SATPC0036174 Tab 04 4 Spec sheet.pdf
PDF 109 KB Posted
- Attached to
- Service Contract for Deionized Water System Federal contract opportunity
- Solicitation number
- 80NSSC24879320Q
About this file
This document contains the specifications for the maintenance of a Deionized water system for the Microfabrication Laboratory at the NASA Glenn Research Center (GRC). The maintenance contract covers the periodic replacement of various filter cartridges, tanks, UV lamps, and sleeves, as well as the provision of bacteria samples and annual system sanitization.
The related federal contract opportunity is a Request for Quote (RFQ) for a Service Contract for Deionized Water System, issued by the NASA Shared Services Center. The RFQ and Statement of Work (SOW) are attached. The contract is for the National Aeronautics and Space Administration, a civilian federal agency.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SATPC0036174 Tab 04 Questions and answers.pdf | ||
| SATPC0036174 Tab 04 attachment.pdf | ||
| SATPC0036174 Tab 10 Open Market RFQ.pdf | ||
| SATPC0036174 Tab 10 Open Market RFQ.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Deionized water system maintenance
The Smart Sensing and Electronics System branch at NASA Glenn Research Center (GRC) at Lewis Field focuses on developing silicon carbide (SiC) integrated circuits (ICs) and sensors for aerospace applications. NASA GRC solicits quotes for maintenance of a Deionized water system for the Microfabrication Laboratory that is used to fabricate the ICs and sensors.
The following items are to be covered by the maintenance contract:
•Replace (1) 20” 5-micron (BB) Pre-filter cartridge every three months.
•Replace (5) 20” 5-micron Pre-filters every three months.
•Replace (1) 20” 1-micron Pre-filter every three months.
•Replace (1) 3.6 Cu. Ft. Activated Carbon tank lead/lag once every six months.
•Replace (12) 3.6 Cu. Ft. Mixbed Deionizer Tanks each year. If exceeded tanks billed as separate.
•Provide and install (8) UV TOC lamp once every year.
•Provide and install (2) UV Gold S UV lamps once every year.
•Provide and install (10) UV Sleeves once every other year.
•Replace (6) 40” 0.2 micron (222)-W/Fin Post-filter cartridge every six months.
•Replace (1) 20” 0.2 micron (222) Post-Filter every six months.
•Replace (1) 20” 0.2 micron (DOE) Post-filter cartridge every six months.
•Take (2) Bacteria samples and report results each month.
•Sanitize system once a year.