasking:
1. P/N 130B-16 Gage Block Comparator for Long Gage Blocks (up to 24"/600mm) with independent controller Qty 3 To include the following:
A. Measuring Range: 1.100-24.000” 2.50-600mm B. Resolution: 1μ” or 0.1μ/ 0.1μm or 1.01μm (selectable) C. Repeatability: 6σ ≤ 1µ (25nm) D. Linearity: within 1µin (0.025μm) in the central ±50µ” (± 1.25μm) and within 1µ (0.025μm) in any 50µ” (1.25μm) elsewhere E. Compressed air requirement: None F. System base weight: 400lbs/180kg G. System base dimensions: 15” Lx15” Wx30”H/ 381mm L x 381mm W x 762mm H H. Lab temperature requirements:
a. Ideal range: -68±0.5ºF/20±0.25ºC
b. Max range: -68±1.0ºF/20±0.5ºC A. Gradient -<0.1ºC/hr <0.18ºF/hr I. P/N 2240154 Tungsten Carbide upper contact tip installed Qty 3 J. P/N 2239733 Tungsten Carbide lower contact tip installed Qty 3 K. P/N 2239299 Foot Pedal for measurement data acquisition Qty 3 L. P/N 2239651 Instruction Manual Qty 3 M. P/N CV-26 Storage cover Qty 3 N. P/N 2263475 Label 2.0x3.0x.02 Alum Qty 3 O. P/N 9967676 Application training precimar by an application engineer for 1-3 participants P. P/N 9968430 Travel Time and Expenses Qty 3 Q. P/N 2253440 Thermometer with two precision probes Qty 3 R. P/N 2240602 Gage Block Measurement Accessories Kit Qty 3 To include:
i. Fitted case
ii. Forceps
iii. Tongs
iv. Brush
v. Blower
vi. Chamois clothes
vii. Deburring stone
viii. Optical flat
ix. Force gage
x. Hex wenches
xi. Rust inhibiting grease C. Deliverables Delivery of the equipment to the following address:
Qty 1 to:
NAVAIR Calibration Standards Attn: Nelson Dale BLDG 4533 Code 6.2.2.4.
MCAS Cherry Point NC 28533-0021
Qty 1 to:
NCLPR
Attn: DAVE MAHONEY 22113 Fortin Circle Bldg. 1403 NAWC Patuxent River, MD 20670-1118
Qty 1 to:
NAVAIR North Island, NPSL ATTN: Al Teruel BLDG. 469 South, Dispatch B-9 San Diego, CA 92135-7058
PH: 619-545-2857
D. Delivery Date Delivery shall be no later than 20 weeks after award of contract *****Early delivery accepted**** E. Invoicing
a. All invoicing shall be submitting into Wide Area Workflow for payment after all material is received in the condition as stated in section B.