SATPC0041463 Tab 04 4 SOW..pdf
PDF 167 KB Posted
- Attached to
- Scanivalve Calibration System Federal contract opportunity
- Solicitation number
- 80NSSC26936068Q
About this file
This is a Statement of Work for NASA Glenn Research Center to procure a complete Scanivalve pressure calibration system. The primary objective is the purchase of a Scanivalve SPC4050 2-Channel Pressure Calibrator with IS-33 transducers providing 0.008% full-scale accuracy and an internal barometer. Channel A includes Low Pressure Servo Regulators with transducers calibrated to 5.75 psig and 17 psig, while Channel B includes Medium Pressure Servo Regulators with transducers calibrated to 56 psig and 115 psig. The system features a color touch screen, removable/interchangeable transducers, and is designed for automated DSA5002 calibrations in either bench or rack-mount configurations.
The complete deliverable system comprises four line items: the SPC4050 pressure calibrator, a Scanivalve SPC4050PLU-2 Pneumatic Logic Unit (2-channel, 90 psi control) with capabilities for positive and vacuum calibrations, a Scanivalve 70380-02 Pneumatic Distribution Panel (2-channel, 90 psi control) providing supply and control pressure, and a Scanivalve 73680 4-Range Portable Calibration System with all three components installed in a durable protective case for transport. The vendor must provide a one-year warranty against defects in materials and workmanship from original delivery, with any identified nonconformities handled as non-conforming deliverables. The period of performance is 14 weeks from authorization to proceed, with the place of performance at NASA Glenn Research Center in Cleveland, Ohio.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SATPC0041463 Tab 07 Capability Statement.pdf | ||
| SATPC0041463 Tab 06 RDSS..pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements
The NASA Glenn Research Center requires the purchase of a Scanivalve Pressure Calibrator.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 SPC4050 SPC4050 2-Channel Pressure Calibrator (SPECIAL: with IS-33, 0.008% FS Accuracy), with internal barometer Channel A: Low Pressure Servo Regulator
(LPSVR)
Transducer 1: 5psig (Cal'd 5.75psig) Transducer 2: 15psig (Cal'd 17psig) Channel B: Medium Pressure Servo Regulator (MPSVR) Transducer 1: 50psig (Cal'd 56psig) Transducer 2: 100psig (Cal'd 115psig)
2 SPC4050PLU-2 Pneumatic Logic Unit, 2-Channel (90psi
CTL)
3 70380-02 Pneumatic Distribution Panel, SPC4050, 2 Channel, 90 psi Control
4 73680 4-Range Portable Calibration System with items 8-10 installed. Including Pneumatic / Electrical interface panel, internal pneumatic and electrical connections, testing and documentation. REF: Sales Dwg 80068-01
System shall support the following pressure controller/ calibrator:
Scanivalve SPC4050 with the following salient specifications:
• ChA and ChB transducers - IS-33, 0.008% FS Accuracy
• Internal barometer channel
• Channel A: Low Pressure Servo Regulator (LPSVR) Transducer 1: 5psig (Cal
5.75psig)
• Transducer 2: 15psig (Cal 17psig)
• Channel B: Medium Pressure Servo Regulator (MPSVR) Transducer 1: 50psig (Cal
56psig)
• Transducer 2: 100psig (Cal 115psig)
• Removeable/ Interchangeable transducers
• Color touch screen
• Used for automated DSA5002 calibrations
• Bench or rack-mount
System shall support the following pneumatic logic unit:
Scanivalve SPC4050PLU-2 Pneumatic Logic Unit with the following salient specifications:
• 2 – channel, 90 psi control pressure
• Positive or below atmospheric (vacuum) calibrations
• Zero offset correction logic
• Leak and response testing on pressure scanners
System shall support the following pneumatic distribution panel:
Scanivalve 70380-02 Pneumatic Distribution Panel with the following salient specifications:
• 2 – channel, 90 psi control pressure
• Provides all necessary supply and control pressure to SPC4050 and SPC4050PLU-2
System shall support the following portable calibration unit/ case:
Scanivalve 73680 4-range Portable Calibration System with the following salient specifications:
• SPC4050 installed
• SPC4050PLU-2 installed
• 70380-02 installed
• Durable case for the above items to allow for ease of transport
• Protection for sensitive instrumentation in transit
Warranty:
The vendor shall provide a warranty against defects in materials and workmanship for a period of 1 year following completion of work. Any nonconformities identified by NASA as defects in materials and workmanship from the original delivery shall be handled as Non- Conforming Deliverables as described above.
3. Place of Performance
NASA Glenn Research Center 21000 Brookpark Rd Cleveland OH 44135-3127
4. Period of Performance
14 weeks ARO.
File details come from the government source that posted it. Updated .