SATPC0041463 Tab 04 4 SOW..pdf

PDF 167 KB Posted

Attached to
Scanivalve Calibration System Federal contract opportunity
Solicitation number
80NSSC26936068Q
Issued by
National Aeronautics and Space Administration Shared Services Center

About this file

This is a Statement of Work for NASA Glenn Research Center to procure a complete Scanivalve pressure calibration system. The primary objective is the purchase of a Scanivalve SPC4050 2-Channel Pressure Calibrator with IS-33 transducers providing 0.008% full-scale accuracy and an internal barometer. Channel A includes Low Pressure Servo Regulators with transducers calibrated to 5.75 psig and 17 psig, while Channel B includes Medium Pressure Servo Regulators with transducers calibrated to 56 psig and 115 psig. The system features a color touch screen, removable/interchangeable transducers, and is designed for automated DSA5002 calibrations in either bench or rack-mount configurations.

The complete deliverable system comprises four line items: the SPC4050 pressure calibrator, a Scanivalve SPC4050PLU-2 Pneumatic Logic Unit (2-channel, 90 psi control) with capabilities for positive and vacuum calibrations, a Scanivalve 70380-02 Pneumatic Distribution Panel (2-channel, 90 psi control) providing supply and control pressure, and a Scanivalve 73680 4-Range Portable Calibration System with all three components installed in a durable protective case for transport. The vendor must provide a one-year warranty against defects in materials and workmanship from original delivery, with any identified nonconformities handled as non-conforming deliverables. The period of performance is 14 weeks from authorization to proceed, with the place of performance at NASA Glenn Research Center in Cleveland, Ohio.

View the file

Other files for this federal contract opportunity

Other files attached to Scanivalve Calibration System, newest first.
File Type Posted
SATPC0041463 Tab 07 Capability Statement.pdf PDF
SATPC0041463 Tab 06 RDSS..pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

1. Objective/Requirements

The NASA Glenn Research Center requires the purchase of a Scanivalve Pressure Calibrator.

2. Characteristics, Scope, and Specs

LINE PART NO. DESCRIPTION QTY.

1 SPC4050 SPC4050 2-Channel Pressure Calibrator (SPECIAL: with IS-33, 0.008% FS Accuracy), with internal barometer Channel A: Low Pressure Servo Regulator

(LPSVR)

Transducer 1: 5psig (Cal'd 5.75psig) Transducer 2: 15psig (Cal'd 17psig) Channel B: Medium Pressure Servo Regulator (MPSVR) Transducer 1: 50psig (Cal'd 56psig) Transducer 2: 100psig (Cal'd 115psig)

2 SPC4050PLU-2 Pneumatic Logic Unit, 2-Channel (90psi

CTL)

3 70380-02 Pneumatic Distribution Panel, SPC4050, 2 Channel, 90 psi Control

4 73680 4-Range Portable Calibration System with items 8-10 installed. Including Pneumatic / Electrical interface panel, internal pneumatic and electrical connections, testing and documentation. REF: Sales Dwg 80068-01

System shall support the following pressure controller/ calibrator:

Scanivalve SPC4050 with the following salient specifications:

• ChA and ChB transducers - IS-33, 0.008% FS Accuracy

• Internal barometer channel

• Channel A: Low Pressure Servo Regulator (LPSVR) Transducer 1: 5psig (Cal

5.75psig)

• Transducer 2: 15psig (Cal 17psig)

• Channel B: Medium Pressure Servo Regulator (MPSVR) Transducer 1: 50psig (Cal

56psig)

• Transducer 2: 100psig (Cal 115psig)

• Removeable/ Interchangeable transducers

• Color touch screen

• Used for automated DSA5002 calibrations

• Bench or rack-mount

System shall support the following pneumatic logic unit:

Scanivalve SPC4050PLU-2 Pneumatic Logic Unit with the following salient specifications:

• 2 – channel, 90 psi control pressure

• Positive or below atmospheric (vacuum) calibrations

• Zero offset correction logic

• Leak and response testing on pressure scanners

System shall support the following pneumatic distribution panel:

Scanivalve 70380-02 Pneumatic Distribution Panel with the following salient specifications:

• 2 – channel, 90 psi control pressure

• Provides all necessary supply and control pressure to SPC4050 and SPC4050PLU-2

System shall support the following portable calibration unit/ case:

Scanivalve 73680 4-range Portable Calibration System with the following salient specifications:

• SPC4050 installed

• SPC4050PLU-2 installed

• 70380-02 installed

• Durable case for the above items to allow for ease of transport

• Protection for sensitive instrumentation in transit

Warranty:

The vendor shall provide a warranty against defects in materials and workmanship for a period of 1 year following completion of work. Any nonconformities identified by NASA as defects in materials and workmanship from the original delivery shall be handled as Non- Conforming Deliverables as described above.

3. Place of Performance

NASA Glenn Research Center 21000 Brookpark Rd Cleveland OH 44135-3127

4. Period of Performance

14 weeks ARO.

File details come from the government source that posted it. Updated .